WallFlex Colonic Stent System with Anchor Lock Delivery System; WallFlex Duodenal Stent System with Anchor Lock Delivery System; WallFlex Colonic Soft Stent System with Anchor Lock Delivery System; WallFlex Duodenal Soft Stent System with Anchor Lock Delivery System
K234046 · Boston Scientific Corporation · MQR · Mar 8, 2024 · General, Plastic Surgery
Device Facts
Record ID
K234046
Device Name
WallFlex Colonic Stent System with Anchor Lock Delivery System; WallFlex Duodenal Stent System with Anchor Lock Delivery System; WallFlex Colonic Soft Stent System with Anchor Lock Delivery System; WallFlex Duodenal Soft Stent System with Anchor Lock Delivery System
Applicant
Boston Scientific Corporation
Product Code
MQR · General, Plastic Surgery
Decision Date
Mar 8, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.3610
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The WallFlex Colonic Stent System with Anchor Lock Delivery System is indicated for palliative treatment of colonic strictures caused by malignant neoplasms and to relieve large bowel obstructions prior to colectomy in patients with malignant strictures. The WallFlex Duodenal Stent System with Anchor Lock Delivery System is indicated for the palliative treatment of gastroduodenal obstructions produced by malignant neoplasms. The WallFlex Colonic Soft Stent System with Anchor Lock Delivery System is indicated for palliative treatment of colonic strictures caused by malignant neoplasms and to relieve large bowel obstructions prior to colectomy in patients with malignant strictures. The WallFlex Duodenal Soft System with Anchor Lock Delivery System is indicated for the palliative treatment of gastroduodenal obstructions produced by malignant neoplasms.
Device Story
Device consists of implantable Nitinol braided stent and coaxial Anchor Lock delivery system. Stent is compressed on flexible catheter; exterior tube constrains stent until deployment via retraction. Radiopaque marker bands facilitate placement and identify deployment limits. Anchor Lock holder at trailing end allows reconstrainment for repositioning. Device inserted through working channel of 9 French endoscope (min 3.2mm channel) over 0.035 in. guidewire. Used by physicians in clinical settings to treat malignant strictures/obstructions. Stent provides mechanical patency to bowel/duodenum, relieving obstruction and improving patient quality of life.
Clinical Evidence
Bench testing only. MR safety evaluation conducted per ASTM F2503-20 and FDA guidance to support updated labeling.
Technological Characteristics
Materials: Nitinol wires. Form factor: Braided cylindrical stent with flared ends; coaxial delivery catheter. Dimensions: 18mm-25mm diameters, 6cm-12cm lengths. Connectivity: None. Sterilization: Not specified. Software: None.
Indications for Use
Indicated for palliative treatment of malignant colonic strictures, relief of large bowel obstructions prior to colectomy, and palliative treatment of malignant gastroduodenal obstructions.
Regulatory Classification
Identification
An esophageal prosthesis is a rigid, flexible, or expandable tubular device made of a plastic, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the esophagus. The metal esophageal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.
Special Controls
The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”
Predicate Devices
Wallflex Colonic Stent System with Anchor Lock Delivery System (K201159)
WallFlex Colonic Soft Stent System with Anchor Lock Delivery System (K200257)
Related Devices
K200257 — WallFlex Colonic Soft Stent System with Anchor Lock Delivery System, WallFlex Duodenal Soft Stent System with Anchor Lock Delivery System · Boston Scientific Corporation · Apr 17, 2020
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 8. 2024
Boston Scientific Corporation Tijana Prodanovic Sr. Regulatory Affairs Specialist 300 Boston Scientific Way Marlborough, Massachusetts 01752
#### Re: K234046
Trade/Device Name: WallFlex Colonic Stent System with Anchor Lock Delivery System; WallFlex Duodenal Stent System with Anchor Lock Delivery System; WallFlex Colonic Soft Stent System with Anchor Lock Delivery System; WallFlex Duodenal Soft Stent System with Anchor Lock Delivery System Regulation Number: 21 CFR 878.3610 Regulation Name: Esophageal Prosthesis Regulatory Class: Class II Product Code: MQR, MUM Dated: February 9, 2024 Received: February 9, 2024
Dear Tijana Prodanovic:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Anthony Lee -S
Anthony Lee, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
Submission Number (if known)
K234046
Device Name
WallFlex Colonic Stent System with Anchor Lock Delivery System WallFlex Duodenal Stent System with Anchor Lock Delivery System
WallFlex Colonic Soft Stent System with Anchor Lock Delivery System
WallFlex Duodenal Soft Stent System with Anchor Lock Delivery System
Indications for Use (Describe)
The WallFlex Colonic Stent System with Anchor Lock Delivery System is indicated for palliative treatment of colonic strictures caused by malignant neoplasms and to relieve large bowel obstructions prior to colectomy in patients with malignant strictures.
