K250766 · Thirona BV · JAK · Oct 2, 2025 · Radiology
Device Facts
Record ID
K250766
Device Name
LungQ 4
Applicant
Thirona BV
Product Code
JAK · Radiology
Decision Date
Oct 2, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Thirona LungQ software provides reproducible CT values for pulmonary tissue and specified endobronchial implants which is essential for providing quantitative support for diagnosis, treatment planning and follow up examination. The LungQ software can be used to support physician in the diagnosis and documentation of pulmonary tissues images (e.g., abnormalities) from CT thoracic datasets. Three-D segmentation and isolation of sub-compartments, volumetric analysis, density evaluations, estimated chronic perfusion defect analysis, fissure evaluation and reporting tools are provided.
Device Story
LungQ 4 is a standalone command-line software tool for analyzing thoracic CT datasets. It processes DICOM images to perform 3D segmentation of lungs, lobes, and airways; volumetric and density analysis; fissure evaluation; and endobronchial implant detection. It also estimates chronic perfusion defects. The software is operated by clinicians in a clinical setting to support diagnosis and treatment planning. It outputs quantitative reports that assist physicians in assessing pulmonary abnormalities and treatment efficacy. The device benefits patients by providing reproducible, quantitative data for pulmonary assessment and treatment monitoring.
Clinical Evidence
No clinical trials were conducted. Substantial equivalence was demonstrated via non-clinical bench testing, including software verification/validation, usability testing, and performance studies. Valve detection sensitivity was ≥95% with ≥90% precision. Chronic perfusion defect analysis was validated against planar perfusion scintigraphy and SPECT-CT, achieving ICC >0.75, bias <5%, and limits of agreement between -10% and 10%. Pulmonologist review confirmed quantitative support for treatment planning was equivalent to SPECT-CT.
Technological Characteristics
Software-based device; standalone command-line application; DICOM input; 3D segmentation and volumetric analysis algorithms; density evaluation; fissure analysis; endobronchial implant detection; estimated chronic perfusion defect analysis. Operates on standard computing hardware. No specific materials or energy sources as it is a software-only device.
Indications for Use
Indicated for use by physicians to support diagnosis, treatment planning, and follow-up of pulmonary tissue abnormalities using CT thoracic datasets, including analysis of pulmonary tissue and endobronchial implants.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 2, 2025
Thirona BV
% Eva Van Rikxoort
CEO
Toernooiveld 300
NIJMEGEN, 6525 EC
NETHERLANDS
Re: K250766
Trade/Device Name: LungQ 4
Regulation Number: 21 CFR 892.1750
Regulation Name: Computed Tomography X-Ray System
Regulatory Class: Class II
Product Code: JAK
Dated: March 10, 2025
Received: September 2, 2025
Dear Eva Van Rikxoort:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250766 - Eva Van Rikxoort
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250766 - Eva Van Rikxoort
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

for
Lu Jiang Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiologic Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250766 | |
| Device Name LungQ 4 | |
| Indications for Use (Describe) The Thirona LungQ software provides reproducible CT values for pulmonary tissue and specified endobronchial implants which is essential for providing quantitative support for diagnosis, treatment planning and follow up examination. The LungQ software can be used to support physician in the diagnosis and documentation of pulmonary tissues images (e.g., abnormalities) from CT thoracic datasets. Three-D segmentation and isolation of sub-compartments, volumetric analysis, density evaluations, estimated chronic perfusion defect analysis, fissure evaluation and reporting tools are provided. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K250766
LungQ 4
Premarket Notification:
Traditional 510(k)
# 510(k) Summary
## 1. SUBMITTER
Submitted by:
Thirona BV
Toernooiveld 300
6525 EC Nijmegen
the Netherlands
Contact Person:
Eva van Rikxoort
Telephone Number: +31 6 47 14 28 38
Email: evavanrikxoort@thirona.eu
Date Prepared: 1st October 2025
## DEVICE
| Trade Name | LungQ 4 |
| --- | --- |
| Common Use/Usual Name | Computer Tomography X-ray system |
| Product Code | JAK |
| Classification | Class II, 21 CFR 892.1750 |
| Device Panel | Radiology |
## PREDICATE DEVICE
| Predicate Device | LungQ v3.0.0 |
| --- | --- |
| Predicate Classification | Class II, 21 CFR 892.1750 |
## 2. REFERENCE DEVICE
| Reference Device | Thoracic VCAR with GSI Pulmonary Perfusion |
| --- | --- |
| Reference Classification | Class II, 21CFR 892.2050 |
Thirona
{5}
LungQ 4
Premarket Notification: Traditional 510(k)
# DEVICE DESCRIPTION
The LungQ software is designed to aid in the interpretation of Computed Tomography (CT) scans of the thorax that may contain pulmonary abnormalities. LungQ is a docker image with a standalone command-line software which must be run from a command-line interpreter and does not have a graphical user interface.
