← Product Code [KPR](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR) · K242678

# Definium Pace Select ET (K242678)

_Ge Hualun Medical Systems Co. , Ltd. · KPR · Oct 1, 2024 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR/K242678

## Device Facts

- **Applicant:** Ge Hualun Medical Systems Co. , Ltd.
- **Product Code:** [KPR](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR.md)
- **Decision Date:** Oct 1, 2024
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The Definium Pace Select ET is intended to generate digital radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts in patients of all ages. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position and the system is intended for use in all routine radiography exams. Optional image pasting function enables the operator to stitch sequentially acquired radiographs into a single image. This device is not intended for mammographic applications.

## Device Story

Definium Pace Select ET is a stationary digital radiographic X-ray system; modular design includes 2-axis motorized tube stand, elevating table, motorized wall stand, high voltage generator, wireless detectors, and acquisition workstation. System captures X-ray signals; transforms inputs into digital diagnostic images for local review or DICOM network transmission. Operated by professionals in clinical settings. New features include motorized tube stand (vertical/angulation), auto-tracking for wall stand, auto-angulation, camera-assisted workflow (depth camera for patient positioning/size suggestion), and software-based DAP calculation. Image pasting function stitches sequential radiographs. Enhancements improve patient positioning, workflow efficiency, and exposure parameter management. Benefits include reduced manual effort for tube/table alignment and improved diagnostic image acquisition.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including risk analysis, requirements/design reviews, module verification, system integration testing, performance/safety verification, and simulated use validation.

## Technological Characteristics

Stationary X-ray system; 2-axis motorized tube stand; elevating table (max load 305kg); wireless 17x17 inch detectors (100um resolution); Siemens AL01CIIeL LED collimator; PSG-HR80S5 high voltage generator (50-80kW). Connectivity via DICOM 3.0 and 802.11 wireless. Software runs on Linux (SUSE).

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Definium Pace Select ([K231892](/device/K231892.md))

## Reference Devices

- Discovery XR656 HD with VolumeRad ([K191699](/device/K191699.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 1, 2024

GE Hualun Medical Systems Co., Ltd % Kenny Ma Manager, Regulatory Affairs - WHXR No.1, Yong Chang North Road, Beijing Economic Technological Development Zone Beijing, 100176 CHINA

Re: K242678

Trade/Device Name: Definium Pace Select ET Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: KPR, MQB Dated: September 3, 2024 Received: September 6, 2024

Dear Kenny Ma:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

Submission Number (if known)

K242678

Device Name

Definium Pace Select ET

### Indications for Use (Describe)

The Definium Pace Select ET is intended to generate digital radiographic images of the skull, spinal column, chest, abdomen, extremities, and other body parts in patients of all ages. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position and the system is intended for use in all routine radiography exams. Optional image pasting function enables the operator to stitch sequentially acquired radiographs into a single image. This device is not intended for mammographic applications.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image contains the GE HealthCare logo. The logo consists of the GE monogram enclosed in a circular design on the left, followed by the text "GE HealthCare" on the right. The text is in a simple, sans-serif font and is colored in the same shade of purple as the monogram.

# 510K Summary Definium Pace Select ET

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image shows the GE HealthCare logo. The logo consists of a purple circular emblem with a stylized "GE" inside, followed by the text "GE HealthCare" in a matching purple color. The text is simple and modern.

## 510(k) Summary:

In accordance with 21 CFR 807.92 the following summary information is provided:

