Lux HD 2530 detector (Lux HD 2530)

K252911 · Iray Imaging Technology (Haining) Limited · MQB · Oct 9, 2025 · Radiology

Device Facts

Record IDK252911
Device NameLux HD 2530 detector (Lux HD 2530)
ApplicantIray Imaging Technology (Haining) Limited
Product CodeMQB · Radiology
Decision DateOct 9, 2025
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1680
Device ClassClass 2
AttributesPediatric

Indications for Use

Lux HD 2530 Detector is indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients. It is intended to replace film/screen systems in all general-purpose diagnostic procedures. Lux HD 2530 Detector is not intended for mammography or dental applications.

Device Story

Lux HD 2530 is a digital flat panel detector for general radiography. It uses a CsI scintillator to convert X-rays to visible light, which is then transformed into electronic signals by a TFT/diode capacitor array. The device transmits digital image data to a PC via wired (Gigabit Ethernet) or wireless (IEEE 802.11a/b/g/n/ac) connection. Operated by clinicians in clinical settings, the detector integrates with a DR software platform for image acquisition, enhancement, and viewing. It features Automatic Exposure Detection (AED) to trigger image acquisition upon X-ray detection and Detector Exposure Control (DEC) to signal the high-voltage generator to terminate exposure once a target dose is reached. The output assists physicians in diagnosing disease or injury, potentially reducing patient radiation dose through precise exposure control.

Clinical Evidence

Bench testing only. Performance evaluated via electrical safety, EMC, biological evaluation, wireless coexistence, and cybersecurity testing. No clinical data provided.

Technological Characteristics

Flat panel detector with CsI scintillator, TFT readout, 16-bit ADC. Active area 25cm x 30cm, 100μm pixel size, 5.0 lp/mm spatial resolution. Connectivity: Gigabit Ethernet, IEEE 802.11a/b/g/n/ac. IP67 rating. Carbon fiber construction. Air-cooled. Powered via 10-pin port.

Indications for Use

Indicated for digital radiographic diagnosis of human anatomy in adult and pediatric patients. Not for mammography or dental applications.

