K211270 · CIVCO Medical Instruments Co., Inc. · ITX · Sep 13, 2021 · Radiology
Device Facts
Record ID
K211270
Device Name
CIV-Clear cover
Applicant
CIVCO Medical Instruments Co., Inc.
Product Code
ITX · Radiology
Decision Date
Sep 13, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1570
Device Class
Class 2
Indications for Use
The cover is intended as a microbial barrier between the patient and medical imaging equipment.
Device Story
CIV-Clear cover acts as a microbial barrier between patients and ultrasound transducers. It is used in clinical environments during diagnostic imaging and minimally invasive procedures (e.g., abdominal, vascular, surgical, transesophageal, transrectal, transvaginal). The device consists of an Ethyl Methyl Acrylate (EMA) and Polyethylene blend cover. During use, the clinician applies ultrasound gel inside the cover or on the transducer face to facilitate acoustic coupling. The cover prevents the transmission of microbes 27 nanometers or larger. It is provided in sterile and non-sterile configurations. The device does not alter the ultrasound signal but provides a protective barrier, ensuring patient safety during procedures. It is intended for use by healthcare professionals.
Clinical Evidence
No clinical data. Substantial equivalence was demonstrated through bench testing, including biocompatibility (ISO 10993-5, -10, -11), viral penetration (ASTM F1671-13), water leak testing, acoustic impedance measurement, simulated use testing, and shelf-life/packaging validation (ISO 11607-2, ASTM F1980-16).
Technological Characteristics
Material: Ethyl Methyl Acrylate (EMA) and Polyethylene blend. Microbial barrier performance meets ASTM F1671-13. Sterilization: Ethylene Oxide (ISO 11135). Biocompatibility: ISO 10993-1. Shelf-life: 3 years (ASTM F1980-16). Connectivity: None (passive barrier).
Indications for Use
Indicated for adult patients of all body sizes in sterile and non-sterile fields for abdominal, small parts, surgical, transesophageal, transrectal, and transvaginal diagnostic imaging and minimally invasive puncture procedures; regional anesthesia minimally invasive puncture procedures; and vascular access vessel identification and catheter placement.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
K153212 — Sheathes Ultrasound Probe Covers · Sheathing Technologies, Inc. · Mar 4, 2016
K970513 — GENERAL PURPOSE TRANDUCER COVER · CIVCO Medical Instruments Co., Inc. · Jun 20, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it. The full name reads "U.S. Food & Drug Administration".
CIVCO Medical Instruments Co., Inc. % Jim Leong Regulatory Affairs Manager 102 First Street South KALONA IA 52247
September 13 , 2021
Re: K211270
Trade/Device Name: CIV-Clear cover Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: Class II Product Code: ITX Dated: July 20, 2021 Received: July 28, 2021
Dear Jim Leong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211270
Device Name CIV-Clear Cover
Indications for Use (Describe)
The cover is intended as a microbial barrier between the patient and medical imaging equipment.
The transducer covers are used for all body sizes in sterile and non-sterile fields and for the following applications.
- · Abdominal Diagnostic imaging and minimally invasive puncture procedures.
- · Small Parts Diagnostic imaging and minimally invasive puncture procedures.
- · Regional Anesthesia Minimally invasive puncture procedures.
- · Vascular Access Vessel identification and catheter placement.
- · Surgical Diagnostic imaging and puncture procedures.
- · Transesophageal Diagnostic imaging and monitoring of heart chamber, valves and vessels.
- · Transrectal Diagnostic imaging and minimally invasive puncture procedures.
- · Transvaginal Diagnostic imaging and minimally invasive puncture procedures
When conducting an ultrasound procedure, place an appropriate amount of gel inside cover and/or on transducer face.
Type of Use (*Select one or both, as applicable*)
| <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|--------------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being prepared in accordance with the requirements of 21 CFR 807.92.
