K243675 · Inkspace Imaging, Inc. · MOS · Dec 20, 2024 · Radiology
Device Facts
Record ID
K243675
Device Name
InkSpace Imaging Small Body Array
Applicant
Inkspace Imaging, Inc.
Product Code
MOS · Radiology
Decision Date
Dec 20, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The InkSpace Imaging Small Body Array for Siemens 1.5T MRI systems is indicated for use as an alternative to a traditional MRI receive array. The conformal fit and blanket-like construction are designed to decrease anxiety and optimize convenience for both the patient and the technologist during the MRI scan. Use of the device during these procedures should be performed by appropriately trained MRI technologists, and interpretation of the images produced by these scans should be performed only by board certified radiologists.
Device Story
Receive-only phased array MRI coil; 24-channel configuration consisting of two flexible 12-channel pads. Designed to conform to patient anatomy for improved comfort and reduced anxiety. Operates with Siemens 1.5T MRI systems to capture hydrogen signals (63.6 MHz). Used by MRI technologists in clinical settings; images interpreted by radiologists to assist in diagnosis. Device is reusable and non-sterile. Provides diagnostic image data for cardiac, spine, and musculoskeletal applications.
Clinical Evidence
Bench testing only. Performance evaluated per FDA guidance for MR receive-only coils, including SNR, image uniformity, surface heating, decoupling, EMC, and electrical/mechanical safety. Biocompatibility, usability, and transit testing leveraged from predicate device via risk analysis. Clinical image quality reviewed by a board-certified radiologist.
Technological Characteristics
24-channel receive-only phased array coil; two 12-channel flexible pads. Operates at 1.5T (63.6 MHz). Materials are soft and pliable. Biocompatibility per ISO 10993. Reusable, non-sterile, prescription-only. Compatible with Siemens 1.5T MRI systems.
Indications for Use
Indicated for obtaining diagnostic images of general human anatomy in adult and pediatric patients, including cardiac, spine, shoulder, elbow, knee, foot, and prostate, using Siemens 1.5T MRI systems.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
K172222 — 8ch Flex Suite · Shenzhen RF Tech Co., Ltd. · Dec 8, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 20, 2024
InkSpace Imaging, Inc. % Taras Bouzakine Director, Regulatory Affairs Veranex, Inc. 5420 Wade Park Blvd, Suite 204 Raleigh, North Carolina 27607
Re: K243675
Trade/Device Name: InkSpace Imaging Small Body Arrav Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: MOS Dated: November 27, 2024 Received: November 27, 2024
Dear Taras Bouzakine:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
FDA
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K243675
Device Name
InkSpace Imaging Small Body Array
Indications for Use (Describe)
The InkSpace Imaging Small Body Array is a receive-only coil, used for obtaining diagnostic images of general human anatomy in adult and pediatric patients such as cardiac, spine, shoulder, elbow, knee, foot, and prostate in Siemens 1.5T magnetic resonance imaging systems. These images, when interpreted by a trained physician, vield information that may assist in diagnosis.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
This summary of the Special 510(k) premarket notification for the InkSpace Imaging Small Body Array is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR\$807.92.
{5}------------------------------------------------
# GENERAL INFORMATION [807.92(A)(1)]
#### Applicant:
InkSpace Imaging, Inc. 5635 West Las Positas Blvd, Suite 403/404 Pleasanton, CA 94588 USA Phone: 925-425-7410
### Contact Person:
Taras Bouzakine Director, Regulatory Affairs, Veranex, Inc. 5420 Wade Park Blvd, Suite 204 Raleigh, NC 27607 USA Phone: 603-205-4874 Email: taras.bouzakine@yeranex.com
Date Prepared: December 18, 2024
# DEVICE INFORMATION [807.92(A)(2)]
### Classification:
Class II per 21 CFR§892.1000
#### Product Code:
MOS
Trade Name:
InkSpace Imaging Small Body Array
### Generic/Common Name:
Coil, Magnetic Resonance, Specialty
### PREDICATE DEVICE(S) [807.92(A)(3)]
InkSpace Imaging Small Body Array (K233444)
### DEVICE DESCRIPTION [807.92(A)(4)]
The InkSpace Imaging Small Body Array is a phased array, receive-only coil intended to work with Siemens 1.5T MRI scanners for body MR imaging examinations. The coil is comprised of 24 total channels, with two individual flexible pads, each containing 12 individual elements. The elements are flexible to conform to patients' anatomies.
