← Product Code [KPR](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR) · K242478

# GF85 (models GF85-3P, GF85-SP) (K242478)

_Samsung Electronics Co., Ltd. · KPR · Sep 19, 2024 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR/K242478

## Device Facts

- **Applicant:** Samsung Electronics Co., Ltd.
- **Product Code:** [KPR](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR.md)
- **Decision Date:** Sep 19, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1680
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** AI/ML

## Indications for Use

The GF85 Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

## Device Story

GF85 is a stationary digital X-ray imaging system; captures radiographic images of human anatomy. X-rays pass through patient to detector; converted to electrical signals; processed by S-station (Operation Software) for amplification and digital conversion; saved as DICOM files. Image Post-processing Engine (IPE) tunes images before transmission to PACS for clinical review. Used in clinical settings by physicians or technicians. Output enables diagnostic visualization of anatomy. Benefits include high-quality radiographic imaging for clinical decision-making. Models GF85-3P and GF85-SP differ by high voltage generator capacity (80/65/50 kW 3-phase vs 40 kW single-phase).

## Clinical Evidence

No clinical data. Bench testing only. Phantom image evaluations performed per FDA guidance for Solid State X-ray Image Devices; evaluated by three radiologists and found equivalent to predicate devices. Dosimetric performance (exposure output, Half Value Layer) compared favorably to predicates.

## Technological Characteristics

Stationary X-ray system; floor-mounted tube stand. Materials: CFRP/HPL patient table. Energy: High voltage generator (40-80 kW). Connectivity: DICOM, PACS integration. Software: S-station OS with IPE. Standards: ES 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, ISO 14971, 21 CFR 1020.30/31.

## Regulatory Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- GC85A ([K181629](/device/K181629.md))
- GM85 ([K222353](/device/K222353.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font below.

September 19, 2024

Samsung Electronics Co., Ltd. % Jaesang Noh Regulatory Affairs, Principal Professional 129, Samsung-ro, Yeongtong-gu Suwon-si, Gyeonggi 16677 SOUTH KOREA

Re: K242478

Trade/Device Name: Gf85 (models gf85-3p, gf85-sp) Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: KPR Dated: August 19, 2024 Received: August 21, 2024

Dear Jaesang Noh:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lu Jiang

Lu Jiang, Ph.D. Assistant Director Diagnostoc X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

### Indications for Use

510(k) Number (if known) K242478

Device Name GF85 (models GF85-3P, GF85-SP)

#### Indications for Use (Describe)

"The GF85 Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications."

| Type of Use (Select one or both, as applicable)                                                                         |                                                                                                                        |
|-------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)         </span> </div> | <div> <span> <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)         </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

### SAMSUNG

# SAMSUNG ELECTRONICS Co., Ltd.

510(k) Premarket Notification - Traditional

## K242478 510(k) Summary

This 510k Summary is being submitted in accordance with the requirements of 21 CFR 807.92.

- 1. Date: August 19, 2024
### 2. Submitter

- A. Company Name: SAMSUNG ELECTRONICS Co., Ltd.
- B. Address: 129, Samsung-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, 16677, Republic of Korea

### 3. Primary Contact Person

- A. Name: JAESANG NOH
- B. Title: Regulatory Affairs, Principal Professional
- C. Phone Number: +82-2-2193-2444
- D. FAX Number: +82-2-2194-0284
- E. E-Mail: jaesang.noh@samsung.com

### 4. Secondary Contact Person

- A. Name: Ninad Gujar
- B. Title: Vice President, Requlatory Affairs & Quality Control
- C. Phone Number: 978-564-8503
- D. FAX Number: 978-560-0602
- E. E-Mail: ngujar@neurologica.com

#### Proposed Device 5.

- A. Trade Name: GF85 (models GF85-3P, GF85-SP)
- B. Device Model Names: GF85-3P, GF85-SP
- C. Common Name: Stationary X-ray System
- D. Classification Name: System, X-ray, Stationary
- E. Product Code: KPR
- ட் Regulation: 21 CFR 892.1680

