K252249 · Mars Bioimaging , Ltd. · JAK · Mar 13, 2026 · Radiology
Device Facts
Record ID
K252249
Device Name
Extremity CT Imaging System
Applicant
Mars Bioimaging , Ltd.
Product Code
JAK · Radiology
Decision Date
Mar 13, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
The Extremity CT Imaging System is intended for x-ray computed tomography imaging of upper extremities of adult patients. The device is intended to be operated in a professional healthcare environment by qualified and trained healthcare professionals only.
Device Story
Extremity CT Imaging System acquires volumetric spectral CT images of adult upper limbs; utilizes helical rotation of single x-ray source and photon-counting detectors (Medipix 3RX/CdTe sensor) with stationary limb. Device operates in professional healthcare environments; used by trained medical professionals. System processes raw data via MARS Vision software for post-processing, including multiplanar reformats (MPR) and 3D visualization (volume rendering, maximum intensity projection). Output consists of high-quality DICOM images uploaded to facility PACS for review. Images support diagnosis, treatment planning, and monitoring of hand/wrist pathologies; system does not perform automated diagnosis. Benefits include high-precision photon counting for detailed tissue composition assessment compared to conventional energy-integrating CT.
Clinical Evidence
Visual grading study conducted; American Board Certified Radiologists reviewed representative hand, wrist, finger, and thumb MSK images. Study confirmed reconstructed images are of diagnostic quality.
Technological Characteristics
Photon-counting detector (CdTe sensor, Medipix 3RX readout); single fixed-anode x-ray source (60-120 kVp); helical scanning geometry; iterative reconstruction (statistical, non-linear). Connectivity via Ethernet/PACS. Biocompatibility per ISO 10993-1. Compact, mobile gantry with 12 cm bore. Software: MARS Vision (Windows-based) for post-processing/visualization.
Indications for Use
Indicated for x-ray computed tomography imaging of upper extremities of adult patients. Intended for use in professional healthcare environments by qualified, trained healthcare professionals.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
K173478 — OnSight 3D Extremity System · Carestream Health, Inc. · Jan 19, 2018
K121418 — PLANMED VERITY · Planmed OY · Feb 1, 2013
K243398 — CT Scanner TSX-501R/1 V11.1 · Canon Medical Systems Corporation · Jun 20, 2025
K233657 — NAEOTOM Alpha · Siemens Medical Solutions USA, Inc. · Mar 28, 2024
K160723 — OnSight 3D Extremity System · Carestream Health, Inc. · Aug 30, 2016
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 13, 2026
MARS Bioimaging Ltd.
% Melissa Dehass
Representative/Consultant
MEDIcept
200 Homer Ave,
ASHLAND, MA 01721
Re: K252249
Trade/Device Name: Extremity CT Imaging System
Regulation Number: 21 CFR 892.1750
Regulation Name: Computed Tomography X-Ray System
Regulatory Class: Class II
Product Code: JAK
Dated: February 11, 2026
Received: February 12, 2026
Dear Melissa Dehass:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252249 - Melissa Dehass
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252249 - Melissa Dehass
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252249 | |
| Device Name Extremity CT Imaging System | |
| Indications for Use (Describe) The Extremity CT Imaging System is intended for x-ray computed tomography imaging of upper extremities of adult patients. The device is intended to be operated in a professional healthcare environment by qualified and trained healthcare professionals only. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Traditional
510(k)
Premarket Notification
Extremity
CT Imaging System™
K252249
# 510(k) Summary
## DATE PREPARED
11 March 2026
## MANUFACTURER AND 510(k) OWNER
MARS Bioimaging Ltd.
