Deepsight NeedleVue LC1 Ultrasound System

K250381 · DeepSight Technology, Inc. · IYO · Aug 1, 2025 · Radiology

Device Facts

Record IDK250381
Device NameDeepsight NeedleVue LC1 Ultrasound System
ApplicantDeepSight Technology, Inc.
Product CodeIYO · Radiology
Decision DateAug 1, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1560
Device ClassClass 2
AttributesPediatric

Indications for Use

The DeepSight NeedleVue LC1 Ultrasound System is intended for use in Abdominal, Pediatric, Small Organ, Peripheral Vascular, Fetal, Urological, and Musculoskeletal Conventional and Superficial Imaging. This device is indicated for Prescription Use Only and by qualified healthcare professionals (HCPs). The DeepSight NeedleVue LC1 ultrasound system is intended to be used by qualified and trained physicians or sonographers with at least basic ultrasound knowledge. The operator must have read and understood the user manual. The DeepSight NeedleVue LC1 ultrasound system is intended to be used in medical practices and clinical environments including healthcare facilities, hospitals and clinics.

Device Story

General-purpose diagnostic ultrasound system mounted on a movable cart; includes mobile console for digital acquisition, processing, and display. System comprises three functional blocks: front end (curved array transducer, analog signal processing, A/D conversion, noise filtering), back end (user interface, memory, USB export, LCD touch screen for parameter control/annotation/measurements), and power systems. Operated by physicians or sonographers in hospitals/clinics. Transducer transmits acoustic energy into body; receives reflections; performs signal processing. Healthcare providers use real-time B-mode/harmonic imaging to visualize anatomy; output aids in diagnostic evaluation. Benefits include non-invasive imaging for various clinical applications.

Clinical Evidence

No clinical data. Bench testing only. Compliance demonstrated with IEC 60601-1, IEC 60601-2-37, IEC 62359, IEC 62127-1/2, and IEC 62304. Performance testing followed FDA guidance for diagnostic ultrasound systems and software functions.

Technological Characteristics

General-purpose diagnostic ultrasound system; curved (convex) array transducer; 120 VAC power. Features LCD touch screen interface. Beamforming on transmit/receive; digital image reconstruction/processing. Software lifecycle per IEC 62304. Electromagnetic compatibility per IEC 60601-1-2. Thermal/mechanical safety per IEC 62359 and IEC 60601-2-37.

Indications for Use

Indicated for adult and pediatric patients requiring ultrasound imaging in abdominal, pediatric, small organ, peripheral vascular, fetal, urological, and musculoskeletal conventional and superficial applications. For use by qualified physicians or sonographers in clinical settings.