### 6. Predicate Devices

|                      | Predicate Device #1     | Predicate Device #2 |
|----------------------|-------------------------|---------------------|
| Device Name          | GC85A                   | GM85                |
| Classification Name  | Stationary X-ray System | Mobile X-ray System |
| Product Code         | KPR                     | IZL                 |
| Regulation           | 21 CFR 892.1680         | 21 CFR 892.1720     |
| 510(K)#              | K181629                 | K222353             |
| 510(K) Decision Date | July 20, 2018           | September 29, 2022  |

### 7. Device Description

The GF85 digital X-ray imaging system is used to capture images by transmitting X-ray

{5}------------------------------------------------

#### 510(k) Premarket Notification - Traditional

to a patient's body. The X-ray passing through a patient's body is sent to the detector and then converted into electrical signals. These signals go through the process of amplification and digital data conversion in the signal process on the S-station, which is the Operation Software (OS) of Samsung Digital Diagnostic X-ray System, and save in DICOM file, a standard for medical imaging. The captured images are tuned up by an Image Post-processing Engine (IPE) which is exclusively installed in S-station, and sent to the Picture Archiving & Communication System (PACS) sever for reading images. The GF85 Digital X-ray Imaging System includes two models, namely the GF85-3P and

GF85-SP. These two models differ primarily in terms of their respective configurations of High Voltage Generators (HVGs). Specifically, the GF85-3P features capacities of 80 kW, 65 kW and 50 kW with 3 phases, whereas the GF85-SP provides a capacity of 40 kW with a single phase. However, all other specifications remain consistent across both models.

#### 8. Intended Use

The GF85 Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.

### 9. Summary of Technological characteristic of the proposed device compared with the predicate device

Though the gantry of the proposed device, GF85, is newly designed, it has the same detectors and image processing as predicate devices. Therefore, it does not have significant changes in image characteristics compared to the predicate device #1 and #2, GC85A and GM85.

Comparisons of technological characteristics were executed and demonstrate the substantial equivalence to the predicate devices.

| Attribute     | Predicate Device #1                                                    | Proposed Device                                                                   | Predicate Device #2                                                          | Comparison Results |
|---------------|------------------------------------------------------------------------|-----------------------------------------------------------------------------------|------------------------------------------------------------------------------|--------------------|
| Name          | GC85A                                                                  | GF85<br>(GF85-3P/GF85-SP)                                                         | GM85                                                                         |                    |
| Manufacturer  | SAMSUNG<br>ELECTRONICS<br>Co., Ltd.                                    | SAMSUNG<br>ELECTRONICS<br>Co., Ltd.                                               | SAMSUNG<br>ELECTRONICS<br>Co., Ltd.                                          |                    |
| 510(k) Number | K181629                                                                | None                                                                              | K222353                                                                      |                    |
| Appearances   | Image: GC85A                                                           | Image: GF85                                                                       | Image: GM85                                                                  | Similar as PD #1   |
| Intended Use  | The GC85A digital<br>X-ray imaging<br>system is intended<br>for use in | The GF85 Digital X-<br>ray Imaging System<br>is intended for use<br>in generating | The GM85 Digital<br>Mobile X-ray<br>imaging System is<br>intended for use in | Same as PD #1      |

### A. Comparing with Predicate Devices

{6}------------------------------------------------

## SAMSUNG

| generating<br>radiographic images<br>of human anatomy<br>by a<br>qualified/trained<br>doctor or technician.<br>This device is not<br>intended for<br>mammographic<br>applications. | radiographic images<br>of human anatomy<br>by a<br>qualified/trained<br>doctor or technician.<br>This device is not<br>intended for<br>mammographic<br>applications. | generating<br>radiographic images<br>of human anatomy<br>by a<br>qualified/trained<br>doctor or technician.<br>This device is not<br>intended for<br>mammographic<br>applications. |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

