PET-DOSE

K042746 · Comecer S.R.L. · KPT · Nov 22, 2004 · Radiology

Device Facts

Record IDK042746
Device NamePET-DOSE
ApplicantComecer S.R.L.
Product CodeKPT · Radiology
Decision DateNov 22, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1360
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

PET-Dose is a radionuclide dose calibrator designed to measure the amount of radioactity i Dr Dose is a racience and in any other user-defined geometry samples. It is intended to assay nr vials, syringos, and maily onlinistration to patients, by measuring the residual of radiopharmaceutical in vials, the activity of the injectable in syringes, the activity of capsules before they are administered to patients, etc. The best applications of PET-Dose are activity measurements of beta emitters used in PET diagnostics such as. 18F, 12N, 150, etc. PET-Dose is best fitted for use in nuclear medicine departments and, particularly, in PET centers. The PET-Dose is intended to be used by trained medical physicist or qualified experts or nuclear medicine technicians to perform activity measurements in the nuclear medicine departments and, particularly, in PET centers.

Device Story

PET-Dose is a radionuclide dose calibrator used in nuclear medicine departments and PET centers. It measures radioactivity in various geometries, including vials, syringes, and capsules. Operators (medical physicists, qualified experts, or technicians) place samples into the device to assay radiopharmaceuticals, including beta emitters like 18F, 13N, and 15O, prior to patient administration. The device measures residual activity in vials and injectable activity in syringes. Output provides quantitative activity measurements, allowing clinicians to verify correct dosage before administration, ensuring patient safety and diagnostic accuracy.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Radionuclide dose calibrator; designed for measurement of beta emitters (18F, 13N, 15O). Supports multiple sample geometries (vials, syringes, capsules). Class II device (21 CFR 892.1360).

Indications for Use

Indicated for activity measurement of radiopharmaceuticals (beta emitters like 18F, 13N, 15O) in vials, syringes, and capsules prior to patient administration in nuclear medicine and PET centers. Intended for use by trained medical physicists, qualified experts, or nuclear medicine technicians.

Regulatory Classification

Identification

A radionuclide dose calibrator is a radiation detection device intended to assay radionuclides before their administration to patients.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines. ## Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## NOV 2 2 2004 COMECER S.r.l. ,% Mr. N. E. Devine, Jr. Responsible Third Party Official Entela, Inc. 3033 Madison Ave. SE GRAND RAPIDS MI 49548 Re: K042746 Trade/Device Name: PET-Dose Regulation Number: 21 CFR 892.1360 Regulation Name: Radionuclide dose calibrator Regulatory Class: II Product Code: 90 KPT Dated: November 11, 2004 Received: November 12, 2004 . Dear Mr. Devine: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reasy by 17 is, accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally premarket nedicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you atte Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmaldsmamain.html Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K 042746 Device Name: PET-Dose Indications for Use: PET-Dose is a radionuclide dose calibrator designed to measure the amount of radioactity i Dr Dose is a racience and in any other user-defined geometry samples. It is intended to assay nr vials, syringos, and maily onlinistration to patients, by measuring the residual of radiopharmaceutical in vials, the activity of the injectable in syringes, the activity of capsules before they are administered to patients, etc. The best applications of PET-Dose are activity measurements of beta emitters used in PET diagnostics such as. 18F, 12N, 150, etc. PET-Dose is best fitted for use in nuclear medicine departments and, particularly, in PET centers. The PET-Dose is intended to be used by trained medical physicist or qualified experts or nuclear medicine technicians to perform activity measurements in the nuclear medicine departments and, particularly, in PET centers. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over the Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NIEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Nancy C. Gordon (Division (Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number K042746
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...