68 St. Asaph Street
Christchurch 8011
New Zealand
Official Contact: Homi Dalal, Head of Regulatory Affairs
Telephone: +64 21 1707631
## REPRESENTATIVE /CONSULTANT
Melissa DeHass
MEDIcept
Telephone: (717)476-0702
Email: mdehass@medicept.com
## DEVICE INFORMATION
Proprietary Name/Trade Name: Extremity CT Imaging System
Regulation Name: Computed Tomography X-Ray System
Regulation Number: 892.1750
Class: II
Product Code: JAK
Review Panel: Radiology (OHT8)
## PREDICATE DEVICE IDENTIFICATION
| 510(k) Number | Predicate Device Name / Manufacturer | Predicate/ Reference |
| --- | --- | --- |
| K160723 | OnSight 3D Extremity System (Carestream Health, Inc.) | Predicate |
| K233657 | NAEOTOM Alpha (Siemens Medical Systems) | Reference |
The predicate and reference devices have not been subject to a design related recall.
## DEVICE DESCRIPTION
The Extremity CT Imaging System™ is a computed tomography x-ray system designed for imaging of the upper limb. The device acquires volumetric spectral CT imaging using helical rotation of a single x-ray source and photon counting detectors with the limb stationary throughout the scan.
The use of photon-counting technology in the Medipix X-ray detector enables the scanner to detect individual X-ray photons with greater precision in comparison to conventional CT scanners, which typically rely on energy-integrating detectors. By counting the number of
{5}
Traditional 510(k) Premarket NotificationExtremity CT Imaging System™
photons that pass through the body part and measuring their energy levels, the scanner can capture more detailed information about the composition of the tissues being imaged.
The Extremity CT Imaging System is intended for use in professional healthcare facility environments, excluding oxygen-rich environments, by trained medical professionals. The imaging data produced by the system is intended to support diagnosis, treatment planning, and monitoring of patient conditions. It is to be noted however that actual diagnosis is not performed using any part of this software. The images produced by the system serve as a valuable tool for assessing pathologies related to the hand and wrist of adult patients.
The device is meant for prescription use only by qualified and trained medical professionals.
The scanner is compact, with a small footprint and an entry port of diameter 12 cm (4.8 inches) ideally suited for imaging the hand and wrist. The scanner is mobile but must be used in a stationary position.
The system is designed to provide a streamlined workflow from patient scanning to the delivery of high-quality images, to ultimately support patients' diagnosis. Once the data has been processed within the scanner's computer, it is accessed from the Workstation using the specialized image processing software, MARS Vision, for post-processing. This stage involves advanced image manipulation, including multiplanar reformats (MPR) and 3D visualization (direct volume rendering and maximum intensity projection). Following the data review and post-processing in MARS Vision, high-quality images in DICOM format are created and uploaded to the facility's Picture Archiving and Communication System (PACS) system, where they are securely stored and made available for review.
Overall, the device description for the Extremity CT Imaging System™ above addresses recommendations outlined in the device-specific guidance documents and meets requirements appropriate for the device classification and applicable special controls.
## INDICATIONS FOR USE
The Extremity CT Imaging System is intended for x-ray computed tomography imaging of upper extremities of adult patients.
The device is intended to be operated in a professional healthcare environment by qualified and trained healthcare professionals only.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
Mars Bioimaging believes that the Extremity CT Imaging System is substantially equivalent to the predicate devices based on the information summarized here:
The subject device has the same intended use/indications for use as the predicate device as both devices are intended to be used as x-ray computed tomography for imaging of upper extremities in adult patients..
{6}
Traditional 510(k) Premarket NotificationExtremity CT Imaging System™
The Extremity CT Imaging System and the predicate device are based on the same or similar technological characteristics, as outlined in the table below.
- Both systems utilize the same or similar power generator.
- Both systems have the same or similar scanner and the same method for scanning acquisition.
- Both systems connect to the network for data display and transmission in the same or similar manner.
- Both systems offer the same or similar image analysis and reconstruction.
| Characteristic | Subject Device Extremity CT Imaging System™ | Predicate Device Carestream OnSight 3D Extremity System K160723 | Equivalence Comparison |
| --- | --- | --- | --- |
| Intended Use/Indications for Use | The Extremity Imaging System is intended for x-ray computed tomography imaging of upper extremities of adult patients | The device is intended to be used for x-ray computed tomography and projection x-ray imaging of upper and lower extremities of adult and pediatric patients aged 12 and over. | Similar |
| X-ray source | Low power, x-ray tube Fixed anode.
