K201000 · Cook Biotech Incorporated · FTM · Jul 13, 2020 · General, Plastic Surgery
Device Facts
Record ID
K201000
Device Name
Biodesign Staple Line Reinforcement
Applicant
Cook Biotech Incorporated
Product Code
FTM · General, Plastic Surgery
Decision Date
Jul 13, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Biodesign Staple Line Reinforcement is intended for use as a prosthesis for the surgical repair of soft tissue deficiencies using surgical staplers. The Reinforcement may be used for buttressing and reinforcing staple lines during lung resection (e.g., wedge resection, blebectomy, bullectomy, bronchial resection, segmentectomy, pneumonectomy, pneumoreduction) and other incisions and excisions of the lung and bronchus. The Staple Line Reinforcement can be used for the reinforcement of the gastric staple line during bariatric surgical procedures, and for reinforcement of staple lines during small bowel, mesentery, colon and colorectal procedures. The Staple Line Reinforcement may be used with anastomotic staplers or with non-anastomotic staplers.
Device Story
Biodesign Staple Line Reinforcement is a multi-layer porcine Small Intestine Submucosa (SIS) strip; used as a prosthesis for soft tissue repair. Device features pre-coated, ready-to-use contact adhesive; eliminates need for rehydration or separate hydrogel. Implanted via surgical stapler; buttresses staple lines to prevent tissue tearing. Remodels into body over time; leaves no permanent graft material. Used in OR by surgeons during lung, bariatric, and colorectal procedures. Provided in alignment tray with protective liner; E-beam sterilized. Output is mechanical reinforcement of tissue; benefits patient by reducing staple line leakage and tissue trauma.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993 (genotoxicity, cytotoxicity, systemic toxicity, irritation, sensitization, implantation). Mechanical performance testing confirmed staple line leakage prevention. Packaging integrity testing supports 6-month shelf life.
Technological Characteristics
Materials: Porcine Small Intestine Submucosa (SIS). Dimensions: 0.05mm to 0.6mm thickness; 76mm to 176mm length; up to 12mm width. Energy: None (passive implant). Connectivity: None. Sterilization: E-beam. Packaging: Foil pouch with integrated humidity control.
Indications for Use
Indicated for patients undergoing surgical procedures requiring staple line reinforcement, including lung resection (wedge, blebectomy, bullectomy, bronchial, segmentectomy, pneumonectomy, pneumoreduction), bariatric gastric procedures, and small bowel, mesentery, colon, or colorectal procedures.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
K051048 — SURGISIS STAPLE LINE REINFORCEMENT II · Cook Biotech, Inc. · Sep 23, 2005
K070405 — MODIFICATION TO SURGISIS SLR STAPLE LINE REINFORCEMENT · Cook Biotech, Inc. · Oct 17, 2007
K040119 — PERI-STRIPS STAPLE LINE REINFORCEMENT-SLEEVE CONFIGURATION, PERI-STRIPS STAPLE LINE REINFORCEMENT-STRIP CONFIGURATION · Synovis Surgical Innovations · Mar 17, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 13, 2020
Cook Biotech Incorporated Chris Lotzow Regulatory Affairs Specialist 1425 Innovation Place West Lafayette, Indiana 47906
Re: K201000
Trade/Device Name: Biodesign Staple Line Reinforcement Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTM Dated: April 16, 2020 Received: April 16, 2020
Dear Chris Lotzow:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201000
Device Name Biodesign® Staple Line Reinforcement
#### Indications for Use (Describe)
The Biodesign Staple Line Reinforcement is intended for use as a prosthesis for the surgical repair of soft tissue deficiencies using surgical staplers. The Reinforcement may be used for buttressing and reinforcing staple lines during lung resection (e.g., wedge resection, blebectomy, bullectomy, bronchial resection, segmentectomy, pneumonectomy, pneumoreduction) and other incisions and excisions of the lung and bronchus. The Staple Line Reinforcement can be used for the reinforcement of the gastric staple line during bariatric surgical procedures, and for reinforcement of staple lines during small bowel, mesentery, colon and colorectal procedures. The Staple Line Reinforcement may be used with anastomotic staplers or with non-anastomotic staplers.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span> <span style="font-size: 14px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 14px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cook Biotech. The logo consists of the word "COOK" in large, white, sans-serif font on a red background. Below the word "COOK" is the word "BIOTECH" in a smaller, white, sans-serif font on a red background. The word "BIOTECH" is centered below the word "COOK".
