← Product Code [KNW](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/KNW) · K213232

# NovaCore Semi-Automatic Biopsy Instrument (K213232)

_Argon Medical Devices · KNW · Jul 1, 2022 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/KNW/K213232

## Device Facts

- **Applicant:** Argon Medical Devices
- **Product Code:** [KNW](/submissions/GU/subpart-b%E2%80%94diagnostic-devices/KNW.md)
- **Decision Date:** Jul 1, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1075
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Semi-Automatic Biopsy Instrument and associated Co-Axial Introducer Needle are used to obtain samples from soft tissue such as lung, liver, spleen, kidney, prostate, lymph nodes, breast, thyroid, pancreas, and other masses.

## Device Story

Semi-automatic, spring-loaded biopsy instrument for soft tissue specimen collection; used in clinical settings by healthcare professionals. Device features a needle with echogenic tip and centimeter markings for depth guidance; adjustable specimen notch size (10mm or 20mm); integrated co-axial adapter. Operated manually via single-handed handle; needle position visualized via X-ray, CT, or ultrasound. Stylet advances prior to cannula firing to capture tissue sample. Provides clinical flexibility for various tissue types; aids in diagnostic decision-making through high-quality tissue sampling.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance testing included dimensional analysis, peak tensile strength, simulated use (charging/firing force, sample collection), shipping qualification, radiopacity, echogenicity, corrosion resistance, needle bending/breakage resistance, aging studies, penetration force, and tissue sampling. Biocompatibility testing performed per ISO 10993-1:2018, including cytotoxicity, sensitization, irritation/intracutaneous reactivity, acute systemic toxicity, and material-mediated pyrogenicity.

## Technological Characteristics

Semi-automatic spring-loaded biopsy instrument. Materials: 304 stainless steel (stylet/cannula), plastic handle. Features: echogenic needle tip, numerically ordered centimeter markings, adjustable notch size (10mm/20mm), integrated co-axial adapter. Dimensions: 14ga-20ga, lengths 6cm-25cm. Sterilization: Not specified. Connectivity: None (mechanical).

## Regulatory Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Predicate Devices

- Argon Medical SuperCore™ Biopsy Instrument ([K974814](/device/K974814.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.

July 1, 2022

Argon Medical Devices Daniel Lanois Senior Regulatory Affairs Specialist 1445 Flat Creek Road Athens. Texas 75751

Re: K213232

Trade/Device Name: Semi-Automatic Biopsy Instrument Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: June 15, 2022 Received: June 16, 2022

Dear Daniel Lanois:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K213232

Device Name Semi-Automatic Biopsy Intrument

Indications for Use (Describe)

The Semi-Automatic Biopsy Instrument and associated Co-Axial Introducer Needle are used to obtain samples from soft tissue such as lung, liver, spleen, kidney, prostate, lymph nodes, breast, thyroid, pancreas, and other masses.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| Date Prepared:                         | June 30, 2022                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company:                               | Argon Medical Devices, Inc.<br>1445 Flat Creek Road<br>Athens, Texas 75751 USA<br>Facility Registration number: 1625425                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Contact:                               | Scott Bishop<br>Director Regulatory Affairs<br>Phone: 469-731-1413<br>Fax: 903-677-9396<br>Email: scott.bishop@argonmedical.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Device Trade Name<br>(Subject Device): | Semi-Automatic Biopsy Instrument                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Device Common<br>Name:                 | Biopsy Instrument                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Device Classification:                 | Instrument, Biopsy<br>Product code: KNW<br>21 CFR 876.1075<br>Class II<br>Review Panel: Gastroenterology/Urology                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Predicate Device(s):                   | K974814 Argon Medical SuperCore™ Biopsy Instrument                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Description of the<br>Device:          | The Semi-Automatic Biopsy Instrument is used to obtain soft<br>tissue biopsies. It is composed of a spring-loaded biopsy<br>needle fitted into a plastic handle permitting single handed<br>specimen collection. The position of the needle may be<br>visualized by x-ray, CT, or ultrasound.<br><br>The needle has numerically ordered centimeter markings to<br>facilitate precise depth placement. The adjustable instrument<br>allows for a specimen notch size of 10mm or 20mm, providing<br>clinical flexibility. The device has an adjustable co-axial<br>adapter built into the device for either the 10mm or 20mm<br>notch settings. The stylet travels up to 27mm from when the<br>device is in a fully charged position (prepared for a collection<br>with the 20 mm notch size) to when the stylet is fully extended.<br>The needle has an echogenic tip. The semi-automatic biopsy<br>instrument is available in several needle gauge sizes and |

