Machine learning model generated data point must be within the range of the human operator data points on each test image
Full Pass (model generated data points fell within the range of manual operated data points)
18,550 images total: 90% training, 5% validation, 5% test
Human vs AI System Output Validation: Test cases where AI-Assisted Landmark feature output was compared against manual human operator selections
Femoral Offset
Machine learning model generated data point must be within the range of the human operator data points on each test image
Full Pass (model generated data points fell within the range of manual operated data points)
18,550 images total: 90% training, 5% validation, 5% test
Human vs AI System Output Validation: Test cases where AI-Assisted Landmark feature output was compared against manual human operator selections
Total Offset
Machine learning model generated data point must be within the range of the human operator data points on each test image
Full Pass (model generated data points fell within the range of manual operated data points)
18,550 images total: 90% training, 5% validation, 5% test
Human vs AI System Output Validation: Test cases where AI-Assisted Landmark feature output was compared against manual human operator selections
Cup Inclination
Machine learning model generated data point must be within the range of the human operator data points on each test image
Full Pass (model generated data points fell within the range of manual operated data points)
18,550 images total: 90% training, 5% validation, 5% test
Human vs AI System Output Validation: Test cases where AI-Assisted Landmark feature output was compared against manual human operator selections
Cup Anteversion
Machine learning model generated data point must be within the range of the human operator data points on each test image
Full Pass (model generated data points fell within the range of manual operated data points)
18,550 images total: 90% training, 5% validation, 5% test
Human vs AI System Output Validation: Test cases where AI-Assisted Landmark feature output was compared against manual human operator selections
Indications for Use
VELYS™ Hip Navigation is an image-processing software indicated to assist in the positioning of total hip replacement components. It is intended to assist in precisely positioning total hip replacement components intra-operatively by measuring their positions relative to the bone structures of interest provided that the points of interest can be identified from radiology images. VELYS™ Hip Navigation is also indicated for assisting healthcare professionals in preoperative planning and postoperative analysis of orthopaedic surgery in Total Hip Replacement and Total Knee Replacement. The device allows for overlaying of prosthesis templates on radiological images and includes tools for performing measurements on the image and for positioning the template. Clinical judgment and experience are required to properly use the software. The software is not for primary image interpretation. The software is not for use on mobile phones.
Device Story
VELYS™ Hip Navigation (VHN) is a Software as a Medical Device (SaMD) for orthopedic surgery. It processes preoperative and intraoperative radiographic/fluoroscopic pelvic images. The device assists surgeons in preoperative templating and intraoperative positioning of hip replacement components. It features a machine learning (ML) model that automatically suggests default anatomical landmark positions; the surgeon retains full control to confirm or manually adjust these landmarks. The software calculates leg length, femoral offset, total offset, cup inclination, and cup anteversion. Used in physician offices and operating rooms by healthcare professionals. The output provides quantitative data and visual overlays to guide implant placement, supporting clinical decision-making for component alignment. Benefits include improved workflow efficiency and precise intraoperative measurements compared to manual-only methods.
Clinical Evidence
No clinical data was necessary. Bench testing included a Human vs. AI System Output Validation study comparing manual landmark selection to AI-assisted selection across 18,550 images (90% training, 5% validation, 5% test). Results showed AI-generated outputs for leg length, femoral offset, total offset, inclination, and anteversion were within the range of human operator data points. Workflow efficiency testing demonstrated reduced completion time with AI assistance.
Technological Characteristics
SaMD; Windows/iOS platform. ML model for automated landmark detection. Inputs: X-ray/fluoroscopy. Outputs: Anatomical measurements (leg length, offset, inclination, anteversion). Connectivity: Standalone software. Software verification and validation performed per 21 CFR 820.
Indications for Use
Indicated for patients who are candidates for total hip replacement and total knee replacement. Assists healthcare professionals in preoperative planning, intraoperative positioning of total hip replacement components, and postoperative analysis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
K211429 — NAVIPLAN - CT Planning Software for Total Hip Replacement · Naviswiss AG · Oct 5, 2021
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 6, 2026
Depuy Ireland UC
Anuradha Moholkar
Sr. Regulatory Affairs Program Lead
Loughbeg Ringaskiddy Ireland
Re: K253551
Trade/Device Name: VELYS™ Hip Navigation
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH, LLZ
Dated: February 9, 2026
Received: February 9, 2026
Dear Anuradha Moholkar:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253551 - Anuradha Moholkar
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K253551 - Anuradha Moholkar
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Jessica Lamb, Ph.D.