The WallFlex Duodenal Stent System with Anchor Lock Delivery System is indicated for the palliative treatment of gastroduodenal obstructions produced by malignant neoplasms.
The WallFlex Colonic Soft Stent System with Anchor Lock Delivery System is indicated for palliative treatment of colonic strictures caused by malignant neoplasms and to relieve large bowel obstructions prior to colectomy in patients with malignant strictures.
The WallFlex Duodenal Soft System with Anchor Lock Delivery System is indicated for the palliative treatment of gastroduodenal obstructions produced by malignant neoplasms.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
K234046 - Page 1/2 510(k) Summary 510(k) #: K234046 Prepared on: 2024-02-09 Contact Details 21 CFR 807.92(a)(1) Boston Scientific Corporation Applicant Name Applicant Address 300 Boston Scientific Way Marlborough MA 01752 United States 5083820440 Applicant Contact Telephone Mrs. Tijana Prodanovic Applicant Contact Applicant Contact Email tijana.prodanovic@bsci.com Device Name 21 CFR 807.92(a)(2) WallFlex Colonic Stent System with Anchor Lock Delivery System WallFlex Duodenal Stent System with Anchor Lock Delivery System Device Trade Name WallFlex Colonic Soft Stent System with Anchor Lock Delivery System WallFlex Duodenal Soft Stent System with Anchor Lock Delivery System Expandable, metallic colonic stent Common Name Expandable, metallic duodenal stent Classification Name Esophageal Prosthesis Regulation Number 21 CFR 878.3610 Product Code(s) MQR,MUM Legally Marketed Predicate Devices 21 CFR 807.92(a)(3) Predicate Trade Name (Primary Predicate is listed first) Product Code Predicate # Wallflex Colonic Stent System with Anchor Lock Delivery System K201159 MQR, MUM K200257 WallFlex Colonic Soft Stent System with Anchor Lock Delivery S MQR, MUM
## Device Description Summary
The WallFlex Colonic Stent System with Anchor Lock Delivery System and WallFlex Duodenal Stent System with Anchor Lock Delivery System each consist of two components: the implantable metal stent and the anchor lock delivery system.
21 CFR 807.92(a)(4)
The proposed stent is manufactured of Ntinol. The stent will be offered in two diameters, 22mm with a 30mm flare. Each diameter will be available in three lengths, 6cm, 9cm, and 12cm. The WallFlex Duodenal Stent will only be offered in the 22mm/27mm stent diameter (all lengths) preloaded on the 230cm Anchor Lock Delivery System. The WallFlex Colonic Stent will be offered in the 22mm/27mm and 25mm/30mm diameters (all lengths) preloaded on the 230cm or 135cm Anchor Lock Delivery System. The Anchor Lock delivery system consist of a coaxial tubing assembly that constrains the stent on the until the stent is released by retracting the exterior tube.
The WallFlex™ Colonic Soft Stem with Anchor Lock Delivery System and WallFlex™ Duodenal Soft Stent System with Anchor Lock Delivery System each consist of two components: the implantable metal stent and the Anchor Lock Delivery System.
The stent is manufactured from Nitinol wires braided together to form a cylinder with a flared end. The wires are looped at both stent ends and welded at the non-flared end. The WallFlex Colonic Soft Stent and the WallFlex Duodenal Soft Stent will be offered in three diameters: 22 mm with a 27 mm flare, 20 mm flare, and 18 mm with a 23 mm flare. Each diameter will be available in three
{5}------------------------------------------------
lengths, 6cm, 9cm, and 12cm. The stent sizes that are shared between the WallFlex Colonic Soft Stent and the WallFlex Duodenal Soft Stent are identical in all aspects of their design; the only difference is in indication.