{6}------------------------------------------------
# INDICATIONS FOR USE [807.92(A)(5)]
The InkSpace Imaging Small Body Array is a receive-only coil, used for obtaining diagnostic images of general human anatomy in adult and pediatric patients such as cardiac, spine, shoulder, elbow, knee, foot, and prostate in Siemens 1.5T magnetic resonance imaging systems. These images, when interpreted by a trained physician, yield information that may assist in diagnosis.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES [807.92(A)(6)]
The technological characteristics of the modified InkSpace Imaging Small Body Array for 1.5T scanners are substantially equivalent to the predicate device, the InkSpace Imaging Small Body Array for 3.0T scanners (K233444). Table 1 lists the technological characteristics of the predicate and modified devices and provides rationale to support a determination of substantial equivalence. Any differences between the devices do not raise any new or different questions of safety or effectiveness.
{7}------------------------------------------------
| | | | | Table 1: Summary of Technological Characteristics |
|--|--|--|--|---------------------------------------------------|
|--|--|--|--|---------------------------------------------------|
| Characteristics | Subject Device:<br>InkSpace Imaging Small Body Array - 1.5T | Predicate Device:<br>InkSpace Imaging Small Body Array - 3.0T<br>(K233444) | Substantial Equivalence<br>Rationale |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | Magnetic Resonance Diagnostic Device per 21<br>CFR§892.1000 | Magnetic Resonance Diagnostic Device per 21<br>CFR§892.1000 | N/A (Same). |
| Product Code | MOS (Coil, Magnetic Resonance, Specialty) | MOS (Coil, Magnetic Resonance, Specialty) | N/A (Same). |
| Intended Use | The InkSpace Imaging Small Body Array for<br>Siemens 1.5T MRI systems is indicated for use<br>as an alternative to a traditional MRI receive<br>array. The conformal fit and blanket-like<br>construction are designed to decrease anxiety<br>and optimize convenience for both the patient<br>and the technologist during the MRI scan. Use<br>of the device during these procedures should be<br>performed by appropriately trained MRI<br>technologists, and interpretation of the images<br>produced by these scans should be performed<br>only by board certified radiologists. | The InkSpace Imaging Small Body Array for<br>Siemens 3.0T MRI systems is indicated for use as<br>an alternative to a traditional MRI receive array.<br>The conformal fit and blanket-like construction are<br>designed to decrease anxiety and optimize<br>convenience for both the patient and the<br>technologist during the MRI scan. Use of the<br>device during these procedures should be<br>performed by appropriately trained MRI<br>technologists, and interpretation of the images<br>produced by these scans should be performed only<br>by board certified radiologists. | Intended use of the modified<br>device maintains the same general<br>purpose and function as the<br>predicate device, differing only<br>by target MRI system field<br>strength (1.5T vs. 3.0T). |
| Indications for<br>Use | The InkSpace Imaging Small Body Array is a<br>receive-only coil, used for obtaining diagnostic<br>images of general human anatomy in adult and<br>pediatric patients such as cardiac, spine,<br>shoulder, elbow, knee, foot, and prostate in<br>Siemens 1.5T magnetic resonance imaging<br>systems. These images, when interpreted by a<br>trained physician, yield information that may<br>assist in diagnosis. | The InkSpace Imaging Small Body Array is a<br>receive-only coil, used for obtaining diagnostic<br>images of general human anatomy in adult and<br>pediatric patients such as cardiac, spine,<br>shoulder, elbow, knee, foot, and prostate in<br>Siemens 3.0T magnetic resonance imaging<br>systems. These images, when interpreted by a<br>trained physician, yield information that may assist<br>in diagnosis. | Indications for Use of the<br>modified device maintains the<br>same general purpose and<br>function as the predicate device,<br>differing only by target MRI<br>system field strength (1.5T vs.<br>3.0T). |
| Patient Anatomy<br>and Population | Imaging of general human anatomy, such as<br>cardiac, spine, shoulder, elbow, knee, foot, and<br>prostate. | Imaging of general human anatomy, such as<br>cardiac, spine, shoulder, elbow, knee, foot, and<br>prostate. | N/A (Same). |
{8}------------------------------------------------
| Table 1: Summary of Technological Characteristics (cont.) | |
|-----------------------------------------------------------|--|
|-----------------------------------------------------------|--|
