K080133 · A.M. Surgical, Inc. · EMF · Mar 13, 2008 · General, Plastic Surgery
Device Facts
Record ID
K080133
Device Name
MOUNTABLE ENDOSCOPIC BLADE
Applicant
A.M. Surgical, Inc.
Product Code
EMF · General, Plastic Surgery
Decision Date
Mar 13, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4800
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The Mountable Endoscopic Blade is indicated for use in minimally invasive ligament or fascia release. - Carpal tunnel release for carpal tunnel syndrome . - Plantar fasciotomy for plantar fasciitis . - Lateral release of the knee for extensor mechanism malalignment . - Forearm fasciotomy for compartment syndrome . - Morton's neuroma for interdigital neuroma - Gastrocnemius release for heel cord contracture (equinus) .
Device Story
Mountable Endoscopic Blade is a manual surgical instrument designed to mount onto a 4mm 30° endoscope. Used by surgeons in clinical settings for minimally invasive ligament or fascia release procedures. Device facilitates precise cutting during endoscopic surgery; benefits include reduced invasiveness compared to open procedures. No electronic, software, or algorithmic components.
Clinical Evidence
No clinical data provided; device is a manual surgical instrument.
Technological Characteristics
Manual surgical instrument; stainless steel construction; mounts to 4mm 30° endoscope; no energy source; no software.
Indications for Use
Indicated for minimally invasive ligament or fascia release in patients with carpal tunnel syndrome, plantar fasciitis, extensor mechanism malalignment, compartment syndrome, interdigital neuroma, or heel cord contracture (equinus).
Regulatory Classification
Identification
A manual surgical instrument for general use is a nonpowered, hand-held, or hand manipulated device, either reusable or disposable, intended to be used in various general surgical procedures. The device includes the applicator, clip applier, biopsy brush, manual dermabrasion brush, scrub brush, cannula, ligature carrier, chisel, clamp, contractor, curette, cutter, dissector, elevator, skin graft expander, file, forceps, gouge, instrument guide, needle guide, hammer, hemostat, amputation hook, ligature passing and knot-tying instrument, knife, mallet, disposable or reusable aspiration and injection needle, disposable or reusable suturing needle, osteotome, pliers, rasp, retainer, retractor, saw, scalpel blade, scalpel handle, one-piece scalpel, snare, spatula, stapler, disposable or reusable stripper, stylet, suturing apparatus for the stomach and intestine, measuring tape, and calipers. A surgical instrument that has specialized uses in a specific medical specialty is classified in separate regulations in parts 868 through 892 of this chapter.
K982142 — MOUNTABLE ENDOSCOPIC KNIFE · A.M. Surgical, Inc. · Dec 30, 1998
K211297 — SMARTRELEASE Endoscopic Soft Tissue Release System · Microaire Surgical Instruments · Jun 24, 2021
K970826 — SELECTA-TIP LAPAROSCOPIC SCISSORS · Microline Pentax, Inc. · Mar 28, 1997
K223406 — SmartRelease Endoscopic Soft Tissue Release System · Microaire Surgical Instruments · Feb 2, 2023
K241855 — SB Knife Jr2 (MD-47702 and MD-47702L) · Sb-Kawasumi Laboratories, Inc. · Mar 26, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
K080133
Page 1 of 2
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CRF 807, this information serves as a Summary of Safety and Effectiveness for the use of AM Surgical's Mountable Endoscopic Blade.
| Submitted By: | A.M. Surgical, Inc. |
|-------------------------------------|----------------------------------------------------------------|
| Date: | January 14, 2008 |
| Contact Person: | Vincent Pascale<br>Phone: 800-437-9653<br>Fax: 631-980-4369 |
| Proprietary Name: | Mountable Endoscopic Blade |
| Common Name: | Surgical Knife |
| Classification Name and Reference: | 21 CFR 878.4800, Manual Surgical Instrument<br>for General Use |
| Device Product Code and Panel Code: | General and Plastic Surgery/EMF |
### DEVICE INFORMATION
#### INTENDED USES/ INDICATIONS A.
The Mountable Endoscopic Blade is indicated for use in minimally invasive ligament or fascia release.
- Carpal tunnel release for carpal tunnel syndrome .
- Plantar fasciotomy for plantar fasciitis .
- Lateral release of the knee for extensor mechanism malalignment .
- Forearm fasciotomy for compartment syndrome
- . Morton's neuroma for interdigital neuroma
- Gastrocnemius release for heel cord contracture (equinus) .
#### DEVICE DESCRIPTION B.
The Mountable Endoscopic Blade is a device that mounts to a 4mm 30° endoscope.
{1}------------------------------------------------
Ko 80133
Page 2 of 2
#### C. MATERIALS
The blade is manufactured from stainless steel. There is no change in material for the Mountable Endoscopic Blade.
#### SUBSTANTIAL EQUIVALENCE INFORMATION D.
The intended use, material composition, and design features of the Mountable Endoscopic Blade are substantially equivalent to previously cleared 510(k) Mountable Endoscopic Blade (K982142). The safety and effectiveness of Mountable Endoscopic Blade is adequately supported by the substantial equivalence information provided within this Premarket Notification.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" are arranged in a circular pattern around the eagle. The logo is black and white and appears to be a scanned image.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 3 2008
A.M. Surgical, Inc. % Mr. Vincent Pascale 290 East Main Street, Suite 200 Smithtown, New York 11787
Re: K080133
Trade/Device Name: Mountable Endoscopic Blade Regulation Number: 21 CFR 878.4800 Regulation Name: Manual surgical instrument for general use Regulatory Class: II Product Code: EMF Dated: March 5, 2008 Received: March 5, 2008
Dear Mr. Pascale:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
### Page 2 - Mr. Vincent Pascale
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N Millman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Ko80133
## Indications for Use Statement
510(k) Number: Not yet assigned.
Device Name: Mountable Endoscopic Blade
Indications For Use:
The Mountable Endoscopic Blade is indicated for use in minimally invasive ligament or fascia release.
- · Carpal tunnel release for carpal tunnel syndrome
- · Plantar fasciotomy for plantar fasciitis
- · Lateral release of the knee for extensor mechanism malalignment
- · Forearm fasciotomy for compartment syndrome
- · Morton's neuroma for interdigital neuroma
- · Gastrocnemius release for heel cord contracture (equinus)
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number
Prescription Use _
OR
Over-The Counter Use
(Optional Format 1-2-
(Per21 CFR 801.109) તેર)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil Rl Oglin for mxn
Division of General, Restorative. and Neurological Devices
510(k) Number. K080133
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches