GUS ASTRA TEE Transesophageal Probe Reprocessor, GUS ASTRA VR Endovaginal/Endorectal Probe Reprocessor

K150504 · Pci Medical, Inc. · PSW · Nov 16, 2015 · Radiology

Device Facts

Record IDK150504
Device NameGUS ASTRA TEE Transesophageal Probe Reprocessor, GUS ASTRA VR Endovaginal/Endorectal Probe Reprocessor
ApplicantPci Medical, Inc.
Product CodePSW · Radiology
Decision DateNov 16, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1570
Device ClassClass 2

Indications for Use

The GUS ASTRA TEE™ automated reprocessor facilitates high-level disinfection and rinsing of 1 or 2 transesophageal ultrasound probes using FDA cleared and PCI Medical approved high-level liquid disinfectants. The GUS ASTRA VR™ automated reprocessor facilitates high-level disinfection and rinsing of 1 or 2 endovaginal and/or endorectal ultrasound probes using FDA cleared and PCI Medical approved high-level liquid disinfectants.

Device Story

ASTRA Product Family (ASTRA TEE and ASTRA VR) are floor-standing electromechanical reprocessors for enzymatically pre-cleaned ultrasound probes. Devices control high-level liquid disinfectant (HLD) temperature, soak time, and rinse cycles. Operator scans HLD barcode and probe ID; system loads preset parameters. Sensors monitor HLD/rinse levels, temperature, and probe presence. Door sensor initiates cycle. Output includes automated disinfection and rinsing; data retrievable via USB. Used in clinical settings by trained personnel. Benefits include standardized, automated disinfection process with tracking of critical parameters, improving disinfection outcomes compared to manual methods.

Clinical Evidence

Bench testing only. Simulated use testing demonstrated ≥6 Log reduction of M. terrae. In-use testing confirmed complete kill. Residual testing per ISO 10993-5 showed cytotoxic effect ≤ 2. Electrical safety and EMC testing passed per UL 61010-1/CSA C22.2 and EN 61326-1.

Technological Characteristics

Floor-standing electromechanical unit. Materials: PVC, polyethylene, polypropylene, EPDM, stainless steel, polycarbonate, polyurethane, polysulfone, ABS, silicone, silver, polyester. Sensing: temperature sensors, microswitches for probe detection, optical sensors for HLD/rinse levels. Energy: electrical. Connectivity: USB port for data retrieval. Software: microprocessor-controlled, fixed parameters based on HLD. Sterilization: high-level liquid disinfection. Standards: UL 61010-1, IEC 62304, ISO 10993-5.

Indications for Use

Indicated for high-level disinfection and rinsing of 1 or 2 transesophageal, endovaginal, or endorectal ultrasound probes in non-sterile healthcare settings by trained personnel.

