KODAK DIGITAL SCIENCE 3600 DISTRIBUTED MEDICAL IMAGER, KODAK DIGITAL SCIENCE 1200 DISTRIBUTED MEDICAL IMAGER, MODEL 3600
K983905 · Eastman Kodak Company · LMC · Jan 25, 1999 · Radiology
Device Facts
Record ID
K983905
Device Name
KODAK DIGITAL SCIENCE 3600 DISTRIBUTED MEDICAL IMAGER, KODAK DIGITAL SCIENCE 1200 DISTRIBUTED MEDICAL IMAGER, MODEL 3600
Applicant
Eastman Kodak Company
Product Code
LMC · Radiology
Decision Date
Jan 25, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2040
Device Class
Class 2
Indications for Use
KODAK DIGITAL SCIENCE Distributed Medical Imagers are computer peripheral printing devices used to make hard copy of medical images and reports on paper and/or film.
Device Story
KODAK DIGITAL SCIENCE 3600 and 1200 Distributed Medical Imagers are desktop computer peripheral printing devices; used in referral imaging and distributed medical imaging applications. Devices accept digital medical image data from connected computer systems; transform data into hard copy output via inkjet printing technology. 3600 model supports up to 11x14 inch media with dual supply trays; 1200 model supports up to 8.5x11 inch media with single supply tray. Operated by clinical staff in medical environments. Output provides clinicians with physical copies of medical images and reports for diagnostic review and patient records. Benefits include distributed access to high-resolution medical hard copies.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Inkjet printing technology; IEEE 1284 parallel interface (optional Ethernet/10Base2 RJ45 for 3600). 3600 model: 11x14 inch max media, dual trays, 8-64 MB RAM, optional hard drive. 1200 model: 8.5x11 inch max media, single tray. Print resolution 1200 dpi. Standards: UL, CSA, FCC, EN50082, CISPR 22, VCCI, CE Mark, C-Tick, Energy Star.
Indications for Use
Indicated for use as a computer peripheral printing device to produce hard copies of medical images and reports on paper or film for clinical use.
Regulatory Classification
Identification
A medical image hardcopy device is a device that produces a visible printed record of a medical image and associated identification information. Examples include multiformat cameras and laser printers.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
KODAK EKTASCAN Medical Color Imager 2000PS (K951948)
Related Devices
K011617 — LIFEJET PRINTER 400 · Ferrania S.P.A. Stabilimento DI Ferrania · Jun 8, 2001
K061302 — DIGITAL PRINTER P6000D SERIES, MODEL P6000DU · Mitsubishi Electric Corporation Kyoto Works · Jun 23, 2006
K962364 — CODONICS MEDICAL PRINTER (NP-1600M,NP1600MD) · Codonics, Inc. · Nov 8, 1996
K032635 — DRYSTAR 5300 MEDICAL IMAGE PRINTER, MODEL 5300 · Agfa Corp. · Sep 24, 2003
K011949 — ELK LASER IMAGER, MODEL EL-DRY 4000 · Nishimoto Sangyo Co., Ltd. · Aug 29, 2001
Submission Summary (Full Text)
{0}------------------------------------------------
K983905
JAN 25 1999
Appendix G -- Summary of Safety and Effectiveness
# I. Date Prepared
October 30, 1998
# II. Submitter
Eastman Kodak Company Health Imaging Division 343 State Street Rochester, New York 14650
### III. Contact Person
Judith A. Wallace Regulatory Affairs (716) 724-2314
### IV. Device Names
Trade Names: KODAK DIGITAL SCIENCE 3600 Distributed Medical Imager KODAK DIGITAL SCIENCE 1200 Distributed Medical Imager
Common Names: 3600 DMI; 3600 Imager 1200 DMI; 1200 Imager
### V. Device Classification
FDA has classified the predicate device as Regulatory Class II under CFR 892.1750
#### VI. Predicate Device
KODAK EKTASCAN Medical Color Imager 2000PS ('KEMCI 2000'); 510(k) number K951948
### VII. Description of Devices
KODAK DIGITAL SCIENCE 3600 Distributed Medical Imager is designed for mid to high end desktop Referral Imaging and Distributed Medical Imaging applications printing color or monochrome images on Medical Paper and/or Film Media up to 11x14 inch, with dual media supply trays and optional network interfacing.
KODAK DIGITAL SCIENCE 1200 Distributed Medical Imager is designed for low end to mid range desktop Referral Imaging and Distributed Medical
{1}------------------------------------------------
Imaging applications printing color or monochrome images on Medical paper and/or film up to 8.5x11 inch, with a single media supply tray.
#### Accessories: Media & Inks
- KODAK DIGITAL SCIENCE Distributed Medical Imaging Paper / 8.5x11 in.
- · KODAK DIGITAL SCIENCE Distributed Medical Imaging Paper / A4
- KODAK DIGITAL SCIENCE Distributed Medical Imaging Paper / 4x6 in.
- KODAK DIGITAL SCIENCE Distributed Medical Imaging Film / 8x10 in.
- KODAK DIGITAL SCIENCE Distributed Medical Imaging Film / 11x14 in.
