Orthopantomograph OP 3D LX (PAN 3D); Orthopantomograph OP 3D LX (PAN CEPH 3D)
K241249 · Palodex Group OY · OAS · Sep 12, 2024 · Radiology
Device Facts
Record ID
K241249
Device Name
Orthopantomograph OP 3D LX (PAN 3D); Orthopantomograph OP 3D LX (PAN CEPH 3D)
Applicant
Palodex Group OY
Product Code
OAS · Radiology
Decision Date
Sep 12, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Pediatric
Indications for Use
ORTHOPANTOMOGRAPH™ OP 3D LX is an X-ray device that is intended to be used for imaging of adult and pediatric patients. The device can be configured to take panoramic, cephalometric or 3D images of cranio-maxillofacial complex including the ear, nose and throat (ENT) and airway regions. and cervical spine. The device can be configured to take carpus images. The device is operated and used by qualified healthcare professionals.
Device Story
Digital panoramic, cephalometric, and cone beam computed tomography (CBCT) X-ray system. Inputs: X-ray radiation passed through patient anatomy. Transformation: Cone beam X-ray technology acquires volumetric data; reconstruction software processes raw data into 2D/3D images. Output: Diagnostic images of craniomaxillofacial complex, ENT, airway, cervical spine, and carpus. Used in clinical settings by qualified healthcare professionals for diagnostic and treatment planning. Includes Automatic Exposure Control (ADC) feature for standard/child panoramic and 3D imaging programs. Output viewed by clinicians to inform diagnosis and treatment decisions. Benefits: Provides high-quality diagnostic imaging for various dental and maxillofacial applications.
Clinical Evidence
No clinical data. Substantial equivalence established via non-clinical bench testing to international standards (IEC/ISO) for safety, electromagnetic compatibility, radiation protection, and imaging performance.
Technological Characteristics
Digital CBCT X-ray system. Components: X-ray source, focal spot, image detector scintillator. Connectivity: Networked/digital. Software: IEC 62304 compliant. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-63, IEC 61223-3-7, IEC 61223-3-4, ISO 10993-1, ISO 14971. Features: Automatic Exposure Control (ADC).
Indications for Use
Indicated for imaging of adult and pediatric patients, including panoramic, cephalometric, 3D images of the cranio-maxillofacial complex (ENT, airway, cervical spine), and carpus images.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
K150354 — PAPAYA 3D Plus · Genoray Co., Ltd. · Dec 10, 2015
K220423 — PAPAYA & PAPAYA Plus · Genoray Co., Ltd. · May 19, 2022
K220392 — PAPAYA 3D & PAPAYA 3D Plus · Genoray Co., Ltd. · May 19, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 12, 2024
Palodex Group Oy % Riika Sarviaho Regulatory Affairs Specialist Nahkelantie 160 Tuusula, 04300 FINLAND
Re: K241249
Trade/Device Name: Orthopantomograph™ OP 3D LX (PAN 3D); Orthopantomograph™ OP 3D LX (PAN CEPH 3D) Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: OAS Dated: May 3, 2024 Received: August 13, 2024
Dear Riika Sarviaho:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostoc X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K241249
Device Name
Orthopantomograph™ OP 3D LX (PAN 3D); Orthopantomograph™ TM OP 3D LX (PAN CEPH 3D)
Indications for Use (Describe)
ORTHOPANTOMOGRAPH™ OP 3D LX is an X-ray device that is intended to be used for imaging of adult and pediatric patients. The device can be configured to take panoramic, cephalometric or 3D images of cranio-maxillofacial complex including the ear, nose and throat (ENT) and airway regions. and cervical spine. The device can be configured to take carpus images.
The device is operated and used by qualified healthcare professionals.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) #: | K241249 | 510(k) Summary | Prepared on: 2024-08-13 |
|-----------|---------|----------------|-------------------------|
|-----------|---------|----------------|-------------------------|
| Contact Details | | | 21 CFR 807.92(a)(1) |
|-----------------------------|---------------------------------------|--|---------------------|
| Applicant Name | Palodex Group Oy | | |
| Applicant Address | Nahkelantie 160 Tuusula 04300 Finland | | |
| Applicant Contact Telephone | +358505727161 | | |
| Applicant Contact | Dr. Riika Sarviaho | | |
| Applicant Contact Email | riika.sarviaho@envistaco.com | | |
| Device Name | | 21 CFR 807.92(a)(2) | |
|---------------------|--------------------------------------------------------------------------------------------------------------|---------------------|--|
| Device Trade Name | Orthopantomograph <sup>TM</sup> OP 3D LX (PAN 3D);<br>Orthopantomograph <sup>TM</sup> OP 3D LX (PAN CEPH 3D) | | |
| Common Name | Computed tomography x-ray system | | |
| Classification Name | Computed tomography x-ray system | | |
| Regulation Number | 892.1750 | | |
| Product Code(s) | OAS | | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) | |
|------------------------------------|----------------------------------------------------------|---------------------|--------------|
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | | Product Code |
| K230505 | Orthopantomograph <sup>TM</sup> OP 3D LX | | OAS |
| Device Description Summary | | 21 CFR 807.92(a)(4) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|---------------------|--|
| ORTHOPANTOMOGRAPH <sup>TM</sup> OP 3D LX is a digital panoramic, cephalometric and cone beam computed tomography (CBCT) x-ray device. OP 3D LX is used for imaging of the craniomaxillofacial complex and neck areas including the ear, nose and throat (ENT) for diagnostic and treatment planning purposes. | | | |
| Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|---------------------|--|
| ORTHOPANTOMOGRAPH <sup>TM</sup> OP 3D LX is an X-ray device that is intended to be used for imaging of adult and pediatric patients. The device can be configured to take panoramic, cephalometric or 3D images of cranio-maxillofacial complex including the ear, nose and throat (ENT) and airway regions, and cervical spine. The device can be configured to take carpus images. | | | |
The device is operated and used by qualified healthcare professionals.
