MEDI-6000 MEDICAL IMAGE DIGITIZER

K091288 · Microtek International, Inc. · LMA · Aug 14, 2009 · Radiology

Device Facts

Record IDK091288
Device NameMEDI-6000 MEDICAL IMAGE DIGITIZER
ApplicantMicrotek International, Inc.
Product CodeLMA · Radiology
Decision DateAug 14, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2030
Device ClassClass 2

Indications for Use

The Medi-6000 consists of a transparency film digitizer and paper CCD scanner that is used to digitize radiographs or X-ray film as well as paper reports or doctor's orders. When the Medi-6000 is used to digitize radiographic films, the digital image is intended for use in primary, secondary and over reading applications. This device is not to be used for primary image diagnosis in mammography. The target users of the device are medical professionals or trained staffs.

Device Story

Medi-6000 is a transparency film digitizer and paper CCD scanner; converts analog X-ray films and paper medical documents into digital format for electronic storage and review. Operated by medical professionals or trained staff in clinical settings. Device captures image data from films; provides digital output for integration into PACS or teleradiology systems. Enables medical professionals to view digitized radiographs for primary, secondary, or over-reading applications; facilitates efficient data management and clinical decision-making by providing high-resolution digital access to patient records.

Clinical Evidence

Bench testing only. Performance testing demonstrated that the device meets its intended use specifications and requirements for integration as a component of a PACS or teleradiology system.

Technological Characteristics

Optical resolution: 600 dpi; dynamic range: 4.0 Dmax; 16-bit grayscale. Technology: Transparency film digitizer and paper CCD scanner. Intended for integration into PACS or teleradiology systems.

Indications for Use

Indicated for digitizing radiographs, X-ray films, paper reports, or doctor's orders for medical professionals or trained staff. Intended for primary, secondary, and over-reading applications. Contraindicated for primary image diagnosis in mammography.

Regulatory Classification

Identification

A medical image digitizer is a device intended to convert an analog medical image into a digital format. Examples include systems employing video frame grabbers, and scanners which use lasers or charge-coupled devices.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std.). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) Summary K091288 [As required by section 807.92(c)] 1. Submitter: Microtek International, Inc. 6, Industry East Road 3, Science-Based Industrial Park, Hsinchu, 30077, Taiwan, TEL: +886-3-5772155 FAX: +886-3-5772598 - 2. Official Correspondent: Mr. George Sun - 3. Date of 510(k) Submittal: 29. April, 2009 - 4. Device Trade Name Medi-6000 Medical Image Digitizer - 5. Common Name: Film digitizer - 6. Classification Name: Image digitizer was classified in class II (21 CFR 892.2030) - 7. Device Product Code: LMA - 8. Predicate Device: Manufacturer: Howtek Incorporated Device name: Film digitizer Model name: Fulcrum 510(k) No.: K021949 Manufacturer: VIDAR Systems Corporation Device name: X-ray film digitizer Model name: P111, TeleRADPRO, VXR-12 plus 510(k) No.: K993597 - 9. Device Description: The Medi-6000 is a digitizer that can easily transfer x-ray films into a digital format for patients, hospitals, and office records. It can capture -3- P. 1 of 2 {1}------------------------------------------------ details in bright and dark areas of x-ray films and provide the medical professionals a convenient method to digitize the roentgenogram for the electric data storage. - 10.Intended Use: The Medi-6000 consists of a transparency film digitizer and paper CCD scanner that is used to digitize radiographs or X-ray film as well as paper reports or doctor's orders. When the Medi-6000 is used to digitize radiographic films, the digital image is intended for use in primary, secondary and over reading applications. The target users of the device are medical professionals or trained staffs. - 11 Technological Characteristics: The Medi-6000 with a high optical resolution of 600 dpi, a dynamic range of 4.0 Dmax, and the 16-bit grayscale, the Medi-6000 allows users to get better image details in just seconds. - 12.Performance Testing: The performance testing results for the Medi-6000 digitizer demonstrated that the device meets its intended use specifications and therefore meets the requirements necessary for its intended use as a component of a PAC or Teleradiology system. - 13. Substantial Equivalence to Predicate Device: Medi-6000 is substantially equivalent to Fulcrum and P111. TeleRADPRO, and VXR-12 plus. Medi-6000 employs the resolution values same as that of Fulcrum, and better than P111, TeleRADPRO, and VXR-12 plus. Comparison table of the principal characteristics of these three devices is shown in the Section II and specification data for the Medi-6000 is included in Section I-3. 14.Conclusions: In terms of intended use, construction, function, safety, operating environmental conditions and effectiveness of the Medi-6000 film digitizer is substantially equivalent to the predicate devices used for this application. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 Microtek International, Inc. % Mr. Hui-chen Kai, Supervisor Electronics Testing Center, Taiwan No. 8, Lane 29, Wenming Rd. Guisan, Taoyuan, 33383 TAIWAN AUG 1 4 2009 Re: K091288 Trade/Device Name: Medi-6000 Medical Image Digitizer Regulation Number: 21 CFR 892.2030 Regulation Name: Medical image digitizer Regulatory Class: II Product Code: LMA Dated: July 17, 2009 Received: July 22, 2009 #### Dear Mr. Hui-chen Kai: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {3}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Lammy M. Moure Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use # 510(k) Number (if known): Not known K091288 Device Name: Medi-6000 Medical Image Digitizer ### Indications For Use: The Medi-6000 consists of a transparency film digitizer and paper CCD scanner that is used to digitize radiographs or X-ray film as well as paper reports or doctor's orders. When the Medi-6000 is used to digitize radiographic films, the digital image is intended for use in primary, secondary and over reading applications. This device is not to be used for primary image diagnosis in mammography. The target users of the device are medical professionals or trained staffs. Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number Lopui Nh Nhay vision of Reproductive, Abdominal and P. 1 of|
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...