Game Ready system (550500-04); Game Ready Control Unit (550550-04); Game Ready Full Leg Boot Wrap - Medium (590923-04); Game Ready Full Leg Boot Wrap - Large (590922-04); Game Ready C-T Spine Wrap (590502-04); Game Ready Half Leg Boot (590906-04); Game Ready Back Wrap (590500-04); Game Ready Hand/Wrist Wrap (590811-04); Game Ready Flexed Elbow Wrap (590903-04); Game Ready Straight Elbow Wrap (590200-04); Game Ready Shoulder Wrap, Medium, Right (590422-04); Game Ready Shoulder Wrap, La
K243572 · Avanos Medical, Inc. · IRP · Jan 31, 2025 · Physical Medicine
Device Facts
Record ID
K243572
Device Name
Game Ready system (550500-04); Game Ready Control Unit (550550-04); Game Ready Full Leg Boot Wrap - Medium (590923-04); Game Ready Full Leg Boot Wrap - Large (590922-04); Game Ready C-T Spine Wrap (590502-04); Game Ready Half Leg Boot (590906-04); Game Ready Back Wrap (590500-04); Game Ready Hand/Wrist Wrap (590811-04); Game Ready Flexed Elbow Wrap (590903-04); Game Ready Straight Elbow Wrap (590200-04); Game Ready Shoulder Wrap, Medium, Right (590422-04); Game Ready Shoulder Wrap, La
Applicant
Avanos Medical, Inc.
Product Code
IRP · Physical Medicine
Decision Date
Jan 31, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Game Ready system is indicated for the temporary relief of minor muscle aches and pains. The device is indicated for temporary increase in circulation of the treated areas in people who are in good health, and simulate kneading and stroking of tissues using an inflatable wrap. The Game Ready system is indicated for localized therapy in situations where cold temperature therapy is necessary or desirable.
Device Story
Game Ready system provides localized cold therapy and intermittent pneumatic compression; consists of a control unit and various anatomical wraps (e.g., leg, shoulder, back, knee). Control unit regulates temperature and pressure; inflates/deflates wraps to simulate kneading and stroking of tissues. Used in clinical or home settings; operated by patients or healthcare providers. Output is physical compression and cooling of the target area. Benefits include pain relief and improved circulation in treated tissues.
Technological Characteristics
Powered inflatable tube massager; includes control unit and anatomical wraps; provides cold therapy and pneumatic compression.
Indications for Use
Indicated for temporary relief of minor muscle aches and pains, temporary increase in local circulation, and localized cold temperature therapy in healthy individuals.
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
Related Devices
K072620 — GAME READY CLASSIC SYSTEM, MODEL 550550 · Coolsystems, Inc. · Oct 31, 2007
K071050 — GAME READY PROFESSIONAL THERAPY SYSTEM, MODEL 550100, GAME READY PRE-PROGRAMMED PROFESSIONAL THERAPY SYSTEM · Coolsystems, Inc. · Jun 29, 2007
K230524 — Cold Compression Wrap Pro · Chengdu Cryo-Push Medical Technology Co.,Ltd · Jan 19, 2024
K192114 — Game Ready GRPro 2.1 System · Cool Systems, Inc. (Dba Game Ready) · Oct 29, 2019
K193079 — ICE COMPRESSION FIRST, DUO, & MOOVE Systems · Mks Paris · Jan 11, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a recognizable symbol of the agency's role in regulating and overseeing food and drug products in the United States.
January 31, 2025
Avanos Medical, Inc. Dana Goussev Principal Regulatory Affairs Specialist 5405 Windward Parkway Alpharetta, Georgia 30004
#### Re: K243572
Trade/Device Name: Game Ready system (550500-04); Game Ready Control Unit (550550-04); Game Ready Full Leg Boot Wrap - Medium (590923-04); Game Ready Full Leg Boot Wrap - Large (590922-04); Game Ready C-T Spine Wrap (590502-04); Game Ready Half Leg Boot (590906-04); Game Ready Back Wrap (590500-04); Game Ready Hand/Wrist Wrap (590811-04); Game Ready Flexed Elbow Wrap (590903-04); Game Ready Straight Elbow Wrap (590200-04); Game Ready Shoulder Wrap, Medium, Right (590422-04); Game Ready Shoulder Wrap, Large, Right (590432-04); Game Ready Shoulder Wrap, Medium, Left (590424-04); Game Ready Shoulder Wrap, Large, Left (590434-04); Game Ready Hip/Groin Wrap, Right (590602-04); Game Ready Hip/Groin Wrap, Left (590604-04); Game Ready Straight Knee Wrap (590100-04); Game Articulated Knee Wrap (590160-04); Game Ready Ankle Wrap, Large (590330-04); Game Ready Ankle Wrap, Extra Large (590340-04) Regulation Number: 21 CFR 890.5650 Regulation Name: Powered Inflatable Tube Massager Regulatory Class: Class II
Product Code: IRP, ILO Dated: January 30, 2025 Received: January 30, 2025
Dear Dana Goussev:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration,
{1}------------------------------------------------
listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part
{2}------------------------------------------------
803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Tushar Bansal -S
for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
## K243572
Device Name
Game Ready system (550500-04); Game Ready Control Unit (550550-04): Game Ready Full Leg Boot Wrap - Medium (590923-04); Game Ready Full Leg Boot Wrap - Large (590922-04); Game Ready C-T Spine Wrap (590502-04); Game Ready Half Leg Boot (590906-04): Game Ready Back Wrap (590500-04): Game Ready Hand/Wrist Wrap (590811-04); Game Ready Flexed Elbow Wrap (590903-04); Game Ready Straight Elbow Wrap (590200-04); Game Ready Shoulder Wrap, Medium, Right (590422-04); Game Ready Shoulder Wrap, Large, Right (590432-04); Game Ready Shoulder Wrap, Medium, Left (590424-04); Game Ready Shoulder Wrap, Large, Left (590434-04); Game Ready Hip/Groin Wrap, Right (590602-04): Game Ready Hip/Groin Wrap, Left (590604-04): Game Ready Straight Knee Wrap (590100-04); Game Articulated Knee Wrap (590160-04); Game Ready Ankle Wrap, Large (590330-04); Game Ready Ankle Wrap, Extra Large (590340-04) Indications for Use (Describe)
The Game Ready system is indicated for the temporary relief of minor muscle aches and pains. The device is indicated for temporary increase in circulation of the treated areas in people who are in good health, and simulate kneading and stroking of tissues using an inflatable wrap. The Game Ready system is indicated for localized therapy in situations where cold temperature therapy is necessary or desirable.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.