← Product Code [LXJ](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/LXJ) · K131660

# SMART (SMART-D, SMART-DX) (K131660)

_Bts S.P.A. · LXJ · Apr 17, 2014 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/LXJ/K131660

## Device Facts

- **Applicant:** Bts S.P.A.
- **Product Code:** [LXJ](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/LXJ.md)
- **Decision Date:** Apr 17, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5360
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine

## Indications for Use

SMART-D is a system for motion analysis, intended for the recording and analysis of human movement patterns in the fields of rehabilitation, sports medicine, ergonomics.

## Device Story

SMART-D is an optoelectronic motion capture system using passive reflective markers, infrared cameras, and strobes. It reconstructs 3D trajectories of markers placed on a subject's body. Cameras are positioned around the performance volume and calibrated to acquire movement data. The system consists of a workstation, cameras, and software (SMARTcapture, SMARTtracker, SMARTanalyzer, SMARTclinic). It is used in clinical or research settings (rehabilitation, sports medicine, ergonomics) by trained operators. The software provides real-time signal visualization, 3D kinematic reconstruction, analysis protocol construction, and multimedia reporting. Healthcare providers use these outputs to assess movement patterns, aiding in clinical decision-making for rehabilitation or performance training. The system benefits patients by providing objective, quantitative data on limb and body motion.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance was verified through subtests (SMARTcapture, SMARTtracker, SMARTanalyzer, SMARTclinic) comparing functionality against predicate devices. Electrical safety and EMC testing confirmed compliance with AAMI/ANSI ES60601-1:2005 and IEC 60601-1-2.

## Technological Characteristics

Optoelectronic motion capture system. Components: workstation, digital cameras (up to 16), infrared strobes (880 nm). Connectivity: workstation-based. Power: 110V, 60Hz. Acquisition frequency: 250-500 Hz. Resolution: 800x600 pixels. Software: SMARTcapture, SMARTtracker, SMARTanalyzer, SMARTclinic. Standards: AAMI/ANSI ES60601-1:2005, IEC 60601-1-2.

## Regulatory Identification

Measuring exercise equipment consist of manual devices intended for medical purposes, such as to redevelop muscles or restore motion to joints or for use as an adjunct treatment for obesity. These devices also include instrumentation, such as the pulse rate monitor, that provide information used for physical evaluation and physical planning purposes., Examples include a therapeutic exercise bicycle with measuring instrumentation, a manually propelled treadmill with measuring instrumentation, and a rowing machine with measuring instrumentation.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a measuring exerciser or an interactive rehabilitation exercise device for prescription use only, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

## Predicate Devices

- PEAK MOTUS ([K030714](/device/K030714.md))
- CODA CX1 MOTION ANALYSIS SYSTEM ([K033514](/device/K033514.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### 510(k) NOTIFICATION

APR 1 7 2014

# 1 Summary of safety and effectiveness

| Company Name:    | BTS SPA                                  |
|------------------|------------------------------------------|
| Company Address: | Via della Croce Rossa 11, PD Italy 35129 |
| Company Phone:   | +39 049 981 5520                         |
| Company Fax:     | +39 049 792 9260                         |

| Contact Person: | Mr. Enrico Bisson                         |
|-----------------|-------------------------------------------|
| Title:          | Quality and Regulatory Affairs Consultant |
| Email:          | enrico.bisson@isoplan.org                 |
| Company Phone:  | +39 049 981 5520                          |
| Company Fax:    | +39 049 792 9260                          |
|                 |                                           |

April 17th 2014

Date Summary Prepared:

#### Device Identification

| A. | Generic Device Name:    | System, optical position/movement recording |
|----|-------------------------|---------------------------------------------|
| B. | Trade/Proprietary Name: | SMART-D                                     |
| C. | Classification:         | Unclassified                                |
| D. | Product Code:           | LXJ                                         |

#### Device description

SMART-D is a motion capture and analysis system. It is based on optoelectronic technology and uses passive reflective markers, infrared cameras and strobes.

SMART-D executes the three-dimensional reconstruction of trajectories of a certain number of small reflective markers on the body to be analyzed.

