K111566 · Carestream Health, Inc. · LMC · Oct 6, 2011 · Radiology
Device Facts
Record ID
K111566
Device Name
DRYVIEW CHROMA IMAGING SYSTEM
Applicant
Carestream Health, Inc.
Product Code
LMC · Radiology
Decision Date
Oct 6, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2040
Device Class
Class 2
Indications for Use
The DRYVIEW CHROMA Imaging System is intended to provide hard copy images from digital imaging source output signals. The device is intended for use with DRY VIEW CHROMA film and reflective media. The printer will interface with a variety of digital modalities, including, but not limited to, CR (Computed Radiology), DR (Digital Radiology), CT (Computerized Tomography) and MRI (Magnetic Resonance Imaging). The images are to be used for medical diagnosis and referral to physicians and their patients. The DRYVIEW CHROMA Imaging System is not intended for use with FFDM or CR Mammography systems.
Device Story
The DRYVIEW CHROMA Imaging System is an inkjet printer producing hard copy medical images and reports from digital modality inputs (CR, DR, CT, MRI). It receives data via DICOM network interface. The device uses piezoelectric inkjet technology to propel ink droplets onto transparent polyester film or reflective paper media. A built-in test pattern generator ensures consistency between input signals and output density. Operated by medical personnel in clinical settings, the system facilitates medical diagnosis and patient referral. It does not monitor or affect patients directly. Healthcare providers review the printed output for clinical decision-making. The system is designed for consistency with predicate hardcopy imagers.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through adherence to voluntary standards including DICOM, IEC 62304, UL 60950, IEC 60601-1-2, and ISO 14971. Consistency verified via test pattern generator.
Technological Characteristics
Inkjet printing system using piezoelectric nozzles. Media: transparent polyester film and reflective paper. Connectivity: DICOM network interface. Standards: IEC 62304 (software), UL 60950 (safety), IEC 60601-1-2 (EMC), ISO 14971 (risk management).
Indications for Use
Indicated for providing hard copy medical images from digital modalities (CR, DR, CT, MRI) for diagnosis and referral. Not for use with FFDM or CR Mammography systems.
Regulatory Classification
Identification
A medical image hardcopy device is a device that produces a visible printed record of a medical image and associated identification information. Examples include multiformat cameras and laser printers.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Codonics Horizon Ci Medical Image Multimedia Imager (K021054)
Related Devices
K042159 — KODAK COLOR MEDICAL IMAGER 1000 · Eastman Kodak Company · Sep 24, 2004
K083522 — SONY UP-DF750 DIGITAL FILM IMAGER · Sony Electronics, Inc. · Jan 9, 2009
K032635 — DRYSTAR 5300 MEDICAL IMAGE PRINTER, MODEL 5300 · Agfa Corp. · Sep 24, 2003
K972660 — RAVEN DRY PRINTER · Sterling Diagnostic Imaging, Inc. · Nov 10, 1997
K024188 — UP-DF500 FILMSTATION DIGITAL FILM IMAGER · Sony Medical Systems · Feb 4, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
#### Carestream Health, Inc.
#### Traditional 510k DRYVIEW CHROMA Imaging System
OCT - 6 2011
## Section B
1.
# 510(k) Summary
June 3, 2011
CARESTREAM Health, Inc. 150 Verona Street Rochester, NY 14608
Contact: Christine Ehmann
Phone: 585-627-6473 FAX: 585-323-7643
#### Device
| Trade name: | DRYVIEW CHROMA Imaging System |
|----------------------|-------------------------------------------------|
| Common name: | Inkjet Printer |
| Classification name: | Medical Image Hardcopy Device (21 CFR 892.2040) |
Codonics Horizon Ci Medical Image Multimedia Imager Predicate device (K021054)
#### Description and Intended Use of Device
The DRYVIEW CHROMA Imaging System is intended to provide hard copy images from digital imaging source output signals. The device is intended for use with DRYVIEW CHROMA film and reflective media. The device will interface with a variety of digital modalities, including, but not limited to, CR (Computed Radiology), DR (Digital Radiology), CT (Computerized Tomography), MRI (Magnetic Resonance Imaging). The images are to be used for medical diagnosis and referral to physicians and their patients. The DRYVIEW CHROMA Imaging System is not intended for use with FFDM or CR Mammography svstems.
