TrueBeam, TrueBeam STX, Edge and VitalBeam

K252948 · Varian Medical Systems, Inc. · IYE · Oct 7, 2025 · Radiology

Device Facts

Record IDK252948
Device NameTrueBeam, TrueBeam STX, Edge and VitalBeam
ApplicantVarian Medical Systems, Inc.
Product CodeIYE · Radiology
Decision DateOct 7, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

TrueBeam-TrueBeam STx-Edge: The TrueBeam™, TrueBeam STx and Edge™ Systems are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients. The TrueBeam radiotherapy system can produce CBCT images that can be used in image guided radiation therapy, and the simulation and planning for adaptive radiation therapy. The TrueBeam, TrueBeam STx and Edge Systems may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, larynx, oropharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, glomus tumors, hemangiomas, and medically refractory essential tremor (indicated for adults only)), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin), pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation, and skull base tumors), and low-dose radiotherapy for adults with medically refractory osteoarthritis. VitalBeam: VitalBeam® is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients. VitalBeam may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, larynx, oropharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin), pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation, and skull base tumors), and low-dose radiotherapy for adults with medically refractory osteoarthritis.

Device Story

Medical linear accelerator (linac) system delivering therapeutic radiation (photon, electron, kV X-ray) for stereotactic radiosurgery and precision radiotherapy. System comprises radiation-shielded vault component and external control console. Operated by authorized medical practitioners in clinical settings. Inputs include physician-prescribed treatment plans; system utilizes integrated imaging (CBCT) for image-guided radiation therapy (IGRT) and adaptive planning. Multi-leaf collimators shape radiation beams to target lesions/tumors while sparing healthy tissue. Output is controlled radiation delivery. Benefits include precise tumor treatment and management of benign conditions. Modifications in this submission primarily involve updated indications for use; device principle of operation, hardware, and software remain substantially equivalent to predicate.

Clinical Evidence

No clinical or animal testing submitted. Substantial equivalence established via bench testing, verification, and validation of hardware and software modifications, and conformance to recognized consensus standards (e.g., IEC 60601 series, ISO 14971, ISO 13485).

Technological Characteristics

Medical linear accelerator (linac) for photon, electron, and kV X-ray delivery. Features multi-leaf collimators, integrated imaging/treatment consoles, and patient support couch. Software-controlled. Conforms to IEC 60601-1, IEC 60601-2-1, IEC 62304, ISO 10993-1, ISO 14971, and IEC 62366-1. No changes to patient-contact materials or biocompatibility.