# 510(k) Summary

{4}------------------------------------------------

|                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                    | K213232                         |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|
|                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                         | lengths. The plunger is color coded according to the various<br>gauge sizes, e.g., green=14-gauge, purple=16-gauge,<br>pink=18-gauge, and yellow=20-gauge.                                                         |                                 |
| Indication for Use:                | The Semi-Automatic Biopsy Instrument and associated Co-<br>Axial Introducer Needle are used to obtain samples from soft<br>tissue such as lung, liver, spleen, kidney, prostate, lymph nodes,<br>breast, thyroid, pancreas, and other masses.                                                                                                                                                                                           |                                                                                                                                                                                                                    |                                 |
| Technological<br>Characteristics:  | A comparison of the technological characteristics of the<br>Subject device and the Predicate device demonstrates the<br>Subject device to be substantially equivalent to the current<br>marketed Predicate device.                                                                                                                                                                                                                      |                                                                                                                                                                                                                    |                                 |
|                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                         | The Subject device and the Predicate device are based on the<br>following same technological elements:                                                                                                             |                                 |
|                                    | ● Use of an adjustable specimen notch size<br>● Use of a semi-automatic spring-loaded biopsy needle<br>permitting the stylet to advance prior to firing the<br>cannula<br>● Use of an echogenic needle tip<br>● Numerically ordered centimeter markings to facilitate<br>precise depth placement of the needle                                                                                                                          |                                                                                                                                                                                                                    |                                 |
|                                    | The following technological differences exist between the<br>Subject device and the Predicate device:<br>● Use of a co-axial adapter built into the Subject Device<br>and is not built into the Predicate Device.                                                                                                                                                                                                                       |                                                                                                                                                                                                                    |                                 |
|                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                    |                                 |
|                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                         | SUBJECT DEVICE                                                                                                                                                                                                     | PREDICATE DEVICE                |
|                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                         | Semi-Automatic<br>Biopsy Instrument                                                                                                                                                                                | SuperCore™ Biopsy<br>Instrument |
|                                    | Manufacturer                                                                                                                                                                                                                                                                                                                                                                                                                            | Argon Medical Devices,<br>Inc - TX                                                                                                                                                                                 | SAME                            |
|                                    | Gauge Size                                                                                                                                                                                                                                                                                                                                                                                                                              | 14ga, 16ga, 18ga, 20ga                                                                                                                                                                                             | 14ga, 16ga, 18ga,<br>20ga       |
|                                    | Lengths                                                                                                                                                                                                                                                                                                                                                                                                                                 | 6cm, 10cm, 16cm,<br>20cm, 25cm (18ga only)                                                                                                                                                                         | 6cm, 9cm, 15cm,<br>20cm         |
|                                    | Notch Size                                                                                                                                                                                                                                                                                                                                                                                                                              | 10mm and 20mm                                                                                                                                                                                                      | 9.5mm and 19mm                  |
|                                    | Tip Geometry<br>(Stylet and<br>Cannula)                                                                                                                                                                                                                                                                                                                                                                                                 | Bevel                                                                                                                                                                                                              | SAME                            |
| Echogenicity<br>(Tip)              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                     | SAME                                                                                                                                                                                                               |                                 |
| Needle<br>Manufacturer             | Argon Medical Devices,<br>Inc. - IL                                                                                                                                                                                                                                                                                                                                                                                                     | SAME                                                                                                                                                                                                               |                                 |
| Patient<br>Contacting<br>Materials | Stylet: 304 SS<br>Cannula: 304 SS                                                                                                                                                                                                                                                                                                                                                                                                       | SAME                                                                                                                                                                                                               |                                 |
| Function Type                      | Semi-Automatic                                                                                                                                                                                                                                                                                                                                                                                                                          | SAME                                                                                                                                                                                                               |                                 |
| Disposable                         | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                     | SAME                                                                                                                                                                                                               |                                 |
| Performance<br>Testing             | • Dimensional<br>• Peak Tensile Strength<br>• Simulated Use<br>(Charging Force,<br>Firing Force, Minimum<br>Firing Force, Dry<br>Firing, Sample<br>Collection)<br>• Shipping Qualification<br>(Product and<br>Packaging Integrity)<br>• Radiopacity<br>• Echogenicity<br>• Corrosion Resistance<br>• Resistance to<br>Breakage / Needle<br>Bending<br>• Accelerated and Real-<br>Time Aging<br>• Penetration Force<br>• Tissue Sampling | • Dimensional<br>• Peak Tensile<br>Strength<br>• Durability of<br>Instrument<br>• Charging Force<br>• Firing Force<br>• Dry Firing (minimum<br>firing force)<br>• Sample Collection<br>• Shipping<br>Qualification |                                 |
| Biological<br>Comparison           | Complies to ISO 10993-<br>1:2018<br>• Cytotoxicity (ISO<br>10993-5)<br>• Sensitization (ISO<br>10993-10)<br>• Irritation or -<br>Intracutaneous<br>Reactivity (ISO 10993-<br>10)<br>• Acute Systemic<br>Toxicity (ISO 10993-<br>11)<br>• Material Mediated<br>Pyrogenicity (ISO<br>10993-11)                                                                                                                                            | Complies to ISO<br>10993-1:2009<br>• Cytotoxicity (ISO<br>10993-5)<br>• Sensitization (ISO<br>10993-10)<br>• Irritation -<br>Intracutaneous<br>Reactivity (ISO<br>10993-10)                                        |                                 |
| Intended<br>Shelf-Life             | 6 months                                                                                                                                                                                                                                                                                                                                                                                                                                | 5 years                                                                                                                                                                                                            |                                 |
| Packaging /<br>Shipping            | Biopsy Instrument<br>Only                                                                                                                                                                                                                                                                                                                                                                                                               | Biopsy Instrument<br>Only                                                                                                                                                                                          |                                 |