| Date:                                       | September 6, 2024                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Owner/Submitter:<br>21 CFR 807.92(a)(1)     | GE Hualun Medical Systems Co., Ltd.<br>No.1, Yong Chang North Road, Beijing Economic Technological Development Zone, 100176 Beijing P.R. China                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Primary Contact<br>Person:                  | Kenny Ma<br>Manager, Regulatory Affairs, WHXR<br>GE Hualun Medical Systems Co., Ltd.<br>Email: Kenny.Ma@gehealthcare.com<br>Contact Phone Number: +86 (181) 01130591                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Secondary Contact<br>Person:                | Christopher Paulik<br>Senior Regulatory Affairs Manager<br>GE HealthCare (GE Medical Systems, LLC)<br>Email: Christopher.A.Paulik@gehealthcare.com<br>Contact Phone Number: +1 (262) 8945415                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Device Trade Name:<br>21 CFR 801.92(a)(2)   | Definium Pace Select ET                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Common/Usual<br>Name:                       | Digital Radiographic System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Regulation Name:                            | Stationary X-Ray System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Regulation:                                 | 21 CFR 892.1680                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Classification:                             | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Product Code:                               | KPR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Subsequent Product<br>Code(s):              | MQB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Predicate Device:<br>21 CFR 807.92(a)(3)    | Definium Pace Select (K231892)<br>21CFR 892.1680 (KPR, MQB)<br>Class 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Predicate Device<br>Manufacturer            | GE Hualun Medical Systems Co., Ltd.<br>No.1, Yong Chang North Road, Beijing Economic Technological Development Zone,<br>100176 Beijing P.R. China                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Reference Device:                           | Discovery XR656 HD with VolumeRad (K191699)<br>21CFR 892.1680 (KPR, MQB)<br>Class 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Reference Device<br>Manufacturer:           | GE Hualun Medical Systems Co., Ltd.<br>No.1, Yong Chang North Road, Beijing Economic Technological Development Zone,<br>100176 Beijing P.R. China                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Device Description:<br>21 CFR 807.92(a)(4)  | The Definium Pace Select ET Radiography X-ray System is designed as a modular<br>system with components that include a 2-axis motorized tube stand with tube and<br>auto collimator assembled on an elevating table, a motorized wall stand, a cabinet<br>with X-ray high voltage generator, a wireless access point, wireless detectors, an<br>acquisition workstation including a monitor and control box with hand-switch. The<br>system generates diagnostic radiographic images which can be reviewed or<br>managed locally and sent through a DICOM network for reviewing, storage and<br>printing.<br><br>By leveraging platform components / design, Definium Pace Select ET is similar to<br>the predicate Definium Pace Select (K231892) and the reference Discovery XR656<br>HD (K191699) with regards to the user interface layout, patient worklist refresh and<br>selection, protocol selection, image acquisition, and image processing based on the<br>raw image. This product introduces motorized tube stand (vertical and tube<br>angulation) instead of manual tube stand of the predicate. The high voltage<br>generator is new and is backwards compatible to the predicates high voltage<br>generator. This product also introduced Image Pasting on Table and Wall Stand<br>Mode, Auto tracking for Wall Stand, Auto Angulation, Camera Workflow, DAP<br>software calculation, Siemens LED collimator and LCD touch screen console. The<br>other minor changes include updates to components due to obsolescence. |
| Indications for Use:<br>21 CFR 807.92(a)(5) | The Definium Pace Select ET is intended to generate digital radiographic images<br>of the skull, spinal column, chest, abdomen, extremities, and other body parts in<br>patients of all ages. Applications can be performed with the patient sitting,<br>standing, or lying in the prone or supine position and the system is intended for<br>use in all routine radiography exams. Optional image pasting function enables the<br>operator to stitch sequentially acquired radiographs into a single image.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |

{6}------------------------------------------------

Image /page/6/Picture/2 description: The image shows the GE HealthCare logo. The logo consists of a circular emblem on the left and the text "GE HealthCare" on the right. The emblem is purple and features a stylized "GE" monogram. The text "GE HealthCare" is also purple and is written in a clean, sans-serif font.

are

This device is not intended for mammographic applications.

## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE: 21CFR807.92(a)(6)

A comparison of the indications for use and technological features of the subject and predicate device is provided in Table 1 below.