The stent is compressed on a flexible catheter for delivery system consists in part of coaxial tubes. The exterior tube serves to constrain the stent until retracted during deployment. There are 3 radiopaque marker bands on the diftate placement; the exterior tube marker band, the deployment limit marker band, and the post deployment marker band. The exterior tube marker band is located adjacent to the leading end the deployment limit marker band is located adjacent to the trailing end of the stent. The deployment limit marker band identifies the stent beyond which the stent cannot be reconstrained. The post deployment marker band, situated 4cm from the trailing end of the constrained stent, helps to facilitate accurate placement of the stent. The only design difference between product offerings is how the stent is loaded onto the delivery system. The colonic stent is loaded with the flare is distal to the operator, and the duodenal stent is loaded with the flare is proximal to the operator.
The Anchor Lock stent holder is located at the trailing end of the stent on the delivery system to aid in reconstrainment when repositioning the stent. The interior tube of the coaxial system contains a central lumen that accommodates a 0.035 in. / 0.89 mm guidewire. The device may be inserted through the working channel of a 9 French endoscope (minimum channel diameter 3.2mm scope). The WallFlex Colonic/Duodenal Soft Stent with Anchor Lock Delivery System will be offered in 230cm.
# Intended Use/Indications for Use
## 21 CFR 807.92(a)(5)
The WallFlex Colonic Stent System with Anchor Lock Delivery System is indicated for palliative treatment of colonic strictures caused by malignant neoplasms and to relieve large bowel obstructions prior to colectomy in patients with malignant strictures.
The WallFlex Duodenal Stent System with Anchor Lock Delivery System is indicated for the palliative treatment of gastroduodenal obstructions produced by malignant neoplasms.
The WallFlex Colonic Soft Stent System with Anchor Lock Delivery System is indicated for palliative treatment of colonic strictures caused by malignant neoplasms and to relieve large bowel obstructions prior to colectomy in patients with malignant strictures.
The WallFlex Duodenal Soft System with Anchor Lock Delivery System is indicated for the palliative treatment of gastroduodenal obstructions produced by malignant neoplasms.
# Indications for Use Comparison
The proposed WallFlex Colonic Stent System with Anchor Lock Delivery System and WallFlex Duodenal Stent System with Anchor Lock Delivery System, WallFlex Colonic Soft System with Anchor Lock Delivery System and WallFlex Duodenal Soft System with Anchor Lock Delivery System are identical to the predicate WallFlex Colonic Stent System with Anchor Lock Delivery System and WallFlex Duodenal Stent System with Anchor Lock Delivery System (K201159), WallFlex Colonic Soft Stent System with Anchor Lock Delivery System and WallFlex Duodenal Soft System with Anchor Lock Delivery System (K200257) respectively, in intended use, indications for use, classification, principles of operation, technical characteristics, performance, and materials.
# Technological Comparison
The proposed WallFlex Colonic Stent System with Anchor Lock Delivery System and WallFlex Duodenal Stent System with Anchor Lock Delivery System, WallFlex Colonic Soft Stem with Anchor Lock Delivery System and WallFlex Duodenal Soft System with Anchor Lock Delivery System are identical to the predicate WallFlex Colonic Stent System with Anchor Lock Deliyery System and WallFlex Duodenal Stent System with Anchor Lock Delivery System (K201159), WallFlex Colonic Soft Stent System with Anchor Lock Delivery System and WallFlex Duodenal Soft System with Anchor Lock Delivery System (K200257) respectively, in intended use, indications for use, classification, principles of operation, technical characteristics, performance, and materials.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
The proposed WallFlex Colonic Stent System with Anchor Lock Delivery System and WallFlex Duodenal Stent System with Anchor Lock Delivery System, WallFlex Colonic Soft Stem with Anchor Lock Delivery System and WallFlex Duodenal Soft System with Anchor Lock Delivery System are compliant with the FDA's quidance document Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment (issued on October 10, 2023). MR evaluation was conducted following ASTM F2503e2020: Standard Practice for Marking Medical Devices and Other tems for Safety in the Magnetic Resonance, and the resultant data supports the proposed updated labeling.
## 21 CFR 807.92(a)(6)
21 CFR 807.92(a)(5)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.