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K253551 | ? |
| Please provide the device trade name(s). | | ? |
| VELYS™ Hip Navigation | | |
| Please provide your Indications for Use below. | | ? |
| VELYS™ Hip Navigation is an image-processing software indicated to assist in the positioning of total hip replacement components. It is intended to assist in precisely positioning total hip replacement components intra-operatively by measuring their positions relative to the bone structures of interest provided that the points of interest can be identified from radiology images.
VELYS™ Hip Navigation is also indicated for assisting healthcare professionals in preoperative planning and postoperative analysis of orthopaedic surgery in Total Hip Replacement and Total Knee Replacement. The device allows for overlaying of prosthesis templates on radiological images and includes tools for performing measurements on the image and for positioning the template. Clinical judgment and experience are required to properly use the software. The software is not for primary image interpretation. The software is not for use on mobile phones. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
DePuy Synthes
K253551
# 510(k) Summary
| Submitter Information | |
| --- | --- |
| Applicant Name | DePuy Ireland UC |
| Applicant Address | Loughbeg Ringaskiddy Ireland |
| Applicant Contact Telephone | +1 574-404-8348 |
| Applicant Contact | Ms. Erin Combs |
| Applicant Contact Email | ecombs@its.jnj.com |
| Correspondent Name | DePuy Ireland UC |
| Correspondent Address | Loughbeg Ringaskiddy Ireland |
| Correspondent Contact Telephone | +1 267-889-5354 |
| Correspondent Contact | Ms. Anuradha Moholkar |
| Correspondent Contact Email | AMoholk1@its.jnj.com |
| Name of Device | |
| Device Trade Name | VELYSTM Hip Navigation |
| Common Name | Medical image management and processing system |
| Classification Name | System, Image Processing, Radiological |
| Regulation Number | 892.2050 |
| Product Code(s) | QIH, LLZ |
| Legally Marketed Predicate Devices | |
| Predicate | K160284 |
| Predicate Trade Name | JointPoint |
| Product Code | LLZ; HAW |
Page 1 of 8
{5}
DePuy Synthes
# Device Description Summary
VELYS™ Hip Navigation (VHN) is a Software as a Medical Device that provides the clinician with intra-operative measurements and visuals of acetabular cup orientation, femoral component leg length and offset calculations, and implant constructs based on user-defined, but machine learning (ML) default-positioned, bony-anatomy landmark points.
VHN includes a machine learning model that places the default position of the landmark based on the output of the model; the user has full control to manipulate the landmark positions after placement. The model inputs the x-ray or fluoroscopy images and outputs a default location for the landmark annotation tool. This machine learning model is Human-in-the-Loop, as the user is expected to position the annotation as they see fit.
# Intended Use/Indications for Use
VELYS™ Hip Navigation is an image-processing software indicated to assist in the positioning of total hip replacement components. It is intended to assist in precisely positioning total hip replacement components intra-operatively by measuring their positions relative to the bone structures of interest provided that the points of interest can be identified from radiology images.
VELYS™ Hip Navigation is also indicated for assisting healthcare professionals in preoperative planning and postoperative analysis of orthopaedic surgery in Total Hip Replacement and Total Knee Replacement. The device allows for overlaying of prosthesis templates on radiological images and includes tools for performing measurements on the image and for positioning the template. Clinical judgment and experience are required to properly use the software. The software is not for primary image interpretation. The software is not for use on mobile phones.
# Technological Comparison
The subject device and the predicate device have similar technological characteristics with respect to principles of operation, system design, and performance. The subject and predicate device achieve the same outcomes, however the main design difference of the subject device includes a machine learning model that places the default position of a set of landmarks, where the user maintains
{6}
DePuy Synthes
control and final confirmation of the landmark annotations. Performance testing on the subject device and comparison to the predicate device support that there are no new questions of safety or effectiveness.
