AccuFFRangio is software intended to be used for performing calculations in X-ray angiographic images of the coronary arteries. AccuFFRangio enables interventional cardiologists to obtain quantifications of one or more lesions in the analyzed coronary vessel segment. In particular, AccuFFRangio provides: Quantitative results of coronary vessel segments based on a 3D reconstructed model; Dimensions of the cardiovascular vessels and lesions; Quantification of the pressure drop in coronary vessels.
Device Story
Stand-alone PC software; processes DICOM X-ray angiographic images. Workflow: image loading; frame selection; 3D vessel reconstruction from two angiographic views; QCA vessel quantification; AccuFFRangio calculation. User inputs aortic pressure; software calculates pressure drop and AccuFFRangio (FFR) value. Used by interventional cardiologists in clinical settings. Output: 3D reconstructed vessel model with color-coded FFR values, vessel dimensions, and stenosis quantification. Assists cardiologists in visualizing lesions and assessing optimal viewing angles for PCI treatment. Provides quantitative imaging output to support clinical assessment; not for independent diagnosis.
Clinical Evidence
Bench testing only. Phantom study using brass models validated 3D QCA parameters (lesion length, diameter/area stenosis, minimal lumen/reference diameter). AccuFFRangio calculation validated using X-ray angiographic datasets with known pressure drops; results demonstrated accuracy, sensitivity, specificity, PPV, and NPV comparable to the predicate device QAngio XA 3D.
Technological Characteristics
Stand-alone software; runs on Windows (7, 8.1, 10) or Linux (Ubuntu 16.04, 18.04). Inputs: DICOM X-ray angiographic images. Processing: 3D vessel reconstruction from two 2D views; centerline/contour detection; QCA; pressure drop/FFR calculation. Output: 3D model visualization, graphs, and PDF reports. Software level of concern: moderate.
Indications for Use
Indicated for use in clinical settings where validated and reproducible quantified results are needed to support the assessment of coronary vessels in X-ray angiographic images, for use on individual patients with coronary artery disease.
Regulatory Classification
Identification
An angiographic x-ray system is a device intended for radiologic visualization of the heart, blood vessels, or lymphatic system during or after injection of a contrast medium. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
K182611 — Qangio XA 3D · Medis Medical Imaging Systems BV · May 30, 2019
K251355 — X1-FFR · Spectrawave, Inc. · Oct 17, 2025
K232147 — CAAS Workstation · Pie Medical Imaging BV · Apr 9, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
September 10, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
ArteryFlow Technology Co., Ltd. % Ashley Fu International Registeration Engineer 459 Oianmo Road, Suite C1-501, Binjiang District Hangzhou. Zhejiang 310051 CHINA
## Re: K210093
Trade/Device Name: AccuFFFRangio Regulation Number: 21 CFR 892.1600 Regulation Name: Angiographic x-ray system Regulatory Class: Class II Product Code: QHA, LLZ Dated: August 12, 2021 Received: August 16, 2021
## Dear Ashley Fu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
, for
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K210093
Device Name
AccuFFRangio
Indications for Use (Describe)
AccuFFRangio is indicated for use in clinical settings where validated and reproducible quantified results are needed to support the assessment of coronary vessels in X-ray angiographic images, for use on individual patients with coronary artery disease.
When the quantified results provided by AccuFFRangio are used in a clinical setting on X-ray images of an individual patient, the results are only intended for use by the responsible clinicians.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Tab #06 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
## 1. Submitter's Information
Submitter: ArteryFlow Technology Co., Ltd.
Address: 459 Qianmo Road, Suite C1-501, Binjiang District, 310051 Hangzhou City, Zhejiang Province, China
Phone Number: +86-571-86772567
Primary correspondent: XIANG Jianping, PhD, General Manager
Email: jianping.xiang@arteryflow.com
Secondary correspondent: Ashley Fu, Registration engineer
Email: fang.fu@arteryflow.com
Date of preparation: January 05, 2021
## 2. Device Information
Trade/ Device Name: AccuFFRangio
Common Name: Radiological Image Processing Software
Regulatory Class: Class II
Regulation Description: Angiographic X-ray system
Regulation number: 892.1600
Classification Product Code: QHA
Subsequent Product Code: LLZ
### 3. Predicate Device Information
Manufacturer: Medis medical imaging systems bv
Device Name: QANGIO XA 3D
Regulatory Class: Class II
Regulation Number: 892.1600
Classification Product Code: QHA
Subsequent Product Code: LLZ
510(k) number: K182611
### 4. Device Description
{4}------------------------------------------------
ArteryFlow® AccuFFRangio is designed as a stand-alone software package to run on a PC. This software can read traditional x-ray angiographic images with DICOM format from the local file directory.