Single source.
Tube voltage: 60 to 120 kVp
Tube current: 0.01 mA – 0.35 mA | Low power x-ray tube. Fixed anode
Three sources.
Tube voltage 50 to 90 kVp
Tube current: 2 – 10 mA | Similar |
| Target Angle | 20° | 20° | Same |
| Focal spot | Fixed. Small (0.07 mm) | Fixed. Small (0.50 mm) | Similar |
| Collimator type | Fixed to match active area of detector | Fixed to match active area of detector | Same |
| Exposure type | Continuous exposure | Pulsed exposure | Different, does not raise new questions of safety and effectiveness |
{7}
Traditional 510(k) Premarket NotificationExtremity CT Imaging System™
| Characteristic | Subject Device Extremity CT Imaging System™ | Predicate Device Carestream OnSight 3D Extremity System K160723 | Equivalence Comparison |
| --- | --- | --- | --- |
| Energy settings (kV) | Scan acquired using one pre-programmed energy setting. Information captured in multiple energy bins, to provide spectral data – see detector technology) | Scan acquired using one pre-programmed energy setting (eg. 90 kV). No spectral data provided. | Similar |
| Detector technology | Photon counting detector with CdTe sensor and Medipix 3RX readout technology. Direct conversion of photon energies into electrical signal. | Flat panel detector with amorphous silicon (a-Si:H) sensor layer. Indirect conversion of photon energies into electrical signal. | Different, does not raise new questions of safety and effectiveness |
| Detector active area | 128 x 1536 pixels
Pixel pitch 110mm | 1782 x 2166 pixels
Pixel pitch 139mm | Different, does not raise new questions of safety and effectiveness |
| Localization | Scout scan | Scout scan | Same |
| Principles of operation | Acquires projection images sequentially as the x-ray source and detector rotate around the limb (typically 373 projections per rotation)
Covers the region of interest by helical rotation and z-axis translation of the x-ray tube and detector.
The scanner cabinet and patient remain stationary throughout the scan. | Acquires projection images sequentially as the x-ray source and detector rotate around the limb (typically 600 projections per rotation)
Covers the region of interest by a single rotation of the x-ray tube and detector.
The scanner cabinet and patient remain stationary throughout the scan. | Similar |
| Weightbearing capability/Gantry tilt | No | Upper and lower extremity imaging | Different, does not raise new questions of safety and |
{8}
Traditional 510(k) Premarket NotificationExtremity CT Imaging System™
| Characteristic | Subject Device Extremity CT Imaging System™ | Predicate Device Carestream OnSight 3D Extremity System K160723 | Equivalence Comparison |
| --- | --- | --- | --- |
| | | | effectiveness |
| Tube modulation / AEC | No | Yes. Smart Dose. (User selected) | Different, does not raise new questions of safety and effectiveness |
| Dose check | System includes dose notifications and alerts based on pre programmed thresholds. | System includes dose notifications and alerts based on pre programmed thresholds. | Same |
| Emergency stop buttons | Available at gantry and control panel | Available at gantry and control panel | Same |
| | | | |
| Power requirements | 120 V or 230 V | 120 V or 230 V | Same |
| Scanner dimensions | Small footprint. 748mm x 1120 mm | Small footprint. 1524mm x 1830mm | Similar |
| Gantry Bore Size | Small | Small | Similar |
| Patient positioning accessories | Static arm support manually aligned in the gantry | Static arm support, manually aligned in gantry. | Similar |
| | Immobilization strapping applied (single use fabric with re-usable anchors) | Immobilization strapping applied (re-usable) | |
| | Patient sat in chair (Standard off-the-shelf office chair, not provided by manufacturer) | Patient sat in chair (Chair provided by manufacturer) | |
| Radiation shielding | Internal shielding in scanner cabinet | Internal shielding in scanner cabinet and additional lead curtain. | Similar |
| Portability | Scanner mounted on roll wheels | Scanner mounted on roll wheels | Same |
{9}
Traditional 510(k) Premarket NotificationExtremity CT Imaging System™
| Characteristic | Subject Device Extremity CT Imaging System™ | Predicate Device Carestream OnSight 3D Extremity System K160723 | Equivalence Comparison |