Regulatory Classification

Identification

A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wing-like shapes extending from its body. The bird is positioned above the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the bird. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 30, 2017 PCI Medical, Inc. Mr. Kevin Mader Director of Quality and Regulatory 6 Winter Avenue Deep River, CT 06417 Re: K150504 Trade/Device Name: GUS ASTRA TEE™ Transesophageal Probe Reprocessor GUS ASTRA VRTM Endovaginal/Endorectal Probe Reprocessor Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic Ultrasonic Transducer Regulatory Class: Class II Product Code: PSW Dated: October 16, 2015 Received: October 16, 2015 Dear Mr. Kevin Mader: This letter corrects our substantially equivalent letter of November 16, 2015. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ # Page 2 - Mr. Kevin Mader Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. Tina Kiang -S Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K150504 Device Name GUS ASTRA TEE™ Transesophageal Probe Reprocessor Indications for Use (Describe) The GUS ASTRA TEE™ automated reprocessor facilitates high-level disinfection and rinsing of 1 or 2 transsophageal ultrasound probes using FDA cleared and PCI Medical approved high-level liquid disinfectants. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Indications for Use 510{k} Number (if known) K | 50504 Device Name GUS ASTRA VR™ Endovaginal/Endorectal Probe Reprocessor Indications for Use (Describe) The GUS ASTRA VRTM automated reprocessor facilitates high-level disinfortion and rinsing of 1 or 2 endovaginal and/or endorectal ultrasound probes using FDA cleared and PCI Medical approved high-level liquid disinfectants. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) 区 Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows a logo for PCI Medical. The logo consists of the letters "PCI" stacked on top of the word "MEDICAL". A plus sign is located above the letter "I" in "PCI". The text and symbol are in black and white. # 510(k) SUMMARY | Submitter | PCI Medical, Inc.<br>6 Winter Avenue<br>Deep River, CT 06417 | |-------------------------|------------------------------------------------------------------------------------------------------------------------------| | Contact Person | Kevin Mader<br>Director of Quality & Regulatory<br>k.mader@pcimedical.com<br>(800) 862-3394 ext. 247<br>(860) 526-3081 (fax) | | Date Prepared | November 13th 2015 | | Trade Name(Common Name) | | | Regulation Name | Diagnostic Ultrasonic Transducer, 21 CFR 892.1570 | | Classification | Class II | | Product Code | ITX | | Predicate Device(s) | CS MEDICAL TD-100 Transesophageal Probe Disinfector<br>K051305 | Intended Use cilitates high-level disinfection and rinsing of 1 or 2 transesophageal ultrasound probes using FDA cleared and PCI Medical approved high-level liquid disinfectants. ates high-level disinfection and rinsing of 1 or 2 endovaginal/endorectal ultrasound probes using FDA cleared and PCI Medical approved high-level liquid disinfectants. The intended use statements use similar language to the predicate TD-100 device (K051305). The differences in wording used does not raise new questions of safety or efficacy. {5}------------------------------------------------ # Device Description The ASTRA Product Family of automated reprocessors are floor standing electromechanical lab equipment that facilitate high-level liquid disinfection of ultrasound probes that have been enzymatically pre-cleaned. # Model types The GUS ASTRA family of products contai # Physical Characteristics The two models are floor standing units with the following physical characteristics: - . ● . . Weight: 73 lbs. (33.1 kg) . . ● - . Weight: 73 lbs. (33.1 kg) ### Materials materials. Only the shape and length of the disinfection chamber and the probe cable restraining designs differ in order to properly accommodate the differences between TEE and VR ultrasound probes. Materials for various components include: - ◆ ● (PVC), Polyethylene, Polypropylene, EPDM, . Stainless Steel, Polycarbonate, Polyurethane, Polysulfone . ● Polycarbonate, ABS, Polypropylene. ● Sulfide resin, Acryl resin, Polycarbonate resin, stainless steel, silicone, silver, polyester. 