Sonosite iLOOK Ultrasound System

K260595 · FUJIFILM Sonosite, Inc. · IYN · Mar 10, 2026 · Radiology

Device Facts

Record IDK260595
Device NameSonosite iLOOK Ultrasound System
ApplicantFUJIFILM Sonosite, Inc.
Product CodeIYN · Radiology
Decision DateMar 10, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric, 3rd-Party Reviewed

Indications for Use

The Sonosite iLOOK Ultrasound System is a general purpose ultrasound system and non-continuous patient monitoring platform intended for use by qualified physicians and healthcare professionals for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: •Pediatric •Peripheral vessel •Needle Guidance Modes of operation include: B Mode (B), Color Power Doppler, Color Velocity Doppler, Multi-beam imaging, and combined modes, including duplex imaging: B+C. This device is indicated for Prescription Use Only. The Sonosite iLOOK system is intended to be used in medical practices, clinical environments, including healthcare facilities, hospitals, clinics, and clinical point-of-care for diagnosis of patients. The system includes a transducer and a display connected together wirelessly. The transducer is powered by battery, while the display is powered either by battery or by AC electrical power. The clinician is positioned next to the patient and places the transducer onto the patient's body where needed to obtain the desired ultrasound image.

Device Story

Portable, all-digital ultrasound system; consists of 10.1-inch tablet display and wireless L15-5 linear transducer. Clinician operates device at point-of-care; places transducer on patient body to acquire real-time 2D, Color Power Doppler, or Color Velocity Doppler images. System processes acoustic signals to display anatomical structures and fluid flow; supports pediatric biometry and needle guidance. Data transferred via wireless networking; DICOM-compatible for storage and review. Output assists clinicians in diagnostic decision-making. Powered by battery or AC; includes optional stand with wireless charger.

Clinical Evidence

No clinical studies required for substantial equivalence. Device performance verified through bench testing, including electrical, thermal, mechanical, and EMC safety, biocompatibility, and acoustic output evaluation. Successfully tested in a clinical environment to confirm expected performance.

Technological Characteristics

All-digital architecture; 10.1-inch LCD tablet (1920x1200); wireless L15-5 linear transducer (5-15 MHz). Patient contact materials: m-PPE, SEBS elastomer, silicone rubber (Shin-Etsu KE-752U/KE-45T). Connectivity: 802.11 wireless, DICOM 3.0. Power: Battery/AC. Standards: ISO 10993-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62304, ISO 14971, IEC 62359.