Regulatory Classification

Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION DeepSight Technology, Inc. David Surber RA/QA Manager 2953 Bunker Hill Ln SANTA CLARA, CA 95054 August 1, 2025 Re: K250381 Trade/Device Name: DeepSight NeedleVue LC1 Ultrasound System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic Pulsed Echo Imaging System Regulatory Class: Class II Product Code: IYO, ITX Dated: July 15, 2025 Received: July 15, 2025 Dear David Surber: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250381 - David Surber Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. {2} K250381 - David Surber Page 3 For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Paramita Sengupta -S For Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K250381 Device Name DeepSight NeedleVue LC1 Ultrasound System Indications for Use (Describe) The DeepSight NeedleVue LC1 Ultrasound System is intended for use in Abdominal, Pediatric, Small Organ, Peripheral Vascular, Fetal, Urological, and Musculoskeletal Conventional and Superficial Imaging. This device is indicated for Prescription Use Only and by qualified healthcare professionals (HCPs). The DeepSight NeedleVue LC1 ultrasound system is intended to be used by qualified and trained physicians or sonographers with at least basic ultrasound knowledge. The operator must have read and understood the user manual. The DeepSight NeedleVue LC1 ultrasound system is intended to be used in medical practices and clinical environments including healthcare facilities, hospitals and clinics. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov > "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} K250381 510(k) Summary 21 CFR 807.92 Submitter: DeepSight Technology, Inc. 2953 Bunker Hill Lane Santa Clara, CA 95054 Company/Submission Contact: David Surber Tel: 1.415.225.9937 Email: dsurber@deepsight.com Date Summary Prepared: February 10, 2025 Device Trade/Proprietary Name: DeepSight NeedleVue LC1 Ultrasound System Classification name, Class II Predicate Device: GE Healthcare Logiq E9 Ultrasound System (K152309) Subject Device Description: The DeepSight NeedleVue LC1 Ultrasound System is a general-purpose diagnostic ultrasound system that is mounted on a movable cart and has a mobile console that provides digital acquisition, processing, and display capabilities. The subject device has three major functional blocks that are consistent across ultrasound systems including the predicate: - A front end, which includes a single curved (a.k.a. convex) array transducer and analog signal processing functions. This transducer transmits acoustic energy into the body and receives the resulting reflections, and performs the signal processing functions on them required to produce an ultrasound image (e.g., analog to digital conversion, noise filtering); - A back end, which includes a user interface, memory, and export via USB. The user interfaces include a computer keyboard, standard ultrasound parameter controls, an LCD touch screen, acoustic output display and an LCD image display. The touch screen is divided into a display area and a user interaction area, and allows for patient data entry, transmit voltage adjustment, transducer/preset selection, depth and focus adjustment, annotation, freeze, image and clip capture, and measurements. - Power systems which generate, regulate and supply the required voltages to the system parts. {5} Subject Device Intended Use: The DeepSight NeedleVue LC1 Ultrasound System is intended for use in Abdominal, Pediatric, Small Organ, Peripheral Vascular, Fetal, Urological, and Musculoskeletal Conventional and Superficial Imaging. This device is indicated for Prescription Use Only and by qualified healthcare professionals (HCPs). The DeepSight NeedleVue LC1 Ultrasound System is intended to be used by qualified and trained physicians or sonographers with at least basic ultrasound knowledge. The operator must have read and understood the user manual. The DeepSight NeedleVue LC1 Ultrasound System is intended to be used in medical practices and clinical environments including healthcare facilities, hospitals and clinics. Substantial Equivalence Comparison | | Subject Device: DeepSight NeedleVue LC1 Ultrasound System | Predicate Device: GE LOGIQ E9 (K152309) | | --- | --- | --- | | Product Codes | IYO - Ultrasonic Pulsed Echo Imaging SystemITX - Diagnostic Ultrasonic Transducer | IYO - Ultrasonic Pulsed Echo Imaging SystemITX - Diagnostic Ultrasonic TransducerIYN - Ultrasonic Pulsed Doppler Imaging System | | Classification Regulations | 892.1560 - Ultrasonic Pulsed Echo Imaging System; biopsy needle guide kit892.1570 - Diagnostic Ultrasound Transducer | 892.1550 - Ultrasonic Pulsed Doppler Imaging System892.1560 - Ultrasonic Pulsed Echo Imaging System892.1570 - Diagnostic Ultrasound Transducer | | General Intended Use | Ultrasound imaging and evaluation of adult and pediatric patients | Ultrasound imaging and evaluation of adult and pediatric patients | | Indications for Use | The DeepSight NeedleVue LC1 Ultrasound System is intended for use in Abdominal, Pediatric, Small Organ, Peripheral Vascular, Fetal, Urological, and Musculoskeletal Conventional and Superficial Imaging. This device is indicated for Prescription Use Only and by qualified healthcare professionals (HCPs). The DeepSight NeedleVue LC1 Ultrasound System is intended to be used by qualified and trained physicians or sonographers with at least basic ultrasound knowledge. The operator must have read and understood the user manual. The DeepSight NeedleVue LC1 Ultrasound System is intended to be used in medical practices and clinical environments including | Intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculoskeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic and vascular). | {6} | | Subject Device: DeepSight NeedleVue LC1 Ultrasound System | Predicate Device: GE LOGIQ E9 (K152309) | | --- | --- | --- | | | healthcare facilities, hospitals and clinics. | | | Number of transducers provided | 1 | 34 | | Imaging