The AccuFFRangio is composed of the following analysis workflows: Image Loading, Frame Selection, Vessel Reconstruction, QCA Vessel Quantification, and AccuFFRangio Calculation for visualization of the target coronary segment, quantification of the stenosis and pressure drop of the coronary seqment. The AccuFFRangio parameter is only for quantitative imaging output but not for diagnosis and the AccuFFRanigo product has a moderate level of concern.
The user can calculate the pressure drop and AccuFFRangio (FFR) value for the coronary vessel. To obtain these values for a specific lesion in a coronary vessel, the user has to start with Frame Selection using two angiographic images from different views. In each of these images, a classic 2D coronary contour detection is performed, after which a reconstruction of the coronary segment is obtained in 3D space. Based on the 3D reconstruction and user input of the aortic pressure, the pressure drop and AccuFFRangio value can be calculated.
AccuFFRangio enables interventional cardiologists to obtain accurate anatomical quantifications of one or more lesions in the analyzed coronary segment, and to assess the best viewing angles which can be helpful for optimal visualization of the lesion during percutaneous coronary intervention (PCI) treatment.
Results can be displayed and generated by the software, which contains patient information, imaging of actual and reference vessel boundaries, dimensions of the vessel sizing, pressure drop, and AccuFFRangio value. The results can be export in PDF format. This functionality is independent of the type of vendor acquisition equipment.
#### 5. Intended Use
AccuFFRangio is software intended to be used for performing calculations in X-ray angiographic images of the coronary arteries. AccuFFRangio enables interventional cardiologists to obtain quantifications of one or more lesions in the analyzed coronary vessel segment. In particular, AccuFFRangio provides:
- Quantitative results of coronary vessel segments based on a 3D reconstructed model;
- Dimensions of the cardiovascular vessels and lesions;
- Quantification of the pressure drop in coronary vessels.
#### 6. Indication for Use
AccuFFRangio is indicated for use in clinical settings where validated and reproducible quantified results are needed to support the assessment of coronary vessels in X-ray angiographic images, for use on individual patients with coronary artery disease.
When the quantified results provided by AccuFFRangio are used in a clinical setting on X-ray images of an individual patient, the results are only intended for use by the responsible clinicians.
{5}------------------------------------------------
# 7. Technological Characteristic Comparison
A comparison of the technological characteristics of the predicate and subject device is given in the table below.
| Item | New device | Predicate device |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | AccuFFRangio | QANGIO XA 3D |
| Manufacturer | ArteryFlow Technology Co., Ltd. | Medis medical imaging systems bv |
| 510(k) No. | - | K182611 |
| Product Code | QHA, LLZ | QHA, LLZ |
| Regulation No. | 892.1600 | 892.1600 |
| Class | II | II |
| Level of Concern<br>of the software | Moderate | Moderate |
| Intended Use and Indications for Use | | |
| Intended Use | AccuFFRangio is software<br>intended to be used for performing<br>calculations in X-ray angiographic<br>images of the coronary arteries.<br>AccuFFRangio enables<br>interventional cardiologists to<br>obtain quantifications of one or<br>more lesions in the analyzed<br>coronary vessel segment. In<br>particular, AccuFFRangio<br>provides:<br>Quantitative results of<br>coronary vessel segments<br>based on a 3D reconstructed<br>model;Dimensions of the<br>cardiovascular vessels and<br>lesions;Quantification of the pressure<br>drop in coronary vessels. | QANGIO XA 3D is software<br>intended to be used for performing<br>calculations in X-ray angiographic<br>images of the coronary arteries.<br>QANGIO XA 3D enables<br>interventional cardiologists and<br>researchers to obtain<br>quantifications of one or more<br>lesions in the analyzed coronary<br>vessel segment. In particular,<br>QANGIO XA 3D provides:<br>Quantitative results of<br>coronary vessel segments<br>based on a 3D reconstructed<br>model;Dimensions of the<br>cardiovascular vessels and<br>lesions;Quantification of the pressure<br>drop in coronary vessels. |