# 510(k) Summary
| Submission Date: | 04/16/2020 |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| K201000 | |
| <b>SUBMITTER INFORMATION:</b> | |
| Company Name: | Cook Biotech Incorporated |
| Company Address: | 1425 Innovation Place<br>West Lafayette, IN 47906 |
| Contact Person: | Chris Lotzow<br>Regulatory Affairs Specialist<br>Phone: 765-497-3355<br>Fax: 765-807-7709<br>Email: chris.lotzow@cookbiotech.com |
| <b>Device Trade Name:</b> | Biodesign® Staple Line Reinforcement |
| <b>Device Common Name:</b> | Surgical Mesh |
| <b>Class:</b> | Class II |
| <b>Classification:</b> | 21 CFR 878.3300 – Surgical Mesh |
| <b>Product Code:</b> | FTM |
#### Predicate Devices:
Biodesign® Staple Line Reinforcement (K170945, Cook Biotech Incorporated)
### Device Description:
The subject device consists of a thin multi-layer strip of SIS, pre-coated with a ready-to-use contact adhesive that eliminates the need for a separate adhesive (hydrogel), or activation of the adhesive through rehydration to affix the device to surgical stapler jaws. The single-use implant device is provided to the point of use with a protective liner covering the adhesive which has been assembled onto a foam carrier which is suspended in an alignment tray sized for the jaws of a surgical stapler.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Cook Biotech. The logo is set against a red background. The word "COOK" is written in white, block letters. Below the word "COOK" is a smaller red rectangle with the word "BIOTECH" written in white letters.
The completed device, liner, carrier, and alignment tray unit is then sealed in a foil pouch with an integrated humidity control feature and terminally E-beam sterilized. In order to accommodate a variety of commercially available surgical staplers the device is available in configurations measuring from 0.05mm to 0.6mm in thickness, and in nominal (unfolded) device lengths ranging from 76 mm to 176 mm in length and up to 12 mm wide.
Upon implantation, the Biodesign® Staple Line Reinforcement device will provide mechanical reinforcement of the staple line by buttressing the soft tissue and preventing the surgical staples from tearing through the affected tissue. In addition, the Biodesign® Staple Line Reinforcement device will incorporate (remodel) into the body over time such that no graft material is left behind.
## Indications for Use:
The Biodesign® Staple Line Reinforcement is intended for use as a prosthesis for the surgical repair of soft tissue deficiencies using surgical staple Line Reinforcement may be used for buttressing and reinforcing staple lines during lung resection (e.g., wedge resection, blebectomy, lobectomy, bullectomy, bronchial resection, segmentectomy, pneumonectomy/pneumectomy, pneumoreduction) and other incisions and excisions of the lung
and bronchus. The Staple Line Reinforcement can be used for the reinforcement of the gastric staple line during bariatric surgical procedures, and for reinforcement of staple lines during small bowel, mesentery, colon and colorectal procedures. The Staple Line Reinforcement may be used with anastomotic staplers or with non-anastomotic staplers.
## Comparison of Technological Characteristics with the Predicate Device:
The intended use and mode of action of the Biodesign® Staple Line Reinforcement device is the same as the predicate K170945 device in that they are both intended for use as a prosthesis for the surgical repair of soft tissue deficiencies using surgical staplers. The porcine Small Intestine Submucosa (SIS) material that functionally buttresses the staple line is identical to that of the predicate. Both devices are single use devices which remodel or are absorbed over time. The technological differences between the subject device and the predicate primarily involve the adhesive which is supplied in a ready-to-use formulation that adheres on contact with the surgical stapler. Other changes include design and packaging improvements made possible by this change.
| Device | Biodesign® Staple Line Reinforcement<br>(subject device) | Biodesign® Staple Line Reinforcement<br>(predicate device) |
|--------------------------|----------------------------------------------------------|------------------------------------------------------------|
| Manufacturer | Cook Biotech, Incorporated<br>**unchanged** | Cook Biotech, Incorporated |
| 510 (k) Number | K201000 | K170945 |
| Product Code | FTM<br>**unchanged** | FTM |
| Materials/<br>Components | SIS buttressing strip<br>**Identical material** | SIS buttressing strip |
#### Device Comparison Table
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Cook Biotech. The logo is set against a red background. The word "COOK" is written in large, white, sans-serif font. Below "COOK" is a smaller red rectangle with the word "BIOTECH" written in white, sans-serif font.
