The algorithm was trained during software development on images of the pathology; each image in the training dataset was tagged based on the presence of the critical finding.
Retrospective, blinded, multicenter pivotal study: 509 cases from 5 US-based clinical sites. The cases were distinct in time or center from the cases used to train the algorithm.
Indications for Use
BriefCase-Triage is a radiological computer aided triage and notification software indicated for use in the analysis of CT chest, abdomen, or chest/abdomen exams with contrast (CTA and CT with contrast) in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive findings of Aortic Dissection (AD) pathology. BriefCase-Triage uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device. The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/ prioritization.
Device Story
BriefCase-Triage is a radiological computer-aided triage and notification software; operates on a Linux-based server in a cloud environment. Input: DICOM CT images (chest, abdomen, or chest/abdomen with contrast). Processing: Deep-learning AI algorithm analyzes images for suspected Aortic Dissection (AD). Output: Pop-up notification on a standalone desktop application with a compressed, low-quality, unannotated grayscale preview image. Usage: Hospital networks/specialists; operates in parallel to standard-of-care (SOC) workflow. The device does not alter original images, remove cases from the SOC queue, or provide diagnostic markings. Benefit: Facilitates efficient triage by alerting clinicians to suspected critical cases, potentially reducing turnaround time for diagnosis. Clinicians use the notification to prioritize the review of original images within their PACS.
Clinical Evidence
Retrospective, blinded, multicenter study (5 US sites, 509 cases). Primary endpoints: sensitivity and specificity (goal >80%). Results: Sensitivity 92.7% (95% CI: 88.2%-95.8%); Specificity 92.8% (95% CI: 89.2%-95.4%). Secondary endpoint: Time-to-notification (mean 10.7s vs 38.0s for predicate). NPV 99.8%, PPV 24.7%. Performance goals achieved for default and four additional operating points.
Technological Characteristics
Software-based deep learning AI algorithm; runs on Linux-based server in cloud environment. Inputs: DICOM CT images (0.5-5.0 mm slice thickness). Outputs: Desktop application notifications with compressed, low-quality, unannotated grayscale preview images. Connectivity: Integrated with image routing module via image communication platform (ICP).
Indications for Use
Indicated for adults and transitional adolescents aged 18+ undergoing CT chest, abdomen, or chest/abdomen exams with contrast (CTA/CT with contrast) to assist hospital networks and specialists in triage/prioritization of suspected Aortic Dissection (AD) pathology.
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
K230020 — BriefCase · Aidoc Medical , Ltd. · Feb 1, 2023
K250248 — BriefCase-Triage · Aidoc Medical , Ltd. · Feb 14, 2025
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 30, 2025
Aidoc Medical, Ltd.
% John Smith
Partner
Hogan Lovells US LLP
555 Thirteenth Street NW
Washington, District of Columbia 20004
Re: K251406
Trade/Device Name: BriefCase-Triage
Regulation Number: 21 CFR 892.2080
Regulation Name: Radiological computer aided triage and notification software
Regulatory Class: Class II
Product Code: QAS
Dated: May 6, 2025
Received: May 6, 2025
Dear John Smith:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251406 - John Smith
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251406 - John Smith
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Jessica Lamb, Ph.D.
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251406 | |
| Device Name BriefCase-Triage | |
| Indications for Use (Describe) BriefCase-Triage is a radiological computer aided triage and notification software indicated for use in the analysis of CT chest, abdomen, or chest/abdomen exams with contrast (CTA and CT with contrast) in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive findings of Aortic Dissection (AD) pathology. BriefCase-Triage uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device. The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/ prioritization. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
aidoc
K251406
510(k) Summary
Aidoc Medical, Ltd.'s Briefcase-Triage
Submitter:
Aidoc Medical, Ltd.
3 Aminadav St.
Tel-Aviv, Israel
Phone: +972-73-7946870
Contact Person: Amalia Schreier, LL.M
Date Prepared: May 28, 2025
Name of Device: Briefcase-Triage
Classification Name: Radiological computer-assisted triage and notification software device (21 CFR 892.2080)
Regulatory Class: Class II
Product Code: QAS
Primary Predicate Device: Briefcase-Triage for AD (K222329)
Reference Device: BriefCase-Triage for RibFx (K243548)
Device Description
Briefcase-Triage is a radiological computer-assisted triage and notification software device.