# Indications for Use Comparison
The Indications for Use is identical to the Indications for Use of the predicate device.
Technological Comparison
21 CFR 807.92(a)(6)
21 CFR 807.92(a)(5)
{5}------------------------------------------------
This section provides the substantial equivalence rationale for ORTHOPANTOMOGRAPH™ OP 3D LX, and the Primary Predicate Device, ORTHOPANTOMOGRAPH™ OP 3D LX (K23050) with reqards to Indications for Use, Technology, and Performance Testing, As described in this section, the differences between the predicate do not raise different questions of substantial equivalence.
Details of the similarities between subject and predicate devices:
The Subject Device shares the same architectural components and utilizes the same X-ray generation as the Predicate Device. Both devices utilize cone beam x-ray technology to acquire volumetric data. Both the Subject Device have the same image modes, X-ray source, focal spot, image detector scintillator, 2D image performance – DQE / MTF 70kV RAQS, 3D image technique, 3D field of view, 3D total viewing andel size, reconstruction software, 3D's effective exposure time, Ceph exposure time, Patient position, system footprint, 3D image programs, and material patient contacting components.
Details of the differences between subject and predicate devices:
There are no major differences between the Subject Device. There is one minor difference, the Automatic Exposure Control (Automatic Dose Control) between the Subject Device. The Automatic Exposure Control (Automatic Dose Control) is identical to the Predicate #2 device is available in standard/child panoramic and in all 3D imaging programs as "ADC" feature.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Non-Clinical performance bench testing to international standards (and FDA recognized consensus standards) for Computed tomography x-ray system has been conducted to determine conformance in regard to:
· IEC 60601-1:2005 / A1:2012 + A2:2020 (Ed. 3.2): Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
+ ANSI AAM E560601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 (Consolidated Text): Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
· IEC 60601-1-2:2014/A1:2020 (Ed. 4.1): CONSOLIDATED VERSION: Medical electrical equipments for basic safety and essential performance - Collateral standard: Electromagnetic disturbances -Reguirements and tests
• IEC 60601-1-3:2008/A1:2013/A2:2021 Ed. 2.2; Medical electrical equipment – Part 1-3: General requirements for basic safety and essential performance – Collateral Standard: Radiation protection in diagnostic X-ray equipment
· IEC 62304:2006/A1:2015 Edition 1.1 CONSOLIDATED VERSION: Medical device software life-cycle processes
• IEC 60601-1-6:2010/A1:2013/A2:2020 (Ed. 3.2) CONSOLIDATED VERSION: Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
• IEC 60601-2-63:2012/A1:2017/A2:2021 (Ed 1.2) CONSOLIDATED VERSION: Medical electrical equipment - Part 2-6 : Particular requirements for the basic safety and essential performance of dental extra-oral X-ray equipment
· IEC 62366-1:2015/A1:2020 (Ed 1.1) CONSOLIDATED VERSION: Medical devices - Part 1: Application of usability engineering to medical devices
· ISO 10993-1:2018 Fifth edition: Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
• ISO 20417:2021 (First edition) CORRECTED VERSION 2021-12: Medical devices – Information to be supplied by the manufacturer «ISO 15223-1:2021 (Ed. 4.0): Medical devices — Symbols to be used with information to be supplied by the manufacturer — Part 1: General requirements
- IEC 61223-3-7:2021 (Ed. 1.0): Evaluation and routine testing in medical imaging departments Part 3-7: Acceptance and constancy tests - Imaging performance of X-ray equipment for dental cone beam computed tomography
- IEC 61223-3-4:2000 [Ed.1.0): Evaluation and routine testing in medical imaging departments – Part 3-4: Acceptance tests – Imaging performance of dental X-ray equipment
· IEC 60336:2020 (Ed. S.0): Medical electrical equipment - X-ray tube assemblies for medical diagnosis - Focal spot dimensions and related characteristics- test methods only
• ISO 14971:2019 (3rd Edition): Medical devices — Application of risk management to medical devices
• IEC 81001-5-1:2021 (Ed 1.0): Health software and health IT systems safety, effectiveness and security
The following list of FDA Guidance Documents has been utilized in the development of the ORTHOPANTOMOGRAPH™ OP 3D LX:
• Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing Within a Risk Management Process", September 2023
· Recommended Content and Format of Non-Clinical Bench Performance Testing Information in Premarket Submissions, December 2019
· Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions, September 2023
{6}------------------------------------------------
· Content of Premarket Submissions for Device Software Functions, June 2023
- · Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices, September 2016
- Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices, September 2018
- · Medical X-Ray Imaging Devices Conformance with IEC Standards, February 2023.
- Electromagnetic Compatibility (EMC) of Medical Devices, June 2022
- Off-The-Shelf Software Use in Medical Devices, August 2023
- · Pediatric Information for X-ray Imaging Device Premarket Notifications, November 2017
Clinical data is not needed to characterize performance and establish substantial equivalence. The non-clinical test data characterize all performance aspects of the device based on well-established scientific and engineering principles. Clinical testing has not been conducted on this product.
Based on a comparison of intended use, indications, material composition, technological characteristics, principle of operation, features and performance data, the Orthopantomograph™ OP 3D LX is deemed to be substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.