| Attribute                                          | Predicate Device #1<br>(GC85A, K181629) |                                                        | GF85                                                      |               | Comparison Results |
|----------------------------------------------------|-----------------------------------------|--------------------------------------------------------|-----------------------------------------------------------|---------------|--------------------|
|                                                    |                                         | GF85-3P                                                | GF85<br>-SP                                               |               |                    |
| (1)High Voltage Generator                          |                                         |                                                        |                                                           |               |                    |
| Name                                               | GXR-82                                  | GXR-52<br>ZR75 PN80<br>ZR75 PN50                       | Cetus 80R01<br>Cetus 65R01<br>Cetus 50R01<br>Cetus 40R03  | Different (1) |                    |
| Max. Power                                         | 82kW                                    | 52kW<br>80kW<br>50kW                                   | 80kW<br>65kW<br>50kW<br>40kW                              | Different (1) |                    |
| Output<br>RANGE                                    | Tube Voltage                            | 40-150kV                                               |                                                           | Same          |                    |
|                                                    | Tube Current                            | 10-<br>1000 mA<br>10-640m A<br>10-1000 mA<br>10-630m A | 10 ~ 1000 mA<br>10 ~ 800 mA<br>10 ~ 630 mA<br>10 ~ 500 mA | Different (1) |                    |
|                                                    | Exposure Time                           | 1msec-10sec                                            |                                                           | Same          |                    |
| AEC<br>(Automatic<br>Exposure<br>Control)          |                                         | Yes                                                    |                                                           | Same          |                    |
| APR<br>(Anatomically<br>Programmed<br>Radiography) |                                         | Yes                                                    |                                                           | Same          |                    |

| Attribute       | Predicate Device #1<br>(GC85A, K181629) | GF85                                 |                   |                  | Comparison Results |               |
|-----------------|-----------------------------------------|--------------------------------------|-------------------|------------------|--------------------|---------------|
| (2) X-ray tube  |                                         |                                      |                   |                  |                    |               |
| Name            | E7869X                                  | E7252X                               | H2100X<br>E7254FX | H1080X<br>E7252X | H1086X<br>E7843X   | Different (2) |
| Focal spot      | 0.6/1.2mm                               | 0.6/1.2mm                            |                   |                  |                    | Same          |
| Target Angle    | 12°                                     | 12°                                  |                   |                  |                    | Same          |
| Target Material | Rhenium-tungsten faced<br>molybdenum    | Rhenium-tungsten faced<br>molybdenum |                   |                  |                    | Same          |

{7}------------------------------------------------

## SAMSUNG

| Attribute                       | Predicate Device #1<br>(GC85A, K181629)   | GF85                                      | Comparison Results    |               |
|---------------------------------|-------------------------------------------|-------------------------------------------|-----------------------|---------------|
| Max. Anode HU                   | 600kHU                                    | 400kHU 300kHU 150KHU                      | Different (2)         |               |
| Attribute                       | Predicate Device #1<br>(GC85A, K181629)   | GF85                                      | Comparison Results    |               |
| (3) Tube Stand                  |                                           |                                           |                       |               |
| Moving Range (mm)               | Longitudinal                              | 1,680~4,180<br>(Varies with room size)    | 2,100                 | Different (3) |
|                                 | Lateral                                   | 1,030~3,030<br>(Varies with room size)    | 150                   | Different (3) |
|                                 | Vertical                                  | 1,800                                     | 1,500                 | Different (3) |
|                                 | Vertical Tube Moving method               | Motorized                                 | Manual & Motorized    | Different (3) |
|                                 | Tube Assembly Rotation                    | -157 ~ +183                               | ±90°                  | Different (3) |
|                                 | Brake locking Method                      | Electromagnetic                           | Electromagnetic       | Same          |
|                                 | Automatic Centering                       | O                                         | O                     | Same          |
|                                 | Moving Rail Type                          | Al Extrusion                              | Steel                 | Different (3) |
|                                 | Image Preview                             | O                                         | O                     | Same          |
|                                 | Display Type                              | Color LCD                                 | Color LCD             | Same          |
|                                 | Control Switch Type                       | Button + Touch Screen                     | Button + Touch Screen | Same          |
| Vertical Sync.                  | With Table                                | O                                         | O                     | Same          |
|                                 | With Stand                                | O                                         | O                     | Same          |
|                                 | Tube Support Mounting                     | Ceiling-mounted                           | Floor-mounted         | Different (3) |
| Attribute                       | Predicate Device #1<br>(GC85A, K181629)   | GF85                                      | Comparison Results    |               |
| Move<br>ment                    | Range<br>(mm)                             | 280~1,850                                 | 400~1,900             | Different (4) |
| Detector/tube<br>servo coupling | Yes                                       | Yes                                       | Same                  |               |
| Detect<br>or                    | Tilting                                   | Motorized                                 | Motorized             | Same          |
|                                 |                                           | -20~+90 / N/A                             | -20~+90 / N/A         | Same          |
|                                 | AEC                                       | Conventional                              | Conventional          | Same          |
|                                 | Lines/<br>cm                              | 85/92                                     | 85                    | Same          |
| Grid                            | Grid<br>mecha<br>nism                     | Stationary                                | Stationary            | Same          |
|                                 | Remo<br>vability                          | Removable                                 | Removable             | Same          |
|                                 |                                           |                                           |                       |               |
| Detector<br>Support<br>Mounting | Floor                                     | Floor                                     | Same                  |               |
| Patient Support<br>Device       | Patient handgrips, lateral<br>support bar | Patient handgrips, lateral<br>support bar | Same                  |               |