| Indication for use | AccuFFRangio is indicated for use<br>in clinical settings where validated<br>and reproducible quantified results<br>are needed to support the<br>assessment of coronary vessels in<br>X-ray angiographic images, for use<br>on individual patients with<br>coronary artery disease | QANGIO XA 3D is indicated for<br>use in clinical settings where<br>validated and reproducible<br>quantified results are needed to<br>support the assessment of<br>coronary vessels in X-ray<br>angiographic images, for use on<br>individual patients with coronary |
| | | |
| | When the quantified results<br>provided by AccuFFRangio are<br>used in a clinical setting on X-ray<br>images of an individual patient, the<br>results are only intended for use by<br>the responsible clinicians. | artery disease.<br>When the quantified results<br>provided by QANGIO XA 3D are<br>used in a clinical setting on X-ray<br>images of an individual patient, the<br>results are only intended for use by<br>the responsible clinicians. |
| Technological Characteristics | | |
| Data type | ● X-ray angiographic data in<br>DICOM format<br>(vendor-independent) | ● X-ray angiographic data in<br>DICOM format<br>(vendor-independent) |
| Import of patient<br>data | ● Automatically load information<br>from patient's DICOM file;<br>● Manual through keyboard | ● Automatically load information<br>from patient's DICOM file;<br>● Manual through keyboard |
| Image display | ● 2D X-ray image visualization<br>with embedded ECG signal;<br>● 3D reconstruction of the<br>vessel segment based on 2<br>X-ray images;<br>● Graph for actual and<br>reference vessel diameter<br>with vessel longitudinal<br>position;<br>● Graph for AccuFFRangio<br>(FFR) value with vessel<br>longitudinal position;<br>● AccuFFRangio (FFR) value is<br>colored on the 3D<br>reconstructed vessel segment<br>model. | ● 2D X-ray image visualization<br>with embedded ECG signal;<br>● 3D reconstruction of the<br>vessel segment based on 2<br>X-ray images;<br>● Graph for actual and<br>reference vessel diameter<br>with vessel longitudinal<br>position;<br>● Graph for QFR (FFR) value<br>with vessel longitudinal<br>position;<br>● QFR (FFR) value is colored<br>on the 3D reconstructed<br>vessel segment model. |
| Centerline and<br>contour definition | ● Manual and semi-automatic<br>centerline definition based<br>contour detection of coronary<br>vessel;<br>● Contour correction and<br>restriction | ● Manual and semi-automatic<br>centerline definition based<br>contour detection of coronary<br>and peripheral vessel;<br>● Contour correction and<br>restriction |
| Image assessment | ● Manual and automatic<br>calibration;<br>● Vessel dimensions<br>(diameters, areas, lengths);<br>● Automatic stenosis<br>assessment;<br>● Pressure drop calculation in<br>coronary vessel | ● Manual and automatic<br>calibration;<br>● Vessel dimensions<br>(diameters, areas, lengths);<br>● Automatic stenosis<br>assessment;<br>● Pressure drop calculation in<br>coronary vessel |
| | | |
| FFR calculation | AccuFFRangio provides a<br>quantification parameter called<br>AccuFFRangio. | QANGIO XA 3D provides a<br>quantification parameter called<br>QFR. |
| Storage of results | ● Printout<br>● Images<br>● PDF | ● Printout<br>● Images<br>● PDF |
| Different technological Characteristics | | |
| Gap 1 | With ECG display in the image<br>pane | ECG display in a separate pane |
| Gap 2 | The vessel is colored by stenosis<br>percentage in red to blue | The vessel is colored by stenosis<br>percentage in red to white |
| Gap 3 | When visualizing the reconstructed<br>vessel segment, there is a body<br>icon to show the relative position<br>for the vessel segment. | In QAngio XA 3D, there is a box<br>icon to show the relative position<br>for the vessel segment. |
| | Image: vessel segment | Image: vessel segment |
| Gap 4 | In the calculated vessel FFR<br>distribution, users can easily point<br>the position by the curser and the<br>specific AccuFFRangio value will<br>show up. | In QAngio XA 3D, the user needs<br>to move four "checking disk" and<br>look at the FFR value in the<br>left-bottom window. |