The software is based on an algorithm programmed component and is intended to run on a linux-based server in a cloud environment.
The Briefcase-Triage receives filtered DICOM Images, and processes them chronologically by running the algorithms on each series to detect suspected cases. Following the AI processing, the output of the algorithm analysis is transferred to an image review software (desktop application). When a suspected case is detected, the user receives a pop-up notification and is presented with a compressed, low-quality, grayscale image that is captioned "not for diagnostic use, for prioritization only" which is displayed as a preview function. This preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification.
Presenting the users with worklist prioritization facilitates efficient triage by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone.
{5}
The algorithm was trained during software development on images of the pathology. As is customary in the field of machine learning, deep learning algorithm development consisted of training on labeled ("tagged") images. In that process, each image in the training dataset was tagged based on the presence of the critical finding.
## Intended Use / Indications for Use
BriefCase-Triage is a radiological computer aided triage and notification software indicated for use in the analysis of CT chest, abdomen, or chest/abdomen exams with contrast (CTA and CT with contrast) in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive findings of Aortic Dissection (AD) pathology.
BriefCase-Triage uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.
The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/ prioritization.
## Summary of Technological Characteristics
The subject Briefcase-Triage for AD and the algorithm analysis module for the primary predicate Briefcase-Triage for AD (K222329) are identical in most aspects and differ mostly with respect to changes in algorithm training process and their algorithm performance.
Both the primary predicate and subject devices are radiological computer-aided triage and notification software programs. Both devices are artificial intelligence, deep-learning algorithms incorporated in software components for use with DICOM compliant CT scanners, PACS, and radiology workstations.
Both devices are intended to aid in triage and prioritization of radiological images and utilize the same design of deep learning algorithm trained on medical images. Both devices are intended to provide the specialists with notifications and unannotated, compressed, low-quality, and grayscale preview images of suspect studies for the purpose of preemptive triage.
The subject and predicate Briefcase-Triage devices raise the same types of safety and effectiveness questions, namely, accurate triage of findings within the processed study. It is important to note that, like the predicate, the subject device neither removes cases from the standard of care reading queue nor de-prioritized cases. Both devices operate in parallel with the standard of care, which remains the default option for all cases.
A table comparing the key features of the subject and the primary predicate devices is provided
{6}
below.
Table 1. Key Feature Comparison
| | Subject Device Aidoc Briefcase-Triage for AD | Predicate Device Aidoc Briefcase-Triage for AD (K222329) |
| --- | --- | --- |
| Intended Use / Indications for Use | BriefCase-Triage is a radiological computer aided triage and notification software indicated for use in the analysis of CT chest, abdomen, or chest/abdomen with contrast (CTA and CT with contrast) in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive findings of Aortic Dissection (AD) pathology. BriefCase-Triage uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device. The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/ prioritization. | BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of CT exams with contrast (CTA and CT with contrast) that include the chest in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive findings of Aortic Dissection (AD) pathology. BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected findings. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device. The results of BriefCase are intended to be used in conjunction with other patient information and based on their professional judgment, to assist with triage/ prioritization. |
{7}
| | Subject Device
Aidoc Briefcase-Triage for AD | Predicate Device
Aidoc Briefcase-Triage for AD
(K222329) |
| --- | --- | --- |
| User population | Hospital networks and appropriately trained medical specialists | Hospital networks and appropriately trained medical specialists |
| Anatomical region of interest | Chest, abdomen and thoraco-abdominal | Chest, abdomen and thoraco-abdominal |
| Data acquisition protocol | CT chest, abdomen, or chest/abdomen exams with contrast (CTA and CT with contrast) | CT exams with contrast (CTA and CT with contrast) that include the chest |
| Notification-only (/notification alerts), parallel workflow tool | Yes | Yes |
| Images format | DICOM | DICOM |
| Interference with standard workflow | No. No cases are removed from desktop app or deprioritized | No. No cases are removed from desktop app or deprioritized |
{8}
| | Subject Device
Aidoc Briefcase-Triage for AD | Predicate Device
Aidoc Briefcase-Triage for AD
(K222329) |
| --- | --- | --- |
| Inclusion/
Exclusion
criteria for
clinical
performan
ce testing | **Inclusion Criteria**
• Scans performed on adults/transitional adolescents ≥ 18 years of age.
