K190185 · C.R. Bard, Inc. · OWT · Jun 12, 2019 · General, Plastic Surgery
Device Facts
Record ID
K190185
Device Name
Phasix ST Mesh with Open Positioning System
Applicant
C.R. Bard, Inc.
Product Code
OWT · General, Plastic Surgery
Decision Date
Jun 12, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Phasix ST Mesh with Open Positioning System is indicated for use in the reinforcement of soft tissue, where weakness exists, in procedures involving soft tissue repair, such as for the repair of hernias. The open positioning system is intended to facilitate the placement, positioning and fixation of the mesh during open ventral hernia repair.
Device Story
Phasix ST Mesh with Open Positioning System is a sterile, single-use, bi-layer surgical mesh designed for open ventral hernia repair. The mesh consists of a posterior layer (Phasix ST Mesh) and an anterior layer (Phasix Mesh) stitched together with P4HB monofilament to form a pocket. A removable accessory, comprising a polypropylene handle and PTFE guide, is preinserted into this pocket to assist the surgeon in placing, positioning, and fixing the mesh over the hernia defect. Once the mesh is initially fixed, the accessory is removed and discarded. The device is used by surgeons in an operating room setting. By providing a temporary structural guide, the system simplifies the handling and orientation of the mesh, potentially improving surgical efficiency and accuracy during soft tissue reinforcement.
Clinical Evidence
No clinical data was required. Evidence consists of bench testing (mesh pore size, ball burst, tear strength, suture retention, recoil, grip strength, and accessory removal force) and a 4-week GLP porcine study comparing the subject device to the Ventrio ST Hernia Patch and Phasix ST Mesh. The animal study evaluated mesh conformance, tissue attachments, contracture, histological response, and mechanical tissue ingrowth, showing comparable results.
Technological Characteristics
Bi-layer mesh composed of Phasix ST Mesh and Phasix Mesh stitched with 10 mil P4HB monofilament. Accessory consists of polypropylene handle and PTFE guide. Sterilization via ethylene oxide. Device is non-electromechanical; no software or electrical components.
Indications for Use
Indicated for reinforcement of soft tissue where weakness exists, specifically for soft tissue repair procedures such as hernia repair. The open positioning system is indicated to facilitate mesh placement, positioning, and fixation during open ventral hernia repair.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
K182008 — Phasix ST Mesh with Echo 2 Positioning System · C.R. Bard, Inc. · Dec 14, 2018
K170294 — VENTRALIGHT ST Mesh with ECHO 2 Positioning System 11cm (4.5) Circle, VENTRALIGHT ST Mesh with ECHO 2 Positioning System 10cm x 15cm (4x 6) Ellipse, VENTRALIGHT ST Mesh with ECHO 2 Positioning System 15cm (6) Circle, VENTRALIGHT ST Mesh with ECHO 2 Positioning System 15cm x 20cm (6x8) Ellipse, VENTRALIGHT ST Mesh with ECHO 2 Positioning System 15cm x 25cm (6x10) Oval · C. R. Bard · May 1, 2017
K130968 — VENTRALIGHT ST MESH WITH ECHO PS POSITIONING SYSTEM · C.R. Bard, Inc. · May 1, 2013
K193144 — T-Line Hernia Mesh · Deep Blue Medical Advances, Inc. · Mar 30, 2020
K243315 — ProGrip Self-Gripping Polypropylene Mesh · Sofradim Production · Jan 17, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 12, 2019
C.R. Bard, Inc. Shirin Mate Regulatory Affairs Specialist 100 Crossings Boulevard Warwick, Rhode Island 02886
Re: K190185
Trade/Device Name: Phasix ST Mesh with Open Positioning System Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: OWT, OOD, FTL Dated: May 13, 2019 Received: May 14, 2019
Dear Shirin Mate:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For David Krause, Ph.D. Acting Division Director Division of Infection Control and Plastic Surgery Devices Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190185
Device Name
Phasix ST Mesh with Open Positioning System
Indications for Use (Describe)
Phasix ST Mesh with Open Positioning System is indicated for use in the reinforcement of soft tissue, where weakness exists, in procedures involving soft tissue repair, such as for the repair of hernias.