| Characteristics | Predicate Device: JointPoint K160284 | Subject Device: VELYSTM Hip Navigation K253551 | Discussion |
| --- | --- | --- | --- |
| Medical Specialty | Radiology | Radiology | Identical |
| Regulation | 21 CFR 892.2050 Picture archiving and communications system | 21 CFR 892.2050 Picture archiving and communications system | Identical |
| Product Code | LLZ System, image processing, radiological HAW Neurological stereotaxic instrument | QIH Automated radiological image processing software LLZ System, image processing, radiological | Equivalent. Both devices use anatomic landmarks identified from radiology images to assist in positioning total hip replacement components intraoperatively, and to support preoperative planning in total hip and total knee procedures |
| Intended Use | JointPoint is a non-invasive software (Software as a Medical Device) intended to provide preoperative templating for orthopedic procedures and intraoperative data for total hip arthroplasties (THA). | VELYSTM Hip Navigation is a non-invasive software (Software as a Medical Device) intended to provide preoperative templating for orthopedic procedures and intraoperative data for total hip arthroplasties (THA). | Identical |
{7}
DePuy Synthes
| Indications for Use | JointPoint is an image-processing software indicated to assist in the positioning of total hip replacement components. It is intended to assist in precisely positioning total hip replacement components intra-operatively by measuring their positions relative to the bone structures of interest provided that the points of interest can be identified from radiology images. JointPoint is also indicated for assisting healthcare professionals in preoperative planning and postoperative analysis of orthopedic surgery in Total Hip Replacement, Total Knee Replacement, and Intertrochanteric Fracture Reduction. The device allows for overlaying of prosthesis templates on radiological images and includes tools for performing measurements on the image and for positioning the template. Clinical judgment and experience are required to properly use the software. The software is not for primary image interpretation. The software is not for use on mobile phones. | VELYS™ Hip Navigation is an image-processing software indicated to assist in the positioning of total hip replacement components. It is intended to assist in precisely positioning total hip replacement components intra-operatively by measuring their positions relative to the bone structures of interest provided that the points of interest can be identified from radiology images. VELYS Hip Navigation is also indicated for assisting healthcare professionals in preoperative planning and postoperative analysis of orthopedic surgery in Total Hip Replacement and Total Knee Replacement. The device allows for overlaying of prosthesis templates on radiological images and includes tools for performing measurements on the image and for positioning the template. Clinical judgment and experience are required to properly use the software. The software is not for primary image interpretation. The software is not for use on mobile phones. | Identical. The indications for use of the subject device fall within the intended use of the predicate device. |
| --- | --- | --- | --- |
| Anatomical Site and Target Population | Patients who are candidates for orthopedic procedures and total hip replacement | Patients who are candidates for total hip replacement and total knee replacement | Identical. The target population of the subject device is within that of the predicate device. There are no new questions of safety or efficacy. |
| Environment of Use | Physician’s office and operating room | Physician’s office and operating room | Identical |
Page 4 of 8
{8}
DePuy Synthes
| PRINCIPLES OF OPERATION | | | |
| --- | --- | --- | --- |
| Image Inputs | Preoperative and intraoperative radiographic patient images of the pelvis | Preoperative and intraoperative radiographic patient images of the pelvis | Identical |
| Pre-operative Analysis | Preoperative templating:
• Image import
• Image calibration
• User selected templates of implants for THA or TKA
• Digital annotations
• Leg length analysis tool (THA)
• Manual landmark selections | Preoperative templating:
• Image import
• Image calibration
• User selected templates of implants for THA or TKA
• Digital annotations
• Leg length analysis tool (THA)
• Manual landmark selections | Identical |
| Intraoperative Landmark Selection | • User places anatomical landmarks on patient images using digital tools (manual landmark selection)
• Surgeon confirms placement of landmarks | • When enabled, an ML model determines the default landmark position. The user confirms and/or adjusts the landmark position using digital tools. This feature is available in Windows Version only.
• Manual landmark selection is available as an option
• Surgeon confirms placement of landmarks | Equivalent. With the subject device, when enabled, the default landmark location is determined by an ML model. In both the subject device and the predicate device, the user can move the landmark selections using digital tools, and the surgeon must confirm placement of the landmarks. Performance testing did not raise new issues of safety or efficacy. |
| Intraoperative Workflow Outputs | • Software provides leg length and offset data for selected implant constructs and alternatives
• Software provides inclination and anteversion of the trial or final acetabular implant
• Manual landmark selection | • Software provides leg length and offset data for selected implant constructs and alternatives
• Software provides inclination and anteversion of the trial or final acetabular implant
• Manual landmark selection or AI assisted landmarking | Identical |
Page 5 of 8
{9}
DePuy Synthes
| Other Workflows | • Contralateral overlay allows an anterior-posterior (AP) image of the non-operative side to be compared to the operative side.
