← Product Code [KPQ](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/KPQ) · K072445

# INTEGRATED BRACHYTHERAPY UNIT - DIGITAL (K072445)

_Nucletron Corp. · KPQ · Sep 14, 2007 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/KPQ/K072445

## Device Facts

- **Applicant:** Nucletron Corp.
- **Product Code:** [KPQ](/submissions/RA/subpart-f%E2%80%94therapeutic-devices/KPQ.md)
- **Decision Date:** Sep 14, 2007
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.5840
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The Integrated Brachytherapy - Digital (IBU-D) is intended to be used for the visualization, localization and confirmation of the volume and the size of the brachytherapy irradiation field(s), using a fluoroscopic and/or radiographic system.

## Device Story

The Integrated Brachytherapy Unit - Digital (IBU-D) is a radiation therapy simulation system used in brachytherapy departments. It features a gantry with an isocentrically rotating L-arm and C-arm, supporting an X-ray tube and a 41x41 cm amorphous silicon flat panel digital image detector. The system is manually operated by clinicians. It captures fluoroscopic and radiographic images, which are processed and managed via a PC-based workstation. The workstation software handles image acquisition, display, annotation, DICOM import/export, gantry position readout, and beam-limiting device control. By providing visualization and localization of the irradiation field, the device assists clinicians in confirming treatment volumes, thereby supporting accurate brachytherapy planning and delivery.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

System consists of a gantry with L-arm and C-arm, X-ray tube, and 41x41 cm amorphous silicon flat panel digital detector. Operates as a fluoroscopic/radiographic simulation system. PC-based workstation runs software for image management and system control. Connectivity includes DICOM import/export. Movements are manually driven with electrical locks.

## Regulatory Identification

A radiation therapy simulation system is a fluoroscopic or radiographic x-ray system intended for use in localizing the volume to be exposed during radiation therapy and confirming the position and size of the therapeutic irradiation field produced. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- Integrated Brachytherapy Unit ([K973848](/device/K973848.md))

## Reference Devices

- Simulix Evolution (K03347)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K072445

Image /page/0/Picture/1 description: The image shows the word "Nucletron" in bold, black font. To the left of the word is a square logo. The logo contains a black circle with two white circles on either side of it.

NUCLETRON B.V. Waardgelder 1 3905 TH Veenendaal P.O.Box 930 3900 AX Veenendaal The Netherlands Phone +31 318 557133 +31 318 550485 Fax

Department of Health and Human Services Centre of Device and Radiological Health Office of Device Evaluation Special 510(k) section

.

・・

### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION as required by section 807.92(c)

#### Submitter of 510(k):

| Company name:        | Nucletron Corporation                                   |
|----------------------|---------------------------------------------------------|
| Registration number: | 1121753                                                 |
| Address:             | 8671 Robert Fulton Drive<br>Columbia, MD 21046          |
| Phone:               | 410-312-4100                                            |
| Fax:                 | 410-312-4197                                            |
| Correspondent:       | Lisa Dimmick<br>Director Assurance & Regulatory Affairs |

### Modified Device Name:

| Trade/Proprietary Name: | Integrated Brachytherapy Unit - Digital                           |
|-------------------------|-------------------------------------------------------------------|
| Common/Usual Name:      | Fluoroscopic / radiographic radiation treatment simulation system |
| Classification Name:    | System, Simulation, Radiation Therapy                             |
| Classification:         | 21Cfr892.5840 Class II                                            |
| Product Code            | KPQ                                                               |

### Legally Marketed Device(s)

Our device is substantially equivalent to the legally marketed predicate device cited in the table below:

| <br>                                 | 110 - 11 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 -<br>Carlo Corporator<br>1 1 0 11 11 11 11 11 11 15 6 7 A 2 A 4 1 1 1 4 4 1 4 1 1 1 1 4 4 4 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1<br><br>. Toursely and and and and seat and the may be and<br>.<br>-<br>" " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " "<br>A WASSALE VARPER FOR SES CAMARS PAS ARS ANAS SAMARATS SA<br>1150 ml . Alliade to Pac 11, 15, 11, 15, 1959 (Page 200 million 1988)<br>YATAKA ZAY=YAYA N<br> |  |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| . LA MINILING . A . A . A . A . MINI |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |  |

