uMR 680

K252371 · Shanghai United Imaging Healthcare Co., Ltd. · LNH · Sep 25, 2025 · Radiology

Device Facts

Record IDK252371
Device NameuMR 680
ApplicantShanghai United Imaging Healthcare Co., Ltd.
Product CodeLNH · Radiology
Decision DateSep 25, 2025
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1000
Device ClassClass 2

Indications for Use

The uMR 680 system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces sagittal, transverse, coronal, and oblique cross sectional images, and spectroscopic images, and that display internal anatomical structure and/or function of the head, body and extremities. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of interest of the scan.

Device Story

The uMR 680 is a 1.5T superconducting MRI system with a 70cm bore. It acquires anatomical and spectroscopic images of the head, body, and extremities. The system comprises a magnet, RF power amplifier, gradient coils, patient table, and spectrometer. This submission adds a Tx/Rx Head Coil and a mobile trailer configuration. Operated by trained clinicians in clinical or mobile settings, the device processes raw RF signals into cross-sectional images. These images are interpreted by physicians to assist in diagnosis. The mobile configuration allows for flexible deployment while maintaining the same diagnostic performance as the fixed-site predicate. The device benefits patients by providing non-invasive diagnostic imaging.

Clinical Evidence

No clinical trials were conducted. Evidence consists of bench testing and clinical image quality evaluation. A U.S. board-certified radiologist reviewed sample images from the new Tx/Rx Head Coil and mobile configuration, confirming diagnostic quality equivalent to the predicate device.

Technological Characteristics

1.5T superconducting magnet; 70cm bore; 45mT/m max gradient amplitude; 200T/m/s max slew rate; 1 transmit channel; up to 96 receive channels; 18 kW amplifier peak power per channel. Conforms to NEMA MS 1, 3, 6, 8, 9, 14; IEC 60601-1, 60601-1-2, 60601-2-33, 60601-4-2; ANSI/AAMI ES60601-1. Mobile configuration installed in a trailer.