Regulatory Classification

Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION October 9, 2025 iRay Imaging Technology (Haining) Limited c/o Junjie Qian No. 2, Caohejing RD. Haining, Jiaxing, Zhejiang 314499 CHINA Re: K252911 Trade/Device Name: Lux HD 2530 Detector (Lux HD 2530) Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: MQB Dated: August 1, 2025 Received: September 12, 2025 Dear Junjie Qian: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252911 - Junjie Qian Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. {2} K252911 - Junjie Qian See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ![img-0.jpeg](img-0.jpeg) Lu Jiang Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K252911 | | | Device Name Lux HD 2530 detector (Lux HD 2530) | | | Indications for Use (Describe) | | | Lux HD 2530 Detector is indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients. It is intended to replace film/screen systems in all general-purpose diagnostic procedures. Lux HD 2530 Detector is not intended for mammography or dental applications. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} iRay Imaging Technology (Haining) Limited [510(k)] Application # 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS (As Required by 21 CFR 807.92) K252911 1. Date Prepared [21 CFR 807.92(a)(1)] August 8, 2025 2. Submitter's Information [21 CFR 807.92(a)(1)] Company Name: iRay Imaging Technology (Haining) Limited Company Address: No. 2, Caohejing RD., Haining 314499, Jiaxing, Zhejiang, China Contact Person: Junjie Qian Phone: 0573-87399739 Email: Junjie.qian@iraygroup.com 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)] Trade Name: Lux HD 2530 Detector (Lux HD 2530) Common Name: Solid State X-Ray Imager Model Name: Lux HD 2530 Classification Name: Stationary X-Ray System Product Code: MQB Regulation Number: 21 CFR 892.1680 Device Class: Class II 1 of 10 {5} iRay Imaging Technology (Haining) Limited [510(k)] Application # 4. Identification of Predicate Devices(s) [21 CFR 807.92(a)(3)] The identification predicates within this submission are as follows: Manufacturer: iRay Imaging Technology (Haining) Limited Trade Name: Lux HD 35 Detector (Lux HD 35) Lux HD 43 Detector (Lux HD 43) Model Name: Lux HD 35 Lux HD 43 Product Code: MQB Classification Name: Stationary X-Ray System Regulation Number: 21 CFR 892.1680 Device Class: Class II FDA 510 (k) #: K243556 # 5. Description of the Device [21 CFR 807.92(a)(4)] Lux HD 2530 Detector is digital flat panel detector. They support the single frame mode, with the key component of TFT/PD image sensor flat panel of active area: 25cm × 30cm (Lux HD 2530 Detector). The differences between two models are overall change in the dimensions of the image receptor The sensor plate of Lux HD 2530 Detector is direct-deposited with CsI scintillator to achieve the conversion from X-ray to visible photon. The visible photons are transformed to electron signals by diode capacitor array within TFT panel, which are composed and processed by connecting to scanning and readout electronics, consequently to form a panel image by transmitting to PC through the user interface. The major function of the Lux HD 2530 Detector is to convert the X-ray to digital image, with the application of high resolution X-ray imaging. 2 of 10 {6} iRay Imaging Technology (Haining) Limited [510(k)] Application The Digital Radiographic Imaging Acquisition Software Platform - DR is part of the system, it is used to acquire, enhance, view image from Lux HD 2530 Detector. Based on the risks and intended use, documentation level of the software is basic ## 6. Intended Use [21 CFR 807.92(a)(5)] ### 6.1. Indications for use Lux HD 2530 Detector is indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients. It is intended to replace film/screen systems in all general-purpose diagnostic procedures. Lux HD 2530 Detector is not intended for mammography or dental applications. ### 6.2. Suitable patient Lux HD 2530 Detector is suitable for providing digital X-ray imaging for DR system to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients, but not intended for mammography or dental applications. ### 6.3. Processing of input and output When Lux HD 2530 Detector work continuously, it can automatically distinguish X-ray and output an imaging for diagnosis of disease, injury, or of any applicable health problem. 3 of 10 {7} iRay Imaging Technology (Haining) Limited [510(k)] Application 7. Technological Characteristic [21 CFR 807.92(a)(6)] | Item | Predicate Device: Lux HD 43 Detector Lux HD 35 Detector | Proposed Device: Lux HD 2530 Detector | | --- | --- | --- | | 510(K) Number | K243556 | K252911 | | Intended Use | The Lux HD detectors are indicated for digital imaging solution designed for providing general radiographic system in all general-purpose diagnostic procedures. | Same | | Indications for Use | Lux HD Detectors are indicated for digital imaging solutions designed to provide general radiographic diagnosis for human anatomy including both adult and pediatric patients. They are intended to replace film/screen systems in all general-purpose diagnostic procedures. The devices are not intended for mammography or dental applications. | same | | Classification Name | Stationary X-ray system | Same | | Product Code | MQB | Same | {8} iRay Imaging Technology (Haining) Limited [510(k)] Application | Item | Predicate Device: Lux HD 43 Detector Lux HD 35 Detector | Proposed Device: Lux HD 2530 Detector | | --- | --- | --- | | Regulation Number | 21 CFR 892.1680 | Same | | Panel | Radiology | Same | | Classification | II | Same | | X-Ray Absorber (Scintillator) | CsI | Same | | Installation Type | Wireless or Wired | Same | | Readout