- KODAK DIGITAL SCIENCE Distributed Medical Imaging Cartridge / Black 1
- KODAK DIGITAL SCIENCE Distributed Medical Imaging Cartridge / Black 2
- KODAK DIGITAL SCIENCE Distributed Medical Imaging Cartridge / Color 1
- KODAK DIGITAL SCIENCE Distributed Medical Imaging Cartridge / Color 2
# VIII. Indications for Use
KODAK DIGITAL SCIENCE Distributed Medical Imagers are computer peripheral printing devices used to make hard copy of medical images and reports on paper and/or film.
# IX. Substantial Equivalence:
The purpose and functionality of the KODAK DIGITAL SCIENCE Distributed Medical Imagers is substantially equivalent to the KODAK EKTASCAN Medical Color Imager 2000PS. The basis for the equivalence is that the 3600 and 1200 Imagers, like the KEMCI 2000PS, are computer peripheral printing devices which print monochrome or color images on paper or transparency media. The primary difference is that the 3600 and 1200 Imagers directly deposit black or color inks onto the receiving media to render the image whereas the KEMCI 2000PS uses heat to transfer black or color dyes from a substrate material onto the receiving media.
Although technologically different, the function and output of the SE devices is similar to that of the predicate device and we conclude that the SE devices are as safe and effective as the predicate device.
| Specification | KODAK DIGITAL SCIENCE<br>3600 Distributed Medical<br>Imager | KODAK DIGITAL SCIENCE<br>1200 Distributed Medical<br>Imager | KODAK EKTASCAN<br>Medical Color Imager<br>2000PS |
|-----------------|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|------------------------------------------------------------------------------|
| Printing Method | Inkjet | Inkjet | Thermal Dye Sublimation |
| Interfaces | Parallel: IEEE 1284<br>Optional: Ethernet,<br>10Base2 RJ45 | Parallel: IEEE 1284 | Parallel: IEEE 1284<br>Optional: Ethernet,<br>10Base2 RJ45<br>SCSI (2 ports) |
| Standards | UL, CSA, FCC, EN50082,<br>CISPR 22, VCCI, CE Mark,<br>C-Tick, Energy Star, SETI<br>Mark, SEMKO | UL, CSA, FCC, EN50082,<br>CISPR 22, VCCI, CE Mark,<br>C-Tick, Energy Star | UL, CSA, TUV-GS, FCC,<br>ICES, EN500, VCCI, CE<br>Mark |
| Sheet Sizes | A3 (11.69x16.54 in.)<br>Tabloid (11x17 in.)<br>11x14 in.<br>8x10 in. | 8x10 in.<br>Letter<br>Legal<br>A4 | 8.5x11.0 in.<br>8.5x11.7 in.<br>8.5x14.0 in.<br>9.5x14.0 in. |
Tabular Comparison of Features and Specifications of the Devices:
{2}------------------------------------------------
| | Letter<br>Legal<br>A4<br>Executive<br>B5<br>A5<br>Custom:<br>3x5 in. min.<br>12.6x22 in. max. | Executive<br>B5<br>A5<br>A6 | |
|---------------------|-----------------------------------------------------------------------------------------------|------------------------------------------------------------------|--------------------------------------|
| Print Media | Coated Paper<br>Transparency (blue)<br>Plain Paper<br>Card Stock | Coated Paper<br>Transparency (blue)<br>Plain Paper<br>Card Stock | Coated Paper<br>Transparency (clear) |
| Media Supplies | 1 or 2 | 1 | 1 |
| Monochrome Printing | Yes | Yes | Yes |
| Color Printing | Yes | Yes | Yes |
| Print Resolution | 1200 dots per inch | 1200 dots per inch | 300 pixels per inch |
| Hard Drive | optional | none | none |
| Memory | 8-64 MB RAM | n/a | 48 MB RAM |
$$\begin{array}{c} \text{G}^{\text{'}} \text{3} \\ \text{10/30/98} \end{array}$$
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a bold, serif font. The text is black and the background is white. The text is centered in the image.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 25 1999
Judith A. Wallace Regulatory Affairs Associate Eastman Kodak Company Health Imaging Division 901 Elmgrove Road Rochester. NY 14653
Re:
K983905 Kodak Digital Science 1200 & 3600 Distributed Medical Imagers Dated: October 30, 1998 Received: November 3, 1998 Regulatory class: II 21 CFR 892.2040/Procode: 90 LMC
Dear Ms. Wallace:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled. "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Capt. Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### Facsimile of CDRH's "Indication for Use" Page 3.
510(k) Number (if known): K983905
# Device Name:
KODAK DIGITAL SCIENCE 3600 Distributed Medical Imager KODAK DIGITAL SCIENCE 1200 Distributed Medical Imager
# Indication of Use:
KODAK DIGITAL SCIENCE Distributed Medical Imagers are computer peripheral printing devices used to make hard copy of medical images and reports on paper and/or film.
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT
Concurrence of CDRH, Office of Device Evaluation
and Radiological Devices
510(k) Number K983905
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter ***_***
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.