← Product Code [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN) · K100186

# MYSONO U5 DIAGNOSTIC ULTRASOUND SYSTEM (K100186)

_Medison Co., Ltd. · IYN · Mar 24, 2010 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K100186

## Device Facts

- **Applicant:** Medison Co., Ltd.
- **Product Code:** [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN.md)
- **Decision Date:** Mar 24, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric, 3rd-Party Reviewed

## Indications for Use

The MySono U5 system and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body. The clinical applications include: Abdomen, Obstetrics, Gynecology, Musculoskeletal, Small Parts, Vascular, Cardiac, Pediatric Cardiology, TCD and Urology applications.

## Device Story

MySono U5 is a hand-held, software-controlled diagnostic ultrasound system. It acquires ultrasound data via transducers and displays it in 2D, M, Color Doppler, Power Doppler, PW Spectral Doppler, and 3D modes. Operated by healthcare professionals in clinical settings, the system provides real-time acoustic output display (mechanical and thermal indices). It includes measurement and analysis packages for anatomical structures and fluid flow. Clinicians use the displayed images and measurements to assist in diagnostic decision-making. Benefits include non-invasive visualization of internal structures and blood flow for various clinical applications.

## Clinical Evidence

Bench testing only. The device complies with safety standards including UL 60601-1, CSA C22.2 No. 601.1, IEC 60601-2-37, NEMA UD 2-2004, NEMA UD 3-2004, IEC 61157, and ISO 10993 for biocompatibility.

## Technological Characteristics

Hand-held diagnostic ultrasound system. Modes: 2D, M, Color Doppler, Power Doppler, PW Spectral Doppler, 3D. Transducers: 3D2-6, C3-7, EV4-9, L5-12, P2-4. Features: Real-time acoustic output display (mechanical/thermal indices), Tissue Harmonic Imaging (THI). Software-controlled. Biocompatibility per ISO 10993. Safety compliance: IEC 60601-2-37, NEMA UD 2/3.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- SONOACE X8 Diagnostic Ultrasound System ([K063580](/device/K063580.md))
- SONOACE PICO Diagnostic Ultrasound System ([K061213](/device/K061213.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Kiwiso

MySono U5 Diagnostic Ultrasound System MAR 2 4 2010

# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.

1. Submitter's Information: 21 CFR 807.92(a)(1)

MEDISON CO., LTD. 1003, Daechi-dong, Gangnam-gu, . Seoul 135-280, Korea

Contact Person: Mr. Kyung-Am, Shim Regulatory Affairs Manager

| Telephone: | 82.2.2194.1381     |
|------------|--------------------|
| Facsimile: | 82.2.2194.1399     |
| Email:     | kashim@medison.com |

Data Prepared: July 15, 2009

### 2. Name of the device:

Common/Usual Name:

Diagnostic Ultrasound System and Accessories

Proprietary Name:

MySono U5 Diagnostic Ultrasound System

| Classification Names:                    | FR Number | Product Code |
|------------------------------------------|-----------|--------------|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550  | IYN          |
| Ultrasound Pulsed Echo Imaging System    | 892.1560  | IYO          |
| Diagnostic Ultrasound Transducer         | 892.1570  | ITX          |

### 3. Identification of the predicate or legally marketed device:

K063580, 12/14/2006, SONOACE X8 Diagnostic Ultrasound System K061213, 05/16/2006, SONOACE PICO Diagnostic Ultrasound System

# 4. Device Description:

The MySono U5 is a hand-held, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as 2D Mode, M Mode, Color Doppler Mode (C Mode), Power Doppler Mode (PD Mode), and PW Spectral Doppler Mode (D Mode), 3D imaging mode or as a combination of these modes.

The MySono U5 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed. The 

{1}------------------------------------------------

system also provides for the measurement of anatomical structures and for analysis packages that provide information used for clinical diagnostic purposes by competent health care professionals.