• CT chest, abdomen, or chest/abdomen exams with contrast (CTA and CT with contrast)
• Slice thickness 0.5 mm - 5.0 mm
**Exclusion Criteria**
• All studies that have an inadequate field of view | **Inclusion Criteria**
• Scans performed on adults/transitional adolescents ≥ 18 years of age.
• CT exams with contrast (CTA and CT with contrast) that include the chest
• Slice thickness 0.5 mm - 5.0 mm
**Exclusion Criteria**
• All studies that are technically inadequate, such as severe metal artifacts, or inadequate field of view. |
| Additional Operating Points | 4 Additional Operating Points | 2 Additional Operating Points |
| Algorithm | Artificial intelligence algorithm with database of images. | Artificial intelligence algorithm with database of images. |
| Structure | - Integrated with image routing module via image communication platform (ICP) (image acquisition).
- Algorithm module (image processing)
- Integrated with desktop application for workflow integration (feed and non-diagnostic Image Viewer). | - AHS module (image acquisition);
- ACS module (image processing);
- Aidoc Desktop Application for workflow integration (Feed/Worklist (alternate names) and non-diagnostic Image Viewer). |
## Performance Data
### Pivotal Study Summary
Aidoc conducted a retrospective, blinded, multicenter study with the Briefcase-Triage software to evaluate the software's performance in identifying CT chest, abdomen, or chest/abdomen exams
{9}
with contrast (CTA and CT with contrast) in 509 cases from 5 US-based clinical sites. The study compared the software's performance to the ground truth, as determined by three senior board-certified radiologists. The cases collected for the pivotal dataset were all distinct in time or center from the cases used to train the algorithm, as was used for the most recent clearance (K222329). Test pivotal study data was sequestered from algorithm development activities, and use of the data is managed by appropriate Quality Management System procedures.
Primary endpoints were sensitivity and specificity with an 80% performance goal. Secondary endpoints were Briefcase-Triage time-to-notification compared to the predicate device. Positive Predictive Value (PPV), Negative Predictive Value (NPV), Positive Likelihood Ratio (PLR), and Negative Likelihood Ratio (NLR) were also assessed.
## Primary Endpoint
Sensitivity was 92.7% (95% CI: 88.2%, 95.8%) and Specificity was 92.8% (95% CI: 89.2%, 95.4%). Because the lower bound of each 95% CI exceeded 80%, the study's primary endpoints were achieved.
## Secondary Endpoint
In addition, the time-to-notification metric observed for the Briefcase-Triage software, when integrated with a compatible image communication platform, was compared to the equivalent metric of the predicate devices. The Briefcase-Triage time-to-notification includes the time to get the DICOM exam, de-identify it, upload it to the cloud, analyze and send a notification on a positive suspect case back to the desktop application.
The Briefcase-Triage time-to-notification was measured for all True Positive cases (i.e., identified as positive both by the reviewers as well as the Briefcase-Triage device) and is given in Table 2 below. The Table also displays the same metric reported for the predicate Briefcase-Triage AD.
The time-to-notification results obtained for the subject Briefcase-Triage device show comparability with the primary predicate with regard to time savings to the standard of care review. The Briefcase-Triage mean time-to-notification for the subject AD triage was 10.7 seconds (95% CI: 10.5-10.9). The time-to-notification for the predicate AD was 38.0 seconds (95% CI: 35.5-40.4).
Table 2. Time-to- notification comparison for Briefcase-Triage devices (Seconds)
{10}
| Time -to-notification | Mean Estimate (seconds) | N | 95% Lower CL | 95% Upper CL | Median |
| --- | --- | --- | --- | --- | --- |
| Predicate K222329 Processing Time | 38.0 | 499 | 35.5 | 40.4 | 31.1 |
| Briefcase-Triage + Image Communication Platform Time-To-Notification | 10.7 | 212 | 10.5 | 10.9 | 10.4 |
NPV was 99.8% (95% CI: 99.7%- 99.9%) and PPV was 24.7% (95% CI: 18.0%- 33.0%).
