ILX · Diathermy, Shortwave, For Use Other Than Applying Therapeutic Deep Heat

Physical Medicine · 21 CFR 890.5290 · Class 2

Overview

Product CodeILX
Device NameDiathermy, Shortwave, For Use Other Than Applying Therapeutic Deep Heat
Regulation21 CFR 890.5290
Device ClassClass 2
Review PanelPhysical Medicine

Identification

A nonthermal shortwave therapy is a prescription device that applies to the body pulsed electromagnetic energy in the RF bands of 13.56 MHz or 27.12 MHz and that is intended for adjunctive use in the palliative treatment of postoperative pain and edema of soft tissue by means other than the generation of deep heat within body tissues as described in paragraph (a) of this section.

Classification Rationale

Class II (special controls). The device is classified as class II.

Special Controls

*Classification: Class II (special controls).* The device is classified as class II. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. (ii) Appropriate analysis/testing must demonstrate that the device is electrically safe and electromagnetically compatible in its intended use environment. (iii) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Non-clinical performance testing must characterize the output waveform of the device and demonstrate that the device meets appropriate output performance specifications. The output characteristics and the methods used to determine these characteristics, including the following, must be determined: (A) Peak output power; (B) Pulse width; (C) Pulse frequency; (D) Duty cycle; (E) Characteristics of other types of modulation that may be used; (F) Average measured output powered into the RF antenna/applicator; (G) Specific absorption rates in saline gel test load or other appropriate model; (H) Characterization of the electrical and magnetic fields in saline gel test load or other appropriate model for each RF antenna and prescribed RF antenna orientation/position; and (I) Characterization of the deposited energy density in saline gel test load or other appropriate model. (iv) A detailed summary of the clinical testing pertinent to use of the device to demonstrate the effectiveness of the device in its intended use. (v) Labeling must include the following: (A) Output characteristics of the device; (B) Recommended treatment regimes, including duration of use; and (C) A detailed summary of the clinical testing pertinent to the use of the device and a summary of the adverse events and complications. (vi) Nonthermal shortwave therapy devices marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with these special controls.

Recent Cleared Devices (15 of 15)

RecordDevice NameApplicantDecision DateDecision
K241395Active System; Avenue8Caerus CorporationDec 18, 2024SESE
K202337Pulsed electromagnetic field wrapHi-Dow International, Inc.Dec 15, 2020SESE
K190251RecoveryRxBioelectronics CorporationJun 28, 2019SESE
K131979PROVANT THERAPY SYSTEMRegenesis Biomedical, Inc.Dec 13, 2013SESE
K121702ORTHOCOR ACTIVE DEVICE ORTHOCOR GENERAL USE ACTIVE DEVICEOrthocor MedicalMay 6, 2013SESE
K121338ZEOBIIvivi Health Science, LLCJul 27, 2012SESE
K091791PROVANT SYSTEM, MODEL 4201Regenesis Biomedical, Inc.Apr 7, 2010SESE
K092044ORTHOCOR KNEE SYSTEM, MEDIAN, LARGE, X-LARGEOrthocor MedicalDec 18, 2009SESE
K091996ORTHOCOR KNEE SYSTEM BASICOrthocor MedicalDec 15, 2009SESE
K070541IVIVI SOFPULSE, MODELS 912-M10, ROMA3 AND TORINO IIIvivi Technologies, Inc.Dec 11, 2008SESE
K070931MODEL PMT850Promedtek, Inc.May 24, 2007SESE
K022404ACTIBANDPaw, LLCAug 8, 2002SESE
K972093REGENESIS MODEL 42Regenesis Biomedical, Inc.Oct 21, 1997SESE
K903675MAGNETIC RESONANCE THERAPY DEVICEMrt, Inc.Jan 17, 1991SESE
K882816PULSAR EMApex Medical, Inc.Apr 27, 1989SESE

Top Applicants

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...