The MySono U5 has been designed to meet the following product safety standards:

- UL 60601-1, Safety requirements for Medical Equipment
- CSA C22.2 No. 601.1. Safety requirements for Medical Equipment
- IEC60601-2-37, Diagnostic Ultrasound Safety Standards
- EN/IEC60601-1, Safety requirements for Medical Equipment
- EN/IEC60601-1-2, EMC requirements for Medical Equipment y
- NEMA UD 2-2004 Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
- NEMA UD 3-2004 Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
- IEC 61157, Declaration of the acoustic output
- · ISO10993, Biocompatibility

#### 5. Intended Uses:

The MySono U5 system and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

The clinical applications include:

Abdomen, Obstetrics, Gynecology, Musculoskeletal, Small Parts, Vascular, Cardiac, Pediatric Cardiology, TCD and Urology applications.

### 6. Technological Characteristics:

The MySono U5 is substantially equivalent to the SONOACE X8 Diagnostic Ultrasound System, cleared via K063580, and the SONOACE PICO Diagnostic Ultrasound System, cleared via K061213. All systems transmit ultrasonic energy into patients, then perform post processing of received echoes to generate on-screen display of anatomic structures and fluid flow within the body. All system allow for specialized measurements of structures and flow, and calculations.

END of 510(K) Summary

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized symbol that resembles a human figure embracing another person.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

MAR 2 4 2010

Medison Co., Ltd. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

Re: K100186

Trade/Device Name: MySono U5 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed Doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: March 8, 2010 Received: March 9, 2010

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the MySono U5 Diagnostic Ultrasound System, as described in your premarket notification:

## Transducer Model Number

| 3D2-6 |
|-------|
| C3-7  |
| EV4-9 |
| L5-12 |
| P2-4  |

.

{3}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

If you have any questions regarding the content of this letter, please contact Shahram Vaezy at (301) 796-6242.

Sincerely vours.

Sincerely, yours,

Donald

Donald St. Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure(s)

{4}------------------------------------------------

. '•

MySono U5 Diagnostic Ultrasound System

# Indications for Use

510(k) Number (if known):

Device Name: MySono U5 Diagnostic Ultrasound System

Indications for Use:

The MySono US system and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body. The clinical applications include: Abdomen, Obsteries, Gynecology, Musculoskeletal, Small Parts, Vascular, Cardiac, Pediatric Cardiology, U.C.D. and Urology applications.

ﮯ Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

(Division Sign-Off)

Page 1 of 7

Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

510K K100186

{5}------------------------------------------------

MySono U5 Diagnostic Ultrasound System

System:

MySono U5 Diagnostic Ultrasound System

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                              |   | Mode of Operation (*includes simultaneous B-mode) |     |     |                   |                      |                  |  |
|---------------------------|------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M                                                 | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |  |
| Ophthalmic                | Ophthalmic                   |   |                                                   |     |     |                   |                      |                  |  |
|                           | Fetal (See Note 3)           | N | N                                                 | N   | N   | N                 | Note 1               | Notes 2, 7, 8    |  |
|                           | Abdominal                    | N | N                                                 | N   | N   | N                 | Note 1               | Notes 2, 7, 8    |  |
|                           | Intra-operative (See Note 6) |   |                                                   |     |     |                   |                      |                  |  |
|                           | Intra-operative (Neuro.)     |   |                                                   |     |     |                   |                      |                  |  |
| Fetal Imaging<br>& Other  | Laparoscopic                 |   |                                                   |     |     |                   |                      |                  |  |
|                           | Pediatric                    | N | N                                                 | N   | N   | N                 | Note 1               | Note 2, 5, 6, 7  |  |
|                           | Small Organ (See Note 5)     | N | N                                                 | N   | N   | N                 | Note 1               | Note 2, 5, 6, 7  |  |
|                           | Neonatal Cephalic            |   |                                                   |     |     |                   |                      |                  |  |
|                           | Adult Cephalic               | N | N                                                 | N   | N   | N                 | Note 1               | Note 7           |  |
|                           | Trans-rectal                 | N | N                                                 | N   | N   | N                 | Note 1               | Note 2, 8        |  |
|                           | Trans-vaginal                | N | N                                                 | N   | N   | N                 | Note 1               | Note 2, 8        |  |
|                           | Trans-urethral               |   |                                                   |     |     |                   |                      |                  |  |
|                           | Trans-esoph. (non-Cardiac)   |   |                                                   |     |     |                   |                      |                  |  |
|                           | Musculo-skel. (Convent.)     | N | N                                                 | N   | N   | N                 | Note 1               | Note 2, 5, 6, 7  |  |
|                           | Musculo-skel. (Superfic.)    | N | N                                                 | N   | N   | N                 | Note 1               | Note 2, 5, 6, 7  |  |
|                           | Intra-luminal                |   |                                                   |     |     |                   |                      |                  |  |
|                           | Other (spec.)                |   |                                                   |     |     |                   |                      |                  |  |
|                           | Cardiac Adult                | N | N                                                 | N   | N   | N                 | Note 1               | Note 7           |  |
| Cardiac                   | Cardiac Pediatric            | N | N                                                 | N   | N   | N                 | Note 1               | Note 7           |  |
|                           | Trans-esophageal (Cardiac)   |   |                                                   |     |     |                   |                      |                  |  |
|                           | Other (spec.)                |   |                                                   |     |     |                   |                      |                  |  |
| Peripheral<br>Vessel      | Peripheral vessel            | N | N                                                 | N   | N   | N                 | Note 1               | Note 2, 5, 6, 7  |  |
|                           | Other (spec.)                |   |                                                   |     |     |                   |                      |                  |  |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional.Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PWD, B/Color Doppier, B/Color Doppler/PWD, B/Color Doppler/M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Hannonic Imaging (THI)

Note 8: 3D imaging

Prescription Use Only (Per 21 CFR801.109)

)

Division Sign-Off Division of Radiological Devices tion and Safety Office of In

K100186
510K_

{6}------------------------------------------------

#### MySono US Diagnostic Ultrasound System

System:

# MySono U5 Diagnostic Ultrasound System

Transducer: 3D2-6

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use:

| Clinical Application      |                              | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                  |
|---------------------------|------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                   |                                                   |   |     |     |                   |                      |                  |
|                           | Fetal (See Note 3)           | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 7, 8    |
|                           | Abdominal                    | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 7, 8    |
|                           | Intra-operative (See Note 6) |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-operative (Neuro.)     |                                                   |   |     |     |                   |                      |                  |
| Fetal Imaging<br>& Other  | Laparoscopic                 |                                                   |   |     |     |                   |                      |                  |
|                           | Pediatric                    | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 7, 8    |
|                           | Small Organ (See Note 5)     |                                                   |   |     |     |                   |                      |                  |
|                           | Neonatal Cephalic            |                                                   |   |     |     |                   |                      |                  |
|                           | Adult Cephalic               |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-rectal                 |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-vaginal                |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-urethral               |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esoph. (non-Cardiac)   |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Convent.)     |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Superfic.)    |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-luminal                |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |
| Cardiac                   | Cardiac Adult                |                                                   |   |     |     |                   |                      |                  |
|                           | Cardiac Pediatric            |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esophageal (Cardiac)   |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |
| Peripheral<br>Vessel      | Peripheral vessel            |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |

N= new indication; P= previously cleared by FDA K063580; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PWD, B/Color Doppler, B/Color Doppler/PWD, B/Color Doppler/M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

(Division Sign-O Division of Radiological De agnostic Device Evaluation and Safety Office of In

610K K106186

{7}------------------------------------------------

System:

