← Product Code [MOS](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS) · K233652

# Contour Hand/Wrist (Q7000232) (K233652)

_Quality Electrodynamics · MOS · Feb 16, 2024 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS/K233652

## Device Facts

- **Applicant:** Quality Electrodynamics
- **Product Code:** [MOS](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS.md)
- **Decision Date:** Feb 16, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The Contour Hand/Wrist is intended for use with Siemens 0.55T MR systems to produce diagnostic images of hand and wrist anatomy that can be interpreted by a trained physician.

## Device Story

Contour Hand/Wrist is a receive-only, 12-channel phased array RF coil; designed for use with Siemens 0.55T MR systems. Device captures magnetic resonance signals from hand and wrist anatomy; transforms signals into diagnostic images for interpretation by trained physicians. Used in clinical imaging environments; operated by MRI technicians or radiologists. Provides high-quality diagnostic images to assist in clinical decision-making and patient diagnosis. Incorporates active PIN diode switching and passive blocking circuitry for signal management. Enclosed in rigid polycarbonate housing with flexible nylon and aramid fabric cover. Benefits patients by enabling non-invasive visualization of musculoskeletal structures.

## Clinical Evidence

Bench testing only. SNR and image uniformity measured on Siemens 0.55T MR system per NEMA MS-9 standards. Clinical evidence consists of diagnostic quality sample images obtained from volunteer scanning of hand/wrist anatomy to demonstrate performance. No adverse events reported.

## Technological Characteristics

12-channel receive-only phased array RF coil. Materials: Polycarbonate housing, aramid felt, polyurethane-coated nylon fabric. Sensing: Active PIN diode switching and passive blocking circuitry. Connectivity: Compatible with Siemens 0.55T MR systems. Standards: NEMA MS-9, NEMA MS-6, IEC standards for electrical safety and EMC.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Contour Knee ([K223429](/device/K223429.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.

February 16, 2024

Quality Electrodynamics Jovanna Boudreaux Quality/Regulatory Engineer 6655 Beta Drive, Suite 100 Mayfield Village, Ohio 44143

Re: K233652

Trade/Device Name: Contour Hand/Wrist (07000232) Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: MOS Dated: January 18, 2024 Received: January 18, 2024

Dear JoVanna Boudreaux:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

D.G.K.

Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

Device Name

| Contour Hand/Wrist (Q7000232) |  |
|-------------------------------|--|
|-------------------------------|--|

Indications for Use (Describe)

The Contour Hand/Wrist is intended for use with Siemens 0.55T MR systems to produce diagnostic images of hand and wrist anatomy that can be interpreted by a trained physician.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) #: K233652

510(k) Summary

Prepared on: 2024-02-16

| Contact Details                    |                                                                   | 21 CFR 807.92(a)(1) |
|------------------------------------|-------------------------------------------------------------------|---------------------|
| Applicant Name:                    | Quality Electrodynamics                                           |                     |
| Applicant Address:                 | 6655 Beta Drive Suite 100 Mayfield Village OH 44143 United States |                     |
| Applicant Contact Telephone:       | 4404842340                                                        |                     |
| Applicant Contact:                 | Mrs. JoVanna Boudreaux                                            |                     |
| Applicant Contact Email:           | jovanna.boudreaux@qualedyn.com                                    |                     |
| Device Name                        |                                                                   | 21 CFR 807.92(a)(2) |
| Device Trade Name:                 | Contour Hand/Wrist (Q7000232)                                     |                     |
| Common Name                        | Magnetic resonance diagnostic device                              |                     |
| Classification Name                | Coil, Magnetic Resonance, Specialty                               |                     |
| Regulation Number                  | 892.1000                                                          |                     |
| Product Code                       | MOS                                                               |                     |
| Legally Marketed Predicate Devices |                                                                   | 21 CFR 807.92(a)(3) |
| Predicate #                        | Predicate Trade Name (Primary Predicate is listed first)          | Product Code        |
| K223429                            | Contour Knee                                                      | MOS                 |

#### Device Description Summary

The Contour Hand/Wrist is a receive-only, 12-channel phased array coil designed for magnetic resonance imaging (MRI) using the Siemens 0.55T MR systems. The Contour Hand/Wrist is intended to be used for imaging hand and wrist anatomy.

#### Intended Use/Indications for Use

The Contour Hand/Wrist is intended for use with Siemens 0.55T MR systems to produce diagnostic images of hand and wrist anatomy that can be interpreted by a trained physician.

#### Indications for Use Comparison

The Indications for Use statement for the Contour Hand/Wrist is not identical to that of the predicate device (Contour Knee); however, the differences do not affect the safety or effectiveness of the device relative to the predicate device. Both Indications for the proposed Contour Hand/Wrist and predicate Contour Knee indicate that the device is intended to be used in conjunction with a MR system to produce images of human anatomy and that the images can be interpreted by a trained physician. The indications for use statements differ only in that the proposed Contour Hand/Wrist is intended for use to image hand and wrist anatomy instead of knee anatomy.

### Technological Comparison

At a high level, the proposed and predicate device are based on the following same technological elements:

- Receive-only phased array RF coil
- · Active PIN diode switching blocking circuitry. Passive blocking circuitry.
- Materials used for flame retardancy and biocompatibility: Polycarbonate and aramid felt with a polyurethane coated nylon fabric cover

The following technological differences exist between the proposed and predicate device:

- Intended for use (hand and wrist anatomy (proposed device) versus knee anatomy (predicate device))
#### 21 CFR 807.92(a)(4)

# 21 CFR 807.92(a)(5)

21 CFR 807.92(a)(5)

# 21 CFR 807.92(a)(6)

{4}------------------------------------------------

• The posterior coil elements are enclosed in a rigid polycarbonate housing with a flexible nylon and aramid fabric (proposed device) versus flexible blanket-like enclosure for anterior coil elements, rigid plastic housing for posterior coil elements (predicate device)

#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

The signal-to-noise ratio (SNR) and image uniformity of the Contour Hand/Wrist were measured on a 0.55T Siemens MR System, manufactured by Siemens Healthineers. The SNR and uniformity of the Contour Hand/Wrist were analyzed per NEMA MS-9 (using alternate method 2,5 from MS-6) and uniformity was analyzed using NEMA MS-9 (primary method from MS-6) using pre-determined acceptance criteria.

In accordance with the FDA Guidance for Industry: Guidance for the Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices, clinical images from volunteer scanning of hand/wrist anatomy were obtained from a Siemens 0.55T MR system. These images were used to demonstrate that the Contour Hand/Wrist produces diagnostic quality images of the intended anatomy. No adverse events were reported or recorded.

The electrical safety and electromagnetic compatibility and biocompatibility data support the safety of the Contour Hand/Wrist and the bench testing per the IEC standards and diagnostic quality sample clinical images demonstrates the performance and effectiveness of the device under the specified use conditions. This testing demonstrates that the Contour Hand/Wrist performs as well as or better than the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS/K233652](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/MOS/K233652)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
