KODAK COLOR MEDICAL IMAGER 1000

K042159 · Eastman Kodak Company · LMC · Sep 24, 2004 · Radiology

Device Facts

Record IDK042159
Device NameKODAK COLOR MEDICAL IMAGER 1000
ApplicantEastman Kodak Company
Product CodeLMC · Radiology
Decision DateSep 24, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2040
Device ClassClass 2

Indications for Use

The Kodak Color Medical Imager 1000 series medical printers are thermal print engines intended to produce continuous tone photographic quality hard copy output with integrated color management that adjusts printed colors to accurately match medical image monitors or user specified color profiles. The hardcopy output includes, however is not limited to, nuclear medicine, ultrasound, CT (especially 3-D reconstruction), MRI and Radiation Therapy planning. Images are suitable for medical image diagnosis use and referral. Digital images can be acquired from, but not limited to local area networks (LAN), Internet, medical image storage devices, or directly from digital medical image capture modalities. The system of printer and physical image is not a high-resolution device and is not intended for use with FFDM systems.

Device Story

Thermal print engine producing continuous tone photographic quality hard copy output from medical imaging modalities; inputs include digital image signals from LAN, Internet, storage devices, or direct modality connection; utilizes thermal dye sublimation technology to heat color donor ribbon and transfer dyes to receptor sheet; features integrated color management to match monitors or user profiles; operated by medical personnel in clinical settings; output used for diagnosis and referral; no patient contact; no direct control or monitoring of patient-connected devices.

Clinical Evidence

Bench testing only. No clinical data provided. Safety and effectiveness demonstrated through compliance with voluntary standards including DICOM, SMPTE, UL 60950, IEC 60601-1-1, ISO 12207, and ISO 14971.

Technological Characteristics

Thermal dye sublimation print engine; uses color donor ribbon and receptor sheet; integrated color management system; connectivity via LAN/Internet/direct modality interface; supports DICOM standards; software-controlled image management; no patient contact; non-high-resolution output.

Indications for Use

Indicated for producing continuous tone hard copy output of medical images for diagnosis and referral, including nuclear medicine, ultrasound, CT, MRI, and radiation therapy planning. Not intended for use with Full-Field Digital Mammography (FFDM) systems.

Regulatory Classification

Identification

A medical image hardcopy device is a device that produces a visible printed record of a medical image and associated identification information. Examples include multiformat cameras and laser printers.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K042/59 ## SEP 2 4 2004 ## Section B1 # 510(k) Summary September 14, 2004 Eastman Kodak Company 343 State Street Rochester NY 14650 Contact: Stephen Slavens 1 Imation Way, 304-3B-61 Oakdale, MN 55128 Phone: 651-393-1395 FAX: 651-393-1440 #### Device: Trade name: Common name: Classification name: KODAK Color Medical Imager 1000 KPRO 8500 Color Printer Medical Image Hardcopy Device 21 CFR 892.2040 Predicate device: Codonics EP-1000 Medical Color dry imager (K030690) #### Description And Intended Use of Device: The KODAK Color Medical Imager 1000 is intended for use as a hard copy device for output from imaging source modalities used in medical imaging diagnosis and referral. Electronic image information signals are managed in the KODAK Color Medical Imager 1000 and transformed to heat a color donor ribbon and drive color dyes to a receptor sheet using thermal dye sublimation technology. The Intended Use of the Kodak Color Medical Imager 1000 series medical printers is as thermal print engines intended to produce continuous tone photographic quality hard copy output with integrated color management that adjusts printed colors to accurately match medical image monitors or user specified color profiles. The hardcopy output includes, however is not limited to, nuclear medicine, ultrasound, CT (especially 3-D reconstruction), MRI and Radiation Therapy planning. Images are suitable for medical image diagnosis use and referral. Digital images can be acquired from, but not limited to local area networks (LAN), Internet, medical image storage devices, or directly from digital medical image capture modalities. #### Technological Characteristics: The subject device and predicate devices use the same technical design base. The printers receive image data from the modality. User control is performed directly by the modality or through the host control. KODAK DRYVIEW imaging media is removed from a daylight cartridge and transported to the laser imaging station. Image data and media merge at the laser station and the film is scanned. The exposed media is transported through the integrated processor and exits the printer. {1}------------------------------------------------ Software is used to control the image management and machine functions #### Performance Data: Safety and effectiveness are assured via meeting voluntary standards, including: DICOM, SMPTE, UL 60950, IEC 60601-1-1, ISO 12207, and ISO 14971. #### Conclusion: The subject device, like the predicate, has no patient contact. The devices also do not control, monitor or otherwise affect any devices directly connected to or affecting the patient. Medical personnel review images displayed by the subject device and its predicates. This offers ample opportunity for competent human intervention in case of a malfunction or other failure. The subject KODAK Color Medical Imager 1000 and predicate device, the Codonics EP-1000 Medical Color dry imager (K030690) have been designed to the equivalent safety standards. Eastman Kodak therefore concludes that the KODAK Color Medical Imager 1000 is as safe and effective as the predicate device. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or a bird in flight. The symbol is composed of several curved lines that create the shape of the bird's wings and body. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 2 4 2004 Mr. Stephen Slavens Regulatory Affairs Director Eastman Kodak Company Health Imaging Group, Digital Output SPG 1 Imation Wav. 304-3B-61 OAKDALE MN 55128-3414 Re: K042159 Trade/Device Name: KODAK Color Medical Imager 1000 Regulation Number: 21 CFR 892.2040 Regulation Name: Medical image hardcopy device Regulatory Class: II Product Code: 90 LMC Dated: August 5, 2004 Received: August 10, 2004 · Dear Mr. Slavens: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean: that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) This letter whilanow you to ought manies. By substantial equivalence of your device to a legally premarket notification. The PPA mialing sification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice to: your of the following numbers, based on the regulation number at the top of the letter: | 8xx. 1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, for questions on and please note the regulation entitled, "Misbranding Other of Othphanoo at (2001) 21 CFR Part 807.97) you may obtain. Other general of received to promarity in the Act may be obtained from the Division of Small nitormation on your responsion. Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brigdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Section B2 ## Statement of Indications for Use: 510(K) Number (if known): _ Device Name: KODAK Color Medical Imager 1000 Indications for Use: The Kodak Color Medical Imager 1000 series medical printers are thermal print engines intended to produce continuous tone photographic quality hard copy output with integrated color management that adjusts printed colors to accurately match medical image monitors or user specified color profiles. The hardcopy output includes, however is not limited to, nuclear medicine, ultrasound, CT (especially 3-D reconstruction), MRI and Radiation Therapy planning. Images are suitable for medical image diagnosis use and referral. Digital images can be acquired from, but not limited to local area networks (LAN), Internet, medical image storage devices, or directly from digital medical image capture modalities. The system of printer and physical image is not a high-resolution device and is not intended for use with FFDM systems. Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use \checkmark OR Over the Counter Use **__** (Per 21 CFR 801.109) (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number **__** K042159
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...