• OneTrial® Manual landmark selection | • Contralateral overlay allows an anterior-posterior (AP) image of the non-operative side to be compared to the operative side.
• OneTrial® Manual landmark selection | Identical |
| --- | --- | --- | --- |
| Operating System | Windows, iOS | Windows, iOS | Identical. With the subject device, the AI assisted landmark placement is only available in the Windows version. The iOS version of the subject device is identical to the predicate device. |
# Performance Testing Summary
# Machine Learning Model Validation
Data collection: The x-ray and fluoroscopic image data used for model training were extracted from the VELYSTM Hip Navigation software. Therefore, all training data are representative of expected disease state, situation, and demographics in which the clinician determined that a total hip replacement surgery with the assistance of VHN was the appropriate treatment option. Clinical institutions geographically spread across the US were strategically selected in an effort to capture the widest range of patient populations and minimize bias.
Truthing process: Training data was pulled from user data from the production VHN software. Clinicians used the software and annotated the patient images in a clinical setting. Therefore, the reference standard is the clinician's annotations for each image which was pulled with the image.
Page 6 of 8
{10}
DePuy Synthes
Data independence: The deidentified data used for model development consisted of 18,550 images, which was split into groups: 90% for training, 5% for validation, and 5% for test. Validation was conducted through testing on independent datasets, ensuring no overlap between training and testing data.
Model Performance Metrics: The table below summarizes the performance of three AI models evaluated for use within the VHN software. Each model was tested using a standardized dataset to assess its ability to accurately detect and localize relevant features in medical imaging. The metrics reported include:
- mAR (mean Average Recall): Measures how well the model identifies all relevant instances.
- mAP (mean Average Precision): Reflects the model's overall precision in predicting correct locations.
- mAP@0.75: Indicates precision at a higher threshold of localization accuracy.
| Model | mAR | mAP | mAP@0.75 |
| --- | --- | --- | --- |
| OneTrial | 0.9920 | 0.9852 | 0.9900 |
| CupCheckGuidedBilateral | 0.9880 | 0.9798 | 1.0000 |
| CupCheckGuidedUnilateral | 0.9985 | 0.9968 | 1.0 |
# Device Validation
Bench testing was performed on the subject device, VELYS Hip Navigation, in accordance with the product risk analysis and product requirements, in line with 21 CFR Part 820, and to demonstrate substantial equivalence with the predicate device. Where available, FDA-consensus standards and/or guidance documents were used.
Performance testing activity included:
1. Human vs AI System Output Validation
Validation Protocol: Human operators completed the VELYS Hip Navigation workflow for each of the test cases, manually selecting the anatomic landmarks. Output values were recorded for each test case. The AI-Assisted Landmark feature was enabled, and the workflow was completed for each of the test cases. Output values were recorded for each test case.
Page 7 of 8
{11}
DePuy Synthes
Acceptance Criteria: The passing criterion was for each of the 5 respective outputs (leg length, femoral offset, total offset, cup inclination, and cup anteversion), the machine learning model generated data point was within the range of the human operator data points on each of the test images.
Result: The study resulted in a Full Pass that validates the acceptance criteria stating that the machine learning model generated data points fell within the range of manual operated data points for the VELYS Hip Navigation outputs of Leg Length, Femoral Offset, Total Offset, Inclination, and Anteversion.
Workflow Efficiency: When AI-Assisted Landmarks are enabled, Human vs AI System Output Validation testing showed the time to complete the workflows took less time than the manual cases.
2. Usability Equivalency Rationale
OneTrial and Cup Check workflow iteration has been found to be safe and effective for the intended users, uses, and use environments.
DePuy Synthes performed and passed software verification and system validation testing on the device. All software requirements and risk analysis have been successfully verified and traced.
No clinical data was necessary to support the determination of substantial equivalence.
# Conclusions
Both the subject device and predicate device have the same intended use, same indications for use, and similar technological characteristics. Where there are technological differences, performance testing was conducted on the subject device to show that no new questions of safety or effectiveness were raised; therefore, DePuy Synthes concludes that the subject and predicate devices are substantially equivalent.
Page 8 of 8
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.