6 Winter Avenue Deep River Connecticut, 06443 - Tel: (800) 862-3394 Fax: (860) 526-3081 www.pcimedical.com {6}------------------------------------------------ - - ● - . - - . . . - compliant electronics, PVC, Copper - Polysulfone, Polypropylene, Silicone Component selection was carefully performed to select materials that have appropriate electrical safety rating, fire safety rating, and to be compatible with select FDA cleared HLDs. # Differences Both reprocessors are derived from the same base unit (identical software, electronics, plumbing, valves, sensors, warming pad, cabinet, etc.) which is configurable to become either a er demand. To configure a base unit, TEE and VR specific components (e.g. disinfection chamber, chamber door, cable holder, etc.) are used to accommodate the differences between TEE and VR ultrasound probe geometries. # Device Operational Overview The ASTRA Product Family of automated reprocessors are designed to control the temperature of the high-level liquid disinfectant (HLD), the disinfection time, and the rinse cycles used to automatically loads the appropriate preset parameters based on the barcode of the scanned HLD bottle. After the probes are enzymatically pre-cleaned per hospital protocol, reprocessing in the ASTRA is initiated through the use of a simple user interface. The operator presses Enter to start the cycle and uses the onboard barcode reader to scan and load 1 or 2 probes and to s MRC status at the next prompt. The user will on completion of the disinfection cycle. The ASTRA systems use select FDA cleared high-level liquid disinfectant to safely and effectively disinfect medical devices. By automating the disinfection process through validated software and hardware, the ASTRA Product Family of automated reprocessors help users improve their disinfection outcomes by controlling and tracking the critical process parameters and recording the key operational data. The GUS ASTRA TEE and GUS ASTRA VR are automated reprocessors which are identical with the exception of the disinfection chamber configuration necessary to compensate for two different types of ultrasound probes. The GUS ASTRA TEE is an automated reprocessor which enables high-level disinfection and rinsing for transesophageal (TEE) ultrasound probes. The GUS ASTRA VR is an automated reprocessor which enables high-level disinfection and rinsing for endovaginal and/or endorectal ultrasound probes. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for PCI Medical. The logo consists of the letters "PCI" in a bold, sans-serif font, with the word "MEDICAL" written in a smaller font below. A plus sign is located above the letter "i" in "PCI". The logo is black and white. Both ASTRA systems utilize predefined programs based upon the high-level liquid disinfectant (HLD) used during the disinfection and potable water rinsing cycles. The system controls the HLD temperature and cycle times which are fixed, no user configuration is available. The device will perform the defined automated disinfection and rinsing cycles based upon the HLD specifications. The microprocessor controlled interface monitors temperature and liquid levels of the HLD in the disinfection chamber and the reservoir. Optical sensors are used for HLD reservoir presence and probe identification. Additional user inputs include unique probe identification (for disinfection records) and the results entry of the Minimum Recommended Concentration (MRC) test using test strips and minimum concentration levels as defined by the HLD manufacturer. A sensor located on the disinfection chamber door is used to initiate the cycle after the user has input the required information. User retrievable data/information regarding reprocessing is accessible through a USB port. # Performance Characteristics The ASTRA Product Family of automated reprocessors utilizes the same technology and highlevel disinfection approach employed by the predicate device, the TD-100 cleared under K051305. It should be noted that TD-100 system is comprised of the instrument and proprietary TD-5 HLD. The TD-100 controls time, temperature, and rinse parameters using the TD-5 proprietary HLD to achieve a ≥6 Log reduction. The ASTRA controls time, temperature, and rinse parameters to facilitate a ≥6 Log reduction by the FDA approved HLD. As a result, this submission primarily focuses on the performance specification of the instrument and controls (time, temperature, and rinse) common to both devices. The ASTRA system and the TD-100 both control the HLD temperature, soak time, and rinse cycle to facilitate ≥6 Log reduction of M. terrae on enzymatically cleaned probes. | Device Name | Manufacturer | 510(k) | Performance Attribute | |-------------------------------------------------------|------------------|---------|-----------------------------------------------------------------------------------------------------------------------------------------| | TD-100<br>Transesophageal Probe<br>Disinfector System | CS Medical, Inc. | K051305 | Control the HLD temperature,<br>soak time, and rinse cycle to<br>achieve ≥6 Log reduction of M.