← Product Code [ILY](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/ILY) · K162763

# Vevazz LED Heat Lamp (K162763)

_Vevazz, LLC · ILY · Jan 5, 2017 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/ILY/K162763

## Device Facts

- **Applicant:** Vevazz, LLC
- **Product Code:** [ILY](/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/ILY.md)
- **Decision Date:** Jan 5, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5500
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The Vevazz is a device intended to emit energy in the visible and near-IR spectrum to provide topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscular and joint pain and stiffness, minor arthritis pain or muscle spasm; the temporary increase in local blood circulation and temporary relaxation of muscles.

## Device Story

Vevazz LED Heat Lamp is a prescription-use device providing topical heating via infrared energy. It consists of a base unit connecting up to 16 large and 4 small LED paddles. Each paddle emits 650 nm light at 50 mW/cm2. The device operates by elevating tissue temperature to 41°±2°C to provide temporary relief of minor muscle/joint pain, stiffness, arthritis, or spasms, and to increase local blood circulation. It is intended for use in clinical settings. Healthcare providers apply the paddles to the patient's skin; the device maintains a constant temperature during a default 17-minute treatment cycle. The heat application aims to relax muscles and alleviate pain, providing therapeutic benefit through thermal effects.

## Clinical Evidence

Bench testing only. Performance testing verified the device reaches and maintains an idle temperature of 40°C (±2°C) for a minimum of 10 minutes. An IR camera monitored skin temperature and uniformity on the abdominal region across multiple skin types. The device complies with IEC 60601-1, IEC 60601-2-57, IEC 62471, and IEC 60601-1-2 standards.

## Technological Characteristics

Base unit supports up to 16 large (28 LEDs) and 4 small (4 LEDs) paddles. LEDs emit 650 nm (+/- 25 nm) light at 50 mW/cm2. Patient contact material is rigid ABS. Power input 120/240 VAC; output 16 VDC @ 0.5 amps per port. Constant waveform. Complies with IEC 60601-1, IEC 60601-2-57, IEC 62471, and IEC 60601-1-2.

## Regulatory Identification

An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

## Predicate Devices

- HVR Infrared Lamp by HVR, LLC ([K101716](/device/K101716.md))
- Tanda Restore by Pharos Life Corporation ([K090008](/device/K090008.md))
- HEATLUX by Home Skinnovations, LTD. ([K120582](/device/K120582.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 5, 2017

Vevazz, LLC Mr. Jamie Fettig, DC Owner 30520 Rancho California Rd, Suite 107-46 Temecula, California 92591

Re: K162763

Trade/Device Name: Vevazz Led Heat Lamp Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: ILY Dated: November 3, 2016 Received: November 8, 2016

Dear Dr. Fettig:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

# Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) 162763

Device Name Vevazz LED Heat Lamp

### Indications for Use (Describe)

The Vevazz is a device that emits energy in the infrared spectrum and is to be used to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain, arthritis, stiffness or muscle spasms; the temporary increasing of local blood circulation and/or the temporary relaxation of muscle.

| Type of Use (Select one or both, as applicable)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                      |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <span style="font-family: Arial;"> <span style="font-size: 10pt;"> <span style="font-style: normal;"> <span style="font-variant: normal;"> <span style="font-weight: normal;"> <span style="letter-spacing: normal;"> <span style="line-height: normal;"> <span style="orphans: auto;"> <span style="text-align: start;"> <span style="text-indent: 0px;"> <span style="text-transform: none;"> <span style="white-space: normal;"> <span style="widows: auto;"> <span style="word-spacing: 0px;"> <span style="-webkit-text-stroke-width: 0px;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)                </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> </span> | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510-(k) Summary

#### 1. Submitters Information

Name: Vevazz, LLC 30520 Rancho California Road Suite 107-46 Temecula, CA 92591 Tel: 773-665-4005 Contact: Jamie Fettig, Owner 773-620-9500 jamie@healthyhuman.org

#### 2. Date Prepared:

January 3, 2017

#### 3. Device Name and Classification

| Device                 | Vevazz LED Heat Lamp |
|------------------------|----------------------|
| Trade Name             | Vevazz LED Heat Lamp |
| Regulation Description | Infrared lamp        |
| Regulation Number      | 21 CFR 890.5500      |
| Device Class           | II                   |
| Review Panel           | Physical Medicine    |
| Product Code           | ILY                  |

#### 4. Intended Use

The Vevazz is a device intended to emit energy in the visible and near-IR spectrum to provide topical heating for the purpose of elevating tissue temperature for a temporary relief of minor muscular and joint pain and stiffness, minor arthritis pain or muscle spasm; the temporary increase in local blood circulation and temporary relaxation of muscles.