Indications for Use

Indicated for patients requiring diagnostic MR imaging of the head, body, and extremities to visualize internal anatomical structures and/or function. No specific age or gender contraindications provided.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION September 25, 2025 Shanghai United Imaging Healthcare Co., Ltd. Gao Xin RA Manager No.2258 Chengbei Rd. Jiading District Shanghai, 201807 China Re: K252371 Trade/Device Name: uMR 680 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH, MOS, QIH Dated: July 30, 2025 Received: September 5, 2025 Dear Gao Xin: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252371 - Gao Xin Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K252371 - Gao Xin Page 3 Sincerely, NINGZHI LI -S Digitally signed by NINGZHI LI -S for Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252371 | ? | | Please provide the device trade name(s). | | ? | | uMR 680 | | | | Please provide your Indications for Use below. | | ? | | The uMR 680 system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces sagittal, transverse, coronal, and oblique cross sectional images, and spectroscopic images, and that display internal anatomical structure and/or function of the head, body and extremities. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of interest of the scan. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} Shanghai United Imaging Healthcare Co., Ltd. Tel: +86 (21) 67076888 Fax: +86 (21) 67076889 www.united-imaging.com UNITED IMAGING ^{}[] 510(k) SUMMARY K252371 1. Date of Preparation July 25, 2025 2. Sponsor Identification Shanghai United Imaging Healthcare Co., Ltd. No.2258 Chengbei Rd. Jiading District, 201807, Shanghai, China Contact Person: Xin GAO Position: Regulatory Affairs Specialist Tel: +86-21-67076888-5386 Fax: +86-21-67076889 Email: xin.gao@united-imaging.com 3. Identification of Proposed Device Trade Name: uMR 680 Common Name: Magnetic Resonance Imaging System Model: uMR 680 Regulatory Information Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: II Product Code: LNH Secondary Product Code: MOS, QIH Review Panel: Radiology 4. Identification of Primary/Reference Device(s) Predicate Device 510(k) Number: K243397 Device Name: uMR 680 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: II Product Code: LNH, MOS Secondary Product Code: QIH Page 1 of 8 {5} Shanghai United Imaging Healthcare Co., Ltd. Tel: +86 (21) 67076888 Fax: +86 (21) 67076889 www.united-imaging.com UNITED IMAGING Review Panel: Radiology 5. Device Description The uMR 680 is a 1.5T superconducting magnetic resonance diagnostic device with a 70cm size patient bore. It consists of components such as magnet, RF power amplifier, RF coils, gradient power amplifier, gradient coils, patient table, spectrometer, computer, equipment cabinets, power distribution system, internal communication system, and vital signal module etc. The uMR 680 Magnetic Resonance Diagnostic Device is designed to conform to NEMA and DICOM standards. This special 510(k) is to request modifications for the cleared uMR 680(K243397). The modifications performed on the uMR 680 in this submission are due to the following changes that include: (1) Addition of RF coils: Tx/Rx Head Coil. (2) Addition of a mobile configuration. 6. Indications for Use The uMR 680 system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces sagittal, transverse, coronal, and oblique cross sectional images, and spectroscopic images, and that display internal anatomical structure and/or function of the head, body and extremities. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of interest of the scan. 7. Comparison of Technological Characteristics with the Predicate Device uMR 680 employs the same basic operating principles and fundamental technologies, and has the same indications for use as the predicate device. A comparison between the technological characteristics of proposed and predicate/reference devices is provided as below. Table 1 Comparison of Indication for Use | ITEM | Proposed Device uMR 680 | Predicate Device uMR 680 (K243397) | Remark | | --- | --- | --- | --- | | indications for use | The uMR 680 system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces sagittal, transverse, | The uMR 680 system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces sagittal, transverse, | Same | Page 2 of 8 {6} Shanghai United Imaging Healthcare Co., Ltd. Tel: +86 (21) 67076888 Fax: +86 (21) 67076889 www.united-imaging.com UNITED IMAGING | coronal, and oblique cross sectional images, and spectroscopic images, and that display internal anatomical structure and/or function of the head, body and extremities. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of interest of the scan. | coronal, and oblique cross sectional images, and spectroscopic images, and that display internal anatomical structure and/or function of the head, body and extremities. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist the diagnosis. Contrast agents may be used depending on the region of interest of the scan. | | --- | --- | Table 2 Comparison to Predicate device | ITEM | Proposed Device uMR 680 | Predicate Device uMR 680(K243397) | Remark | | --- | --- | --- | --- | | General | | | | | Magnet system | | | | | Field Strength | 1.5 Tesla | 1.5 Tesla | Same | | Type of Magnet | Superconducting | Superconducting | Same | | Patient-accessible bore dimensions | 70cm | 70cm | Same | | Type of Shielding | Actively shielded, OIS technology | Actively shielded, OIS technology | Same | | Gradient system | | | | | Max gradient amplitude | 45mT/m | 45mT/m | Same | | Max slew rate | 200T/m/s | 200T/m/s | Same | | Shielding | active | active | Same | | Cooling | water | water | Same | | RF system | | | | | Resonant frequencies | 63.87 MHz | 63.87 MHz | Same | | Number of transmit channels | 1 | 1 | Same | | Number of receive channels | Up to 96 | Up to 96 | Same | | Amplifier peak power per channel | 18 kW | 18 kW | Same | | RF Coils | | | | | Tx/Rx Head Coil | Yes | No | Note 1 | | Mobile Configuration | | | | | Mobile configuration | Yes | No | Note 2 | | Note 1 | The intended use of Tx/Rx Head Coil is essentially identical to previously cleared Head Coil - 16. The difference is that the Tx/Rx Head Coil is of a 16 Legs High-Pass Birdcage used for transmitter coil. The difference did not raise new safety and effectiveness concerns. | | --- | --- | Page 3 of 8 {7} Shanghai United Imaging Healthcare Co., Ltd. Tel: +86 (21) 67076888 Fax: +86 (21) 67076889 www.united-imaging.com UNITED IMAGING | Note 2 | The proposed device with mobile configuration has the same intended use and same basic technological characteristics than the predicate device system, with respect to the magnetic resonance features and functionalities. There are no