Indications for Use

Indicated for pediatric patients, peripheral vessel evaluation, and needle guidance in medical practices, clinical environments, hospitals, and point-of-care settings. Prescription use only.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION March 10, 2026 FUJIFILM Sonosite, Inc. % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Dr. Suite #510k Saint Paul, Minnesota 55114 Re: K260595 Trade/Device Name: Sonosite iLOOK Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, ITX, IYO Dated: February 23, 2026 Received: February 23, 2026 Dear Prithul Bom: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K260595 - Prithul Bom Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the Medical Device File (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K260595 - Prithul Bom Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # YANNA S. KANG -S Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260595 | ? | | Please provide the device trade name(s). | | ? | | Sonosite iLOOK Ultrasound System | | | | Please provide your Indications for Use below. | | ? | | The Sonosite iLOOK Ultrasound System is a general purpose ultrasound system and non-continuous patient monitoring platform intended for use by qualified physicians and healthcare professionals for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: •Pediatric •Peripheral vessel •Needle Guidance Modes of operation include: B Mode (B), Color Power Doppler, Color Velocity Doppler, Multi-beam imaging, and combined modes, including duplex imaging: B+C. This device is indicated for Prescription Use Only. The Sonosite iLOOK system is intended to be used in medical practices, clinical environments, including healthcare facilities, hospitals, clinics, and clinical point-of-care for diagnosis of patients. The system includes a transducer and a display connected together wirelessly. The transducer is powered by battery, while the display is powered either by battery or by AC electrical power. The clinician is positioned next to the patient and places the transducer onto the patient's body where needed to obtain the desired ultrasound image. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} 510(K) Summary K260595 This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92. 1) Submitter: Manufacturer Name: FUJIFILM SonoSite, Inc. Address: 21919 30th Drive SE Bothell, WA 98021-3904 Corresponding Official: Anoush Frankian Director, Regulatory Affairs Email: anoush.frankian@fujifilm.com Telephone: (425) 951-6824 Facsimile: (425) 951-1201 Date prepared: February 6, 2026 Alternate Contact: Abirami Krishnamurthi Regulatory Affairs Specialist abirami.krishnamurthi@fujifilm.com Telephone: (425) 951-6824 Facsimile: (425) 951-1201 2) Device Trade Name: Sonosite iLOOK Ultrasound System Common Name: Diagnostic Ultrasound System and Transducers with Accessories Regulation Name: Ultrasonic Pulsed Doppler Imaging System Ultrasonic Pulsed Echo Imaging System Diagnostic Ultrasound Transducer Regulation Number: 892.1550 892.1560 892.1570 Primary Product Code: IYN Page 1 of 16 {5} Secondary Product Codes: IYO ITX Device Class: Class II Classification Panel: Radiology ## 3) Predicate Device: Primary Predicate: Sonosite iViz Ultrasound System (K180704) Regulation Name: Ultrasonic Pulsed Doppler Imaging System Ultrasonic Pulsed Echo Imaging System Diagnostic Ultrasound Transducer Regulation Number: 892.1550 892.1560 892.1570 Primary Product Code: IYN Secondary Product Code: IYO, ITX Device Class: II Classification Panel: Radiology Reference Device: Vscan Air (K250087) Regulation Name: Ultrasonic Pulsed Doppler Imaging System Ultrasonic Pulsed Echo Imaging System Diagnostic Ultrasound Transducer Regulation Number: 892.1550 892.1560 892.1570 Primary Product Code: IYN Secondary Product Code: IYO, ITX Device Class: II Classification Panel: Radiology Page 2 of 16 {6} Page 3 of 16 ## 4) Device Description: The Sonosite iLOOK Ultrasound System is a fully featured, general-purpose, software-controlled diagnostic ultrasound system utilizing an all-digital architecture. The system is a portable ultrasound device consisting of a display, adjustment handle, wireless L15-5 linear ultrasound transducer, and a transducer holder with a built-in wireless charger. The system is designed to acquire and display high-resolution, real-time ultrasound data in 2D, Color Power Doppler (CPD), Color Velocity Doppler (CVD), or a combination of these modes. It provides measurement capabilities for anatomical structures and pediatric biometry, offering information for clinical diagnostic purposes. The system includes a variety of accessories, including an optional stand. Additionally, it features Digital Imaging and Communications in Medicine (DICOM) capabilities, along with general computer communication functionalities, including wireless networking. These capabilities enable the acceptance, transfer, display, storage, and