RejuvaKnee™ Collagen Meniscus Implant

K242302 · Collagen Matrix, Inc. · OLC · Oct 2, 2024 · General, Plastic Surgery

Device Facts

Record IDK242302
Device NameRejuvaKnee™ Collagen Meniscus Implant
ApplicantCollagen Matrix, Inc.
Product CodeOLC · General, Plastic Surgery
Decision DateOct 2, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

RejuvaKnee is intended for use in surgical procedures for the reinforcement and repair of soft tissue injuries of the medial meniscus. In repairing and reinforcing medial meniscal defects, the patient must have an intact meniscal rim and anterior and posterior horns for attachment of the mesh. In addition, the surgically prepared site for the RejuvaKnee must extend at least into the red/white zone of the meniscus to provide sufficient vascularization. RejuvaKnee reinforces soft tissue and provides a resorbable scaffold that is replaced by the patient's own soft tissue. The RejuvaKnee is not a prosthetic device and is not intended to replace normal body structure.

Device Story

RejuvaKnee is a resorbable scaffold composed of bovine Type I collagen; used in orthopedic surgical procedures to reinforce and repair medial meniscus defects. Device is semi-lunar with triangular cross-section; implanted by surgeons to provide a scaffold for host tissue ingrowth; eventually replaced by patient's own soft tissue. Not a prosthetic; does not replace normal body structure. Requires intact meniscal rim and horns for attachment; site must reach red/white zone for vascularization. Benefits include medial condyle support and prevention of secondary joint changes post-partial meniscectomy.

Clinical Evidence

No human clinical trials. Evidence includes ISO 10993 biocompatibility testing, bench characterization (dimensions, porosity, pH, tensile strength, suturability, pyrogenicity, thermal stability), and animal studies (goat and canine models). Animal models confirmed safety, biocompatibility, and efficacy in meniscus repair/remodeling, providing biomechanical support equivalent to the predicate device.

Technological Characteristics

Material: 99% bovine Type I collagen. Form: Semi-lunar, triangular cross-section. Sterilization: Gamma irradiation (SAL 10^-6). Biocompatibility: ISO 10993 compliant. Mechanical properties: Tensile strength ≥ 150 kg/cm², suturability ≥ 0.5 kg, thermal stability 60±3°C.

Indications for Use

Indicated for reinforcement and repair of medial meniscus soft tissue injuries in patients with intact meniscal rim and anterior/posterior horns. Requires surgically prepared site extending into the red/white zone for vascularization.