The open positioning system is intended to facilitate the placement, positioning and fixation of the mesh during open ventral hernia repair.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg class="bi bi-check-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> <path d="M10.97 4.97a.75.75 0 0 1 1.071 1.05l-3.992 4.99a.75.75 0 0 1-1.08.02L4.324 8.384a.75.75 0 1 1 1.06-1.06l2.094 2.093 3.473-4.425a.75.75 0 0 1 1.08-.022z" fill-rule="evenodd"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Phasix™ ST Mesh with Open Positioning System 510(k) Summary 21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (D)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
#### 1.Submitter Information:
Davol Inc., C. R. Bard, Inc. 100 Crossings Boulevard Warwick, RI 02886 Phone: (401) 825-8569 Fax: (401) 825-8765
Contact Person: Shirin Mate Title: Regulatory Affairs Specialist Email: shirin.s.mate@crbard.com
Date: June 6, 2019
#### 2.Subject Device:
| 510(k) Type: | Traditional 510(k) |
|-----------------------------|---------------------------------------------------------------------|
| Trade Name: | Phasix™ ST Mesh with Open Positioning System |
| Common or Usual Name: | Surgical Mesh |
| Classification Name: | Mesh, Surgical, Absorbable, Abdominal Hernia |
| Classification Panel: | General and Plastic Surgery |
| Regulatory Class: | Class II (Phasix™ ST Mesh with Open Positioning System) |
| | Class I ( Open Positioning System Accessory) |
| Regulation Number: | 21 CFR § 878.3300 (Phasix™ ST Mesh with Open Positioning<br>System) |
| | 21 CFR § 878.4800 (Open Positioning System Accessory) |
| Product Code: | OWT, OOD, FTL (Phasix™ ST Mesh with Open Positioning<br>System) |
| | MDM (Open Positioning System Accessory) |
| 3.Primary Predicate Device: | |
Name of Device:
K101920 - Ventrio™ ST Hernia Patch
{4}------------------------------------------------
| Common or Usual Name: | Surgical Mesh |
|-----------------------|---------------------------|
| Classification Name: | Mesh, Surgical, Polymeric |
| Classification Panel: | General & Plastic Surgery |
| Regulatory Class: | Class II |
| Regulation Number: | 21 CFR § 878.3300 |
| Product Code: | FTL |
### 4.Reference Device:
| Name of Device: | K173143 - Phasix™ ST Mesh |
|--------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| Common or Usual Name: | Surgical Mesh |
| Classification Name:<br>Classification Panel:<br>Regulatory Class: | Mesh, Surgical, Absorbable, Abdominal Herni<br>General and Plastic Surgery<br>Class II |
| Regulation Number: | 21 CFR § 878.3300 |
| Product Code: | OWT, OOD, FTL, OWZ, OXC |
#### 5.Reference Device:
| Name of Device: | K161424 - Phasix™ Mesh |
|-----------------------|-----------------------------|
| Common or Usual Name: | Surgical Mesh |
| Classification Name: | Surgical Film |
| Classification Panel: | General and Plastic Surgery |
| Regulatory Class: | Class II |
| Regulation Number: | 21 CFR § 878.3300 |
| Product Code: | OOD |
#### 6.Device Description:
Phasix™ ST Mesh with Open Positioning System is a sterile, single-use device for prescription use only. It is a bi-layer mesh comprised of Phasix™ ST Mesh (K173143, forms posterior layer) and Phasix™ Mesh (K161424, forms anterior layer) stitched together with a 10 mil P4HB monofilament. The combination of the two distinct layers forms a pocket to accommodate a preinserted removable accessory. The subject device is designed for reinforcement of soft tissue deficiencies during open ventral hernia repair. The subject device and reference device have identical intended use for the mesh i.e. soft tissue repair/ reinforcement.