Modes | Ultrasound B-Mode, including harmonic imaging. Operating Modes: 2D-mode; 2D-mode with Harmonics Imaging | Ultrasound B, M, Color M, Color & Power Doppler and various combinations: B/M B/PW, Color/Pwr/PW, Harmonic, Coded Pulse, Realtime 3D & Multi-place, Elastography Imaging, Sear wave elastography (SWE) | | User | Qualified physicians and trained healthcare professionals in hospitals, clinics and other point of care locations | Qualified physicians | | Contact Type | Body surface | Body surface, Cavitary (TV, TR, TE) and intraoperative | | Transducer type | Curved array | Support for: Matrix Array, Convex Array, Linear Array, Micro Convex Array, Sector Phased Array, Volumes Probes | | Power source | 120 VAC | 120 VAC | | Beamforming | Beamforming on both the transmit and receive sides, and data undergoes imaging reconstruction and processing | Beamforming on both the transmit and receive sides, and data undergoes imaging reconstruction and processing | | Controls and Display | Touchscreen divided into image display and user interaction area | Control panel with touch screen for user interaction and tilt/swivel LCD type image monitor | | Screen mount | Movable cart | Movable cart | As shown in the comparison table, the predicate device, the Logiq E9 Ultrasound System (K152309) is indicated for use in a broader range of anatomical regions than the subject device. The predicate device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic, vascular and neurosurgical) while the subject device for just a subset of these applications. The subject device and the predicate device have the same intended use and substantially the same indications for use. More specifically, in both the subject and predicate devices, the ultrasound system is intended for general ultrasound evaluation by a qualified health care {7} professional. These differences do not affect the safety and effectiveness of the device when used as labeled. ## Summary of comparison of technological characteristics Like the predicate LOGIQ E9, the DeepSight NeedleVue LC1 Ultrasound System is a general-purpose diagnostic ultrasound system that is mounted on a movable cart and has a mobile console that provides digital acquisition, processing, and display capabilities. Moreover, the subject device includes the same three major functional blocks that are consistent across ultrasound systems including the predicate, the front end, back end and power systems. Although some differences exist between the two ultrasound systems, these differences are limited and do not raise different questions of safety or effectiveness, as they do not introduce any new functionalities or alter how the device is used in any meaningful way. The two systems are technologically equivalent in all material respects. ## Performance Testing: The bench testing conducted consisted of all applicable testing specified in FDA’s guidance documents, *Marketing Clearance of Diagnostic Ultrasound Systems and Transducers* (February 2023) and *Content of Premarket Submissions for Device Software Functions* (June 2023). In addition, compliance data was provided for the following recognized standards and FDA Guidance: ### Software: - IEC 62304 Ed 1.1 Consolidated Version Medical Device Software – Software Life Cycle Processes - FDA guidance *Content of Premarket Submissions for Device Software Functions* - FDA guidance: *Off-The-Shelf Software Use in Medical Devices* - FDA guidance: *Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions* ### Ultrasound System and Transducer **Electromagnetic Compatibility** - IEC 60601-1-2 Ed. 4.1 Consolidated Version Collateral Standard: *Electromagnetic disturbances - Requirements and tests* - FDA guidance: *Electromagnetic Compatibility (EMC) of Medical Devices* **Thermal, Mechanical and Electrical Safety** - ES 60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2] *Medical electrical equipment – Part 1: General requirements for basic safety and essential performance* - IEC 60601-2-37 Edition 2.1, *Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment* 4 {8} 5 Thermal and Mechanical Safety - IEC 62359 Ed. 2.1 Consolidated Version, Ultrasonics – Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields - FDA guidance: Marketing Clearance of Diagnostic Ultrasound Systems and Transducers Acoustic Output - IEC 62127-1 Ed. 2.0, Ultrasonics – Hydrophones – Part 1: Measurement and characterization of medical ultrasonic fields - IEC 62127-2 Ed. 1.2 Consolidated Version, Ultrasonics – Hydrophones -- Part 2: Calibration for ultrasonic fields up to 40 MHz Acoustic Power - FDA guidance: Marketing Clearance of Diagnostic Ultrasound Systems and Transducers Reprocessing - FDA guidance: Marketing Clearance of Diagnostic Ultrasound Systems and Transducers - FDA guidance: Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Summary - Performance Testing (nonclinical) The DeepSight NeedleVue LC1 Ultrasound System was found to have a safety and effectiveness profile that is similar to the predicate device based on the performance testing conducted of all applicable testing specified in FDA’s guidance documents, Marketing Clearance of Diagnostic Ultrasound Systems and Transducers (February 2023) and Content of Premarket Submissions for Device Software Functions (June 2023), along with compliance data provided for the following recognized standards and FDA Guidance. Clinical Testing Clinical data was not required for this submission. Conclusion The non-clinical (performance) data supports that the DeepSight NeedleVue LC1 Ultrasound System is as safe, as effective and performs as well as the predicate device. The hardware and software verification and validation demonstrate that the DeepSight NeedleVue LC1 Ultrasound System should perform as intended in the specified use conditions and demonstrates that the DeepSight NeedleVue LC1 Ultrasound System performs comparable to the predicate device that is currently marketed for the same intended use.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...