# MySono U5 Diagnostic Ultrasound System

C3-7 Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                              | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                  |
|---------------------------|------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                   |                                                   |   |     |     |                   |                      |                  |
|                           | Fetal (See Note 3)           | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 7, 8    |
|                           | Abdominal                    | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 7, 8    |
|                           | Intra-operative (See Note 6) |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-operative (Neuro.)     |                                                   |   |     |     |                   |                      |                  |
| Fetal Imaging<br>& Other  | Laparoscopic                 |                                                   |   |     |     |                   |                      |                  |
|                           | Pediatric                    | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 7, 8    |
|                           | Small Organ (See Note 5)     |                                                   |   |     |     |                   |                      |                  |
|                           | Neonatal Cephalic            |                                                   |   |     |     |                   |                      |                  |
|                           | Adult Cephalic               |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-rectal                 |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-vaginal                |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-urethral               |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esoph. (non-Cardiac)   |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Convent.)     |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Superfic.)    |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-luminal                |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |
|                           | Cardiac Adult                |                                                   |   |     |     |                   |                      |                  |
| Cardiac                   | Cardiac Pediatric            |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esophageal (Cardiac)   |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |
| Peripheral                | Peripheral vessel            |                                                   |   |     |     |                   |                      |                  |
| Vessel                    | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |

N= new indication; P= previously cleared by FDA K063580; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PWD, B/Color Doppler, B/Color Doppler/PWD, B/Color Doppler/M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

D.A.R

(Division Sign-Off) Division of Radiological Devices Office of Ir Vitro Diagnostic Device Evaluation and Safety

510K K106186

{8}------------------------------------------------

System: MySono U5 Diagnostic Ultrasound System

Transducer: EV4-9

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use:

| Clinical Application      |                              | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                  |  |
|---------------------------|------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |  |
| Ophthalmic                | Ophthalmic                   |                                                   |   |     |     |                   |                      |                  |  |
|                           | Fetal (See Note 3)           |                                                   |   |     |     |                   |                      |                  |  |
|                           | Abdominal                    |                                                   |   |     |     |                   |                      |                  |  |
|                           | Intra-operative (See Note 6) |                                                   |   |     |     |                   |                      |                  |  |
|                           | Intra-operative (Neuro.)     |                                                   |   |     |     |                   |                      |                  |  |
| Fetal Imaging<br>& Other  | Laparoscopic                 |                                                   |   |     |     |                   |                      |                  |  |
|                           | Pediatric                    |                                                   |   |     |     |                   |                      |                  |  |
|                           | Small Organ (See Note 5)     |                                                   |   |     |     |                   |                      |                  |  |
|                           | Neonatal Cephalic            |                                                   |   |     |     |                   |                      |                  |  |
|                           | Adult Cephalic               |                                                   |   |     |     |                   |                      |                  |  |
|                           | Trans-rectal                 | N                                                 | N | N   |     | N                 | Note 1               | Note 2, 8        |  |
|                           | Trans-vaginal                | N                                                 | N | N   |     | N                 | Note 1               | Note 2, 8        |  |
|                           | Trans-urethral               |                                                   |   |     |     |                   |                      |                  |  |
|                           | Trans-esoph. (non-Cardiac)   |                                                   |   |     |     |                   |                      |                  |  |
|                           | Musculo-skel. (Convent.)     |                                                   |   |     |     |                   |                      |                  |  |
|                           | Musculo-skel. (Superfic.)    |                                                   |   |     |     |                   |                      |                  |  |
|                           | Intra-luminal                |                                                   |   |     |     |                   |                      |                  |  |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |  |
|                           | Cardiac Adult                |                                                   |   |     |     |                   |                      |                  |  |
| Cardiac                   | Cardiac Pediatric            |                                                   |   |     |     |                   |                      |                  |  |
|                           | Trans-esophageal (Cardiac)   |                                                   |   |     |     |                   |                      |                  |  |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |  |
| Peripheral                | Peripheral vessel            |                                                   |   |     |     |                   |                      |                  |  |
| Vessel                    | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |  |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PWD, B/Color Doppler, B/Color Doppler/PWD, B/Color Doppler/M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

(Division Sign-Off) vision of Radiological Devices agnostic Device Evaluation and Safety

K10018(
510K

{9}------------------------------------------------

MySono U5 Diagnostic Ultrasound System System:

Transducer: L5-12

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use:

| Clinical Application      |                              | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                  |
|---------------------------|------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                   |                                                   |   |     |     |                   |                      |                  |
| Fetal Imaging<br>& Other  | Fetal (See Note 3)           |                                                   |   |     |     |                   |                      |                  |
|                           | Abdominal                    |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-operative (See Note 6) |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-operative (Neuro.)     |                                                   |   |     |     |                   |                      |                  |
|                           | Laparoscopic                 |                                                   |   |     |     |                   |                      |                  |
|                           | Pediatric                    | N                                                 | N | N   |     | N                 | Note 1               | Note 2, 5, 6, 7  |
|                           | Small Organ (See Note 5)     | N                                                 | N | N   |     | N                 | Note 1               | Note 2, 5, 6, 7  |
|                           | Neonatal Cephalic            |                                                   |   |     |     |                   |                      |                  |
|                           | Adult Cephalic               |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-rectal                 |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-vaginal                |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-urethral               |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esoph. (non-Cardiac)   |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Convent.)     | N                                                 | N | N   |     | N                 | Note 1               | Note 2, 5, 6, 7  |
|                           | Musculo-skel. (Superfic.)    | N                                                 | N | N   |     | N                 | Note 1               | Note 2, 5, 6, 7  |
|                           | Intra-luminal                |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |
| Cardiac                   | Cardiac Adult                |                                                   |   |     |     |                   |                      |                  |
|                           | Cardiac Pediatric            |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esophageal (Cardiac)   |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |
| Peripheral<br>Vessel      | Peripheral vessel            | N                                                 | N | N   |     | N                 | Note 1               | Note 2, 5, 6, 7  |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note I: B/M, B/PWD, B/Color Doppler, B/Color Doppler/PWD, B/Color Doppler/M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Prescription Use Only (Per 21 CFR801.109)

---

(Division Sign-Off)

Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

510K K100186

{10}------------------------------------------------

#### MySono U5 Diagnostic Uitrasound System

System: MySono U5 Diagnostic Ultrasound System

P2-4 Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                              | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                  |
|---------------------------|------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                   |                                                   |   |     |     |                   |                      |                  |
|                           | Fetal (See Note 3)           |                                                   |   |     |     |                   |                      |                  |
| Fetal Imaging<br>& Other  | Abdominal                    | P                                                 | P | P   |     | P                 | Note 1               | Note 7           |
|                           | Intra-operative (See Note 6) |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-operative (Neuro.)     |                                                   |   |     |     |                   |                      |                  |
|                           | Laparoscopic                 |                                                   |   |     |     |                   |                      |                  |
|                           | Pediatric                    |                                                   |   |     |     |                   |                      |                  |
|                           | Small Organ (See Note 5)     |                                                   |   |     |     |                   |                      |                  |
|                           | Neonatal Cephalic            |                                                   |   |     |     |                   |                      |                  |
|                           | Adult Cephalic               | P                                                 | P | P   |     | P                 | Note 1               | Note 7           |
|                           | Trans-rectal                 |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-vaginal                |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-urethral               |                                                   |   |     |     |                   |                      |                  |
|                           | Trans-esoph. (non-Cardiac)   |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Convent.)     |                                                   |   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Superfic.)    |                                                   |   |     |     |                   |                      |                  |
|                           | Intra-luminal                |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |
|                           | Cardiac Adult                | P                                                 | P | P   |     | P                 | Note 1               | Note 7           |
| Cardiac                   | Cardiac Pediatric            | P                                                 | P | P   |     | P                 | Note 1               | Note 7           |
|                           | Trans-esophageal (Cardiac)   |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |
| Peripheral<br>Vessel      | Peripheral vessel            |                                                   |   |     |     |                   |                      |                  |
|                           | Other (spec.)                |                                                   |   |     |     |                   |                      |                  |

N= new indication; P= previously cleared by FDA K063580; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B/M, B/PWD, B/Color Doppler, B/Color Doppler/PWD, B/Color Doppier/M

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

510K K100186

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K100186](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K100186)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