# INDICATION FOR USE
The Thirona LungQ software provides reproducible CT values for pulmonary tissue and specified endobronchial implants which is essential for providing quantitative support for diagnosis, treatment planning and follow up examination. The LungQ software can be used to support physicians in the diagnosis and documentation of pulmonary tissues images (e.g., abnormalities) from CT thoracic datasets. Three-D segmentation and isolation of sub-compartments, volumetric analysis, density evaluations, estimated chronic perfusion defect analysis, fissure evaluation and reporting tools are provided.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
Table 1 below compares the Thirona LungQ 4 software to the predicate device.
Thirona
{6}
LungQ 4
Premarket Notification:
Traditional 510(k)
Table 1: Substantial Equivalence Comparison between Subject and Predicate
| Item | LungQ 4
Thirona
(Subject device) | LungQ v3.0.0
Thirona
(Predicate Device) |
| --- | --- | --- |
| 510(k) Number | K250766 | K232412 |
| Product Code | JAK | Same |
| Regulation Number | 21 CFR 892.1750 | Same |
| Device Classification | Class II | Same |
| Common Name | Computed tomography x-ray system | Same |
| Indication for use | The Thirona LungQ software provides reproducible CT values for pulmonary tissue and specified endobronchial implants which is essential for providing quantitative support for diagnosis, treatment planning and follow up examination. The LungQ software can be used to support physician in the diagnosis and documentation of pulmonary tissues images (e.g., abnormalities) from CT thoracic datasets. Three-D segmentation and isolation of sub-compartments, volumetric analysis, density evaluations, estimated chronic perfusion defect analysis, fissure evaluation and reporting tools are provided. | Equivalent |
| Modality | CT | Same |
| Data Loading | DICOM | Same |
| Application | Command-line interface | Same |
| Segmentation | Provides 3D segmentation | Same |
| | Provides Segmentation of the:
• Left Lung
• Right Lung
• Left Upper Lobe
• Left Lower Lobe | Same |
Thirona
{7}
LungQ 4
Premarket Notification: Traditional 510(k)
| | • Right Upper Lobe
• Right Middle Lobe
• Right Lower Lobe
• Pulmonary (sub)segments | |
| --- | --- | --- |
| | Provides Segmentation of the:
Airways | Same |
| | Provides segmentation of specified endobronchial implants | Equivalent |
| Lung Volume Analysis Support | Ability to measure volume
for:
• Both Lungs
• Left Lung
• Right Lung
• Left Upper Lobe
• Left Lower Lobe
• Right Upper Lobe
• Right Middle Lob
• Right Lower Lobe
• Pulmonary (sub)segments | Same |
| Volume Density Analysis | Ability to measure volume at multiple density ranges for:
• Both Lungs
• Left Lung
• Right Lung
• Left Upper Lobe
• Left Lower Lobe
• Right Upper Lobe
• Right Middle Lob
• Right Lower Lobe
• Pulmonary (sub)segments | Same |
Thirona
{8}
LungQ 4
Premarket Notification: Traditional 510(k)
| | Ability to measure the 15th percentile density analysis | Same |
| --- | --- | --- |
| Fissure Analysis | Ability to perform fissure evaluations | Same |
| Endobronchial implant analysis | Ability to report the number of specified endobronchial implants in the lungs. | Equivalent |
| Estimated chronic perfusion defect analysis | Ability to report estimated chronic perfusion defect quantification for:
• Both Lungs
• Left Lung
• Right Lung
• Left Upper Lobe
• Left Lower Lobe
• Right Upper Lobe
• Right Middle Lob
• Right Lower Lobe
• Pulmonary (sub)segments | Equivalent |
| Analyzed Data Output | Provides a report | Same |
Thirona
{9}
LungQ 4
Premarket Notification: Traditional 510(k)
# PERFORMANCE DATA
The performance testing for LungQ 4 consists of norm-compliance testing, design verification and validation testing. For norm-compliance and design verification, the testing was performed per subject device, while the validation, such as non-clinical validation and usability validation, were performed at system level representing a logical clinical workflow following the intended use.