| --- | --- | --- | --- |
| Reconstruction method | Iterative reconstruction (statistical, non-linear) | Iterative reconstruction (algebraic, linear)
Or
Filtered back projection | Similar |
| Reconstructed images | Cross-sectional CT imaging in Hounsfield units (HU) | Cross sectional CT imaging in Hounsfield units (HU). | Similar |
| Reconstructed voxel size | 0.09mm - 0.5mm
User selected | 0.26mm – 0.52 mm
User selected | Similar |
| Metal artifact reduction algorithms | Not required | Optional | Different, does not raise new questions of safety and effectiveness |
| Image visualization platform | Yes (MARS Vision), Microsoft Windows based operating system | Yes (Image View), Microsoft Windows based operating system. | Same |
| Window width / window level adjustment | Yes | Yes | Same |
| Multiplanar reformats | Multiplanar reformats include axial, coronal, sagittal and oblique planes | Multiplanar reformats including axial, coronal, sagittal and oblique planes | Same |
| 3D Volume rendering | Yes | Yes | Same |
| Network connections | Ethernet connections.
PACS connectivity. | Ethernet connections.
RIS and PACS connectivity. | Similar |
| Data exports | DICOM images and structured dose report exported directly to site PACS | DICOM images and structured dose report exported directly to site PACS | Same |
| DICOM conformance | Yes | Yes | Same |
{10}
Traditional 510(k) Premarket NotificationExtremity CT Imaging System™
| Characteristic | Subject Device
Extremity CT Imaging System™ | Predicate Device
Carestream OnSight
3D Extremity System
K160723 | Equivalence
Comparison |
| --- | --- | --- | --- |
| | | | |
Based on the testing performed, including software and performance, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate device.
## SUMMARY OF NON-CLINICAL TESTING
The Extremity CT Imaging System conforms to the FDA Guidance Document 'Medical X-Ray Imaging Devices Conformance with IEC Standards: Guidance for Industry and Food and Administration Staff' and applicable clauses of the applied Recognized Consensus Standards pertinent to Registration Number 892.1750, Product Code JAK.
**Biocompatibility**: Biocompatibility was assessed per ISO 10993-1: 2018 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management system, and in line with FDA Final Guidance on Use of International Standard ISO 10993-1 (Biocompatibility).
**Performance Testing**: Performance testing was conducted to evaluate the performance of the Extremity CT Imaging System in comparison to the predicate OnSight 3D Extremity System (K160723), using the following:
- Water Phantom QA
- Uniformity
- Signal to Noise Ratio
- Noise
- Wire Phantom QA
- Modulation transfer function
- Effective slice thickness
- Contrast Resolution
- High-z materials
- Direct dose comparison
The performance of the Extremity CT imaging system was also characterized through the following tests:
- Detector characterization
- Environmental conditions testing
- Slice Sensitivity Profile
- Noise Power Spectrum
- Pulse pile-up
{11}
Traditional 510(k) Premarket NotificationExtremity CT Imaging System™
- Spectral resolution
- Count rate vs Current and Linearity of radiation output
- Deterministic Effects
- Focal Spot to Skin Distance
- Stray radiation
- Beam quality (half-value layer) and filtration
- Dose Measurements including:
- CTDI air
- CTDI 100
- Reproducibility
- Geometric efficiency
The results of these tests indicate that the Extremity CT Imaging System is substantially equivalent to the predicate and reference devices.
## SUMMARY OF CLINICAL TESTING
Visual grading studies were conducted in which representative hand, wrist, finger and thumb MSK diagnostic images, reviewed by American Board Certified Radiologists, were obtained using the subject device and it was confirmed that the reconstructed images using the subject device were of diagnostic quality.
## CONCLUSION
Based on the testing performed, including non-clinical and clinical performance testing, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate device. The similar indications for use, technological characteristics, and performance characteristics for the proposed Extremity CT Imaging System are assessed to be substantially equivalent to the predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.