digital processing of ultrasound images and loops. The system features a single USB port, allowing the use of compatible storage devices, such as memory sticks and hard drives. An AC adapter is included in the stand module, enabling manual connection to AC power. ## Intended Use/Indications for Use: The Sonosite iLOOK Ultrasound System is a general-purpose ultrasound system and non-continuous patient monitoring platform intended for use by qualified physicians and healthcare professionals for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: - Pediatric - Peripheral vessel - Needle Guidance Modes of operation include: B Mode (B), Color Power Doppler, Color Velocity Doppler, Multi-beam imaging, and combined modes, including duplex imaging: B+C. This device is indicated for Prescription Use Only. The Sonosite iLOOK system is intended to be used in medical practices, clinical environments, including healthcare facilities, hospitals, clinics, and clinical point-of-care for diagnosis of patients. The system includes a transducer and a display connected together wirelessly. The transducer is powered by battery, while the display is powered either by battery or by AC electrical power. The clinician is positioned next to the patient and places the transducer onto the patient's body where needed to obtain the desired ultrasound image. {7} 5) Technological Characteristics: The Sonosite iLOOK Ultrasound System, subject device of this submission, is equivalent to the previously cleared primary predicate Sonosite iViz Ultrasound System (K180704) and the reference device, Vscan Air (K250087) in terms of both the intended use and technological characteristics. The Sonosite iLOOK (subject device) uses the same fundamental scientific technology as the primary predicate device. Page 4 of 16 {8} | Feature | Sonosite iLOOK Ultrasound System (This submission) | Sonosite iViz Ultrasound System (K180704)- Primary Predicate | VScan Air (K250087)- Reference | Evaluation of Differences | | --- | --- | --- | --- | --- | | Intended Use | Diagnostic ultrasound imaging or fluid flow analysis of the human body. | Diagnostic ultrasound imaging or fluid flow analysis of the human body. | Diagnostic ultrasound examinations, image guidance and for measurements of anatomical structures and fluid. | Same as the primary predicate Sonosite iViz Ultrasound System (K180704). | | Indications for Use | Pediatric Peripheral vessel Needle guidance | Fetal – OB/GYN Abdominal Pediatric Small Organ (breast, thyroid, testicles, prostate) Musculoskeletal (Conventional) Musculoskeletal (Superficial) Cardiac Adult Cardiac Pediatric Peripheral vessel Ophthalmic | Vascular/ peripheral vascular musculoskeletal (conventional and superficial) small organs thoracic/lung ophthalmic pediatric neonatal cephalic interventional guidance (includes free-hand needle/ catheter placement, fluid drainage, nerve block, vascular access and biopsy) | Same. The indications for use include the subset of indications previously cleared on the primary predicate, Sonosite iViz Ultrasound system (K180704) and the reference device, Vscan Air (K250087). | | Transducer Types | Linear Array | Curved Array Linear Array Phased Array | Linear array Sector array | Same. Sonosite iLOOK Ultrasound System include the subset of the transducer types that were previously cleared on the primary predicate Sonosite iViz Ultrasound System (K180704) and reference device Vscan Air (K250087). | Page 5 of 16 {9} | Transducer Frequency | 5- 15 MHz | 6-13 MHz | 3 – 12 MHz | The transducer frequency range is similar to both primary predicate Sonosite iViz Ultrasound System (K180704) and reference device Vscan Air (K250087). iLOOK L15-5 and iViz L25v are both linear transducers, while iLOOK L15-5 covers a wider bandwidth and higher frequency. | | --- | --- | --- | --- | --- | | Global Maximum Outputs/Worst Case Setting | Ispta.3=138.41mW/cm2 (Color, ROI Deep) TI= 0.42 MI =0.80 Ipa.3@MI Max: 268.8 mW/cm² | Ispta.3= 47.7 mW/cm² TI Value: 0.42 MI: 0.86 Ipa.3@MI Max: 272 W/cm² | MI <1.9 Ispta<720 mW/cm^2 | Acoustic output is within the FDA established limits. | | Acoustic Output Display & FDA Limits | No output display because MI<1.0, TI<1.0 | MI and TI are displayed | MI and TI values are displayed on the scanning screen. | MI and TI values are less than 1.0. IEC 60601-2-37 Ed 3.0 section 201.12.4.2 "Indication relevant to safety" b) and c) which states no display is required if you meet the criteria. | | Modes of Operation | 2D (B-mode Grayscale Imaging) Color Velocity Doppler Color Power Doppler Combination modes (color mode is combined with 2D) | B-mode Grayscale Imaging Tissue Harmonic Imaging M-mode Color M-Mode Color Power Doppler Zoom Combination Modes SonoHD3 Noise Reduction | B-mode Color Doppler M-mode PW | Modes of operations include the subset of the previously cleared modes on the primary predicate (K180704). | Page 6 of 16 {10} | | | Velocity