← Product Code [OOD](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/OOD) · K091633

# TEPHAFLEX SURGICAL FILM (K091633)

_Tepha, Inc. · OOD · Aug 7, 2009 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/OOD/K091633

## Device Facts

- **Applicant:** Tepha, Inc.
- **Product Code:** [OOD](/submissions/SU/subpart-d%E2%80%94prosthetic-devices/OOD.md)
- **Decision Date:** Aug 7, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

TephaFLEX® Surgical Film is intended to be used wherever temporary wound support is required, to reinforce soft tissues where weakness exists in the urological, or gastrointestinal anatomy, or for the repair of hernia or other fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result. The absorbable protective film also may help minimize the potential for tissue attachment to the device in case of direct contact with the viscera. NOTE: The TephaFLEX Surgical Film is not intended to be placed transvaginally. The film should be placed abdominally when being used to reinforce soft tissues in the urological, gynecological or gastrointestinal anatomy.

## Device Story

TephaFLEX Surgical Film is an absorbable, protective film used for temporary wound support and soft tissue reinforcement in urological, gynecological, or gastrointestinal surgeries; also used for hernia or fascial defect repair. Device acts as a physical barrier to minimize tissue attachment to viscera. Used in clinical settings by surgeons; placed abdominally. Benefits include temporary structural support and reduced adhesion potential. Device is absorbable.

## Clinical Evidence

Bench testing only. Performance characterized via comparative burst strength, suture pull-out strength, tensile strength, and tear resistance. In vivo strength retention evaluated via subcutaneous implantation study. Biocompatibility testing performed.

## Technological Characteristics

Absorbable surgical film. Mechanical properties characterized by burst, suture pull-out, tensile, and tear resistance strength. Biocompatible material. No software or electronic components.

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- TephaFLEX Surgical Film ([K091633](/device/K091633.md))
- TephaFLEX Absorbable Mesh ([K091633](/device/K091633.md))
- MAST Biosurgery, Inc. - Surgi-Wrap Film

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K0916633

RAGE 1/1

99 Hayden Avenue Suite 360 Lexington, MA 02421 tel: (781) 357-1700 fax: (781) 357-1701

Image /page/0/Picture/2 description: The image shows the logo for TEPHA Medical Devices. The logo features a stylized graphic element above the text. The text "MEDICAL DEVICES" is printed in a smaller font size below the word TEPHA.

## Section X Summary of Safety and Effectiveness (Prepared June 22, 2012)

Pursuant to §513(i)(3)(A) of the Food, Drug, and Cosmetic Act, Tepha, Inc. is submitting the following summary of information respecting safety and effectiveness:

| Trade Name:                 | TephaFLEX® Surgical Film                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor:                    | Tepha, Inc.<br>99 Hayden Avenue, Suite 360<br>Lexington, MA 02421<br>Telephone: 781.357.1700<br>Fax: 781.357.1701                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Contact Person:             | Mary P. LeGraw, V.P., Regulatory Affairs                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Device Classification Name: | CFR \$878.3300 - OOD, PAJ<br>Surgical Mesh                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Classification:             | According to Section 13 of the Federal Food. Drug and Cosmetic Act, the device<br>classification is Class II, Performance Standards.                                                                                                                                                                                                                                                                                                                                                                                                          |
| Predicate Devices:          | Tepha, Inc. - TephaFLEX Surgical Film<br>Tepha, Inc. - TephaFLEX Absorbable Mesh<br>MAST Biosurgery, Inc. - Surgi-Wrap Film                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Device Description:         | TephaFLEX surgical film is intended to be used wherever temporary wound<br>support is required, to reinforce soft tissues where weakness exists in the<br>urological, gynecological, or gastrointestinal anatomy, or for the repair of hernia<br>or other fascial defects that require the addition of a reinforcing or bridging<br>material to obtain the desired surgical result. The absorbable protective film also<br>may help minimize the potential for tissue attachment to the device in case of<br>direct contact with the viscera. |
|                             | Note: The TephaFLEX surgical mesh is not intended to be placed transvaginally<br>The film should be placed abdominally when being used to reinforce soft tissues<br>in the urological, gynecological or gastrointestinal anatomy                                                                                                                                                                                                                                                                                                              |
| Safety and Performance:     | Mechanical testing, in vivo animal testing, and biocompatibility testing, were<br>performed based on recommendations identified in the FDA surgical mesh<br>guidance document: The Guidance for the Preparation of a Pre-market<br>Notification Application for a Surgical Mesh. Specifically, comparative burst<br>strength, suture pull-out strength, tensile strength and tear resistance strength<br>was characterized. In vivo strength retention was characterized via a<br>subcutaneous implantation study.                            |
| Conclusion:                 | Based on the indications for use, technological characteristics, and safety and<br>performance testing, the TephaFLEX surgical film has been shown to be<br>substantially equivalent to predicate devices under the Federal Food, Drug and<br>Cosmetic Act.                                                                                                                                                                                                                                                                                   |

{1}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue

Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized emblem. The emblem consists of a blue abstract design that resembles an eagle or a human figure, with flowing lines suggesting movement or support. 

Tepha, Incorporated % Ms. Mary P. LeGrew Vice President, Regulatory Affairs 99 Hayden Avenue, Suite 300 Lexington, Massachusetts 02421

AUG 2 4 2012

Re: K091633

Trade/Device Name: TephaFLEX® Surgical Film Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: OOD, PAJ Dated: June 3, 2009 Received: June 4, 2009

4 2019

Dear Ms. LeGrew:

This letter corrects our substantially equivalent letter of August 7, 2009.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2-Mary P. LeGrew

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm1

Sincerely yours,

cerely yours,

for

k N. Melkerson

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known): K091633

TephaFLEX® Surgical Film Device Name:

Indications for Use:

TephaFLEX® Surgical Film is intended to be used wherever temporary wound support is required, to reinforce soft tissues where weakness exists in the urological, or gastrointestinal anatomy, or for the repair of hernia or other fascial defects that require the addition of a reinforcing or bridging material to obtain the desired surgical result. The absorbable protective film also may help minimize the potential for tissue attachment to the device in case of direct contact with the viscera.

NOTE: The TephaFLEX Surgical Film is not intended to be placed transvaginally. The film should be placed abdominally when being used to reinforce soft tissues in the urological, gynecological or gastrointestinal anatomy.

Prescription Use: X (21 CFR 801 Subpart D)

AND/OR

Over-The-Counter (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Daniel Kimefumm

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K091633

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/OOD/K091633](https://fda.innolitics.com/submissions/SU/subpart-d%E2%80%94prosthetic-devices/OOD/K091633)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