#### Technological Characteristics
The DRYVIEW CHROMA Imaging System is an inkjet printing system. The DRYVIEW CHROMA Imaging System (CHROMA System) receives medical data including image and clinical report data from a digital modality. This data is received from medical image source devices (modalities) over a network and communicated to the CHROMA device via digital communication standard, DICOM. User control is performed directly by the modality or through the host control.
{1}------------------------------------------------
The CHROMA System device prints the information received using piezoelectric inkjet technology. Tiny ink droplets are propelled from pizeoelectric nozzles in inige toenhology. " in media to form the image or report communicated by the digital modality.
The CHROMA System device prints on transparent polyester based (film) as well as reflective (paper) media. Media is removed from a cartridge and transported into the CHROMA System device. Print data and media merge within the device.
The CHROMA System employs the use of test patterns to verify imaging performance. A test pattern generator is incorporated to assure consistency between input signals and output density.
Software is used to control the image management and machine functions. The information sent to the DICOM interface is used by the DICOM Interface to choose the correct set of printing parameters and halftone patterns for optimal image quality.
#### Performance Data
Safety and effectiveness are assured via meeting voluntary standards, including: DICOM, IEC 62304, UL 60950, IEC 60601-1-2 and ISO 14971.
#### Conclusion
The CHROMA System, like the predicate, has no patient contact. The CHROMA System device does not control, monitor or otherwise affect any devices directly connected to or affecting the patient. Medical personnel review images and hardcopy reports printed by the subject device and its predicate. The images or reports may be reprinted, if desired, by the end user since the original image data or report is stored on a connected modality..
As with the predicate device, a test pattern generator is incorporated to assure consistency between input signals and output density. Both are hardcopy printers that employ the use of test patterns to verify imaging performance.
The DRYVIEW CHROMA Imaging System and predicate device Codonics Horizon Ci Medical Image Multimedia Imager (Codonics Horizon Ci) have both been designed to the equivalent or comparable safety standards. The hazards associated with the DRYVIEW CHROMA Imaging System do not pose any new safety or efficacy issues compared to those associated with the predicate device as well as other medical image hardcopy devices.
Carestream Health concludes that the DRYVIEW CHROMA Imaging System is as safe and effective as the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circle with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is an image of a stylized bird. The bird is made up of three curved lines that represent the wings and body.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Ms. Christine Ehmann Regulatory Affairs Director CARESTREAM Health, Inc. 150 Verona Street ROCHESTER NY 14608
- 6 2011
Re: K111566
Trade/Device Name: DRYVIEW CHROMA Imaging System Regulation Number: 21 CFR 892.2040 Regulation Name: Medical image hardcopy device Regulatory Class: II Product Code: LMC Dated: August 10, 2011 Received: August 12, 2011
Dear Ms. Ehmann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
ff your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not inean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reportion of
{3}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This feter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (2) CFR Parts \$01 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours.
Mary Patel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known): K111566
Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications for Use:
The DRYVIEW CHROMA Imaging System is intended to provide hard copy images from digital imaging source output signals. The device is intended for use with DRY VIEW CHROMA film and reflective media. The printer will interface with a variety of digital modalities, including, but not limited to, CR (Computed Radiology), DR (Digital Radiology), CT (Computerized Tomography) and MRI (Magnetic Resonance Imaging). The images are to be (sed for medical diagnosis and referral to physicians and their patients. The DRYVIEW about of maging System is not intended for use with FFDM or CR Mammography systems.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mary Spottl
Office of In Vitro Diagnostic L
510K K11566
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.