Color Doppler | | | | --- | --- | --- | --- | --- | | DICOM | DICOM 3.0 Image and Multi-frame Image Storage; Modality Worklist; BCP 195 RFC 8996 TLS Secure Transport Connection | DICOM 3.0 Image, Multi-frame Image, Secondary Capture, and Basic Text Structured Report Storage; Modality Worklist; USB Media Storage | DICOM 3.0 Modality Worklist Server DICOM Image Server Network Shared Folder DICOM Web Server | Same as the primary predicate (K180704) | | #Transmit Channels | 64 | 64 | No information available publicly | Same as the primary predicate (K180704) | | #Receive Channels | 64 | 64 digital channels using Synthetic Aperture | No information available publicly | Same as the primary predicate (K180704) | | Patient Contact Materials | Transducers: m-PPE SEBS Elastomer Silicone rubber compounds (Shin-Etsu KE-752U) Silicone rubber compounds (Shin-Etsu KE-45T) | Transducers: Polysulfone UDEL P1700 Poly-Vinyl-Chloride (PVC) Silicone Rubber | No information available publicly | The materials introduced in this submission have passed biocompatibility testing in accordance with ISO 10993-1, and the results are provided in the biocompatibility section of the eSTAR. | Page 7 of 16 {11} | | | | | | | --- | --- | --- | --- | --- | | DICOM | DICOM PS3.15 | NEMA PS3.15 | DICOM PS3.15 | Same | Page 8 of 16 {12} | System characteristics | Tablet dimensions: 10.1 inches 1920x1200 pixels LCD Operating system: Android iLOOK ultrasound software running as an “app” on tablet System operates via battery Wireless 802.11 support for image transfer Transducer dimensions: L15-5 transducer • Length: 151 mm (5.94 in) • Width: 56.5 mm (2.22 in) • Depth: 29.5 mm (1.16 in) • Weight: 170 g (6 oz) Stand • Depth: 54.4 cm (21.42 in) wheels turned in; 59.6 cm (23.46 in) wheels turned out • Width: 49.2 cm (19.37 in) wheels | Tablet dimensions: 7" inches 1920 x 1200 pixels LCD Operating system: Android iViz ultrasound software running as an “app” on tablet System operates via battery Wireless 802.11 support for image transfer and over the-air (OTA) software updates. Dimensions: System (without the case) • Length: 7.21 in. (183.1 mm) • Width: 4.59 in. (116.5 mm) • Height: 1.06 in (26.9 mm) System (with the case) • Length: 9.5 in. | Transducer S probe dimensions: • Dimension: 141 x 67 x 33 mm (length, width, height) • Weight: 218 +/- 3g Operating system options: - Android phones and tablets with OS version 11, 12 or 13, device with 0x64 ARM based CPU architecture and 64-bit Kernel, Android open GL ES 3.0, and compatibility with Google Play store - iPad and iPhone devices with iOS 14, 15 or 16 • Screen requirements - Size: from 5 to 20 inches - 960 x 640 (or 640 x 960) pixel or more | Same as the primary predicate, iViz (K180704), iLOOK is a tablet with a 10.1 inches display that runs on an Android Operating system and a transducer connected wirelessly, which is similar to the reference device Vscan Air (K250087). | | --- | --- | --- | --- | --- | Page 9 of 16 {13} Page 10 of 16 | | turned in; 57.7 cm (22.72 in) wheels turned out • Height: 129.4 cm (50.94 in) maximum, 165.1 cm (65 in) minimum • Weight: 15.0 kg (33.07 lbs) • Storage bin capacity: 3 kg (6.6 lbs) • Tray capacity: 3 kg (6.6 lbs) • Total stand weight with system and peripherals: 17.65 kg (38.91 lbs) maximum. System (display, stand mount, and transducer holder) • Width: 36.9 cm (14.53 in) • Height: 23.6 cm (9.29 in) • Depth: 5.2 cm (2.05 in) • Diagonal: 25.65 cm (10.1 in) • Resolution: 1920 x 1200 px • Weight: 1.85 kg (4.08 lbs) | (241.3 mm) • Width: 4.8 in. (123.0 mm) • Height: 1.12 in. (28.6 mm) Display • Length: 6.37 in. (161.87 mm) • Width: 4.11 in. (104.52 mm) • Diagonal: 7.0 in. (177.8 mm) | | | | --- | --- | --- | --- | --- | {14} | | Total system (system plus stand) • Depth: 54- 70 cm (21.26- 27.56 in) • Width: 48.5- 58 cm (19.09- 22.83 in) • Height: 140- 167 cm (55.12- 65.75 in) | | | | | --- | --- | --- | --- | --- | Page 11 of 16 {15} | 510k Track | Track 3 | Track 3 | Track 3 | Same | | --- | --- | --- | --- | --- | Page 12 of 16 {16} 6) Determination of Substantial Equivalence: Summary of Technological Comparison to Predicate Devices: The Sonosite iLOOK Ultrasound System, subject device of this submission, is an enhanced implementation of previous FDA cleared primary predicate device Sonosite iViz Ultrasound System (K180704) and reference device, Vscan Air (K250087). The primary function of Sonosite iLOOK Ultrasound System (this submission) and the predicate device is diagnostic ultrasound imaging or fluid flow analysis of the human body. The Sonosite iLOOK Ultrasound System employs the same fundamental scientific characteristics as the currently marketed primary predicate and reference devices. The Sonosite iLOOK Ultrasound System and both primary predicate (K180704) and reference device (K250087) share indications for use, share modes of operation and have biosafety equivalence. The following lists an