Performance testing data of the proposed devices demonstrate that the subject device is substantially equivalent to the predicate device, and that the design output meets the design input requirements.
# Performance Testing
Software Verification testing was conducted to ensure that the LungQ 4 software met its requirements. The verification testing included white box testing to verify implementation and system integration testing. The LungQ 4 software successfully passed the verification testing.
Software Validation was conducted to ensure the software met the user needs (i.e. input requirements). This validation was based on user scenarios. The LungQ 4 software successfully passed the software validation.
Human factors (HF) engineering process was followed in accordance with the usability standard and FDA guidance. The restricted interface of LungQ 4 with the third party end-user-interface is well controlled by the use of standard input and output formats. The usability is determined by the input requirements, which have been identified during risk management. The human factors validation test (summative usability evaluation) of LungQ 4 was performed.
# Substantial equivalence Study
A head-to-head performance testing was conducted between the subject and the predicate device. The aim of this study was to assess and compare the measurement of lung structure parameters, such as (sub)segment volume and density analysis, density scores (LAA-950HU and LAA-910HU) and airway segmentation analysis between LungQ 4 (subject device) and LungQ v3.0.0 (predicate device, K232412). The results showed that for volume, density scores and airway segmentation, the acceptance criteria were met.
In addition, a study has been performed to provide performance testing of the endobronchial implant analysis feature. The study dataset included multi-center CT scans of patients that underwent endobronchial valve (EBV) treatment and CT scans from negative controls (i.e. patients eligible for EBV treatment but without EBV). Reference valve segmentations have been annotated on the dataset by clinical truthers and output metrics are directly compared to the reference Detection performance was evaluated by calculating the sensitivity and precision of the valve analysis component. A valve detection sensitivity $\geq 95\%$ and a valve detection precision of $\geq 90\%$ was achieved and therefore the performance of the valve analysis component met the acceptance criteria.
Thirona
{10}
LungQ 4
Premarket Notification: Traditional 510(k)
The added CT derived chronic perfusion defect analysis feature estimates chronic perfusion defects from CT images, similarly to GSI in the Reference Device. The feature was evaluated on a clinical dataset of patients who underwent HRCT and planar perfusion scintigraphy or HRCT and SPECT-CT. Two separate analyses were performed to compare reference perfusion scores (either extracted from planar perfusion scintigraphy or SPECT-CT) to the LungQ 4 CT derived chronic perfusion defect scores. Agreement between LungQ 4 and both planar perfusion scintigraphy and SPECT-CT was assessed through intra-class correlation coefficients (ICC) and Bland-Altman analysis. An ICC > 0.75, bias < 5% and limits of agreements between -10% and 10% were required for estimated chronic perfusion defect measurements and reference perfusion scores for the left lung, right lung and each of the individual lobes. In addition to the direct comparison of perfusion values, experienced pulmonologists reviewed whether the use of LungQ 4 CT derived chronic perfusion defect analysis would provide similar quantitative support for treatment planning as SPECT-CT. The results confirmed that the provided quantitative support was the same for LungQ 4 and SPECT-CT. All acceptance criteria for direct comparison and treatment planning support have been met. We conclude that LungQ 4 is similar to the chosen reference device for chronic perfusion defect measurement method in terms of technological characteristics and does not raise new questions on the safety and effectiveness.
## Conclusion on performance testing
All compliance, verification and validation tests have been used to support substantial equivalence of the subject device and to demonstrate that the LungQ 4 device:
- comply with the aforementioned international and FDA recognized consensus standards and FDA guidance documents; and
- meet the acceptance criteria and are adequate for their intended use.
Based on the information provided above, the LungQ 4 device is considered substantially equivalent to the predicate device in terms of safety and effectiveness.
No clinical testing was required as substantial equivalence was demonstrated by the attributes of intended use, technological characteristics, and non-clinical testing.
## CONCLUSIONS
The proposed device, LungQ 4, is substantially equivalent to the above-mentioned predicate device, in terms of intended use, technological characteristics and, safety and effectiveness.
Substantial equivalence was demonstrated by non-clinical performance tests provided in this 510(k) premarket notification. These tests demonstrate that the proposed device comply with the user needs specifications and product requirements, as well as the requirements specified in the international and FDA-recognized consensus standards, and are as safe and effective as the predicate device, and do not raise any new safety and/or effectiveness concerns.
Thirona
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.