<br>terrae using proprietary TD-5<br>HLD | # Performance Specifications The ASTRA Product Family of automated reprocessors share common hardware and software with minor differences necessary to accommodate the different ultrasound probes (TEE and VR). The intended use and function of the automated reprocessors are the same and they are substantially equivalent to the TD-100 probe disinfector in design and function. The ASTRA {8}------------------------------------------------ series of automated reprocessors control the HLD temperature, soak time, and rinse cycle to facilitate ≥6 Log reduction of M. terroe on enzymatically cleaned probes at each site tested, thus achieving equivalent performance to the TD-100 instrument. | Device Name | Manufacturer | 510(k) | Performance Attribute | |--------------------|-------------------|---------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Probe Reprocessors | PCI Medical, Inc. | K150504 | Control the HLD temperature,<br>soak time, and rinse cycle to<br>facilitate ≥6 Log reduction of<br>M. terrae at each site tested<br>using FDA cleared Metricide<br>OPA. | # Clinical/Non-Clinical Performance Testing Summary - . The ASTRA system maintains time and temperature required for disinfection as per HLD - . The ASTRA system passes Safety and EMC testing - The ASTRA system passes simulated use and residual testing ● - . The ASTRA system passes In-use testing ### Technology Comparison to Predicate Devices The ASTRA Product Family of automated reprocessors utilizes the same technology employed by the predicate device, the TD-100 cleared under K051305. The ASTRA system and the TD-100 both control the HLD temperature, soak time, and rinse cycle for the safe and effective disinfection of ultrasound probes. | Summary of Predicate Comparisons | | | | | |-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Characteristic | GUS ASTRA TEE<br>PCI medical | GUS ASTRA VR<br>PCI medical | TD-100<br>CS Medical | | | <b>Regulatory</b> | | | | | | 510(k) Number | K150504 | K150504 | K051305 | | | Product Code | ITX | ITX | ITX | | | Regulation<br>Number | 21 CFR 892.1570 | 21 CFR 892.1570 | 21 CFR 892.1570 | | | | | | N/A<br>ASTRA TEE and ASTRA VR same as<br>predicate. | | | | | | ASTRA TEE and ASTRA VR same as<br>predicate. | | | | | Summary of Predicate Comparisons | | | | Characteristic | GUS ASTRA TEE<br>PCI medical | GUS ASTRA VR<br>PCI medical | TD-100<br>CS Medical | | | Regulation<br>Name | Diagnostic<br>ultrasonic<br>transducer | Diagnostic<br>ultrasonic<br>transducer | Diagnostic<br>ultrasonic<br>transducer | ASTRA TEE and ASTRA VR same as<br>predicate. | | Intended Use/<br>Indications for<br>Use | The GUS ASTRA<br>reprocessor<br>facilitates high-level<br>disinfection and<br>rinsing of 1 or 2<br>transesophageal<br>ultrasound probes<br>using FDA cleared<br>and PCI Medical<br>approved high-level<br>liquid disinfectants. | automated<br>reprocessor<br>facilitates high-level<br>disinfection and<br>rinsing of 1 or 2<br>endovaginal and/or<br>endorectal<br>ultrasound probes<br>using FDA cleared<br>and PCI Medical<br>approved high-level<br>liquid disinfectants. | The TD-100<br>disinfector is<br>designed to provide<br>high-level<br>disinfection of<br>Transesophageal<br>(TEE) ultrasound<br>probes. The system<br>uses the TD-5<br>disinfectant, which<br>is designed to be<br>used only with the<br>TD-100. The<br>disinfectant bottles<br>cannot be reused in<br>the system.<br>TD-5 is intended for<br>use as a single use<br>high-level<br>disinfectant to be<br>used exclusively in<br>the TD-100 for high<br>level disinfection of<br>TEE ultrasound<br>probes.<br>The TD-5 High Level<br>Disinfectant and TD-<br>100 disinfector<br>system is intended<br>for use by qualified<br>individuals trained<br>in its use. | ASTRA TEE and ASTRA VR similar<br>to predicate:<br>Both are intended for<br>use with FDA cleared<br>High-level Liquid<br>Disinfectants (HLDs) to<br>disinfect ultrasound<br>probes and are to be<br>used by trained<br>personnel in non-sterile<br>healthcare settings<br>ASTRA TEE:<br>The ASTRA TEE can<br>reprocess up to 2 TEE<br>probes; TD-100 does 1<br>TEE probe at a time<br>The ASTRA TEE uses<br>select FDA cleared<br>HLDS; TD-100 uses<br>proprietary HLD<br>The ASTRA TEE re-uses<br>HLD; TD-100 uses a<br>single-use bottle<br>ASTRA VR:<br>The ASTRA VR con<br>reprocess up to 2<br>endovaginal/endorectal<br>ultrasound