# Open Positioning System
{5}------------------------------------------------
The removable open positioning system is an accessory with polypropylene (PP) handle attached to a Polytetrafluoroethylene (PTFE) guide. The accessory comes preinserted into the mesh pocket to aid with placement, positioning, and fixation. The center marking on the positioning guide will aid with proper centering and orientation over the defect. The accessory is removed following the initial fixation and then discarded. The intended use of the accessory is similar to the SorbaFlex Memory Technology utilized in the Ventrio™ ST Hernia Patch (K101920).
# 7.Indications for Use:
Phasix™ ST Mesh with Open Positioning System is indicated for use in the reinforcement of soft tissue, where weakness exists, in procedures involving soft tissue repair, such as for the repair of hernias.
The open positioning system is intended to facilitate the placement, positioning and fixation of the mesh during open ventral hernia repair.
# 8.Technological Comparison to Predicate and Reference Devices:
The subject device, Phasix™ ST Mesh with Open Positioning System has the following similarities to the predicate and reference devices:
- The intended use of the mesh, i.e. for the repair / reinforcement of soft tissue, is identical . to the Ventrio™ ST Hernia Patch (K101920), Phasix™ ST Mesh (K173143) and Phasix™ Mesh (K161424).
- The intended use of the open positioning accessory, i.e. to facilitate the ease of insertion, . placement, positioning and fixation of the mesh, is similar to the SorbaFlex Memory Technology (PDO) utilized in the Ventrio™ ST Hernia Patch (K101920 with the exception of the removal of the open positioning accessory).
- The mesh and mesh materials are identical to the Phasix™ ST Mesh (K173143) and ● Phasix™ Mesh (K161424).
- The mesh pocket design is similar to the Ventrio™ ST Hernia Patch (K101920). .
- The principle of operation for open ventral hernia repair is identical to the Ventrio™ ST . Hernia Patch (K101920), Phasix™ ST Mesh (K173143) and Phasix™ Mesh (K161424)
{6}------------------------------------------------
with the exception of the open positioning accessory removal, i.e. insertion, placement/ positioning and fixation.
- The mesh sizes are within the similar range of the Ventrio™ ST Hernia Patch (K101920), . Phasix™ ST Mesh (K173143) and Phasix™ Mesh (K161424).
- The base materials utilized in open positioning accessory, Polypropylene and PTFE are . similar to the Polypropylene mesh material and PTFE monofilament sew line utilized in the Ventrio™ ST Hernia Patch (K101920).
The subject device, the Phasix™ ST Mesh with Open Positioning System incorporates following change as compared to the predicate and reference devices:
- . It consists of a removable accessory instead of the embedded SorbaFlex Memory Technology (PDO) material as compared to Ventrio™ ST Hernia Patch (K101920). None of the cited predicate/reference devices consists of a removable accessory.
The subject device has similar intended use, indication for use, principle of operation, mesh materials/ manufacturing processes, packaging, sterilization (ethylene oxide) method and utilizes the same manufacturing facility. Any differences in the technological characteristics were thoroughly tested and the results demonstrate that there are no new questions of safety and effectiveness.
# 9.Performance Data:
The following performance data is provided in support of substantial equivalence determination.
#### Performance Standards
No performance standards have been established for this device under Section 514 of the Federal Food, Drug, and Cosmetic Act.
# Biocompatibility Testing:
The biocompatibility evaluation for the Phasix™ ST Mesh with Open Positioning System was conducted in accordance with the Guidance for Industry and Food and Drug Administration Staff Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" June 16, 2016, and International
{7}------------------------------------------------
Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process," as recognized by the FDA.