| Specification                                    | Predicate Device<br>Definium Pace Select                                                                                                                                                                                                   | Subject Device<br>Definium Pace Select ET                                                                                                                                                                                                                                                                                                                                    | Discussion of<br>Differences                                                                                                                                                                        |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                                     | General Purpose Digital<br>Radiographic Imaging System                                                                                                                                                                                     | General Purpose Digital<br>Radiographic Imaging System                                                                                                                                                                                                                                                                                                                       | Identical                                                                                                                                                                                           |
| Indications for<br>Use                           | The Definium Pace Select is<br>intended to generate digital<br>radiographic images of the<br>skull, spinal column, chest,<br>abdomen, extremities, and<br>other body parts in patients of<br>all ages. Applications can be                 | The Definium Pace Select ET is<br>intended to generate digital<br>radiographic images of the<br>skull, spinal column, chest,<br>abdomen, extremities, and<br>other body parts in patients of<br>all ages. Applications can be                                                                                                                                                | Equivalent<br>The optional image<br>pasting function has<br>been cleared in the<br>reference product<br>Discovery XR656 HD<br>under K191699.                                                        |
| Specification                                    | Predicate Device<br>Definium Pace Select                                                                                                                                                                                                   | Subject Device<br>Definium Pace Select ET                                                                                                                                                                                                                                                                                                                                    | Discussion of<br>Differences                                                                                                                                                                        |
|                                                  | performed with the patient<br>sitting, standing, or lying in the<br>prone or supine position and<br>the system is intended for use<br>in all routine radiography<br>exams.<br>The device is not intended for<br>mammographic applications. | performed with the patient<br>sitting, standing, or lying in the<br>prone or supine position and<br>the system is intended for use<br>in all routine radiography<br>exams. Optional image pasting<br>function enables the operator<br>to stitch sequentially acquired<br>radiographs into a single<br>image.<br>The device is not intended for<br>mammographic applications. |                                                                                                                                                                                                     |
| Contraindications                                | None .                                                                                                                                                                                                                                     | None .                                                                                                                                                                                                                                                                                                                                                                       | Identical                                                                                                                                                                                           |
| User group                                       | Professional Use Only                                                                                                                                                                                                                      | Professional Use Only                                                                                                                                                                                                                                                                                                                                                        | Identical                                                                                                                                                                                           |
| Patient<br>Population                            | All ages                                                                                                                                                                                                                                   | All ages                                                                                                                                                                                                                                                                                                                                                                     | Identical                                                                                                                                                                                           |
| System Software<br>Architecture                  | Distributed software<br>architecture (Atlas)                                                                                                                                                                                               | Distributed software<br>architecture (Atlas)                                                                                                                                                                                                                                                                                                                                 | Identical                                                                                                                                                                                           |
| Operator I/F                                     | One-LCD monitor, Keyboard,<br>Mouse, Barcode reader                                                                                                                                                                                        | One-LCD monitor, Keyboard,<br>Mouse, Barcode reader                                                                                                                                                                                                                                                                                                                          | Identical                                                                                                                                                                                           |
| X-Ray Tube                                       | Kailong 300kHU Tube: KL<br>H1080Y                                                                                                                                                                                                          | Kailong 300kHU Tube: KL<br>H1080Y                                                                                                                                                                                                                                                                                                                                            | Identical                                                                                                                                                                                           |
| Tube head<br>console                             | LED screen display                                                                                                                                                                                                                         | LCD touch screen console                                                                                                                                                                                                                                                                                                                                                     | Changed<br>Besides the previous<br>angle display. This<br>console provides<br>more positioner<br>information and<br>exposure parameter<br>display and<br>adjustment approach<br>from the exam room. |
| Collimator and<br>Auto FOV                       | Manual collimator<br>Manual FOV change and light<br>on/off                                                                                                                                                                                 | Siemens AL01CIIeL LED lamp<br>Manual and automatic FOV<br>change and light on/off                                                                                                                                                                                                                                                                                            | Changed<br>The device now<br>provides additional<br>automatic<br>collimation (Auto<br>FOV).                                                                                                         |
| DAP                                              | Physical DAP meter                                                                                                                                                                                                                         | DAP software calculation<br>based on FOV and SID<br>measured by Encoder after<br>Dose calibration                                                                                                                                                                                                                                                                            | Changed<br>The system now<br>automatically<br>calculates the DAP<br>instead of providing a<br>physical<br>measurement.                                                                              |
| Specification                                    | Predicate Device<br>Definium Pace Select                                                                                                                                                                                                   | Subject Device<br>Definium Pace Select ET                                                                                                                                                                                                                                                                                                                                    | Discussion of<br>Differences                                                                                                                                                                        |
| Camera Workflow                                  | None                                                                                                                                                                                                                                       | The depth camera provides<br>workflow improvement:<br>1. Live video on main UI<br>2. Patient size suggestion<br>based on camera depth<br>measurement;<br>3. Detector and ion chamber<br>outline on video;<br>4. FOV adjustable from video                                                                                                                                    | New<br>This new feature<br>provides additional<br>observation for<br>patient positioning.                                                                                                           |