### Description:

Integrated Brachytherap Unit - Digital (IBU-D) is a modification to the Integrated Brachytherapy Unit (IBU) in which the Image Intensifier of the IBU is replaced by a Flat Panel image detector. The Flat Panel image detector used in the IBU-D is the same Flat Panel image detector as used in Nucletron's Simulix Evolution product (K03347)

{1}------------------------------------------------

The Integrated Brachytherpay Unit – Digital (IBU-D) is a localization and simulation device for a Brachy radiation therapy department. It consists of a gantry that supports an L-arm and a C-arm which can rotate isocentrically. The C-arm houses an X-ray tube housing assembly with collimator on one side and a flat panel image detector. The movements of the IBU-D are manually driven, after the relevant electrical locks are lifted. The mobile IBU patient table has mechanical motions which can be controlled from a hand pendant affixed to the table. Images are displayed and managed by a PC based workstation running specialized software.

The system makes also use of the same third party X-ray tube and X-ray high tension generator as used in the Simulix Evolution system.

The Flat Panel image detector which replaces the current Image intensifier is a Amorphous silicon, digital detector, with a square image area of 41 x 41 cm.

The PC based workstation runs the same software as the workstation of the Simulix Evolution system. It supports the following functionality:

- . Image acquisition
- Image display .
- Image annotation .
- � Database and DICOM Import / Export functionality
- Position read out and display of the IBU-D gantry. .
- Control of the IBU-D beam limiting device. .

The last two items in this list are specific for the IBU-D.

#### Intended use:

The modified device has the same intended use as the legally marketed predicate device cited:

The Integrated Brachytherapy - Digital (IBU-D) is intended to be used for the visualization, localization and confirmation of the volume and the size of the brachytherapy irradiation field(s), using a fluoroscopic and/or radiographic system.

### Summary of technological considerations:

Integrated Brachytherapy Unit - Digital is substantially equivalent to the cleared predicate device, Integrated Brachytherapy Unit, 510(k)#: K973848.

Signature

13 AUG 2007
Date

Name: Dick van-Waes Title: Business Director Brachytherapy & Imaqing Nucletron B.V. Veenendaal, The Netherlands

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, arranged in a vertical stack. The profiles are depicted with flowing lines, giving a sense of movement or continuity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES U.S.A." is arranged in a circular fashion around the image.

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

#### 2007 SEP 14

Ms. Lisa Dimmick Director, Regulatory Affairs and Quality Assurance Nucletron Corporation 8671 Robert Fulton Drive COLUMBIA MD 21046

Re: K072445

Trade/Device Name: Integrated Brachytherapy Unit - Digital Regulation Number: 21 CFR 892,5840 Regulation Name: Radiation therapy stimulation system Regulatory Class: II Product Code: KPO Dated: August 13, 2007 Received: August 30, 2007

#### Dear Ms. Dimmick:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indived on for use stated in the enclosure) to legally marketed predicate devices marketed in ior un mically in prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You most therefore, market the device, subject to the general controls provisions of the Act, TTo general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adultervition.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/2/Picture/10 description: The image shows a circular logo with the text "1906-2006" at the top. Below the text is a large "FDA" in bold letters. Underneath "FDA" is the word "Centennial" in a cursive font. At the bottom of the logo are three small stars.

otion Publio Hoolts clating and Prom

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements not the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as se( forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the ed ecronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1150

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a logal/v marketed predicate device results in a classification for your device and thus, perceits your evice to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                      | 240-276-0120 |
| Other          |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on vour responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistances at the toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours,

Nancy Chrogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

# Indications for Use Statement

510(k) Number

Device Name

Indications for Use

K072445

Integrated Brachytherapy Unit - Digital

The Integrated Brachytherapy - Digital (IBU-D) is intended to be used for the visualization, localization and confirmation of the volume and the size of the brachytherapy irradiation field(s), using a fluoroscopic and/or radiographic system.

PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

|                                                                  | <img alt="signature" src="signature.png"/> |
|------------------------------------------------------------------|--------------------------------------------|
| (Division Sign-Off)                                              |                                            |
| Division of Reproductive, Abdominal, and<br>Radiological Devices |                                            |
| 510(k) Number                                                    | K072445                                    |

| Prescription Use     | <div style="text-align:center;">✓</div> | OR | Over-The-Counter Use |
|----------------------|-----------------------------------------|----|----------------------|
| (Per 21 CFR 801.109) |                                         |    |                      |

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/KPQ/K072445](https://fda.innolitics.com/submissions/RA/subpart-f%E2%80%94therapeutic-devices/KPQ/K072445)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