| Device | Biodesign® Staple Line Reinforcement<br>(subject device) | Biodesign® Staple Line Reinforcement<br>(predicate device) |
|----------------------------------------|-----------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| | Pre-coated adhesive | Pre-coated adhesive |
| | (ready-to-use contact adhesive) | (requires activation with water) |
| | Center support/carrier<br>(discarded; indirect contact)<br>**identical material** | Center support<br>(discarded; indirect contact) |
| | Protective adhesive liner<br>(discarded; indirect contact) | none |
| | Alignment tray (discarded; non-contact) | Storage tray (discarded; non-contact) |
| SIS Device<br>Dimensions<br>(unfolded) | Length: up to 176 mm<br>Width: up to 12 mm<br>**unchanged** | Length: up to 176 mm<br>Width: up to 12 mm |
| Device Thickness<br>(unfolded) | Overall device thickness: 0.05mm to 0.6mm<br>(depending on layer count) | Overall device thickness: 0.05mm to 0.5mm<br>(depending on layer count) |
| Shelf-Life | 6 months | 18 months |
| Sterile Barrier<br>Packaging | foil pouch | foil pouch |
| Humidity control<br>packaging | Yes<br>(discarded; non-contact) | No |
| Sterilization | E-beam process | E-beam process |
| One-time Use | Yes<br>**unchanged** | Yes |
### Performance Data:
The following performance data were provided in support of the substantial equivalence determination.
#### Biocompatibility testing
Biocompatibility of the predicate device has been established in accordance with the ISO 10993 Biological Evaluation of Medical Devices series of standards. Biocompatibility test data for the subject device has been submitted in support of the substantial equivalence also under the ISO 10993 regime, specifically addressing:
- Evaluation and testing within a risk management process • Part 1:
- Part 3: Tests for genotoxicity, carcinogenicity, and reproductive toxicity .
- Part 5: Tests for in vitro cytotoxicity .
- Tests for local effects after implantation Part 6: ●
- Part 10: Tests for irritation and delayed-type hypersensitivity
- Part 11: Tests for systemic toxicity ●
### Non-Clinical bench testing
Device performance testing:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Cook Biotech. The logo is set against a red background. The word "COOK" is written in white, block letters at the top of the logo. Below "COOK" is a smaller red rectangle with the word "BIOTECH" written in white letters.
Product verification testing was performed on SHS conditioned sterilized finished devices to evaluate the mechanical performance of the subject device for its intended use. Staple line leakage testing confirms that the Biodesign® Staple Line Reinforcement provides adequate staple line buttressing for use as a prosthesis for the surgical repair of soft tissue deficiencies using surgical staplers.
Packaging system and shelf life testing:
Non-clinical bench testing was performed on representative SHS conditioned packaging system samples to confirm that the packaging system is adequate to support a 6-month shelf life claim. Testing at additional timepoints is ongoing and the product shelf life will be extended as stability and package system integrity data justifies.
## Sterilization
The subject Biodesign® Staple Line Reinforcement device undergoes an E-beam sterilization process, with the sterile barrier established by the primary packaging of each device in a reinforced foil pouch.
# Substantial Equivalence:
For purposes of determinations of substantial equivalence under section 513(i) of the FD&C Act (21 U.S.C. § 360c(i)), the Biodesign® Staple Line Reinforcement device has the same intended use and functions under the same mode of action as the predicate device. The only change presented to the patient through implantation of the subject device is from the improved adhesive formulation, while other design and packaging changes are exclusively related to ease-of-use prior to loading the device onto the stapler and do not affect the safety or effectiveness of the implanted device. The physical, functional, biocompatibility, and performance characteristics of the subject device are substantially equivalent to the predicate device and do not indicate any new or increased risks to safety or effectiveness when used according to its labelling.
The supplied verification testing therefore supports the determination that the subject Biodesign® Staple Line Reinforcement device is substantially equivalent to the legally marketed predicate K170945 device.
## Conclusion:
The Biodesign® Staple Line Reinforcement device is substantially equivalent to the predicate devices K170945 in indications for use, fundamental scientific technology, and technological characteristics. It is as safe and effective as the predicate. Cook Biotech considers the Biodesign® Staple Line Reinforcement device to be substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.