overview of differences between the proposed subject device (Sonosite iLOOK Ultrasound System) and its predicates. - The Sonosite iLOOK Ultrasound System offers an all-touch display that is similar to the primary predicate device iViz (K180704) with a slightly larger size on the tablet of 10.1 inches compared to the 7 inches offered on the primary predicate iViz (K180704). Like the primary predicate (K180704), the iLOOK Ultrasound System operates via battery or AC electric power. The tablet is offered with a stand. - The Sonosite iLOOK Ultrasound System utilizes wireless connectivity for its wireless ultrasound transducer, which is the same as the reference device Vscan Air's (K250087) SL probe that has wireless connectivity capability. - The Sonosite iLOOK Ultrasound System also includes the addition of Needle Guidance as an indication, which is the same as the Needle Guidance indication offered on the reference device Vscan Air (K250087). - The Sonosite iLOOK Ultrasound System uses advanced image processing (CViz) that is available on most cleared Sonosite products and is derived from the SonoHD3 image processing feature that is available on the primary predicate Sonosite iViz Ultrasound System (K180704). 7) Summary of Non-Clinical Tests: The Sonosite iLOOK Ultrasound System, a fully featured, general-purpose, software controlled diagnostic ultrasound system, utilizes an all-digital architecture and a wireless transducer (L15-5). It is a portable ultrasound device consisting of a display, adjustment handle, and a transducer holder with a built-in wireless charger. The Sonosite iLOOK Ultrasound System has been evaluated for electrical, thermal, mechanical, and EMC safety. Additionally, cleaning/disinfection, biocompatibility, and acoustic output have been evaluated for Sonosite iLOOK, and the device has been found to conform to applicable medical device safety standards. Assurance of quality was established by employing the following elements of product development: Design Phase Reviews, Risk Assessment, Requirements Development, System and Software Verification, Hardware Verification, Safety Compliance Verification, and Clinical Validation. All patient contact materials are biocompatible. {17} The Sonosite iLOOK Ultrasound System is designed to comply with the following FDA recognized standards. | Reference No. | Recognition No. | Title | | --- | --- | --- | | ISO 10993-1 | 2-258 | ISO 10993-1:2018, Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process | | IEC 60601-1 | 19-46 | AAMI / ANSI ES60601- 1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment -- Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) | | IEC 60601-1-2 | 19-36 | ANSI AAMI IEC 60601-1-2:2014 Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance -- Collateral Standard: Electromagnetic disturbances -- Requirements and tests | | IEC 60601-1-6 | 5-132 | IEC 60601-1-6 Edition 3.1 2013-10 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability | | IEC 60601-2-37 | 12-293 | IEC Edition 3.0 2024 Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment | | IEC 62304 | 13-79 | ANSI AAMI IEC 62304:2006/A1:2016 Medical device software - Software life cycle processes [Including Amendment 1 (2016) | | ISO 14971 | 5-125 | ANSI AAMI ISO 14971:2019 Medical devices - Application of risk management to medical devices | | IEC 62359 | 12-316 | IEC 62359 Edition 2.1 2017-09 CONSOLIDATED VERSION - Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields | {18} 8) Summary of Clinical Tests: The Sonosite iLOOK Ultrasound System and transducers, subject of this submission, a fully featured, general-purpose, software controlled diagnostic ultrasound system, utilizes an all-digital architecture and a wireless transducer (L15-5). It is a portable ultrasound device consisting of a display, adjustment handle, and a transducer holder with a built-in wireless charger. The Sonosite iLOOK Ultrasound System did not require clinical studies to support the determination of substantial equivalence. The Sonosite iLOOK has been successfully tested in a clinical environment and the device performed as expected. 9) Conclusion: Intended uses and other key features are consistent with traditional clinical practice and FDA guidance. The Sonosite iLOOK Ultrasound System and both predicate and reference devices conform to applicable electromedical device safety standards with compliance verified through independent evaluation. The Sonosite iLOOK Ultrasound System and both predicate and reference devices meet FDA requirements for Track 3 devices, share indications for use, have biosafety equivalence and are manufactured using the same ISO 13485 and 21 CFR 820 quality system. FUJIFILM SonoSite, Inc. believes that the Sonosite iLOOK Ultrasound System is substantially equivalent with regards to safety and effectiveness to the predicate devices. 15
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...