ULTRAEKOGEL MODELS EK 2000 & EK 2001

K030889 · Inmed Ltda. · MUI · May 23, 2003 · Radiology

Device Facts

Record IDK030889
Device NameULTRAEKOGEL MODELS EK 2000 & EK 2001
ApplicantInmed Ltda.
Product CodeMUI · Radiology
Decision DateMay 23, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1570
Device ClassClass 2

Intended Use

ULTRAEKOGEL™ is an electroconductive gel media used with ultrasonic pulsed echo imaging system. The gel -once applied on the defined area of the body- facilitates the intended use of the system, which is intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver.

Device Story

ULTRAEKOGEL™ is an electroconductive gel media applied topically to the body to facilitate acoustic coupling between an ultrasonic transducer and patient tissue. It functions as a transmission medium for pulsed sound beams used in diagnostic imaging to determine tissue interface depth and location. The device is used in clinical settings where ultrasonic imaging is performed. It is supplied in 250g and 5kg containers with a 3-year shelf life.

Clinical Evidence

No clinical data provided; substantial equivalence is based on identical technological characteristics and performance to the predicate device.

Technological Characteristics

Electroconductive gel media. Supplied in 250g and 5kg containers. 3-year shelf life. Identical to predicate device in materials, design, and performance.

Indications for Use

Indicated for use as an electroconductive coupling medium for ultrasonic pulsed echo imaging systems to facilitate the projection of sound beams into body tissue for diagnostic imaging.

Regulatory Classification

Identification

A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K030889 # MAY 2 3 2003 510(k) Summary As Required by 21 section 807.92 ( c ) | 1-Submitter Name: | Mansour Consulting LLC | |---------------------------------------------------------------------|------------------------------------------------------| | 2-Address: | 1308 Morningside Park Dr<br>Alpharetta, GA 30022 USA | | 3-Phone: | (678) 908-8180 | | 4-Fax: | (425) 795-9341 | | 5-Contact Person: | Jay Mansour | | 6-Date summary prepared: | March 18th, 2003 | | 7-Device Trade or Proprietary Name: | ULTRAEKOGEL™ | | 8-Device Common or usual name: | ULTRASOUND GEL | | 9-Device Classification Name: | SYSTEM, IMAGING, PULSED ECHO,<br>ULTRASONIC- GEL | | 10-Substantial Equivalency is claimed against the following device: | AQUASONIC 100 from Parker Laboratories, Inc. | - AQUASONIC 100 from Parker Laboratories, Inc 510k # K802146 ## 11-Description of the Device: ULTRAEKOGEL™ is an electroconductive gel media used with ultrasonic pulsed echo imaging system. The gel -once applied on the defined area of the body- facilitates the intended use of the system, which is intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. It is supplied in 250 grams and 5 Kilograms containers with a 3-year shelf life. ## 12-Intended use of the device: (refer to FDA form attached) ULTRAEKOGEL™ is an electroconductive gel media used with ultrasonic pulsed echo imaging system. The gel -once applied on the defined area of the body- facilitates the intended use of the system, which is intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. ### 13-Safety and Effectiveness of the device: This device is safe and effective as the other predicate device cited above. This is better expressed in the tabulated comparison (Paragraph 14 below) ## 14-Summary comparing technological characteristics with other predicate device: Please find below a tabulated comparison supporting that this device is substantially equivalent to other medical devices in commercial distribution. Also, Equivalency overview chart path is attached. Indeed, this device is IDENTICAL to the predicate device. Refer to the explanations within the main submission. {1}------------------------------------------------ | FDA file reference number | 510k # K802146 | |--------------------------------------------------|------------------------| | Attachments inside notification submission file | 510k summary print out | | TECHNOLOGICAL CHARACTERISTICS | Comparison result | | Indications for use | Identical | | Target population | Identical | | Design | Identical | | Materials | Identical | | Performance | Identical | | Sterility | Identical | | Biocompatibility | Identical | | Mechanical safety | Identical | | Chemical safety | Identical | | Anatomical sites | Identical | | Human factors | Identical | | Energy used and/or delivered | Identical | | Compatibility with environment and other devices | Identical | | Where used | Identical | | Standards met | Identical | | Electrical safety | Identical | | Thermal safety | Identical | | Radiation safety | Identical | டு {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is a stylized image of an eagle with three lines extending from its head. Rockville MD 20850 Food and Drug Administration 9200 Corporate Boulevard MAY 2 3 2003 INMED, LTDA % Mr. Jay Mansour President Mansour Consulting, LLC 1308 Morningside Park Dr. ALPHARETTA GA 30022 Re: K030889 Trade/Device Name: ULTRAEKOGEL™ Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: 90 MUI Dated: March 18, 2003 Received: March 21, 2003 Dear Mr. Mansour: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean t that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive. Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K030 889 Page of 510(k) Number (if known): _ K (3088 9 Device Name: ULTRAEKOGEL TM Indications for Use: ULTRAEKOGEL™ is an electroconductive gel media used with ultrasonic pulsed echo imaging system. The gel -----------------------------------------------------------------------------------------------------------------------------------------the body- facilitates the intended use of the system, which is intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. David R. Ingram Division Sign-Off ision of Reproductive, Radiological Device 1 ට *Prescription Use* √ (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurence of CDRH, Office of Device Evaulation (ODE) (Optional Format 3-10-98) ---------------------
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...