← Product Code [MUD](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MUD) · K221010

# Intra.Ox 2.0 Handheld Tissue Oximeter (K221010)

_Vioptix, Inc. · MUD · May 3, 2022 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-c%E2%80%94cardiovascular-monitoring-devices/MUD/K221010

## Device Facts

- **Applicant:** Vioptix, Inc.
- **Product Code:** [MUD](/submissions/CV/subpart-c%E2%80%94cardiovascular-monitoring-devices/MUD.md)
- **Decision Date:** May 3, 2022
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.2700
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Intra.Ox™ 2.0 Handheld Tissue Oximeter is intended to non-invasively estimate the percent oxygen saturation (StO2) in a volume of tissue. The Intra.Ox™ 2.0 Handheld Tissue Oximeter is indicated for use in monitoring patients during circulatory or perfusion examinations. The Intra.Ox™ 2.0 Handheld Tissue Oximeter is intended to be used by physicians, surgeons, nurses, or other skilled users in a medical environment. The Intra.Ox™ 2.0 Handheld Tissue Oximeter should only be used on adult patients.

## Device Story

Handheld, battery-powered, sterile tissue oximeter; estimates StO2 non-invasively. Components: reusable main unit (light sources, detectors, processing electronics), single-use sterile sheath, single-use battery pack. Operation: device placed in direct contact with tissue; uses spatially resolved spectrophotometric oximetry at five wavelengths (730, 760, 810, 845, 895 nm). Output: StO2 percentage displayed on built-in screen; provides both 'live' and 'average' readings. Used by physicians, surgeons, or nurses in medical environments. Assists clinicians in assessing tissue perfusion; aids in clinical decision-making regarding circulatory status.

## Clinical Evidence

No clinical data. Bench testing only. Heterogeneous blood phantom study using swine blood and Intralipid solution compared device performance against a gold-standard blood co-oximeter. Results demonstrated substantial equivalence in limits of agreement, slope, and intercept compared to the predicate device.

## Technological Characteristics

Handheld, battery-powered (4 Lithium AA, 6V). Spectrophotometric oximetry using 5 LED wavelengths (730, 760, 810, 845, 895 nm). Biocompatible materials. Dimensions: 7.3" x 2.7" x 3.1"; weight < 1 lb. Includes single-use sterile sheath and battery pack. Software includes averaging functionality for StO2 readings.

## Regulatory Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

## Predicate Devices

- Intra.Ox™ 2.0 Handheld Tissue Oximeter ([K191676](/device/K191676.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 3, 2022

ViOptix, Inc. % Valerie Defiesta-Ng Vice president, Reuglatory Affairs Veranex, Inc. (formerly Experien Group, LLC) 224 Airport Parkway, Suite 250 San Jose, California 95110

Re: K221010

Trade/Device Name: Intra.Ox 2.0 Handheld Tissue Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: MUD Dated: April 4, 2022 Received: April 5, 2022

Dear Valerie Defiesta-Ng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Purva Pandya, D.Eng. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K221010

Device Name Intra.Ox™ 2.0 Handheld Tissue Oximeter

Indications for Use (Describe)

The Intra. Ox™ 2.0 Handheld Tissue Oximeter is intended to non-invasively estimate the percent oxygen saturation (StO2) in a volume of tissue.

The Intra. Ox™ 2.0 Handheld Tissue Oximeter is indicated for use in monitoring patients during circulatory or perfusion examinations.

The Intra. Ox™ 2.0 Handheld Tissue Oximeter is intended to be used by physicians, surgeons, nurses, or other skilled users in a medical environment.