| High Voltage<br>Generator                        | - Jedi 80 RD 1T<br>- 50KW or 65KW or 80kW                                                                                                                                                                                                  | - PSG-HR80S5<br>- 50KW or 65KW or 80kW                                                                                                                                                                                                                                                                                                                                       | Equivalent<br>Same function,<br>different supplier                                                                                                                                                  |
| DICOM                                            | DICOM 3.0                                                                                                                                                                                                                                  | DICOM 3.0                                                                                                                                                                                                                                                                                                                                                                    | Identical                                                                                                                                                                                           |
| OS                                               | Linux OS - SUSE                                                                                                                                                                                                                            | Linux OS - SUSE                                                                                                                                                                                                                                                                                                                                                              | Identical                                                                                                                                                                                           |
| Detector                                         | IRAY<br>17x17 detector US (K210314)                                                                                                                                                                                                        | IRAY<br>17x17 detector US (K210314)                                                                                                                                                                                                                                                                                                                                          | Identical                                                                                                                                                                                           |
| Detector Wireless<br>Connectivity                | YES – 802.11 wireless (Personal<br>Area Network)                                                                                                                                                                                           | YES – 802.11 wireless (Personal<br>Area Network)                                                                                                                                                                                                                                                                                                                             | Identical                                                                                                                                                                                           |
| Detector size/<br>resolution                     | One detector Size supported at<br>100 um resolutions:<br>17 x 17 inch                                                                                                                                                                      | One detector Size supported at<br>100 um resolutions:<br>17 x 17 inch                                                                                                                                                                                                                                                                                                        | Identical                                                                                                                                                                                           |
| Detector loading                                 | 300kg (distributed loading)<br>150kg (point loading)                                                                                                                                                                                       | 300kg (distributed loading)<br>150kg (point loading)                                                                                                                                                                                                                                                                                                                         | Identical                                                                                                                                                                                           |
| Patient Table                                    | Fixed table height 66cm<br>220 kg static non-elevating<br>table<br>Longitudinal travel 900mm<br>Transversal travel 220mm                                                                                                                   | Elevating range 59 – 90 cm<br>Maximum Loading 305kg<br>(672lbs)<br>Elevation speed< 18 seconds<br>with loading 135kg(397lbs)<br>Longitudinal travel 1110mm<br>Transversal travel 220mm                                                                                                                                                                                       | Changed<br>The table now<br>provides motorized<br>elevation with<br>additional tabletop<br>transverse and<br>longitudinal travel<br>range                                                           |
| Tube Support                                     | Floor-mount tube support with<br>longitudinal vertical and<br>angulation manual motion                                                                                                                                                     | Floor-mount tube support with<br>longitudinal vertical and<br>angulation manual motion.<br>Additional motorization of<br>vertical and angulation are<br>provided                                                                                                                                                                                                             | Changed<br>The vertical and<br>angulation motions<br>are now motorized                                                                                                                              |
| Tube to Table<br>Tray Alignment                  | Alignment between tube stand<br>and table tray is implemented<br>by mechanical connection                                                                                                                                                  | Alignment between tube stand<br>and table tray is implemented<br>by automatic electrical control<br>(also called auto alignment)                                                                                                                                                                                                                                             | Equivalent<br>Same function in tube<br>and table tray<br>alignment.                                                                                                                                 |
| Wall stand                                       | Minimum height from floor to<br>center of panel: 50cm<br>Maximum height from floor to<br>center of panel 176cm                                                                                                                             | Minimum height from floor to<br>center of panel: 28.5 cm<br>Maximum height from floor to<br>center of panel 180.5 cm                                                                                                                                                                                                                                                         | Changed<br>Motorized vertical<br>travel with a minor                                                                                                                                                |
| Specification                                    | Predicate Device<br>Definium Pace Select                                                                                                                                                                                                   | Subject Device<br>Definium Pace Select ET                                                                                                                                                                                                                                                                                                                                    | Discussion of<br>Differences                                                                                                                                                                        |
|                                                  | Vertical travel 126cm<br>No motorization                                                                                                                                                                                                   | Vertical travel: 152 cm.<br>Vertical motorization is<br>provided                                                                                                                                                                                                                                                                                                             | change in travel<br>ranges.                                                                                                                                                                         |
| Image Pasting on<br>Table and Wall<br>Stand Mode | None                                                                                                                                                                                                                                       | Image pasting can be<br>performed at Wall stand mode<br>and Table mode                                                                                                                                                                                                                                                                                                       | New<br>The image pasting<br>function has been<br>cleared in the<br>reference product<br>Discovery XR656 HD<br>under K191699.                                                                        |
| Auto Grid Option                                 | Yes<br>Virtual grid function in Table,<br>WS and DC mode                                                                                                                                                                                   | Yes<br>Virtual grid function in Table,<br>WS and DC mode                                                                                                                                                                                                                                                                                                                     | Identical                                                                                                                                                                                           |
| Auto tracking for<br>Wall Stand                  | None                                                                                                                                                                                                                                       | Auto-tracking between Wall<br>stand and tube stand vertically                                                                                                                                                                                                                                                                                                                | New<br>This feature is new to<br>Predicate. But it is<br>equivalent to the<br>auto-tracking in the<br>reference Discovery<br>XR656 HD.                                                              |
| Auto Angulation                                  | None                                                                                                                                                                                                                                       | Yes<br>Tube can be automatically<br>angulated according to the<br>input on tube head console UI.                                                                                                                                                                                                                                                                             | New<br>The tube angulation<br>is now motion<br>controlled.                                                                                                                                          |