| Attribute      | Predicate Device #1 (GC85A, K181629) |           | GF85               | Comparison Results |
|----------------|--------------------------------------|-----------|--------------------|--------------------|
| (4) Wall Stand |                                      |           |                    |                    |
| Vertical       | Mechanism                            | Motorized | Manual & Motorized | Different (4)      |

{8}------------------------------------------------

### SAMSUNG

| Attribute                        | Predicate Device #1<br>(GC85A, K181629) | GF85         | Comparison Results     |               |
|----------------------------------|-----------------------------------------|--------------|------------------------|---------------|
| (5) Patient Table                |                                         |              |                        |               |
| Table Top                        | Size(mm)                                | 2,410 X 812  | 2,208 X 800            | Different (5) |
|                                  | Range (mm)                              | $\pm$ 140    | $\pm$ 105              | Different (5) |
|                                  | Range (mm)                              | $\pm$ 480    | $\pm$ 400              | Different (5) |
| Table height                     | Mechanism                               | Motorized    | Motorized / Fixed      | Same          |
|                                  | Range (mm)                              | 545 ~ 900    | 500 ~770 / 680 (Fixed) | Different (5) |
| Horizontal range of detector(mm) |                                         | 688          | 600                    | Different (5) |
| AEC                              |                                         | Conventional | Conventional           | Same          |
| Grid                             | Lines/cm                                | 85/92        | 85                     | Same          |
|                                  | Grid mechanism                          | Stationary   | Stationary             | Same          |

{9}------------------------------------------------

## SAMSUNG

| Attribute                                                    | Predicate Device #1<br>(GC85A, K181629) | GF85                   | Comparison Results |
|--------------------------------------------------------------|-----------------------------------------|------------------------|--------------------|
| Removability                                                 | Removable                               | Removable              | Same               |
| Vertical Sync.                                               | O                                       | O / Fixed              | Different (5)      |
| Control Switch<br>Type                                       | Foot switch                             | Foot switch            | Same               |
| Maximum<br>Patient<br>Weight(kg)<br>(Static, Center<br>load) | 350                                     | 350 (CFRP) / 300 (HPL) | Different (5)      |

| Attribute                               | Predicate Device #1<br>(GC85A, K181629)               | GF85                       | Comparison Results |
|-----------------------------------------|-------------------------------------------------------|----------------------------|--------------------|
| (6) Collimator                          |                                                       |                            |                    |
| Name                                    | SDR-OGCL80<br>U<br>SDR-OGCL81<br>U<br>SDR-OGCL83<br>U | SDR-OGCL20P<br>SDR-OGCL21P | Different (6)      |
| Beam Limiting<br>Blade/Moving<br>Method | Motorized<br>/Manual                                  | Motorized<br>/Manual       | Same               |
| Manual<br>Operation<br>Method           | Volume                                                | Volume                     | Same               |
| Collimator<br>Rotation                  | ±45                                                   | -90°/+180°                 | Different (6)      |
| Beam Light<br>Source                    | LED                                                   | LED                        | Same               |
| Light Field<br>Indicator Timer          | O                                                     | O                          | Same               |
| Side Lamp                               | O                                                     | O                          | Same               |
|                                         | Laser Module                                          | Laser Module               | Same               |
| Field Size / SID<br>Display             | Color LCD                                             | Color LCD                  | Same               |

| Attribute    | Predicate Device #2<br>(GM85, K222353)                                                                               | GF85                                                                                | Comparison Results |
|--------------|----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|--------------------|
| (7) Detector |                                                                                                                      |                                                                                     |                    |
| Name         | S4335-W<br>S4343-W<br>S3025-W<br>S4335-AW<br>S4343-AW<br>S4335-AWM<br>S4343-AWM<br>S3025-AW<br>S3025-AWM<br>F4335-AW | S4335-AW<br>S4343-AW<br>S4335-AWM<br>S4343-AWM<br>S3025-AW<br>S3025-AWM<br>F4335-AW | Same               |

