AMS COLLAGEN DERMAL MATRIX

K050445 · American Medical Systems, Inc. · PAG · Jun 17, 2005 · General, Plastic Surgery

Device Facts

Record IDK050445
Device NameAMS COLLAGEN DERMAL MATRIX
ApplicantAmerican Medical Systems, Inc.
Product CodePAG · General, Plastic Surgery
Decision DateJun 17, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The AMS collagen dermal matrix is intended for use in the treatment of hernias where the connective tissue has ruptured or for implantation to reinforce soft tissues where weakness exists in the urological, gynecological and gastrointestinal anatomy. This includes but is not limited to the following procedures: pubourethral support including urethral slings, urethral and vaginal prolapse repair, colon and rectal prolapse repair, reconstruction of the pelvic floor, bladder support, tissue repair, sacral colposuspension and reinforcement in the repair of Peyronie's disease. By providing pubourethral support, the AMS collagen dermal matrix may be used for the treatment of urinary incontinence resulting from urethral hypermobility or intrinsic sphincter deficiency. The Bioarc SP Sling Kit with Pre-connected Collagen Dermal Matrix and Bioarc TO Sublascial Hammock with Pre-connected Collagen Dermal Matrix are intended for the placement of a pubourethral sling for the treatment of stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency. The AMS Perigee System with Pre-Connected Collagen Dermal Matrix is intended for the placement of graft material in the anterior vaginal wall via the obturator foramen for the treatment of anterior wall prolapse. The AMS Apogee System with Pre-Connected Collagen Dermal Matrix is intended for use in vaginal vault suspension to treat pelvic organ prolapse.

Device Story

Device consists of decellularized porcine dermis, lyophilized and terminally sterilized; provided as standalone matrix or pre-connected to polypropylene mesh systems (Bioarc, Perigee, Apogee). Systems include needles and connectors for surgical placement. Used by surgeons in clinical/OR settings for pelvic floor reconstruction, hernia repair, and stress urinary incontinence treatment. Acts as graft material to reinforce weakened soft tissues. Clinical benefit derived from structural support provided by porcine collagen matrix. No software or algorithmic components involved.

Clinical Evidence

Bench testing only; no clinical data presented. Device tested in accordance with FDA's Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh.

Technological Characteristics

Decellularized porcine dermis; lyophilized; terminally sterilized. Available in various sizes. Systems include polypropylene mesh, needles, and connectors. Standalone device; no electronic or software components.

Indications for Use

Indicated for patients requiring soft tissue reinforcement or hernia repair in urological, gynecological, or gastrointestinal procedures, including urethral slings for stress urinary incontinence (urethral hypermobility/intrinsic sphincter deficiency), pelvic organ prolapse repair, and Peyronie's disease repair.