Mechanism | Thin Film Transistor | Same | | Image Matrix Size | Lux HD 35 Detector: 3500 ×4300pixels Lux HD 43 Detector: 4267 × 4267 pixels | 2500 × 3000pixels | | Pixel Size | 100μm | Same | | ADC Digitization | 16 bit | Same | | Effective Imaging Area | Lux HD 35 Detector: 350mm ×430mm Lux HD 43 Detector: 426.7 mm × 426.7mm | 250mm × 300mm | | Spatial Resolution | 5.0 lp/mm | Same | | Detective Quantum Efficiency (DQE) | 0.54 at 1 lp/mm (RQA5, 2.5μGy) | Same | 5 of 10 {9} iRay Imaging Technology (Haining) Limited [510(k)] Application | Item | Predicate Device: Lux HD 43 Detector Lux HD 35 Detector | Proposed Device: Lux HD 2530 Detector | | --- | --- | --- | | Power Consumption | Max. 30W | Same | | Communications | a) Wired: Gigabit Ethernet (1000BASE-T) b) Wireless: IEEE 802.11a/b/g/n/ac (2.4 GHz / 5 GHz) | Same | | Imaging protect Plate | Carbon Fiber Plate | Same | | Cooling | Air cooling | Same | | Dimensions | Lux HD 35 Detector: 384mm×460mm×15mm Lux HD 43 Detector: 460mm×460mm×15mm | 281.5mm×332.3mm×15 mm | | Detector IP grade | IP67 | Same | | Power input port | 10 pin port | Same | | Frame material | Carbon fiber | Same | | Surface pressure | Uniform load: 300 kg over the whole area of the surface; Local load: 100 kg on an area 4 cm diameter of center | Same | | Operation | Temperature: +5 ~ +35°C Humidity: 5 ~ 90% (Non-Condensing) | Same | {10} iRay Imaging Technology (Haining) Limited [510(k)] Application | Item | Predicate Device: Lux HD 43 DetectorLux HD 35 Detector | Proposed Device: Lux HD 2530 Detector | | --- | --- | --- | | | Atmospheric pressure: 70 ~ 106 kPaAltitude: Max. 3000 meters | | | Storage and Transportation: (detector) | Temperature: -10 to + 60°C.(Excluding battery)Relative Humidity: 5 to 95%, non-condensing.Atmosphere: 700mBar~1060mBarAltitude: Max. 3000 meter | Same | | Software | Digital Radiographic Imaging Acquisition Software Platform - DRThe software used for getting Digital X-ray radiography images from the flat panel detectors.The software is used to handle the DICOM protocol (DICOM 3.0).The software is responsible for the DR equipment management, acquisition and processing functions, to | Same | {11} iRay Imaging Technology (Haining) Limited [510(k)] Application | Item | Predicate Device: Lux HD 43 Detector Lux HD 35 Detector | Proposed Device: Lux HD 2530 Detector | | --- | --- | --- | | | provide patient registration, scanning, image processing and other functions. | | # 8. Minimum requirement for connection Lux HD 2530 Detector connected via wireless and wired communication. Operating System: Windows 11 CPU: Intel Core i3- 8100 3.6GHz 4C 65W Memory: 16GB (2x8GB) DDR4 2666 DIMM Hard Disk: 1TB # 9. X-ray exposure mode The AED (Automatic Exposure Detection) trigger module is a unit can connect X-ray signal in the Lux HD 2530 Detector. Once there is X-ray generator exposure exist, the AED trigger module will detect the X-ray radiation and output signal to the detector. Until the exposure finished, the detector will receive a signal which represent the end of exposure from the inner trigger module and begin to acquire the image. The DEC (Detector Exposure Control) trigger module is a unit can cutoff HVG output when patients need specified dose, which avoid extra dose absorbed by patients. Once the dose detected by the detector reaches the target value, the detector will send a cut-off signal to the high-voltage system. This signal indicates that no further exposure from HVG is required. 8 of 10 {12} iRay Imaging Technology (Haining) Limited [510(k)] Application # 10. Non-clinical study 1) Electrical Safety and EMC testing: Electrical, mechanical, environmental safety according to IEC/ES 60601-1 was performed, and EMC testing was also conducted in accordance with IEC 60601-1-2. All test results are meet the standard requirements. 2) Biological Evaluation: The materials of the detector which contact operators' or patients' skin have been evaluated with the FDA guidance "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". And the evaluation result assured the biological safety. 3) Non-clinical Considerations: The non-clinical studies have been performed and the results have shown that sections of the non-clinical consideration mentioned in the 'Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices' are substantially equivalent to the non-clinical consideration of predicate device on the Market (Lux HD 35/43 detector, K243556). 4) Clinical Consideration: Intended use, fundamental scientific technology, regulatory requirement, non-clinical performance, labeling, quality-assurance program keep the same with those of predicate device. Software keep the same with that of reference device except "image forwarding, image printing" function and literal name. There is no any negative change about clinical performance from predicate device. 5) Wireless testing Wireless functionality and wireless coexistence testing in accordance with ANSI IEEE C63.27-2017 was performed. All test results are meet the standard requirements. 9 of 10 {13} iRay Imaging Technology (Haining) Limited [510(k)] Application 6) Cybersecurity testing Cybersecurity threat modeling, risk assessment, and controls and testing were performed to comply with requirements specified in section 524B(b)(2) of the Federal Food, Drug, and Cosmetics Act to provide a reasonable assurance that the subject device with its wireless capabilities are cybersecure. 11. Conclusion [21 CFR 807.92(b)(3)] In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, applicant concludes that Lux HD 2530 Detector is substantially equivalent to predicate device with regards to safety and effectiveness. 10 of 10
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...