differences in technical features compared to the predicate device, but minor modified hardware and an installation in a mobile trailer. The resulting differences have been tested and the conclusions from all verification and validation data suggest that the system bears an equivalent safety and performance profile to that of the predicate device and reference devices. The difference did not raise new safety and effectiveness concerns. | | --- | --- | Page 4 of 8 {8} Shanghai United Imaging Healthcare Co., Ltd. Tel: +86 (21) 67076888 Fax: +86 (21) 67076889 www.united-imaging.com UNITED IMAGING # 8. Performance Data The following performance data were provided in support of the substantial equivalence determination. | Device Change | Risks | Verification/Validation Method(s) | Acceptance Criteria | Summary of Results | | --- | --- | --- | --- | --- | | Add RF coils: Tx/Rx Head Coil | • Biocompatibility • EMC-immunity, electrostatic discharge • SNR and Uniformity • Surface Heating • Clinical image quality • General electrical/mechanical safety • SAR | • Biocompatibility evaluation in agreement with recommendations in Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". • Perform test according to IEC 60601-1-2 and IEC 60601-4-2. • Perform test according to NEMA MS 1, NEMA MS 3, NEMA MS 6 and NEMA MS 9. • Perform test according to NEMA MS 14. • Evaluate the image generated by RF coils: Tx/Rx Head Coil with the same method as the predicate device. • Perform test according to ANSI/AAMI ES60601-1. • Perform test according to IEC 60601-2-33 and NEMA MS 8. | • Materials of construction and manufacturing materials do not introduce chemicals that raise a biocompatibility concern. • Conform with IEC 60601-1-2 and IEC 60601-4-2. • SNR and Uniformity shall fulfill with the design specification. • The maximum temperature of all temperature probes shall not exceed 41°C. • Image quality is sufficient for diagnostic use. • Conform with ANSI/AAMI ES60601-1. • Conform with IEC 60601-2-33. | • Biocompatibility testing is not needed because device does not introduce a biocompatibility risk. • Pass • Pass • Pass • The U.S. Board Certified radiologist confirms that image quality is sufficient for diagnostic use. • Pass • Pass | Page 5 of 8 {9} Shanghai United Imaging Healthcare Co., Ltd. Tel: +86 (21) 67076888 Fax: +86 (21) 67076889 www.united-imaging.com | Add mobile configuration | • EMC • Clinical image quality • General electrical/mechanical safety | • Perform test according to IEC 60601-1-2 and IEC 60601-4-2. • Evaluate the image generated by RF coils: Tx/Rx Head Coil with the same method as the predicate device. • Perform test according to ANSI/AAMI ES60601-1. | • Conform with IEC 60601-1-2 and IEC 60601-4-2. • Image quality is sufficient for diagnostic use. • Conform with ANSI/AAMI ES60601-1. | • Pass • The U.S. Board Certified radiologist confirms that image quality is sufficient for diagnostic use. • Pass | | --- | --- | --- | --- | --- | Page 6 of 8 {10} Shanghai United Imaging Healthcare Co., Ltd. Tel: +86 (21) 67076888 Fax: +86 (21) 67076889 www.united-imaging.com UNITED IMAGING # Non-Clinical Testing Non-clinical testing including surface heating, SAR and image performance tests were conducted for the uMR 680 to verify that the proposed device met all design specifications as it is Substantially Equivalent (SE) to the predicate device. UNITED IMAGING HEALTHCARE claims conformance to the following standards and guidance: ## Electrical Safety and Electromagnetic Compatibility (EMC) - ANSI/AAMIES60601-1: 2005/ (R) 2012+A1:2012+C1:2009/(R)2012+A2:2010/(R)2012) [Including Amendment2(2021)] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - IEC 60601-1-2:2014+A1:2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests - IEC 60601-2-33 Ed. 4.0:2022 Medical Electrical Equipment - Part 2-33: Particular Requirements for The Basic Safety and Essential Performance of Magnetic Resonance Equipment for Medical Diagnostic - IEC 60601-1-6:2010+A1:2013+A2:2020, Edition 3.2, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability. - NEMA MS 1-2008(R2020), Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Images - NEMA MS 3-2008(R2020), Determination of Image Uniformity in Diagnostic Magnetic Resonance Images - NEMA MS 6-2008(R2014, R2020), Determination of Signal-to-Noise Ratio and Image Uniformity for Single-Channel Non-Volume Coils in Diagnostic MR Imaging - NEMA MS 9-2008(R2020), Standards Publication Characterization of Phased Array Coils for Diagnostic Magnetic Resonance Images - NEMA MS 14-2019, Characterization of Radiofrequency (RF) Coil Heating in Magnetic Resonance Imaging Systems - IEC /TS 60601-4-2: 2024, Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems ## Software - IEC 62304:2006+AMD1:2015 CSV Consolidated version, Medical device software - Software life cycle processes - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Page 7 of 8 {11} Shanghai United Imaging Healthcare Co., Ltd. Tel: +86 (21) 67076888 Fax: +86 (21) 67076889 www.united-imaging.com UNITED IMAGING > Content of Premarket Submissions for Management of Cybersecurity in Medical Devices ## Biocompatibility > ISO 10993-5: 2009, Edition 3.0, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity. > ISO 10993-10: 2021, Edition 4.0, Biological evaluation of medical devices - Part 10: Tests for skin sensitization. > ISO 10993-23: 2021, Edition 1.0, Biological evaluation of medical devices - Part 10: Tests for irritation. > Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" ## Other Standards and Guidance > ISO 14971: 2019, Edition 3.0, Medical Devices – Application of risk management to medical devices > Code of Federal Regulations, Title 21, Part 820 - Quality System Regulation > Code of Federal Regulations, Title 21, Subchapter J - Radiological Health ## Performance Verification Non-clinical testing was conducted to verify the features described in this premarket submission. > Sample clinical images for Tx/Rx Head Coil were reviewed by U.S. board-certified radiologist comparing the proposed device and predicate device. It was shown that the proposed device can generate diagnostic quality images in accordance with the MR guidance on premarket notification submissions. ## 9. Conclusion Based on the comparison and analysis above, the proposed device has similar indications for use, performance, safety equivalence, and effectiveness as the predicate device. The differences above between the proposed device and predicate device do not affect the intended use, technology characteristics, safety, and effectiveness. And no issues are raised regarding to safety and effectiveness. The proposed device is determined to be Substantially Equivalent (SE) to the predicate device. Page 8 of 8
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...