17mm, 27mm

12mm, 22mm

Throw

{5}------------------------------------------------

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{6}------------------------------------------------

| • One biopsy device<br>packaged in a Tyvek<br>Pouch                                                                                                                                                                                                                                                   | • One biopsy device<br>packaged in a Tyvek<br>Pouch                                                                                                                                                   |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Co-packaged Kit<br>• One biopsy device,<br>corresponding coaxial<br>introducer needle in a<br>holder packaged in a<br>Tyvek pouch (18ga<br>and 20ga coaxial<br>introducer needles are<br>provided with an<br>additional blunt stylet)<br>10 pouches and one IFU<br>in each labeled shipping<br>carton | Co-packaged Kit<br>• One biopsy device<br>with corresponding<br>coaxial introducer<br>needle, and adaptor<br>packaged in a Tyvek<br>pouch<br>10 pouches and one<br>IFU in one labeled shelf<br>carton |

Equivalence is based upon the product performance, design and intended use. The Subject device and the Predicate device have similar materials of construction, dimensional specifications, design, and sterilization process. This demonstrates that the Subiect device is as safe and effective as the Predicate device, and it does not raise different questions of safety and effectiveness than the Predicate device.

Performance Tests No performance standards have been established under (Non-Clinical): section 514, performance standards, of the Food, Drug and Cosmetic Act for these devices. A series of testing was conducted in accordance with protocols based on requirements outlined in guidance and industry standards and the testing below was shown to meet the acceptance criteria that was determined to demonstrate substantial equivalence.

> The following tests were performed under the specified testing parameters to support the Subject device substantial equivalence.

## Performance Testing, including:

- Dimensional ●
- Peak Tensile Strength
- Simulated Use (Charging Force, Firing Force, Minimum ● Firing Force, Dry Firing, Sample Collection)
- Shipping Qualification (Product and Packaging Integrity) ●
- Radiopacity
- Echogenicity ●

{7}------------------------------------------------

- Corrosion Resistance ●
- Resistance to Breakage / Needle Bending ●
- Accelerated and Real-Time Aging ●
- Penetration Force ●
- Tissue Sampling o

## Biocompatibility Testing, including:

- Cytotoxicity (ISO 10993-5) .
- Sensitization (ISO 10993-10) ●
- Irritation or Intracutaneous Reactivity (ISO 10993-10) ●
- . Acute Systemic Toxicity (ISO 10993-11)
- Material Mediated Pyrogenicity (ISO 10993-11) ●

| Substantial<br>Equivalence: | Based on the Indication for Use, design, and safety and<br>performance testing, the Subject device meets the<br>requirements for its intended use and is substantially<br>equivalent to the Predicate device. |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion:                 | The results of all testing demonstrate that the Subject device is<br>substantially equivalent to the Predicate device.                                                                                        |

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/KNW/K213232](https://fda.innolitics.com/submissions/SU/subpart-b%E2%80%94diagnostic-devices/KNW/K213232)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
