K142219 · Best Theratronics Limited · IWB · Nov 13, 2014 · Radiology
Device Facts
Record ID
K142219
Device Name
GammaBeam 100
Applicant
Best Theratronics Limited
Product Code
IWB · Radiology
Decision Date
Nov 13, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.5750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
A Cobalt Teletherapy unit is a device by which gamma radiation is delivered for the treatment of cancer under the direction of health care professionals in a radiation therapy clinic.
Device Story
GammaBeam 100 is a cobalt-60 teletherapy unit for cancer treatment. System comprises source head, collimator, gantry, main frame, base, and counterweight. Device delivers gamma radiation to target tissues. Operated by healthcare professionals in radiation therapy clinics. Updated control system replaces obsolete components from predicate; new independent jaws collimator added; covers redesigned. Mechanical structure, radiological shielding, and cobalt-60 source mechanism remain unchanged from predicate. Device provides radiation therapy to patients; updated controls improve usability for contemporary treatment techniques. Safety and effectiveness validated through functional testing in simulated clinical environments.
Clinical Evidence
No clinical data. Bench testing only. Performance was validated against functional specifications in a simulated operating environment to ensure safety and effectiveness equivalent to the predicate device.
Technological Characteristics
Cobalt-60 radionuclide radiation therapy system. Components: source head, gantry, main frame, base, pendulum/beamstopper counterweight, independent jaws collimator. Compliance: IEC 60601-1 (General safety), IEC 60601-1-2 (EMC), EN 60601-2-11 (Gamma beam therapy safety), EN 61217 (Coordinates/movements). Updated control system.
Indications for Use
Indicated for the treatment of cancer in patients requiring radiation therapy, administered by healthcare professionals in a clinical setting.
Regulatory Classification
Identification
A radionuclide radiation therapy system is a device intended to permit an operator to administer gamma radiation therapy, with the radiation source located at a distance from the patient's body. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts (including beam-limiting devices), and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is a stylized image of a caduceus, which is a symbol of medicine.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 13, 2014
Best Theratronics Limited % Mr. Mike de van der Schueren Quality and Regulary Manager 413 March Road Ottawa, Ontario K2K 0E4 CANADA
Re: K142219
Trade/Device Name: Gammabeam 100 Regulation Number: 21 CFR 892.5750 Regulation Name: Radionuclide radiation therapy system Regulatory Class: II Product Code: IWB Dated: October 15, 2014 Received: October 15, 2014
Dear Mr. de van der Schueren:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D. 'O'Hara
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Best Theratronics. The word "Best" is written in a stylized, cursive font in a dark blue color. Below "Best", the word "Theratronics" is written in a sans-serif font in a green color. A registered trademark symbol is located to the right of the word "Best".
# INDICATIONS FOR USE
K142219 510(k) Number:
Device Name: GammaBeam 100
### Indications For Use:
A Cobalt Teletherapy unit is a device by which gamma radiation is delivered for the treatment of cancer under the direction of health care professionals in a radiation therapy clinic.
X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Best Theratronics. The word "Best" is written in a cursive, blue font. Below that, the word "Theratronics" is written in a non-cursive, green font. There is a registered trademark symbol next to the word "Best".
## 510(k) SUMMARY
| Date Summary Prepared | October 14, 2014 |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------|
| Submitted by | Best Theratronics<br>413 March Road<br>Ottawa, Ontario K2K 0E4<br>Canada<br>Tel. (613) 591-2100<br>Fax. (613) 592-5680 |
| Contact Person | Mr. Mike de van der Schueren<br>Quality & Regulatory Manager |
| Trade Name | GammaBeam 100 |
| Common Name | Cobalt Teletherapy Device |
| Classification Name | Radionuclide Radiation Therapy System |
| Legally Marketed Predicate Device | Theratron Phoenix (K863180) |
| Device Classification | Class II, 21 CFR 892.5750 |
| Product Code | IWB |
#### Description of Device
The GammaBeam 100 is a radiotherapy treatment unit with a cobalt-60 radiation delivery system. The device consists of a source head, collimator, gantry, main frame, base, controls and a pendulum or beamstopper style counterweight. The design of this device is substantially equivalent to predicate device Theratron Phoenix.
#### Intended Use of Device
A Cobalt Teletherapy unit is a device by which gamma radiation is delivered for the treatment of cancer under the direction of health care professionals in a radiation therapy clinic.
The indications/intended use of the modified device, as described in the labeling, has not changed as a result of the modifications.
#### Summary of Technological Characteristics
The GammaBeam 100 is substantially equivalent to the predicate device (K863180).
This GammaBeam 100 includes several changes to the design of Phoenix teletherapy unit.
- . an updated control system
- new independent jaws collimator
- . new covers
Changes are made due to obsolescence of many components in the existing Phoenix control system as well as to achieve compliance with current regulatory standards.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for "Best Theratronics". The word "Best" is written in a stylized, cursive font in blue. Below "Best", the word "Theratronics" is written in a sans-serif font in green. There is a registered trademark symbol next to the word "Best".
Usability is being improved by adding functions relevant to contemporary cancer treatment techniques and removing those that are not.
The major components of the Theratron Phoenix, including the head, gantry, main frame, base, and pendulum/beam stopper counterweight have had minor modifications to accommodate the above changes.
There are no changes to the mechanical structure or radiological shielding of the head.
The irradiation source and radioactivity of the cobalt-60 source remains unchanged as does the source drawer mechanism.
The control system has been designed to meet the same intended use as the current model.
### Safety & Effectiveness
The safety of the GammaBeam 100 is equivalent or better than the predicate device.
In terms of safety, the GammaBeam 100 is designed to comply with
- IEC 60601-1-2: Edition 3.0 2007-03 Medical electrical equipment Part 1-2: General . requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests (Appendix Q)
- IEC 60601-1: Edition 3.0 2005 + CORR. 1 (2006) + CORR. 2 (2007) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (Appendix R)
- EN 60601-2-11 (2004) Medical electrical equipment Part 2: Particular requirements for the . safety of gamma beam therapy equipment
- EN 61217 (2011) Radiotherapy equipment Coordinates, movements and scales .
The performance of the device was tested against a set of functional specifications in an environment that simulated, as much as possible, the actual operating environment. Validation testing demonstrated that the device is as safe and effective as the predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.