Using a set of cameras placed all around the volume where the action is performed, the system acquires the movement of the subject; cameras position is computed after a calibration step.

#### Intended Use

SMART-D is a system for motion analysis, intended for the recording and analysis of human movement patterns in the fields of rehabilitation, sports medicine, ergonomics.

{1}------------------------------------------------

#### 510(k) NOTIFICATION

#### BTS Spa

#### Substantial equivalents

The SMART-D device is of comparable type and is substantially equivalent to the following predicate devices:

| Predicate Device                   | 510(K) Holder                         | 510(k) No. | Date Cleared |
|------------------------------------|---------------------------------------|------------|--------------|
| PEAK MOTUS                         | PEAK PERFORMANCE<br>TECHNOLOGIES, INC | K030714    | 16 May 2003  |
| CODA CX1 MOTION<br>ANALYSIS SYSTEM | CHARNWOOD DYNAMICS<br>LTD.            | K033514    | 6 Jan 2004   |

In further support of a substantial equivalence determination, Section 12 provides a comparison chart of SMART-D and the predicate devices.

#### Technological characteristics

A comparison of the technological characteristics of SMART-D and the predicate devices has been performed. The results of this comparison demonstrate that the system for motion analysis SMART-D is equivalent to the marketed predicate devices.

The two tables below show the comparison of technical characteristics of SMART-D system compared to the characteristics of predicate devices:

- Table 1 contains the summary of the technical characteristics which are the same for both devices; ー
- Table 2 contains the summary of the different technical characteristics and provides an explanation of how they are comparable to the predicate device.