| Table 1: Summary of Technological Characteristics (cont.) | | | |
|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Characteristics | Subject Device:<br>InkSpace Imaging Small Body Array - 1.5T | Predicate Device:<br>InkSpace Imaging Small Body Array - 3.0T<br>(K233444) | Substantial Equivalence<br>Rationale |
| Comparison of<br>Technological<br>Characteristics | 24 channel, receive-only phased array, designed<br>for 1.5T MR systems (63.6 MHz, hydrogen).<br>Device is reusable, non-sterile, and by<br>prescription only.<br>Device is made of soft, pliable materials, and is<br>composed of two identical 12ch pads. | 24 channel, receive-only phased array, designed<br>for 3.0T MR systems (123.2 MHz, hydrogen).<br>Device is reusable, non-sterile, and by prescription<br>only.<br>Device is made of soft, pliable materials, and is<br>composed of two identical 12ch pads. | Different receive frequencies<br>(63.6 MHz vs. 123.2 MHz) due to<br>operation at 1.5T vs. 3.0T. This<br>difference is fully verified. |
| Compatible MRI<br>Systems | Siemens 1.5T MRI Systems | Siemens 3.0T MRI Systems | Different MR field strength, same<br>OEM. This difference is fully<br>verified. |
| Biocompatibility | ISO 10993 testing | ISO 10993 testing | N/A (Same). |
{9}------------------------------------------------
#### SUBSTANTIAL EQUIVALENCE
The InkSpace Imaging Small Body Array for 1.5T scanners is substantially equivalent to the predicate device, the InkSpace Imaging Small Body Array for 3.0T scanners, with regard to function, intended use, and physical characteristics. There are no differences in the technological characteristics between the devices which would raise any different questions of safety or effectiveness. Thus, the InkSpace Imaging Small Body Array is substantially equivalent to the predicate device.
### PERFORMANCE DATA [807.92(B)]
Performance testing was performed in accordance with the FDA Guidance Document entitled, "Magnetic Resonance (MR) Receive-only Coil – Performance Criteria for Safety and Performance Based Pathway," issued December 11, 2020.
Some of the nonclinical testing conducted on the InkSpace Imaging Small Body Array for 3.0T scanners (predicate device) is still applicable to the modified InkSpace Imaging Small Body Array for 1.5T scanners (subject of this submission).
Additional analysis was conducted based on the differences between 1.5T MR and 3.0T MR systems. These results and representative scan images listed below have been included as part of this submission.
### [807.92(b)(1)] Nonclinical Testing Summary:
The nonclinical. bench tests, per the FDA Guidance Document entitled. "Magnetic Resonance (MR) Receive-only Coil - Performance Criteria for Safety and Performance Based Pathway," were performed on the InkSpace Imaging Small Body Array for 1.5T scanners:
- . Image Signal to Noise (SNR)
- Image Uniformity
- Surface Heating
- Decoupling Circuit
- EMC - Immunity, Electrostatic Discharge
- General Electrical/Mechanical Safety
- Acquired Image Quality ●
In addition to the above listed nonclinical testing, a risk analysis was performed to validate that the following testing performed on the InkSpace Imaging Small Body Array for 3.0T scanners (predicate device) could be applied to the InkSpace Imaging Small Body Array for 1.5T scanners without retesting:
- Biocompatibility Testing (Cytotoxicity, Sensitization, Irritation) ●
- Usability Testing ●
- Transit Testing
The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and design of the InkSpace Imaging Small Body Array meet the
{10}------------------------------------------------
established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that the InkSpace Imaging Small Body Array does not raise different questions of safety or effectiveness for MR imaging examinations with Siemens 1.5T MRI Systems when compared to the predicate device.
# [807.92(b)(2)] Clinical Testing Summary:
Not applicable. Clinical testing was not required to demonstrate substantial equivalence of the modified InkSpace Imaging Small Body Array to the predicate device. Acquired image quality was analyzed based on sample clinical images. These images, as well as a review of their clinical quality by a board-certified radiologist, have been included as part of this submission.
### CONCLUSIONS [807.92(B)(3)]
Based on the results from the nonclinical tests performed in support of the InkSpace Imaging Small Body Array, it is concluded that the modified device is safe, effective, and performs at least as safely and effectively as the legally marketed predicate device.
### SUMMARY
The InkSpace Imaging Small Body Array is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.