K202425 · Tepha, Inc. · OOD · Aug 19, 2021 · General, Plastic Surgery
Device Facts
Record ID
K202425
Device Name
SurgiLattice scaffold
Applicant
Tepha, Inc.
Product Code
OOD · General, Plastic Surgery
Decision Date
Aug 19, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Intended Use
SurgiLattice™ scaffold is indicated for use as a bioabsorbable scaffold for soft tissue support and to repair, elevate and reinforce deficiencies where weakness or voids exist that require the addition of material to obtain the desired surgical outcome. This includes reinforcement of soft tissue in plastic and reconstructive surgery, and general soft tissue reconstruction. SurgiLattice scaffold is also indicated for the repair of hernia and other fascial defects that require the addition of a reinforcing material to obtain the desired surgical result.
Device Story
SurgiLattice scaffold is a bioabsorbable surgical mesh made of poly-butylene succinate (PBS) monofilament fibers in a macroporous, warp-knit construction. Used by surgeons in clinical settings for soft tissue reinforcement and hernia repair, the device provides mechanical support to tissues while healing occurs. The scaffold degrades via hydrolysis into 1,4-butanediol and succinic acid, which are metabolized through the Krebs Cycle and eliminated. Its porous design facilitates native tissue ingrowth and gradual load transfer from the scaffold to the healing tissue. The device is intended to provide support throughout the healing period, with strength retention profiles comparable to existing absorbable meshes. By reinforcing fascial defects or soft tissue voids, the scaffold helps achieve desired surgical outcomes in reconstructive and general surgery.
Clinical Evidence
No human clinical data. Evidence includes bench testing (physical/mechanical properties, shelf life), biocompatibility per ISO 10993, and animal studies. Rabbit model (subcutaneous) compared SurgiLattice to GalaFLEX at 4, 8, 12, and 26 weeks, showing comparable strength retention and degradation. A long-term rabbit study (simulating 116 weeks) showed minimal tissue reaction. A functional porcine hernia repair model confirmed comparable repair site mechanics (burst strength, stiffness) and histology.
Technological Characteristics
Material: Poly-butylene succinate (PBS). Construction: Macroporous, monofilament warp-knit. Principle: Bioabsorbable scaffold via hydrolytic degradation. Properties: Comparable to predicate in thickness, density, pore diameter, burst strength, bending stiffness, tensile strength, suture pull-out, and tear strength. Sterilization: Not specified.
Indications for Use
Indicated for soft tissue support, repair, elevation, and reinforcement of deficiencies, weaknesses, or voids in plastic, reconstructive, and general soft tissue surgery, including hernia and fascial defects requiring reinforcing material.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
K161092 — GalaSHAPE 3D · Tepha, Inc. · Aug 23, 2016
K173143 — Phasix ST Mesh · C.R. Bard, Inc. · Apr 25, 2018
K113723 — TEPHAFLEX MESH · Tepha, Inc. · Feb 15, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 19, 2021
Tepha, Inc. Connie Garrison SVP, Regulatory, Clinical and Quality Systems 99 Hayden Avenue, Suite 360 Lexington, MA 02421
Re: K202425
Trade/Device Name: SurgiLattice scaffold Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTL Dated: July 15, 2021 Received: July 20, 2021
Dear Ms. Garrison:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Deborah Fellhauer Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K202425
Device Name SurgiLattice™ scaffold
#### Indications for Use (Describe)
SurgiLattice™ scaffold is indicated for use as a bioabsorbable scaffold for soft tissue support and to repair, elevate and reinforce deficiencies where weakness or voids exist that require the addition of material to obtain the desired surgical outcome. This includes reinforcement of soft tissue in plastic and reconstructive surgery, and general soft tissue reconstruction. SurgiLattice scaffold is also indicated for the repair of hernia and other fascial defects that require the addition of a reinforcing material to obtain the desired surgical result.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
In accordance with 21 CFR 807.92 of the Federal Code of Regulations, the following information is a summary of safety and effectiveness of SurgiLattice scaffold.
#### SUBMITTED BY:
| Company Name: | Tepha, Inc. | |
|----------------------------|---------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 99 Hayden Avenue Suite 360<br>Lexington, MA 02421 | |
| Telephone:<br>Fax: | 781-357-1777<br>781-357-1701 | |
| CONTACT PERSON: | | Connie Garrison |
| DATE PREPARED: | | 13 August 2021 |
| TRADE NAME: | | SurgiLattice™ scaffold |
| COMMON NAME: | | Surgical mesh |
| CLASSIFICATION NAME/PANEL: | | Mesh, surgical, polymeric<br>CFR §878.3300 / General and Plastic Surgery |
| PROCODE: | | FTL |
| PREDICATE DEVICE | | K140533 |
| TRADE NAME: | | GalaFLEX® scaffold |
| COMMON NAME: | | Surgical mesh |
| CLASSIFICATION NAME/PANEL: | | Absorbable Poly(hydroxybutyrate) Surgical Mesh<br>Produced by Recombinant DNA Technology<br>CFR §878.3300 / General and Plastic Surgery |
| PROCODE: | | OOD |
#### DEVICE DESCRIPTION:
SurgilLattice scaffold is a bioabsorbable surgical mesh manufactured from poly-butylene succinate (PBS). PBS is an absorbable polymer that is processed into monofilament fibers and knitted into a surgical scaffold. PBS degrades through the process of hydrolysis, is absorbed over time, and is ultimately eliminated as CO2 and H2O. It has been developed to optimize absorption rate and prolong strength retention in order to provide support throughout the expected period of healing. Although the scaffold loses strength with time, its porous construction was designed to allow native tissue ingrowth and gradual transfer of load from the scaffold to the tissue.