| Table 1 - Same Technical Characteristics |                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristics                          | SMART-D System                                                                                                                                                                                                                                                                                                                                                                    | PEAK MOTUS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | CODA CX1 MOTION<br>ANALYSIS SYSTEM                                                                                                                                                                                                                                                                                                                                                                                                     |
| CFR Section                              | Unclassified                                                                                                                                                                                                                                                                                                                                                                      | Unclassified                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Unclassified                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Pro-code                                 | LXJ                                                                                                                                                                                                                                                                                                                                                                               | LXJ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | LXJ                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Classification name                      | System, optical position/movement<br>recording                                                                                                                                                                                                                                                                                                                                    | System, optical position/movement<br>recording                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | System, optical position/movement<br>recording                                                                                                                                                                                                                                                                                                                                                                                         |
| Intended / Indications<br>For Use        | SMART-D is a system for<br>motion analysis,<br>intended for the<br>recording and analysis of<br>human movement<br>patterns in the fields of<br>rehabilitation, sports<br>medicine, ergonomics.                                                                                                                                                                                    | Computer and video<br>system used to quantify<br>and graphically display<br>human movement<br>patterns and techniques<br>for uses such as<br>assessment and training<br>of limb or body motion<br>in gait analysis,<br>prosthetic design,<br>pre/post rehabilitation<br>evaluation, physical<br>therapy, and the like.                                                                                                                                                                                                    | The Coda CX I Motion<br>Analysis System has<br>general application to<br>measurement and<br>recording of 3D position<br>and movement,<br>including human<br>movement. It is<br>appropriate for use in<br>assessment of the 3D<br>motion of the limbs and<br>body of patients who<br>have some impairment                                                                                                                               |
|                                          |                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | of movement functions<br>of either a neurological<br>or orthopaedic cause.                                                                                                                                                                                                                                                                                                                                                             |
| System Components                        | Workstation + Cameras<br>+ Software                                                                                                                                                                                                                                                                                                                                               | Workstation + Cameras<br>+ Software                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Workstation + Cameras<br>+ Software                                                                                                                                                                                                                                                                                                                                                                                                    |
| Software tools                           | SMARTcapture,<br>SMARTtracker,<br>SMARTanalyzer,<br>SMARTclinic to achieve:<br>a) Real-time<br>visualization of signals of<br>all integrated devices'<br>b) 3D kinematics<br>reconstruction<br>Validated by<br>international scientific<br>community protocols<br>c) Tool for easy<br>construction of analysis<br>protocols<br>d) Multimedia<br>customizable and web<br>reporting | Vicon Motus<br>a) Advanced video or<br>3D-optical-motion-<br>capture module<br>b) 2D and/or 3D<br>kinematic calculation and<br>display software<br>Analog acquisition<br>module<br>c) 3D pan and tilt<br>module<br>KineCalc mathematical<br>analysis module<br>d) Peak Motus Gait<br>Analysis Template                                                                                                                                                                                                                    | a) Realtime displays to<br>verify marker location<br>Fully labelled data<br>appears immediately<br>after acquisition from<br>Codamotion active<br>marker system<br>b) Switch between<br>different projects in<br>seconds<br>Options for simultaneous<br>EMG and Force Vector<br>Overlay<br>Full standardised gait<br>analysis package<br>available<br>c) Graphs and stick<br>figure displays<br>Highly configurable<br>analysis setups |
| Contraindications                        | Not needed                                                                                                                                                                                                                                                                                                                                                                        | Not needed                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Not needed                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Target population                        | All population                                                                                                                                                                                                                                                                                                                                                                    | All population                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | All population                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Electrical Safety                        | AAMI/ANSI ES60601-<br>1:2005 and further<br>amendments                                                                                                                                                                                                                                                                                                                            | SAME                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | SAME                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| EMC                                      | IEC 60601-1-2                                                                                                                                                                                                                                                                                                                                                                     | IEC 60601-1-2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | IEC 60601-1-2                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Power Supply                             | 110 V, 60 Hz                                                                                                                                                                                                                                                                                                                                                                      | 110 V, 60 Hz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 110 V, 60 Hz                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Frame rate                               | 200 fps                                                                                                                                                                                                                                                                                                                                                                           | VCam/SV Cam<br>200 fps                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Not available                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Acquisition<br>Frequencies               | 250 Hz (up to 500 Hz)                                                                                                                                                                                                                                                                                                                                                             | up to 250 Hz progressive<br>up to 500 Hz progressive<br>up to 1000 Hz<br>progressive                                                                                                                                                                                                                                                                                                                                                                                                                                      | Sampling Rates vs.<br>Marker Numbers<br>100Hz for 56 markers<br>400Hz for 12 markers                                                                                                                                                                                                                                                                                                                                                   |
| Light emission                           | Infrared                                                                                                                                                                                                                                                                                                                                                                          | Infrared                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Not available                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Characteristics                          | SMART-D System                                                                                                                                                                                                                                                                                                                                                                    | PEAK MOTUS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | CODA CX1 MOTION ANALYSIS SYSTEM                                                                                                                                                                                                                                                                                                                                                                                                        |
| Max Input power                          | 950 W                                                                                                                                                                                                                                                                                                                                                                             | 15-20 W for single unit                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Not available                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Number of cameras                        | Up to 16 digital TVC                                                                                                                                                                                                                                                                                                                                                              | up to 12 cameras on one PC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Up to 16 digital inputs                                                                                                                                                                                                                                                                                                                                                                                                                |
| Resolution (pixel)                       | 800 H x 600 V                                                                                                                                                                                                                                                                                                                                                                     | MCam-2 59.94: 1272 H x 1024 V<br>MCam-2 100: 1266 H x 1024 V<br>MCam-2 119.88: 1266 H x 940 V<br>MCam-2 250: 816 H x 656 V<br>MCam-2 500: 498 H x 480 V<br>MCam-2 1000: 498 H x 210 V<br>VCam/SV Cam : 648 H x 493 V<br>MCam (PAL): 1011 H x 1024 V<br>1000 H x 972 V<br>MCam (NTSC): 1012 H x 987 V<br>947 H x 881 V<br>All Vicon MX cameras are built for speed at their full frame resolutions - 370fps for the 4 Megapixel F40, 240fps for the MX-3+ and a massive 500fps for the 2 Megapixel F20. On top of this all | 1/10th of pixel in a 5 Megapixel camera                                                                                                                                                                                                                                                                                                                                                                                                |

{2}------------------------------------------------

Confidential

.

·

.