{10}------------------------------------------------

### SAMSUNG

| Attribute                   | Predicate Device #1<br>(GC85A, K181629) | GF85             | Comparison Results |
|-----------------------------|-----------------------------------------|------------------|--------------------|
| (8) Weight Distribution Cap |                                         |                  |                    |
| W x L x H (mm)              | 505 X 553 X 37.0                        | 505 X 553 X 37.0 | Same               |

| Attribute             | Predicate Device #1<br>(GC85A, K181629) | GF85                         | Predicate Device #2<br>(GM85, (K222353) | Comparison Results |
|-----------------------|-----------------------------------------|------------------------------|-----------------------------------------|--------------------|
| (9) Software Features |                                         |                              |                                         |                    |
| Feature<br>Names      | S-Share                                 | S-Share                      | S-Share                                 | Same               |
|                       | S-Enhance                               | S-Enhance                    | S-Enhance                               | Same               |
|                       | SimGrid                                 | SimGrid                      | SimGrid                                 | Same               |
|                       | -                                       | PEM                          | PEM                                     | Same               |
|                       | QAP                                     | QAP                          | QAP                                     | Same               |
|                       | Bone Suppression                        | Bone Suppression             | Bone Suppression                        | Same               |
|                       | Remote View                             | Remote View                  | Remote View                             | Same               |
|                       | Smart Stitching                         | Smart Stitching              | -                                       | Same               |
|                       | -                                       | Mirror View                  | Mirror View                             | Same               |
|                       | -                                       | RFID                         | RFID.                                   | Same               |
|                       | -                                       | Vision Live                  | -                                       | Different (7)      |
|                       | -                                       | Vision Stitching             | -                                       | Different (7)      |
|                       | -                                       | Vision Capture               | -                                       | Different (7)      |
|                       | -                                       | Virtual Remote<br>Controller | -                                       | Different (7)      |

{11}------------------------------------------------

### SAMSUNG

#### 510(k) Premarket Notification - Traditional

| Attribute                         | Predicate Device #1<br>(GC85A, K181629)                                              | GF85                                                                                 | Predicate Device #2<br>(GM85, (K222353)                                              | Comparison Results |
|-----------------------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|--------------------|
|                                   | -                                                                                    | Lunit INSIGHT CXR Triage                                                             | -                                                                                    | Different (7)      |
| Attribute                         | Predicate Device #1<br>(GC85A, K181629)                                              | GF85                                                                                 | Predicate Device #2<br>(GM85, (K222353)                                              | Comparison Results |
| (10) Image Post-processing Engine |                                                                                      |                                                                                      |                                                                                      |                    |
| Noise reduction<br>algorithm      | Image Post-<br>processing Engine<br>with an advanced<br>noise reduction<br>algorithm | Image Post-<br>processing Engine<br>with an advanced<br>noise reduction<br>algorithm | Image Post-<br>processing Engine<br>with an advanced<br>noise reduction<br>algorithm | Same               |