| Gap 5 | Software operation environment:<br>● Microsoft Windows 7, SP1,<br>32 and 64-bit version<br>● Microsoft Windows 8.1, 32<br>and 64-bit version<br>● Microsoft Windows 10, 32<br>and 64-bit version<br>● Linux Ubuntu16.04 and<br>18.04, 32 and 64-bit version | Software operation environment:<br>● Microsoft Windows 7, SP1,<br>32 and 64-bit version<br>● Microsoft Windows 8.1, 32<br>and 64-bit version<br>● Microsoft Windows 10, 32<br>and 64-bit version<br>● Microsoft Windows Server<br>2008 R2, SP1, 64-bit version<br>● Microsoft Windows Server<br>2012 R2, 64-bit version<br>● Microsoft Windows Server<br>2016, 64-bit version |
# Table 1 General Comparison
{6}------------------------------------------------
{7}------------------------------------------------
AccuFFRangio has the same intended use and indications for use as the predicate devices QANGIO XA 3D (K182611). The basic features in technological characteristics are the same, such as same types of data format, same method for importing patient data, similar
{8}------------------------------------------------
display of image, same definition for centerline and contour, same method for Image assessment, same output for 3D reconstruction and the calculation of the pressure drop, etc.
AccuFFRanqio is a quantitative imaging output device, like the QANGIO XA 3D. Though above five differences concerning the display pattern (gap 1), display interface (gap 2-3), specific operate method on showing result (gap 4) and software operating system (gap 5) do exist, they do not affect the function and safety of AccuFFRangio.
#### 8. Performance Data
System requirements – derived from the intended use and indications for use – as well as risk control measures are verified by system testing. All requirements are tested and all results of the tests performed are summarized in the software test statement and especially the requirements coverage matrix of AccuFFRangio providing traceability between requirements and the tests successfully executed.
The main functionality (including 3D vessel reconstruction, the calculation of the pressure drop and AccuFFRangio) have all been validated:
- 3D vessel reconstruction: A phantom study had been implemented by using three different types stenosis of brass model. This phantom test was to evaluate the accuracy of the 3D quantitative coronary angiography (QCA) of AccuFFRangio software. The following software technical parameters were validated, namely, lesion length accuracy, diameter stenosis accuracy, area stenosis accuracy, minimal lumen diameter accuracy and reference diameter accuracy. This demonstrated that the quantitative coronary angiography for coronary vessel have similar performance compare to the predicate device QAngio XA 3D (K182611).
- AccuFFRanqio calculation: To confirm the performance of the pressure drop and AccuFFRangio (FFR) calculation in the coronary vessel, a series of X-ray angiographic dataset with known pressure drops were analyzed. The functionality of pressure drop and AccuFFRangio calculation for the new device is based on 3D reconstruction out of two angiographic images. The results, including accuracy, sensitivity, specificity, positive predictive value and negative predictive value for per-vessel were calculated. The results demonstrated that the quantification of AccuFFRangio in coronary vessel have similar performance compare to the QFR by the predicate device QAngio XA 3D (K182611).
#### 9. Conclusion
AccuFFRangio product manufactured by our company shares the same intended use, and indications for use with QAnqio XA 3D by Medis.
Testing, validation and verification have produced results consistent with design input requirements. AccuFFRangio is a software-only device for which there are no applicable mandatory performance standards.
During the development, potential hazards were controlled by a risk management report, including risk analysis, risk mitigation, verification and validation.
ArteryFlow technology concludes that the AccuFFRangio is only for quantitative imaging
{9}------------------------------------------------
output but not for diagnosis, and it is a safe and effective medical device as its predicate device. The use of AccuFFRangio does not change the intended use of X-ray angiography acquisition, nor does the use of this software results in any new potential hazards.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.