{3}------------------------------------------------

Confidential

Section5 Rev 4

{4}------------------------------------------------

| run at speeds up to<br>2000fps at reduced,<br>windowed resolutions<br>particularly high<br>i for a |
|----------------------------------------------------------------------------------------------------|
| speed applications                                                                                 |

The Resolution (pixels) does not bring a point of non-substantial equivalence between the SMART-D System and the predicate devices, as it does not influence the intended use, the performance, the safety and effectiveness of the system than the predicate devices.

The only effect that this characteristic produces is a better or worse capability to detect the markers in the space and displaying them on the video.

| Acquisition Frequencies | 250 Hz (up to 500 Hz) | up to progressive | 120Hz   | Sampling Rates vs. Marker Numbers |
|-------------------------|-----------------------|-------------------|---------|-----------------------------------|
|                         |                       | up to progressive | 250 Hz  | 100Hz for 56 markers              |
|                         |                       | up to progressive | 500 Hz  | 200Hz for 28 markers              |
|                         |                       | up to progressive | 1000 Hz | 400Hz for 12 markers              |
|                         |                       |                   |         | 800 Hz for 6 markers              |

The Acquisition frequency does not bring a point of non-substantial equivalence between the SMART-D System and the predicate devices, as it does not influence the intended use, the performance, the safety and effectiveness of the system than the predicate devices.

The only effect that this characteristic produces is a different quantity of data in the time (frames in the time) to reconstruct the trajectory of the markers in the space.

| September 2000 September 2000 September 200 September 200 September 200<br>Waight | Not of Section Section Concession Concession, Concellente of Children, Concession, Concession, Concession, Concession, Concession, Concession, Concession, Concession, of Chil<br>nn<br>t<br>Comments of the first of the first of the first of the first of the first and | rs re<br>i<br>and the comments of the many of the comments of the comments of the country of the county of the county of the county of the county of the county of the county of the county | .<br>140-4-40. |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
|                                                                                   |                                                                                                                                                                                                                                                                            |                                                                                                                                                                                             |                |

The weight of workstation does not bring a point of non-substantial equivalence between the SMART-D System and the predicate devices, as it does not influence the intended use, the performance, the safety and effectiveness of the system than the predicate devices.

The three devices are not portable, so the weight does not influence their use and application.

| Dimensions | 450mm long x 330mm<br>high x 500 mm deep | 443.4 mm long x<br>43.6mm high x<br>348.2mm deep | 800mm long x 112mm<br>high x 81mm deep |
|------------|------------------------------------------|--------------------------------------------------|----------------------------------------|
|------------|------------------------------------------|--------------------------------------------------|----------------------------------------|

The Dimensions of workstation do not bring a point of non-substantial equivalence between the SMART-D System and the predicate devices, as it does not influence the intended use, the performance, the safety and effectiveness of the system than the predicate devices.

The three devices are not portable, so the dimensions do not influence their use and application. Moreover the difference of dimensions between the three workstations is negligible respect to the environment in which they have to be placed.

| Sensor<br>Technical<br>Characteristics | Wave<br>length: | 880 nm                                                                                   | Not available | Not available |
|----------------------------------------|-----------------|------------------------------------------------------------------------------------------|---------------|---------------|
|                                        | Output<br>angle | 40°                                                                                      | Not available | Not available |
|                                        | Modulation      | Square<br>wave<br>synchronized with the<br>acquisition frequency,<br>Duty Cycle 0+0.5 ms | Not available | Not available |
| Lens                                   |                 | C-mount compatible<br>(std. 8 mm)                                                        | Not available | Not available |

{5}------------------------------------------------

The Lenses and Sensor technical characteristics do not bring a point of non-substantial equivalence between the SMART-D System and the predicate devices, as they do not influence the intended use, the performance, the safety and effectiveness of the system than the predicate devices.

The difference brouqht by those differences is in the quality of the image provided to the system.

| Options | None | SAME | Tripod Mount for cx1<br>units                                 |
|---------|------|------|---------------------------------------------------------------|
|         |      |      | Gait Wand Set                                                 |
|         |      |      | Wired Strobe Adapter<br>unit and Drive Box<br>Extension Cable |
|         |      |      | External Strobe Panel<br>for Active Hub or Mini<br>Hub        |

The options do not bring a point of non-substantial equivalence between the SMART-D System and the predicate devices, as they do not influence the intended use, the safety and effectiveness of the system than the predicate devices.