#### Table 1: High-level Comparison of Subject Device to Predicate

{7}------------------------------------------------

Image /page/7/Picture/2 description: The image shows the GE HealthCare logo. The logo consists of the GE monogram in a circle on the left, followed by the words "GE HealthCare" in a sans-serif font. The logo is purple in color. 

Page 4 of 7

{8}------------------------------------------------

Image /page/8/Picture/2 description: The image contains the logo for GE HealthCare. The logo consists of the GE monogram enclosed in a circular design on the left, followed by the text "GE HealthCare" in a sans-serif font. The color of the logo and text is a shade of purple.

{9}------------------------------------------------

Image /page/9/Picture/2 description: The image contains the GE HealthCare logo. The logo consists of the GE monogram enclosed in a circular frame on the left, followed by the text "GE HealthCare" on the right. The color of the logo and text is a shade of purple.

{10}------------------------------------------------

Image /page/10/Picture/2 description: The image shows the GE Healthcare logo. The logo consists of a purple circular emblem with intertwined letters on the left, followed by the text "GE HealthCare" in purple. The text is written in a clean, sans-serif font.

## PERFORMANCE DATA: Determination of Substantial Equivalence 21 CFR807(b)(1)

#### Summary of Non-Clinical Tests:

The following quality assurance measures were applied to the development of Definium Pace Select ET system:

- 1. Risk Analysis
- 2. Requirements Reviews
- 3. Design Reviews
- 4. Testing on unit level (Module verification)
- 5. Integration testing (System verification)
- 6. Performance testing (Verification)
- 7. Safety testing (Verification)
- 8. Simulated use testing (Validation)

Definium Pace Select ET verification and validation testing was performed to confirm that the safety and effectiveness of the device has not been affected. The test plans and results have been executed with acceptable results.

#### Summary of Clinical testing: 21 CFR 807.92(b)(2)

The Definium Pace Select ET does not contain clinical testing data.

#### CONCLUSION:

The results of the testing described above demonstrate that the Definium Pace Select ET is as safe and effective as the predicate device and supports a determination of substantial equivalence.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR/K242678](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR/K242678)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