Indications for Use

Indicated for adults and pediatric patients requiring stereotactic radiosurgery or precision radiotherapy for lesions, tumors, and conditions anywhere in the body, including brain, spine, head/neck, thoracic, gynecologic, gastrointestinal, genitourinary, breast, sarcomas, lymphoid, skin, benign diseases, and metastasis. Includes low-dose radiotherapy for adults with medically refractory osteoarthritis.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION October 7, 2025 Varian Medical Systems, Inc. Lynn Allman Senior Director, Regulatory Affairs 3100 Hansen Way Palo Alto, California 94304 Re: K252948 Trade/Device Name: TrueBeam, TrueBeam STX, Edge and VitalBeam Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: IYE Dated: September 15, 2025 Received: September 16, 2025 Dear Lynn Allman: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252948 - Lynn Allman Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K252948 - Lynn Allman Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ![img-0.jpeg](img-0.jpeg) Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252948 | ? | | Please provide the device trade name(s). | | ? | | TrueBeam, TrueBeam STX, Edge and VitalBeam | | | | Please provide your Indications for Use below. | | ? | | TrueBeam-TrueBeam STx-Edge: | | | | The TrueBeam™, TrueBeam STx and Edge™ Systems are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients. The TrueBeam radiotherapy system can produce CBCT images that can be used in image guided radiation therapy, and the simulation and planning for adaptive radiation therapy. The TrueBeam, TrueBeam STx and Edge Systems may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, larynx, oropharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, glomus tumors, hemangiomas, and medically refractory essential tremor (indicated for adults only)), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin), pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation, and skull base tumors), and low-dose radiotherapy for adults with medically refractory osteoarthritis. VitalBeam: VitalBeam® is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients. VitalBeam may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, larynx, oropharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous | | | {4} malformation, cavernous malformation, trigeminal neuralgia, chordoma, glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin), pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation, and skull base tumors), and low-dose radiotherapy for adults with medically refractory osteoarthritis. Please select the types of uses (select one or both, as applicable). ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) {5} varian # 510(K) SUMMARY K252948 Varian Medical Systems 3100 Hansen Way Palo Alto, CA 94304 The following information is provided as required by 21 CFR 807.92. ## SUBMITTER **Name and Address:** Varian Medical Systems 3100 Hansen Way, m/s E110 Palo Alto, CA 94304 **Contact Person:** Lynn Allman Sr. Director, Regulatory Affairs Phone: 650-424-5369 submissions.support@varian.com **Date Prepared:** 15 September 2025 ## DEVICE **Subject Device Name:** TrueBeam™ /TrueBeam STx™/Edge™/VitalBeam **Common/Usual Name:** Linear accelerator radiation therapy system **Product Code and Classification:** Medical charged-particle radiation therapy system IYE | 21 CFR 892.5050 | Class II ## PREDICATE DEVICE **Predicate Device Name:** TrueBeam™ /TrueBeam STx™/Edge™/VitalBeam (K232870) **Reference Device(s):** No reference devices were used in this submission. ## DEVICE DESCRIPTION The TrueBeam and VitalBeam Radiotherapy System is a medical linear accelerator that delivered therapeutic radiation to patient in accordance with the physician's prescription. The system consists of two major components – a photon, electron and diagnostic kV X-ray radiation beam producing component that is installed in a radiation-shielded vault and a control console area located outside the treatment room. ## INTENDED USE **TrueBeam-TrueBeam STx-Edge:** The TrueBeam™ radiotherapy delivery system is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated. **VitalBeam:** The VitalBeam system is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated. Product: TrueBeam™ /TrueBeam STx™/Edge™/VitalBeam {6} Varian A Siemens Healthineers Company # INDICATIONS FOR USE ## TrueBeam-TrueBeam STx-Edge: The TrueBeam™, TrueBeam STx and Edge™ Systems are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients. The TrueBeam radiotherapy system can produce CBCT images that can be used in image guided radiation therapy, and the simulation and planning for adaptive radiation therapy. The TrueBeam, TrueBeam STx and Edge Systems may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, larynx, oropharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, glomus tumors, hemangiomas, and medically refractory essential tremor (indicated for adults only)), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin), pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation, and skull base tumors), and low-dose radiotherapy for adults with medically refractory osteoarthritis. ## VitalBeam: VitalBeam® is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients. VitalBeam may be used in the delivery of radiation for treatment that includes: brain and spine tumors (such as glioma, meningioma, craniopharyngioma, pituitary tumors, spinal cord tumors, hemangioblastoma, orbital tumors, ocular tumors, optic nerve tumors, and skull based tumors), head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, larynx, oropharynx, nasopharynx, sinonasal, salivary gland, and thyroid cancer), thoracic tumors (such as lung cancer, esophageal cancer, thymic tumors, and mesothelioma), gynecologic