The assigned 510(k) numberis: K211270
- 1. Submitter's Identifications:
| Establishment: | CIVCO Medical Instruments Co., Inc |
|--------------------------------------------|--------------------------------------------|
| Address: | 102 First Street South<br>Kalona, IA 52247 |
| Registration Number: | 1937223 |
| Operations Manufacturer<br>Owner/Operator: | CIVCO Medical Instruments Co., Inc. |
| Owner/Operator Number: | 1937223 |
| Contact Person: | Jim Leong |
| Phone: | 319-248-6502 |
| e-mail: | James.Leong@civco.com |
- 2. Date 510(k) Summary Prepared: April 21, 2021
- 3. Name of the Subject Device and Classification Information:
| Trade/Device Name | CIV-Clear cover |
|---------------------|----------------------------------|
| Regulation Number | 21 CFR 892.1570 |
| Classification Name | Diagnostic ultrasonic transducer |
| Regulatory Class | Class II |
| Product Code | ITX |
### 4. Information for the Predicate Device:
| Tradename/Device Name | Envision™ Ultrasound Cover and Pad |
|-----------------------|-------------------------------------|
| Manufacturer | CIVCO Medical Instruments Co., Inc. |
| 510(k) Number | K190802 |
| Regulation Number | 892.1570 |
| Classification Name | Diagnostic ultrasonic transducer |
| Regulatory Class | II |
| Product Code | ITX |
{4}------------------------------------------------
#### 5. Information for Reference Device:
| Tradename/Device Name | EcoVue® Sterile and Non-Sterile Ultrasound Gels |
|-----------------------|-------------------------------------------------|
| Manufacturer | HR Pharmaceuticals, Inc. |
| 510(k) Number | K181363 |
| Regulation Number | 892.1570 |
| Classification Name | Diagnostic ultrasonic transducer |
| Regulatory Class | II |
| Product Code | MUI |
### 6. Device Description:
The CIV-Clear cover is a microbial barrier for use during ultrasound procedures. The cover is offered in sterile and non-sterile configurations in a variety of sizes to function with different equipment. The material has been tested and proven to be a viral barrier for microbes 27 nanometers and larger in size.
### 7. Intended Use / Indications for Use:
The cover is intended as a microbial barrier between the patient and medical imaging equipment.
The transducer covers are used for adult of all body sizes in sterile and non-sterile fields and for the following applications.
- Abdominal Diagnostic imaging and minimally invasive puncture procedures. .
- Small Parts Diagnostic imaging and minimally invasive puncture procedures. ●
- Regional Anesthesia Minimally invasive puncture procedures. ●
- Vascular Access Vessel identification and catheter placement. ●
- Surgical Diagnostic imaging and puncture procedures. ●
- Transesophageal Diagnostic imaging and monitoring of heart chamber, valves and ● vessels.
- Transrectal Diagnostic imaging and minimally invasive puncture procedures. ●
- . Transvaginal - Diagnostic imaging and minimally invasive puncture procedures
When conducting an ultrasound procedure, place an appropriate amount of gel inside cover and/or on transducer face.
#### 8. Clinical Use Envirionment:
The covers are to be used in the healthcare environment where equipment is being utilized.