{4}------------------------------------------------

#### 5. Device Description

The Vevazz device is a moderately complex device that consists of a base unit that plugs into 120 VAC or 240 VAC main power with Ground (earth) and a 2 amp 250 VAC Fuse. The base unit can support up to 16 large LED paddles and 4 small LED paddles. The connections on the base are female phono jacks that are low voltage having a maximum electrical output of 16 VDC @ .5 amps for a total of 8 Watts per port. The large paddles consist of 28 LEDs in a series/parallel electrical connection that has a cord with a male phono jack. Small paddles consist of 4 LEDs and have matching electrical characteristics as the large paddles. Each LED produces 50 mw/cm2 of light energy with a wavelength of 650 nm: +/- 25 pm uniformly over the treatment area without producing hot spots.

#### Predicate Devices 6.

- HVR Infrared Lamp by HVR, LLC K101716 ●
- Tanda Restore by Pharos Life Corporation K090008 .
- . HEATLUX by Home Skinnovations, LTD. - K120582

#### 7. Technical and Performance Substantial Equivalency

The submission and predicate devices are fully compliant with USFDA recognized standards for electrical safety and emit a substantially equivalent amount of power and energy. Differences where they occur are minor and do not represent any risk or hazard.

| Performance<br>Characteristic  | Vevazz                  | HVR Lamp          | Tanda Restore | HeatLux     |
|--------------------------------|-------------------------|-------------------|---------------|-------------|
| Power                          | 50 mW/cm2               | 50-80 mW/cm2      | 60 mW/cm2     | 60 mW/cm2   |
| Wavelength                     | 650 nm +/-<br>25nm      | 650 nm to 950 nm  | 870 nm        | 645 nm      |
| Waveform                       | Constant                | Constant          | Constant      | Constant    |
| Energy Source                  | LED                     | LED               | LED           | LED         |
| Treatment<br>Area              | 7.5 cm2                 | 1000 mm2 (10 cm2) | 27 cm2        | 7 cm2       |
| Treatment<br>Time              | 17 minutes<br>(default) | 90 seconds        | 3 minutes     | 3-5 minutes |
| Target Skin<br>Temperature     | 41°±2°C                 | 42°±3°C           | 41°±2°C       | 41°±2°C     |
| Patient<br>Contact<br>Material | Rigid ABS               | Rigid ABS         | Rigid ABS     | Rigid ABS   |

# 8. Bench Testing Summary

The Vevazz is in compliance with the following standards:

- IEC 60601-1:2005 +C1:2006, +C2:2007, AM1:2012, Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- IEC 60601-2-57:2011, Medical electrical equipment Part 2-57: Particular requirements for basic safety and essential performance of non-laser lightsource equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
- IEC 62471 Ed. 1, International Standard for Photobiological Safety for Lamps and Lamp Systems - General Requirements

{5}------------------------------------------------

IEC 60601-1-2:2007, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests.

## 9. Clinical Testing Summary

The Vevazz heat performance was tested to verify it will reach an idle temperature of 40℃ and will maintain that temperature throughout the treatment for a minimum duration of 10 minutes ±2℃. An IR camera was used to monitor skin temperature and uniformity on the abdominal region and multiple skin types.

## 10. Conclusion

The Vevazz is substantially equivalent to its predicate devices. Based on the technological comparison between the Vevazz and the predicates plus the human test data, it can be concluded that the Vevazz is equivalent to the identified predicates.

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/ILY/K162763](https://fda.innolitics.com/submissions/PM/subpart-f%E2%80%94physical-medicine-therapeutic-devices/ILY/K162763)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