| No  | Comparison Results     | Discussion                                                                                                                                                                                                                                                                                                                                              |
|-----|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| (1) | High Voltage Generator | The High Voltage Generators are different from the<br>predicate device, which have various Max. Power and<br>Output Range for the Tube Current. They do not contribute<br>any adverse impact to the device's safety and effectiveness.                                                                                                                  |
| (2) | X-ray tube             | The X-ray tube are different from the predicate device,<br>which have different Max. Anode HU. They do not contribute<br>any adverse impact to the device's safety and effectiveness.                                                                                                                                                                   |
| (3) | Tube Stand             | The Tube Stand is different from the predicated device,<br>which have different Moving Range, Vertical Tube Moving<br>method, Tube Assembly Rotation, Moving Rail Type and<br>Tube Support Mounting because it's a floor-mounded type<br>in the proposed device. It does not contribute any adverse<br>impact to the device's safety and effectiveness. |
| (4) | Wall Stand             | The Wall Stand different from the predicate device, which<br>have different Vertical Movement. It does not contribute any<br>adverse impact to the device's safety and effectiveness.                                                                                                                                                                   |
| (5) | Patient Table          | The Patient Table is different from the predicate device,<br>which has different Table Top, Table height, Horizontal<br>range of detector, Vertical Sync. and Maximum Patient<br>Weight. It does not contribute any adverse impact to the<br>device's safety and effectiveness.                                                                         |
| (6) | Collimator             | The Collimator is different from the predicate device, which<br>have different Collimator Rotation. It does not contribute any<br>adverse impact to the device's safety and effectiveness.                                                                                                                                                              |
| (7) | Software               | New features are provided for user convenience. It does not<br>contribute any adverse impact to the device's safety and<br>effectiveness.                                                                                                                                                                                                               |

#### B. Safety, EMC and Performance Data

Electrical, mechanical, environmental safety and performance testing according to standard ES 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, ISO 14971, 21 CFR1020.30 and 21 CFR1020.31 were performed, and EMC testing was conducted in accordance with standard IEC 60601-1-2. Wireless function was tested and

{12}------------------------------------------------

#### 510(k) Premarket Notification - Traditional

verified followed by guidance, Radio frequency Wireless Technology in Medical Devices. All test results were satisfying the standards.

### C. Non-clinical data

In the non-clinical data, the basic dosimetric performance data of the proposed device and predicate device were compared and the proposed device has the similar characteristics in the exposure output and Half Value Layer even for different tube combinations. The phantom image evaluations were performed in accordance with the FDA guidance for the submission of 510(k)'s for Solid State X-ray Image Devices and were evaluated by three different professional radiologists and found to be equivalent to the predicate device. These reports show that the proposed device is substantially equivalent to the proposed devices.

#### D. Clinical data

This submission does not required clinical data. Testing for image quality has been conducted with image phantoms, as described above.

- E. Summary of the Standards and Guidance Compliance
	- 1. ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012 C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical Electrical Equipment -Part 1: General requirements for basic safety and essential performance
	- IEC 60601-1-2 Edition 4.1 Medical Electrical Equipment Part 1-2: General 2. Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic compatibility - Requirements and Tests
	- 3. IEC 60601-1-3 Edition 2.2 Medical Electrical Equipment – Part 1-3: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Radiation protection in Diagnostic X-ray Equipment
	- 4. IEC 60601-2-28 Edition 3 Medical Electrical Equipment – Part 2-28: Particular Requirements for the Basic Safety and Essential Performance of X-ray Tube Assemblies for Medical Diagnosis
	- 5. IEC 60601-2-54 Edition 2.0 Medical Electrical Equipment – Part 2-54: particular Requirements for the Basic Safety and Essential Performance of X-ray Equipment for Radiography and Radioscopy
	- 6. IEC 62220-1-1 Edition 1.0 Medical electrical Equipment - Characteristics of digital X-ray imaging devices Part 1-1: Determination of the detective quantum efficiency Detectors used in radiographic imaging
	- 7. Cybersecurity in Medical Device: Quality System Considerations and Content of Premarket Submissions issued on September 27, 2023
	- 8. Content of Premarket Submissions for Device software Functions issued on June 14, 2023
	- Guidance for the Submission for 510(k) for Solid State X-ray Imaging Devices 9. Guidance for Industry and Food and Drug Administration Staff issued on September 1, 2016
	- 10. Pediatric Information for X-ray Imaging Device Premarket Notifications Guidance for Industry and Food and Drug Administration Staff issued on November 28, 2017

{13}------------------------------------------------

### SAMSUNG

# SAMSUNG ELECTRONICS Co., Ltd.

510(k) Premarket Notification - Traditional

F. Conclusion

The non-clinical data demonstrates that the proposed device is as safe, as effective, and performs as well as the legally marketed devices indicating substantial equivalence to the predicate.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR/K242478](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/KPR/K242478)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