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 2, 2024 Collagen Matrix, Inc. Devin Dragon Sr. Regulatory Specialist 15 Thornton Road Oakland, New Jersey 07436 Re: K242302 Trade/Device Name: RejuvaKnee™ Collagen Meniscus Implant Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: OLC Dated: August 2, 2024 Received: August 5, 2024 Dear Devin Dragon: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. {2}------------------------------------------------ evices and radiation-emitting products, including For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Thomas Mcnamara -S For: Christopher Ferreira, MS Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K242302 #### Device Name RejuvaKnee™ Collagen Meniscus Implant #### Indications for Use (Describe) RejuvaKnee is intended for use in surgical procedures for the reinforcement and repair of the medial meniscus. In repairing and reinforcing medial meniscal defects, the patient must have an intact meniscal rim and anterior and posterior horns for attachment of the mesh. In addition, the surgically prepared site for the RejuvaKnee must extend at least into the red/white zone of the meniscus to provide sufficient vascularization. RejuvaKnee reinforces soft tissue and provides a reasonable scaffold that is replaced by the patient's own soft tissue. The RejuvaKnee is not a prosthetic device and is not intended to replace normal body structure. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ #### 510(k) SUMMARY | 1. | Applicant Information | | |----|----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------| | | Applicant Name:<br>Address: | Collagen Matrix, Inc.<br>15 Thornton Road<br>Oakland, New Jersey 07436 | | | Telephone:<br>Fax: | (201) 405-1477<br>(201) 405-1355 | | | Contact Person: | Devin Dragon<br>Sr. Regulatory Specialist<br>ddragon@regenity.com | | | Date Prepared: | September 30, 2024 | | 2. | Name of the Device | | | | Device Trade Name: | RejuvaKnee™ Collagen Meniscus Implant<br>K242302 | | | Device Common Name(s): | Scaffold, Partial Medial Meniscal Defects Extending into<br>the Red/White Zone, Resorbable Bovine Collagen | | | Device Classification Name: | Surgical Mesh<br>878.3300<br>OLC<br>Class II | | 3. | Legally Marketed Devices to Which Substantial Equivalence is Claimed | | | | Predicate Device: | Collagen Meniscus Implant<br>Stryker Endoscopy<br>(originally Ivy Sports Medicine, LLC)<br>K170364 | # 4. Description of the Device The RejuvaKnee™ Collagen Meniscus Implant is comprised primarily of bovine Type I collagen (nominally 99%) derived from meniscus. The device is provided in a semi-lunar shape with a triangular cross section to be used to reinforce weakened soft tissue and provide a resorbable scaffold that is replaced by the patient's own tissue. RejuvaKnee is supplied sterile and is intended for single use. {5}------------------------------------------------ ## 5. Intended Use RejuvaKnee is intended for use in surgical procedures for the reinforcement and repair of soft tissue injuries of the medial meniscus. In repairing and reinforcing medial meniscal defects, the patient must have an intact meniscal rim and anterior and posterior horns for attachment of the mesh. In addition, the surgically prepared site for the RejuvaKnee must extend at least into the red/white zone of the meniscus to provide sufficient vascularization. RejuvaKnee reinforces soft tissue and provides a resorbable scaffold that is replaced by the patient's own soft tissue. The RejuvaKnee is not a prosthetic device and is not intended to replace normal body structure. #### Summary/Comparison of Technical Characteristics 6. The subject device and predicate device have indications for use. The subject device has substantially equivalent technological characteristics as the predicate device. Differences in tissue source are minimized after processing and purification. | Feature | RejuvaKnee™ Collagen Meniscus Implant | Ivy Sports Medicine Collagen Meniscus Implant | |------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | (This submission) | From Stryker Endoscopy | | Indications for Use | RejuvaKnee is intended for use in surgical<br>procedures for the reinforcement and repair of<br>soft tissue injuries of the medial meniscus. In<br>repairing and reinforcing medial meniscal<br>defects, the patient must have an intact meniscal<br>rim and anterior and posterior horns for<br>attachment of the mesh. In addition, the<br>surgically prepared site for RejuvaKnee must<br>extend at least into the red/white zone of the<br>meniscus to provide sufficient vascularization.<br>RejuvaKnee reinforces soft tissue and provides a<br>resorbable scaffold that is replaced by the<br>patient's own soft tissue. The RejuvaKnee is not a<br>prosthetic device and is not intended to replace<br>normal body structure. | The Collagen Meniscus Implant (CMI) is intended<br>for use in surgical procedures for the<br>reinforcement and repair of soft tissue injuries<br>of the medial meniscus. In repairing and<br>reinforcing medial meniscal defects, the patient<br>must have an intact meniscal rim and anterior<br>and posterior horns for attachment of the mesh.<br>In addition, the surgically prepared site for the<br>CMI must extend at least into the red/white<br>zone of the meniscus to provide sufficient<br>vascularization.<br>The CMI reinforces soft tissue and provides a<br>resorbable scaffold that is replaced by the<br>patient's own soft tissue. The CMI is not a<br>prosthetic device and is not intended to replace<br>normal body structure. | | Chemical and Physical Properties | | | | Material<br>Composition | Primarily Type I bovine collagen | Type I bovine collagen | | Material Origin | Bovine | Bovine | | Tissue Source | Meniscus | Deep flexor Achilles tendon | | Physical Form | Semi-lunar shape with a triangular cross-section | Semi-lunar shape with a triangular cross-section | | Product Sizes (cm) | Small Medial - 7.5 mm<br>Large Medial - 9 mm | Small Medial – 7.5 mm<br>Large Medial – 9 mm<br>XL Medial | | Porosity - SEM of<br>the outer and<br>cross-section (not<br>regular testing) | Provides sufficient porosity to allow for cellular<br>integration from host tissue<br>>5 micron | Provides sufficient porosity to allow for cellular<br>integration from host tissue | | pH | 6 - 9.5 | 7.12 | | Tensile Strength<br>(kg/cm²) | ≥ 150 | 62.5 | | Suturability (kg) | ≥ 0.5 | 0.86 ±0.08 | {6}------------------------------------------------ | Feature | RejuvaKnee™ Collagen Meniscus Implant<br>(This submission) | Ivy Sports Medicine Collagen Meniscus Implant<br>From Stryker Endoscopy | |---------------------------|------------------------------------------------------------|-------------------------------------------------------------------------| | Thermal Stability<br>(°C) | $60 ± 3$ | $61±1$ | | Biocompatibility | Meets ISO 10993 biocompatibility series of<br>testing | Biocompatible | | Sterility | Sterile, SAL 10-6 using Gamma irradiation | Sterile, SAL 10-6 using Gamma irradiation | | Pyrogenicity | Non-pyrogenic - <0.5EU/ml and ≤20.0EU/device | Non-pyrogenic - <0.5EU/ml and ≤20.0EU/device | | Single Use/ Reuse | Single use only | Single use only | #### 7. Performance Data In vivo and in vitro testing of the subject device was conducted to demonstrate substantial equivalence of the subject device to its predicate device. The following performance data are provided in support of the substantial equivalence determination. #### Biocompatibility Testing Biocompatibility testing was conducted in accordance with the testing matrix of ISO 10993-1. #### Bench Testing Collagen characterization testing and product characterization testing was performed to demonstrate substantial equivalence of the subject device to its predicate. Testing included dimensions, porosity, pH, tensile strength, suturability, pyrogenicity and thermal stability. Each lot of finished devices is tested for bacterial endotoxin for lot release. #### Animal Testing The performance of the device in goat and canine models confirmed that RejuvaKnee is safe, biocompatible and associated with meniscus defect tissue repair and remodeling, and provides the necessary medial condyle support and biomechanics to prevent secondary joint changes resulting from an untreated partial meniscectomy. The performance in the canine model was compared to the published performance of the predicate device and confirmed equivalent performance. #### Clinical Studies Clinical performance of the RejuvaKnee™ Collagen Meniscus Implant subject device was evaluated in a simulated clinical setting to demonstrate that the RejuvaKnee maybe successfully implanted equivalent to the predicate device. The subject device is the same intended use, design, and bovine-derived collagen safety profile as the predicate device. Based on the equivalent material and performance specifications, and equivalent principle of operation to the predicate device, clinical studies of the subject device are not needed to support substantial equivalence. #### Conclusions Drawn from Non-clinical Studies 8. The conclusions drawn from the nonclinical tests demonstrate that the device is substantially equivalent to its predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...