{5}------------------------------------------------
## 9. Comparison to Legally Marketed Device
| Item | Subject Device<br>CIVCO CIV-Clear Cover<br>K21XXXX | Predicate Device<br>CIVCO Envision Cover and<br>Pad<br>K190802 | |
|-------------------------|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| Material | N/A | Pad: Polycarbonate, silicone<br>elastomer, polyether<br>polyurethane, coating<br>formulation | |
| | Cover: Ethyl Methyl Acrylate<br>(EMA) and Polyethylene blend | Cover: Polyurethane | |
| Microbial Barrier | Meets requirements of ASTM<br>F1671-13 for prevention of<br>blood-borne pathogens | Meets requirements of ASTM<br>F1671-13 for prevention of<br>blood-borne pathogens | |
| Acoustic<br>Performance | Acoustic<br>Impedance | $1.54 ± 0.13 x 10^5 g/(cm^2 sec)$ | Primary coating: $1.69 ± 0.18 x 10^5 g/(cm^2 sec)$<br>Secondary coating: $1.68 ± 0.29 x 10^5 g/(cm^2 sec)$ |
| | Acoustic<br>Velocity | 1630 m/s | Primary Coating: 1594 m/s<br>Secondary Coating: 1631 m/s |
| | Density<br>of<br>Coating | N/A - no coating | Primary: 1.06 g/mL)<br>Secondary: 1.03 g/mL |
| | pH of<br>coating | N/A - no coating | 7.11 (both primary and<br>secondary coatings) |
| Acoustic Coupling | | Uses gel to facilitate coupling | Hydrated coating used to<br>facilitate coupling |
| Sterilization | | Ethylene Oxide | Ethylene Oxide |
| Shelf-life | | 3 years | 3 years |
| Accessory | | Ultrasound gel packet | Saline wipe to facilitate<br>coupling |
{6}------------------------------------------------
## 10. Comparison of Indications to the Legally Marketed Device:
The proposed CIV-Clear devices have the same intended use, of being a microbial barrier, as the legally marketed device. Any differences in the indications for use statement would not affect the safety or effectiveness of the device because the safety and efficacy related to the indicated procedures is dictated by the use of the underlying ultrasound equipment, and the ultrasound cover is secondary to the ultrasound use. Any questions related to safety and effectiveness of the CIV-Clear cover have been addressed using the same testing performed by the legally marketed device, including biocompatibility, viral penetration, and material strength and elasticity studies.
### 11. Summary of Non-Clinical Tests Performed:
- . Biocompatibility:
The CIV-Clear devices met ISO 10993-1 biocompatibility requirements for limited contact duration for surface devices of breached or compromised surface and external communicating indirect tissue contact:
- Cytotoxicity ISO 10993-5 -
- Sensitization ISO 10993-10 -
- Irritation ISO 10993-10 -
- Acute Systemic Toxicity ISO 10993-11 -
- Material Mediated Pyrogen ISO 10993-11 -
- Viral Penetration: ●
The CIV-Clear cover's ability to withstand viral penetration was performed to evaluate the barrier performance of the material in accordance with ASTM F1671-13. Testing has demonstrated the ability to block microbes of size 27 nanometers and larger.
- Water leak testing: ●
Water leak testing was performed to demonstrate material strength and elasticity of the CIV-Clear device after sterilization, shipping/conditioning, and aging.
- Acoustic Impedance:
The measurement of acoustic impedance was performed to compare the acoustic characteristics of the CIV-Clear cover to the predicate cover with coating.
- Simulated Use Testing
Simulated use testing was performed to ensure the design of the CIV-Clear cover conforms to the user needs and intended use.
- . Packaging and Shelf-life:
CIV-Clear packing uses the same packaging as the predicate device which has been validated per ISO 11607-2 and ISTA 3A parameters to ensure the packaging
{7}------------------------------------------------
maintains its integrity over the course of the device's shelf-life. Accelerated aging in support of a three-year shelf life of product was completed in accordance with ASTM F1980-16.
- Ethylene Oxide Sterilization Validation and Residual Testing: ● The CIV-Clear devices are processed using sterilization cycle validated per ISO 11135 to ensure that the devise meet the required sterility assurance level, so the devices remain adequately free from viable microorganisms following sterilization.
The sterile CIV-Clear covers qualified for 2x Ethylene Oxide sterilization and tested to ensure that residual levels do not exceed the limits per ISO 10993-7 for contact.
#### 12. Clinical Test Performed:
Clinical tests were not required to demonstrate substantial equivalence.
#### 13. Conclusions:
The CIV-Clear device has the same intended use and its technological characteristics do not raise any different questions of safety or effectiveness, as compared to the legally marketed device. Therefore, the CIV-Clear covers are substantially equivalent to the legally marketed Envision device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.