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K050445 1/2 JUN 17 2005 ## 510(k) Summary AMS Collagen Dermal Matrix 510(k) Number Date of Summary Preparation: February 18, 2005 Submitter/Contact Person: Elsa A. Linke Regulatory Affairs Specialist American Medical Systems 10700 Bren Rd. W Minnetonka, MN 55343 > Phone: (952) 930-6000 Fax: (952) 930-6496 Device Name and Classification: Trade Name(s): AMS Collagen Dermal Matrix AMS Apogee System with Pre-Connected Collagen Dermal Matrix AMS Apogee Oyetom with Pre-Connected Collagen Dermal Matrix AMS Fenger Systems TO with Pre-Connected Collagen Dermal Matrix Common/Usual Name: Surgical Mesh Classification Name: Surgical Mesh Product Code: DTP, PAI Classification: Class II Manufacturing Location: American Medical Systems, Inc. 10700 Bren Rd. West Minnetonka, MN 55343 Predicate Devices: For AMS Collagen Dermal Matrix : DermMatrix/InteXen - K021160 Surgisis Sling - K992159 For AMS Bioarc SP & Bioarc TO with Pre-Connected Collagen Dermal Matrix : AMS Bioarc SP - K040538 AMS Bioarc TO -- K041948 For AMS Perigee System with Pre-Connected Collagen Dermal Matrix : AMS Perigee System - K040623 For AMS Apogee System with Pre-Connected Collagen Dermal Matrix : AMS Apogee System - K040537 Indications for Use: The AMS collagen dermal matrix is intended for use in the treatment of hernias where the The AMS collagen ochharmania in to implantation to reinforce soft tissues where weakness Connective lissoc has ruptured of for implanations of this includes but is {1}------------------------------------------------ КОТОШИГ .2/2 not limited to the following procedures: pubourethral support including urethral slings, urethral and vaginal prolapse repair, colon and rectal prolapse repair, reconstruction of the pelvic floor, bladder support, tissue repair, sacral colposuspension and reinforcement in the repair of Peyronie's disease. By providing pubourethral support, the AMS collagen dermal matrix may be used for the treatment of urinary incontinence resulting from urethral hypermobility or intrinsic sphincter deficiency. The Bioarc SP Sling Kit with Pre-connected Collagen Dermal Matrix and Bioarc TO Sublascial Hammock with Pre-connected Collagen Dermal Matrix are intended for the placement of a pubourethral sling for the treatment of stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency. The AMS Perigee System with Pre-Connected Collagen Dermal Matrix is intended for the placement of graft material in the anterior vaginal wall via the obturator foramen for the treatment of anterior wall prolapse. The AMS Apogee System with Pre-Connected Collagen Dermal Matrix is intended for use in vaginal vault suspension to treat pelvic organ prolapse. ## Device Description: The AMS collagen dermal matrix is decellularized porcine dermis that is lyophilized and terminally sterilized. The product is available in a range of sizes. The Bioarc SP Sling Kit with Pre-connected Collagen Dermal Matrix and Bioarc TO Sublascial Hammock with Pre-connected Collagen Dermal Matrix consist of needles and connectors used to pass a polypropylene mesh preconnected to porcine dermis for use as a urethral sling. The Perigee System with Pre-Connected Collagen Dermal Matrix consists of needles and connectors used to pass a polypropylene mesh preconnected to porcine dermis in support of the anterior vaginal wall. The AMS Apoqee System with Pre-Connected Collagen Dermal Matrix consists of needles and connectors used to pass a polypropylene mesh preconnected to porcine dermis in support of the vaginal vault. ## Summary of Testing The AMS collagen dermal matrix and all of the pre-connected devices have been tested in accordance with the requirements of FDA's Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh and has been shown to be equivalent to the listed predicate devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol is a stylized representation of a human figure, with three faces overlapping each other. The logo is simple and clean, and it is easily recognizable. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Ms. Elsa A. Linke Regulatory Affairs Specialist American Medical Systems Incorporated 10700 Bern Road West MINNETONKA MN 55343 SEP 2 8 2012 Re: K050445 > Trade/Device Name: AMS Collagen Dermal Matrix, AMS Apogee System with Pre-Connected AMS Collagen Dermal Matrix, AMS Perigee System with Pre-Connected AMS Collagena Dermal Matrix, AMS BioArc™ SP Sling Kit and BioArc To Subfascial Hammock with Pre-connected AMS Collagen Dermal Matrix Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: OTP, PAI, PAJ, OTN, PAG Dated: May 24, 2005 Received: May 25, 2005 Dear Ms. Linke: This letter corrects our substantially equivalent letter of June 17, 2005. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {3}------------------------------------------------ comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Sincerely yours, Benjamin R. Perkins Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ КОГОЧИГ ## INDICATIONS FOR USE 510(k) Number (if known): Device Name: AMS Collagen Dermal Matrix Indications For Use: The AMS collagen dermal matrix is intended for use in the treatment of The AMO onliagon connective tissue has ruptured or for implantation to reinforce soft tissues where weakness exists in the urological, reinforce contineerological anatomy. This includes but is gynecological and gaol.com procedures: pubourethral support including nor limited to the following prolapse repair, colon and rectal urethnul Silings, are.narruction of the pelvic floor, bladder support, prolapse repair, sacral colposuspension and reinforcement in the repair of the Peyronie's disease. By providing pubourethral support, the AMS collagen demal matrix may be used for the treatment of urinary incontinence resulting from urethral hypermobility or intrinsic sphincter deficiency. Prescription Use______________________________________________________________________________________________________________________________________________________________ Use (Per 21 CFR 801 Subpart D) Over-The Counter (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) AND/OR Concurrence of CDRH, Office of Device Evaluation (ODE) ( Division of General, Restorative and Neurological Devices Page 1 of 4 5.29(c) Number KOS0445
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...