tumors (such as ovarian, cervical, endometrial, vulvar, and vaginal), gastrointestinal tumors (such as gastric, pancreatic, hepatobiliary, colon, rectal, and anal carcinoma), genitourinary tumors (such as prostate, bladder, testicular, and kidney), breast tumors, sarcomas, lymphoid tumors (such as Hodgkin's and non-Hodgkin's lymphoma), skin cancers (such as squamous cell, basal cell, and melanoma), benign diseases (such as schwannoma, arteriovenous malformation, cavernous malformation, trigeminal neuralgia, chordoma, Product: TrueBeam™ /TrueBeam STx™/Edge™/VitalBeam Page | 2 {7} varian A Siemens Healthineers Company glomus tumors, and hemangiomas), metastasis (including all parts of the body such as brain, bone, liver, lung, kidney, and skin), pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformation, cavernous malformation, and skull base tumors), and low-dose radiotherapy for adults with medically refractory osteoarthritis. The modification to the Indications for use do not change the Intended Use of the device, as the general purpose of the device and its function remain the same. Additionally, the intended therapeutic use of the device remains the same - "to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated." As demonstrated by the risk management activities per ISO 14971, the differences do not affect the safety and effectiveness of the device when used as labeled. ## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS Both subject device and the predicate device contain the same technological characteristics and functional scientific technology to deliver radiation therapy and stereotactic radiosurgery by authorized medical practitioners. The Intended Use is unchanged. There are no changes in the principle of operation of the device. The biocompatibility of patient-contacting components remains the same as the predicate device. The results of the verification, validation and safety standards testing demonstrates that there are no changes to the safety profile of the device. Both the predicate device and the subject device are based on the same characteristics: Both the subject device and the predicate device configurations use the same medical linear accelerator (linac) to deliver stereotactic radiosurgery and precision radiotherapy and support the same treatment techniques. - Both devices use the same couch and support the same couch motions. - Both devices use the same integrated treatment and imaging consoles and support external beam radiation delivery with X-RAY, Electron, and PHOTON beam for SRT/SRS. - Both devices use the same muti-leaf collimators and support a range of imaging techniques. The significant difference (based on "Deciding When to Submit a 510(k) for a Change to an Existing Device. (October 2017)") compared to the predicate device is the modification of the indications for use. Non-significant changes include defect fixes, cybersecurity improvements, and minor hardware changes. ## PERFORMANCE DATA The following performance data were provided in support of the cumulative changes since K232870. Product: TrueBeam™ /TrueBeam STx™/Edge™/VitalBeam Page | 3 {8} Varian A Siemens Healthineers Company Hardware and software verification and validation testing was conducted according to the FDA Quality System Regulation (21 CFR §820), ISO 13485 quality Management System standard, ISO 14971 Risk Management Standard and the other FDA recognized consensus standards listed below. Test results showed conformance to applicable requirements specifications and assured hazard safeguards functioned properly. Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern. There was no change to patient-contact materials biocompatibility in this medical device. Therefore, no change occurred in conformance to ANSI/AAMI/ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1". There was no change to human factors or usability in the subject device. Therefore, no change occurred in conformance to the standard IEC 62366. There were no substantial technological changes between the subject device and predicate device, and it continues to comply with the following standards. ## STANDARDS Varian TrueBeam, TrueBeam STx, Edge and VitalBeam conforms to the following FDA recognized standards. | IEC 60601-1 3.2 | IEC 62304:2006+A1:2015 | ISO 10993-1:2018 | | --- | --- | --- | | IEC 60601-1-2:2014 | IEC 62366-1:2015+A1:2020 | ISO 15223-1:2021 | | IEC 60601-1-3:2008/A1:2013 | IEC 61217:2011 | ISO 13485:2016 | | IEC 60601-1-6:2010+A1:2013 | IEC 60976:2007 | ISO 14971:2019 | | IEC 60601-2-1:2020 | IEC 62274:2005 | ISO 20417:2021 | | IEC 60601-2-68:2014 | AAMI RT2:2017 | IEC 62368-1:2014/A1 | ## CLINICAL TESTING No animal or clinical tests are being submitted to establish substantial equivalence with the predicate device. ## CONCLUSION TrueBeam, TrueBeam STx, Edge and VitalBeam 4.1 is substantially equivalent to the predicate device. The intended use is the same. The modifications to the indications for use do not change the Intended Use of the device, as the general purpose of the device and its function remain the same. There are no significant changes to the technological characteristics of the subject device. Therefore it is substantially equivalent to the predicate device, and there are no new questions of safety and effectiveness. The results of verification and validation as well as conformance to relevant safety standards demonstrate that TrueBeam, TrueBeam Product: TrueBeam™ /TrueBeam STx™/Edge™/VitalBeam Page | 4 {9} Varian A Siemens Healthineers Company STx, Edge and VitalBeam meets the safety and performance criteria and is substantially equivalent to the predicate device. Product: TrueBeam™ /TrueBeam STx™/Edge™/VitalBeam Page | 5
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...