Options are optional configurations or accessories to the main system which are not needed for its normal functioning and do not influence it.

#### Performance Data

The performance data indicate that the system for motion analysis SMART-D meets all specified requirements, and is substantially equivalent to the predicate device.

#### Conclusions

The non-clinical tests performed on the SMART-D System provided in Appendix 4, demonstrate that the SMART-D System is as effective and performs as the predicate devices. The acceptance criteria for each test item included the appropriate device specification or expected operation results. All acceptance criteria were met in the testing and the overall performance bench testing concluded that the SMART-D system had passed all the performance requirements defined by the System Requirement Specifications (provided in Appendix 2). The effectiveness and performance of the system are tested through the subtest: SMART capture, SMART tracker, SMART analyser, SMART clinic, which verify the same functionalities of the predicate devices as described in the comparison below:

| Functionality                          | SMART-D System                                                                          | PEAK MOTUS                                                                                                  | CODA CX1 MOTION<br>ANALYSIS SYSTEM                                                                                                                              |
|----------------------------------------|-----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Calibration and Signals<br>acquisition | SMARTcapture: a) Real-<br>time visualization of<br>signals of all integrated<br>devices | Vicon Motus<br>a) Advanced video or<br>3D-optical-motion-<br>capture module<br>Analog acquisition<br>module | a) Realtime displays to<br>verify marker location<br>Fully labelled data<br>appears immediately<br>after acquisition from<br>Codamotion active<br>marker system |
| Trajectory<br>reconstruction, signal   | SMARTtracker: b) 3D                                                                     | b) 2D and/or 3D                                                                                             | b) Switch between                                                                                                                                               |

{6}------------------------------------------------

#### 510(k) NOTIFICATION

| elaboration                                                           | kinematics<br>reconstruction<br>Validated by<br>international scientific<br>community protocols | kinematic calculation and<br>display software                            | different projects in<br>seconds<br>Options for simultaneous<br>EMG and Force Vector<br>Overlay<br>Full standardised gait<br>analysis package<br>available |
|-----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Management of<br>acquired data,<br>customizable analysis<br>protocols | SMARTanalyzer: c) Tool<br>for easy construction of<br>analysis protocols                        | c) 3D pan and tilt<br>module<br>KineCalc mathematical<br>analysis module | c) Graphs and stick<br>figure displays<br>Highly configurable<br>analysis setups                                                                           |
| Report drafting, data<br>visualization                                | SMARTclinic: d)<br>Multimedia customizable<br>and web reporting                                 | d) Peak Motus Gait<br>Analysis Template                                  |                                                                                                                                                            |

The EMC and Electrical Safety Tests performed on the SMART-D System and the predicate devices are the same and demonstrate that the SMART-D System complies with the AAMI/ANSI ES60601-1:2005 and further amendments and IEC 60601-1-2, and that it is as safe, as effective and performs as well as the predicate devices.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three wavy lines representing the department's commitment to health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol.

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002

April 17, 2014

BTS Spa c/o Enrico Bisson ISO plan Via Della Croce Rossa. I I Padova, Italy. 35129

Re: K131660 Trade/Device Name: SMART-D Regulatory Class: Unclassified Product Code: 1_XJ Dated: March 12, 2014 Received: March 18, 2014

Dear Mr. Bisson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements. including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{8}------------------------------------------------

Page 2 - Enrico Bisson

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Felipe Aguel -S

Carlos L. Peña, Ph.D., M.S. for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

510(k) Number (if known)

K131660

Device Name

SMART-D

Indications for Use (Describe)

SMART-D is a system for motion analysis, intended for the recording and analysis of human movement in the fields of rehabilitation, sports medicine, ergonomics.

Type of Use (Select one or both, as applicable)

[x] Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

## FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)

# Date: 2014.04.17 13:35:53 -04'00'

FORM FDA 3881 (1/14)

{10}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/LXJ/K131660](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/LXJ/K131660)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