Pre-clinical implantation studies indicate that SurgiLattice scaffold retains approximately 88% of its strength at 12 weeks, with minimal residual strength remaining at 78 weeks based on in vitro degradation studies.
{4}------------------------------------------------
## INDICATIONS FOR USE/INTENDED USE:
SurqiLattice™ scaffold is indicated for use as a bioabsorbable scaffold for soft tissue support and to repair. elevate, and reinforce deficiencies where weakness or voids exist that require the addition of material to obtain the desired surgical outcome. This includes reinforcement of soft tissue in plastic and reconstructive surgery, and general soft tissue reconstruction. SurgiLattice scaffold is also indicated for the repair of hernia and other fascial defects that require the addition of a reinforcing material to obtain the desired surgical result.
## COMPARISON of TECHNOLOGICAL CHARACTERISTICS:
SurgiLattice scaffold has an indications statement that is the same as the predicate, with the exception of bridging repair which is not included for the subject device. The technological characteristics and principles of operation are the same. The minor technological difference related to the biomaterial composition does not raise new issues of safety or effectiveness. The data demonstrate that SurgiLattice scaffold is substantially equivalent to the predicate device.
Both the subject and predicate devices are absorbable surgical mesh devices that slowly hydrolyze resulting in a gradual loss of strength and mass for the bulk material. In the body, the hydrolytic degradation products of PBS, 1,4-butanediol and succinic acid, are further metabolized through 4-hydroxybutyric acid (4HB) and succinic acid and eliminated from the body primarily by metabolism via the Krebs Cycle. In a similar degradation pathway, the predicate device, composed of poly-4-hydroxybutyrate (P4HB), hydrolyzes to 4-hydroxybutryate (4HB), which is also further metabolized into succinic acid and likewise is eliminated from the body via the Krebs Cycle.
The physical and mechanical properties of the subject and predicate device are also comparable. Results from performance testing based on "Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh", dated March 2, 1999, demonstrate comparable mesh thickness, density, pore diameter, mesh knit characteristics, burst strength, bending stiffness, tensile strength, suture pull-out, and tear strength. Both products have the same macroporous, monofilament warp knit construction.
#### PERFORMANCE DATA:
The performance characteristics of SurgiLattice scaffold were established via comprehensive studies of physical and mechanical properties, biocompatibility testing per ISO 10993 including a Toxicological Risk Assessment based on extractables and leachables data per ISO 10993-18, shelf life testing, lifecycle evaluation in GLP rabbit studies per ISO 10993-6 including implantation of pre-degraded biomaterial representative of longer term in-vivo data, in vitro degradation studies, and functional performance in a large animal model study.
Bench testing of physical and mechanical characteristics, and shelf life studies using real-time and accelerated aging were performed with passing results. Biocompatibility testing of SurgiLattice scaffold was conducted per the categorization principles in ISO 10993-1:2009. Based on the standard, the device was categorized as an implant device in contact with tissue and bone
{5}------------------------------------------------
and having a duration of contact of greater than 30 days (permanent). Results from the panel of testing, specifically cytotoxicity, irritation or intracutaneous reactivity, maximization, acute systemic toxicity, material-mediated pyrogenicity, (bacterial reverse mutation), genotoxicity (mouse lymphoma assay), hemolysis, subacute/subchronic/chronic local toxicity, subcutaneous implantation with histology (4. 12 and 26 weeks), subcutaneous implantation with histology using partially-degraded mesh(representing 116 weeks), coupled with a toxicological assessment conducted per ISO 10993-18, support the biocompatibility and safety of the subject device.
A study was conducted to evaluate dimensional, morphologic, and histological properties of SurgiLattice at 4, 8, 12 and 26 weeks following subcutaneous dorsal implantation in a rabbit model, as compared to GalaFLEX Scaffold. The initial strength, the strength retention profile, and the degradation profile of the SurgiLattice™ scaffold were comparable to the predicate. The results from this study support the in vivo comparability of SurgiLattice™ scaffold with the predicate, GalaFLEX scaffold.
A second study was conducted to evaluate the local tissue reaction using SurgiLattice scaffold which had undergone accelerated hydrolytic degradation followed by 26 weeks implantation which simulated a final tissue assessment timepoint of approximately 116 weeks. This study was designed to assess the host response upon essentially complete mechanical and molecular weight polymer degradation. Microscopically, SurgiLattice scaffold caused minimal or no reaction and was comparable to the predicate device.
SurgiLattice scaffold was further evaluated in a functional porcine model of hernia repair through assessment of repair site mechanics (burst strength and stiffness), morphologic properties (fiber diameter/surface roughness), molecular weight, and histology. The results support substantial equivalence of SurqiLattice scaffold to the predicate.
#### CONCLUSION:
Based on the indications for use, technological characteristics, and the summary of data submitted, Tepha, Inc. has determined that the proposed subject device, SurgiLattice scaffold is substantially equivalent to the currently marketed predicate device, GalaFLEX scaffold. Performance testing, including in vivo data and a comprehensive assessment of biocompatibility, demonstrated that the device functions as intended without raising new questions of safety or effectiveness.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.