PLR was 12.8 (95% CI: 8.5- 19.2) and NLR was 0.08 (95% CI: 0.05- 0.13).
Thus, the reported similar time-to-notification data demonstrates that when using the subject Briefcase-Triage for AD the clinician may have the same benefit in time saving as with the predicate Briefcase-Triage for AD.
As can be seen in Table 3 the mean age of patients whose scans were reviewed for AD was 60.1 years, with a standard deviation of 17.3 years. Gender distribution was 50.9% male, and 49.1% female (Table 4). Scanner distribution can also be found in Table 5 below.
Table 3. Descriptive Statistics for Age
| | Mean | Std | Min | Median | Max | N |
| --- | --- | --- | --- | --- | --- | --- |
| Age (Years) | 60.1 | 17.3 | 18 | 62 | 90 | 212 |
{11}
Table 4. Frequency Distribution of Gender
| Ground Truth Results | Gender | | | | All | |
| --- | --- | --- | --- | --- | --- | --- |
| | Female | | Male | | | |
| | N | % | N | % | N | % |
| Positive | 73 | 14.3% | 132 | 25.9% | 205 | 40.3% |
| Negative | 177 | 34.8% | 127 | 25.0% | 304 | 59.7% |
| All | 250 | 49.1% | 259 | 50.9% | 509 | 100.0% |
Table 5. Frequency Distribution of Manufacturer
| Manufacturer | N | % |
| --- | --- | --- |
| GE MEDICAL SYSTEMS | 141 | 27.7% |
| Philips | 112 | 22% |
| SIEMENS | 174 | 34.2% |
| TOSHIBA | 82 | 16.1% |
| Total | 509 | 100% |
Clinical Subgroups and Confounders:
Pathologies present in negative cases: Inflammatory; Oncology; Trauma; Heart & vascular; Chronic diseases; None of the above and Fully negative.
Additional Operating Points:
In addition to the default operating point that was selected to maximize both sensitivity and specificity, four additional operating points (AOP) were selected to maximize specificity or sensitivity
{12}
while maintaining a lower bound $95\%$ confidence interval of $80\%$ for sensitivity and specificity respectively:
Table 6. Additional Operating Points
| Operating Points | Sensitivity % (95% CI) | Specificity % (95% CI) |
| --- | --- | --- |
| AOP1 | 95.6% (95% CI: 91.8%-98.0%) | 88.2% (95% CI: 84.0%-91.6%) |
| AOP2 | 94.1% (95% CI: 90.0%-96.9%) | 89.8% (95% CI: 85.8%-93.0%) |
| AOP3 | 89.3% (95% CI: 84.2%-93.2%) | 94.7% (95% CI: 91.6%-97.0%) |
| AOP4 | 86.3% (95% CI: 80.9%-90.7%) | 97.7% (95% CI: 95.3%-99.1%) |
In summary, performance goals were achieved for the default and four additional operating points. Combined with the comparison results of time-to-notification metric with the predicate device, these data establish the achievement by the subject Briefcase-Triage of preemptive triage in the range of several minutes.
# Conclusions
The subject Briefcase-Triage for AD and the predicate Briefcase-Triage for AD (K222329) are intended to aid in prioritization and triage of radiological images for the indications for suspected positive findings of incidental pulmonary embolism pathologies. Both devices are software components consisting of deep learning AI algorithms that process images and produce analysis results, which are displayed to the user by a prioritization alert and a compressed, low-quality, grayscale, unannotated preview image. In both devices, the labeling clearly states that the devices are not for diagnostic use and instructs the user to further evaluate and diagnose based only on the original images in the local PACS.
Both devices operate in parallel to the standard of care workflow in the sense that they do not change the original image, do not provide any marking on the output preview, do not remove images from the standard of care FIFO queue and do not de-prioritize cases, thus not disturbing standard interpretation of the images. Both devices notify the radiologist of time-sensitive critical cases within the range of several minutes, and thus contribute similarly to the standard of care workflow turnaround time reduction through preemptive triage.
The subject Briefcase-Triage device for AD is thus substantially equivalent to the primary predicate Briefcase-Triage for AD (K222329).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.