The Intra. Ox™ 2.0 Handheld Tissue Oximeter should only be used on adult patients.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) Notification K221010

#### GENERAL INFORMATION [807.92(A)(1)]

#### Applicant:

ViOptix, Inc. 39655 Eureka Drive Newark, CA 94560-4806 USA Phone: 510-226-5860 Fax: 510-226-5864

#### Contact Person:

Valerie Defiesta-Ng Vice President, Regulatory Affairs, Veranex, Inc. 224 Airport Parkway, Suite 250 San Jose, CA 95110 USA Phone: 408-400-0856 Fax: 408-400-0865 Email: Valerie.Defiesta-Ng@veranexsolutions.com

Date Prepared: April 22, 2022

#### DEVICE INFORMATION [807.92(A)(2)]

Classification: 21 CFR§870.2700

Product Code: MUD

Trade Name: Intra.Ox™ 2.0 Handheld Tissue Oximeter

Generic/Common Name: Oximeter, Tissue Saturation

PREDICATE DEVICE(S) [807.92(A)(3)] Intra.Ox™ 2.0 Handheld Tissue Oximeter (K191676)

{4}------------------------------------------------

### DEVICE DESCRIPTION [807,92(A)(4)]

The ViOptix Intra.Ox 2.0 Handheld Tissue Oximeter is a sterile, cordless, battery-powered device that non-invasively estimates the percent oxygen saturation (StO2) in a volume of tissue. The device includes three components and an accessory:

- Main Unit: a re-usable module consists of light sources, detectors, and processing ● electronics to convert measurements of reflected light into an estimate of StO>:
- Sheath: a single-use, sterile Sheath placed around the Main Unit during device use . (provide in the Disposable Kit); and
- Battery Pack: a single-use battery (provided in the Disposable Kit) that is paired with the . Main Unit to provide power.

The device uses spatially resolved optical measurements at five wavelengths. The device performs measurements on the patient by direct physical contact to the patient's tissue and displays the StO2 estimate on the built-in screen. The Intra.Ox 2.0 Handheld Tissue Oximeter is a single-use disposable constructed from biocompatible materials that can tolerate bodily fluids and other liquids such as disinfectants and marking materials.

The device shares the same indication for use and the same technological characteristics as the predicate device (K191676), including principle of operation, StO2 measured parameters, accuracy and range, energy delivered and power source.

# INDICATIONS FOR USE [807.92(A)(5)]

"The Intra.Ox™ 2.0 Handheld Tissue Oximeter is intended to non-invasively estimate the percent oxygen saturation (StO2) in a volume of tissue.

The Intra.Ox" 2.0 Handheld Tissue Oximeter is indicated for use in monitoring patients during circulatory or perfusion examinations.

The Intra.Ox"" 2.0 Handheld Tissue Oximeter is intended to be used by physicians, surgeons, nurses, or other skilled users in a medical environment.

The Intra.Ox™ 2.0 Handheld Tissue Oximeter should only be used on adult patients."

# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES [807.92(a) (6)]

The technological characteristics of the modified Intra.Ox 2.0 Handheld Tissue Oximeter are substantially equivalent to the predicate device, Intra.Ox 2.0 Handheld Tissue Oximeter (K191676). Table 1 lists the technological characteristics of the predicate and modified devices and provides rationale to support a determination of substantial equivalence. Any differences between the devices do not raise any different questions of safety or efficacy.