probes; TD-<br>100 does 1 TEE probe at<br>a time<br>The ASTRA VR uses<br>select FDA cleared<br>HLDs; TD-100 uses<br>proprietary HLD<br>The ASTRA VR re-uses<br>HLD; TD-100 uses a<br>single-use bottle | | Prescription<br>Use | No | No | No | ASTRA TEE and ASTRA VR same as<br>predicate. | | Types of probes | Transesophageal<br>(TEE) probes | Endorectal,<br>endovaginal (VR)<br>probes | Transesophageal<br>probes | (TD-100)<br>ASTRA VR<br>equivalent to TD-100 as both<br>reprocess non-lumened<br>ultrasound probes of various<br>geometries | | | Summary of Predicate Comparisons | | | | | Characteristic | GUS ASTRA TEE<br>PCI medical | GUS ASTRA VR<br>PCI medical | TD-100<br>CS Medical | | | | | | | | | Theory of<br>Operation | • Control time,<br>temperature<br>and fluid levels<br>to maintain<br>optimal<br>conditions for<br>disinfection<br>and rinsing as<br>established by<br>HLD<br>manufacturers | • Control time,<br>temperature<br>and fluid levels<br>to maintain<br>optimal<br>conditions for<br>disinfection<br>and rinsing as<br>established by<br>HLD<br>manufacturers | • Control time,<br>temperature<br>and fluid levels<br>to maintain<br>optimal<br>conditions for<br>disinfection<br>and rinsing as<br>established by<br>HLD<br>manufacturer | ASTRA TEE and ASTRA VR same as<br>predicate. | | Disinfection<br>time control | • Timing Chip &<br>Software | • Timing Chip &<br>Software | • Electronics &<br>Software | ASTRA TEE and ASTRA VR<br>Substantially equivalent to TD-<br>100. Time control parameters are<br>contingent to HLD Type and<br>manufacturer specifications. | | Disinfection<br>temperature<br>control | • Temperature<br>sensor &<br>Software | • Temperature<br>sensor &<br>Software | • Electronics &<br>Software | ASTRA TEE and ASTRA VR<br>Substantially equivalent to TD-<br>100. Temperature control<br>parameters are contingent to HLD<br>Type and manufacturer<br>specifications. | | HLD warming<br>capability | • Yes (Warmer<br>pad) | • Yes (Warmer<br>pad) | • Yes (Unknown) | ASTRA TEE and ASTRA VR<br>Substantially equivalent to TD-<br>100. | | Liquid Level<br>Control<br>(HLD and rinse) | • Sensors<br>(upper/lower)<br>& Software | • Sensors<br>(upper/lower)<br>& Software | • Sensor (upper)<br>& Software<br>Fixed volume<br>single use<br>bottle (HLD) | ASTRA TEE and ASTRA VR<br>Substantially equivalent to TD-<br>100. HLD Level control<br>TEE and ASTRA VR use sensors to<br>detect when HLD level is met.<br>Predicate uses a pre-measured<br>HLD single use bottle<br>Rinse Water Level: ASTRA TEE and<br>ASTRA VR same as predicate. | | Overflow<br>Protection | • Sensor<br>(overflow) &<br>Software | • Sensor<br>(overflow) &<br>Software | • Sensor<br>(overflow) &<br>Software | ASTRA TEE and ASTRA VR same as<br>predicate. | | Probe<br>Detection | • Sensor<br>(microswitch)<br>& Software | • Sensor<br>(microswitch)<br>& Software | • Sensor<br>(microswitch)<br>& Software | ASTRA TEE and ASTRA VR same as<br>predicate. | | HLD Vapor<br>Control | • Door, fan &<br>filter | • Door, fan &<br>filter | • Door, fan &<br>filter | ASTRA TEE and ASTRA VR same as<br>predicate. | | User Interface | • Text based<br>interface<br>4 button<br>keypad<br>USB port | • Text based<br>interface<br>4 button<br>keypad<br>USB port | • Text based<br>interface<br>15 button<br>keypad<br>Printer | ASTRA TEE and ASTRA VR<br>Substantially equivalent to TD-<br>100. ASTRA TEE and ASTRA VR<br>utilize USB port to download data<br>while predicate uses printable<br>data. ASTRA TEE and ASTRA VR | | Characteristic | GUS ASTRA TEE<br>PCI medical | GUS ASTRA VR<br>PCI medical | TD-100<br>CS Medical | | | | Barcode<br>Scanner | Barcode<br>Scanner | | utilize Barcode Scanner to capture<br>input data while the predicate<br>device allows manual input<br>through keypad. | | HLD MRC<br>Verification | Operator must<br>verify MRC | Operator must<br>verify MRC | Operator must<br>verify HLD<br>expiration date | ASTRA TEE and ASTRA VR<br>Substantially equivalent to TD-<br>100. | | Material<br>Components | Diaphragm<br>Pump Solenoid valves Polyethylene<br>tubing Push Fit fittings Powder coated<br>sheet metal<br>cabinet with<br>ABS plastic<br>components | Diaphragm<br>Pump Solenoid valves Polyethylene<br>tubing Push Fit fittings Powder coated<br>sheet metal<br>cabinet with<br>ABS plastic<br>components | Diaphragm<br>Pump Solenoid valves Polyethylene<br>tubing Fittings ABS Plastic<br>cabinet | ASTRA TEE and ASTRA VR<br>Substantially equivalent to TD-<br>100. | | Compatible<br>HLDS | Uses FDA<br>cleared HLDS<br>(Metricide<br>OPA) per<br>specifications…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...