The subject device mesh materials are identical to the mesh materials utilized in the successful clearance of the Phasix™ ST Mesh (K173143) and Phasix™ Mesh (K161424) and therefore considered safe and biocompatible. Thus, the tests specific to Phasix™ ST Mesh and Phasix™ Mesh were not repeated. However, in order to address any risks associated with interaction between the materials of the subject device mesh and the accessory (guide and handle), a biocompatibility testing was completed to demonstrate the subject device; Phasix™ ST Mesh with Open Positioning System is biocompatible for its intended use. Refer to Table 1.
Table 1: Biocompatibility Testing: Phasix™ ST Mesh with Open Positioning System
| Test Method | Results |
|----------------------------------------------|---------|
| MEM Cell Cytotoxicity Elution | Pass |
| Intracutaneous Reactivity | Pass |
| Acute Systemic Toxicity | Pass |
| ISO Material Mediated Rabbit Pyrogenicity | Pass |
| Guinea Pig Maximization Test (Sensitization) | Pass |
| Material/ Chemical Characterization | Pass |
The biocompatibility test results demonstrate that the subject device is biocompatible and there are no interactions between the subject device mesh and the accessory to affect its safety and effectiveness. Thus the subject device is safe and biocompatible for its intended use.
## Electrical safety and electromagnetic compatibility (EMC)
The Phasix™ ST Mesh with Open Positioning System is not an electro-mechanical medical device nor is it medical system; therefore this section does not apply.
{8}------------------------------------------------
## Software Verification and Validation Testing
The Phasix™ ST Mesh with Open Positioning System does not contain software therefore this section does not apply.
# Product Testing
The performance test results demonstrate that subject device successfully met the established acceptance criteria and is substantially equivalent to the Ventrio™ ST Hernia Patch (K101920), Phasix™ ST Mesh (K173143) and Phasix™ Mesh (K161424). Performance testing completed on the subject device is listed in Table 2.
Table 2: Performance Testing - bench
| Performance Testing- Bench | Test Method |
|----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial equivalency testing | Mesh Pore Size (Major and Minor), Gel Disruption, Ball Burst,<br>Tear Strength (Cross and Machine direction), Suture Retention<br>Strength (Cross and Machine Direction). |
| Additional mechanical testing to<br>evaluate features unique to the<br>subject device. | 15 Degree Recoil, Grip Strength, Pocket Integrity, Open<br>Positioning System (accessory) removal force, Handle<br>Attachment Strength, Three Tack Pluck. |
| Design Validation Usability Test. | Attributes: IFU, Insertion, Positioning and Placement, Fixation,<br>Removal, Open Ventral Hernia Repair. |
## Animal studies:
A comprehensive 4 week GLP study in a porcine model was performed on the subject device as compared to Ventrio™ ST Hernia Patch (K101920) and Phasix™ ST Mesh (K173143) to evaluate the following:
- . Mesh conformance
- . Peritoneal tissue attachments (% area coverage and tenacity)
- Percentage of mesh contracture
- Histological evaluation of the host inflammatory/fibrotic response, vascular integration, Collagen deposition/remodeling and tissue ingrowth
- Mechanical tissue ingrowth properties via T-peel analysis .
{9}------------------------------------------------
The subject device results were found to be comparable and therefore. Phasix™ ST Mesh with Open Positioning System is substantially equivalent to the Ventrio™ ST Hernia Patch (K101920) and Phasix™ ST Mesh (K173143). .
# Clinical Study:
Clinical Study was not required in support of the subject device, Phasix™ ST Mesh with Open Positioning System.
## Conclusion:
The subject device has similar intended use, indications for use, principle of operation, mesh technological characteristics, packaging materials/ components, manufacturing process and sterilization (ethylene oxide). Any differences in the technological characteristics between the subject device and the predicate/reference devices were assessed and evaluated. All test results provided in support of the subject device demonstrate safety and effectiveness similar to the predicate/reference devices. Therefore, Phasix™ ST Mesh with Open Positioning System is substantially equivalent to the predicate device, Ventrio™ ST Hernia Patch (K101920).
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.