{5}------------------------------------------------

| Table 1: Summary of Technological Characteristics |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                  |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature                                           | Predicate Intra.Ox 2.0<br>Handheld Tissue Oximeter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Modified Intra.Ox 2.0<br>Handheld Tissue Oximeter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Substantial<br>Equivalence<br>Rationale                                                                                                                                                                                                                          |
| 510(k) Number                                     | K191676                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | TBD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | --                                                                                                                                                                                                                                                               |
| Indications for<br>Use                            | The Intra.OxTM 2.0 Handheld<br>Tissue Oximeter is intended to<br>non-invasively estimate the<br>percent oxygen saturation (StO2)<br>in a volume of tissue.<br><br>The Intra.OxTM 2.0 Handheld<br>Tissue Oximeter is indicated for<br>use in monitoring patients during<br>circulatory or perfusion<br>examinations.<br><br>The Intra.OxTM 2.0 Handheld<br>Tissue Oximeter is intended to be<br>used by physicians, surgeons,<br>nurses, or other skilled users in a<br>medical environment.<br><br>The Intra.OxTM 2.0 Handheld<br>Tissue Oximeter should only be<br>used on adult patients. | The Intra.OxTM 2.0 Handheld<br>Tissue Oximeter is intended to<br>non-invasively estimate the<br>percent oxygen saturation (StO2)<br>in a volume of tissue.<br><br>The Intra.OxTM 2.0 Handheld<br>Tissue Oximeter is indicated for<br>use in monitoring patients during<br>circulatory or perfusion<br>examinations.<br><br>The Intra.OxTM 2.0 Handheld<br>Tissue Oximeter is intended to be<br>used by physicians, surgeons,<br>nurses, or other skilled users in a<br>medical environment.<br><br>The Intra.OxTM 2.0 Handheld<br>Tissue Oximeter should only be<br>used on adult patients. | N/A (same)                                                                                                                                                                                                                                                       |
| Measured<br>Parameters                            | Tissue oxygen saturation<br>(% StO2)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Tissue oxygen saturation<br>(% StO2)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | N/A (same)                                                                                                                                                                                                                                                       |
| Principle of<br>Operation                         | Spectrophotometric oximetry                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Spectrophotometric oximetry                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | N/A (same)                                                                                                                                                                                                                                                       |
| Energy<br>Delivered                               | Near-infrared light<br>Source: LED chips<br>Wavelengths: 730, 760, 810, 845,<br>and 895 nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Near-infrared light<br>Source: LED chips<br>Wavelengths: 730, 760, 810, 845,<br>and 895 nm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | N/A (same)                                                                                                                                                                                                                                                       |
| Single Patient<br>Use?                            | Yes, the Sheath is single patient<br>use.<br>The Main Unit is re-usable after<br>proper cleaning.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Yes, the Sheath is single patient<br>use.<br>The Main Unit is re-usable after<br>proper cleaning.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | N/A (same)                                                                                                                                                                                                                                                       |
| Power Source                                      | Battery Powered<br>Battery Type: 4 Lithium AA<br>Battery Voltage: 6 V total                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Battery Powered<br>Battery Type: 4 Lithium AA<br>Battery Voltage: 6 V total                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | N/A (same)                                                                                                                                                                                                                                                       |
| StO2<br>Measurement<br>Range                      | StO2 displayed range shall be an<br>integer between 1 to 99%.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | StO2 displayed range shall be an<br>integer between 1 to 99%.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | N/A (same)                                                                                                                                                                                                                                                       |
| Physical Design                                   | Ergonomic hand-held design for<br>range of hand sizes, includes<br>thumb rest, ambidextrous use.<br>Device contains a disposable kit<br>that contains a single-use battery<br>pack and sterile single-use<br>disposable sheath.                                                                                                                                                                                                                                                                                                                                                             | Ergonomic hand-held design for<br>range of hand sizes, includes<br>thumb rest, ambidextrous use.<br>Device contains a disposable kit<br>that contains a single-use battery<br>pack and sterile single-use<br>disposable sheath.                                                                                                                                                                                                                                                                                                                                                             | N/A (same)                                                                                                                                                                                                                                                       |
| Feature                                           | Predicate Intra.Ox 2.0<br>Handheld Tissue Oximeter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Modified Intra.Ox 2.0<br>Handheld Tissue Oximeter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Substantial<br>Equivalence<br>Rationale                                                                                                                                                                                                                          |
| Dimensions and<br>Weight                          | 7.3" x 2.7" x 3.1"<br>L x W x H<br>Less than 1 lb                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 7.3" x 2.7" x 3.1"<br>L x W x H<br>Less than 1 lb                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | N/A (same)                                                                                                                                                                                                                                                       |
| Components                                        | Main Unit, Sheath,<br>Battery Pack and Quality Control<br>Target                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Main Unit, Sheath, and<br>Battery Pack                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Similar components.<br>The principle of<br>operation remains the<br>same for both devices.<br>The removal of the<br>Quality Control Target<br>does not raise different<br>questions of safety and<br>effectiveness as<br>demonstrated by<br>performance testing. |

Table 1: Summary of Technological Characteristics

{6}------------------------------------------------

#### SUBSTANTIAL EQUIVALENCE

The Intra.Ox 2.0 Handheld Tissue Oximeter is substantially equivalent to the predicate device with regard to intended use, Indications for Use, principle of operation and fundamental scientific technology. Any differences in the technological characteristics between the devices do not raise any different questions of safety or effectiveness. Thus, the modified Intra.Ox 2.0 Handheld Tissue Oximeter is substantially equivalent to the predicate device.

#### SUMMARY OF DESIGN MODIFICATION

The following design modifications have been implemented and do not affect the intended use, the Indications for Use, or alter the fundamental scientific technology of the device. ViOptix is proposing modifications to the cleared Intra. Ox 2.0 Handheld Tissue Oximeter (K191676) to optimize the design and improve manufacturability.

- The robustness of the latch-detection functions for sheath closure was improved. ●
- Adhesive was removed from the inside of the sheath window, a sterile pressure-relief vent was added to the lid, and the optical target was removed for the kit.
- An averaging functionality was added with screen modification to display both "live" and ● "average" readings.
- The removal of two non-functional components from a device Printed Circuit Board . Assembly (PCBA).

# PERFORMANCE DATA [807.92(B)]

All necessary bench testing was conducted on the modified Intra.Ox 2.0 Handheld Tissue Oximeter to support a determination of substantial equivalence to the predicate device. Tests performed included:

- Mechanical Testing ●
- Sterility Testing ●
- Latch Testing ●
- Shelf-Life and Transport Testing ●
- Software Verification and Validation ●

{7}------------------------------------------------

- Heterogeneous Blood Phantom Study
The collective performance testing demonstrates that the Intra.Ox 2.0 Handheld Tissue Oximeter does not raise any different questions of safety or effectiveness when compared to the predicate device. The results of performance testing demonstrate that the Intra.Ox 2.0 Handheld Tissue Oximeter performs as intended.

# [807.92(b)(1)]Non-clinical Testing Summary

In a heterogenous phantom study, measurements were made using the Intra.Ox 2.0 Handheld Tissue Oximeter in a heterogeneous phantom prepared with swine whole blood using an Intralipid solution to mimic tissue scattering and compared to a "gold standard" blood cooximeter. The results show that the Intra.Ox 2.0 Handheld Tissue Oximeter is substantially equivalent to the predicate device in limits of agreement to the gold standard, as well as 95% confidence intervals in slope and intercept. These data support the conclusion that the Intra.Ox 2.0 Handheld Tissue Oximeter provides as good or better agreement to the gold standard over the clinically relevant range than the predicate Intra.Ox Handheld Tissue Oximeter.

# [807.92(b)(2)]Clinical Testing Summary

Clinical testing was not required to demonstrate substantial equivalence.

# CONCLUSION [807.92 (b) (3)]

Nonclinical performance testing has been performed on the Intra.Ox 2.0 Handheld Tissue Oximeter to evaluate the overall performance of the device. The results confirm that the Intra. Ox 2.0 Handheld Tissue Oximeter meets its specifications and meets requirements for its intended use. The Intra.Ox 2.0 Handheld Tissue Oximeter is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-c%E2%80%94cardiovascular-monitoring-devices/MUD/K221010](https://fda.innolitics.com/submissions/SU/subpart-c%E2%80%94cardiovascular-monitoring-devices/MUD/K221010)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
