INCARE HOT/ICE LITE MACHINE

K962537 · Hollister, Inc. · ILO · Nov 27, 1996 · Physical Medicine

Device Facts

Record IDK962537
Device NameINCARE HOT/ICE LITE MACHINE
ApplicantHollister, Inc.
Product CodeILO · Physical Medicine
Decision DateNov 27, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5720
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The InCare HOT/ICE Lite Machine is intended to provide cold therapy to body surfaces by pumping and circulating chilled water through a plastic bag or plastic blanket that has been secured to the patient’s anatomy.

Device Story

Device provides localized cold therapy; circulates chilled water/alcohol mixture through plastic blanket or bag secured to patient anatomy. Used in institutional or home environments. Solid-state electronics control pump and circulation. Healthcare provider or patient secures blanket; device maintains therapy temperature range (50° to 72° F). Output is cooling of localized skin surface; intended to reduce skin temperature for therapeutic benefit.

Clinical Evidence

Bench testing only. Evaluation of skin cooling ability performed using temperature probe on test subjects' skin surface over a specific period. Results confirmed device efficacy in reducing patient skin temperature.

Technological Characteristics

Solid-state electronics; 120VAC, 136 watts, 50/60Hz power supply. Liquid media: water/alcohol mixture (500 ml capacity). Nominal flow rate: 4 GPH; nominal operating pressure: 3.5 PSI. Therapy temperature range: 50° to 72° F (10° to 22° C). Weight: approx 7.1 lbs. Form factor: portable unit for institutional/home use.

Indications for Use

Indicated for patients requiring localized cold therapy to body surfaces.

Regulatory Classification

Identification

A water circulating hot or cold pack is a device intended for medical purposes that operates by pumping heated or chilled water through a plastic bag and that provides hot or cold therapy for body surfaces.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 890.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} NOV 27 1996 Hollister Incorporated InCare HOT/ICE Lite Machine K962537 # Safety and Effectiveness Summary ## 1. Submitter's name, Address and Contact Person | Submitter | Contact Person | | --- | --- | | Hollister Incorporated 2000 Hollister Drive Libertyville, IL 60048 | Joseph S. Tokarz Manager, Regulatory Affairs Phone (847)680-2849 Fax (847)918-3860 | Date Summary Prepared - June 26, 1996 ## 2. Name of Device: InCare HOT/ICE Lite Machine ## 3. Name of Predicate Device(s) a. InCare HE500 (K830757) b. InCare System 3 (K915196) ## 4. Description of Device The InCare HOT/ICE Lite Machine incorporates state of the art technology with solid state electronics. The device is designed to deliver cold therapy to a localized area. This is accomplished by circulating cold water through a plastic blanket or bag that has been secured to the patient’s anatomy. The device is intended to be used in an institution or home environment. ## 5. Statement of Intended Use The InCare HOT/ICE Lite Machine is intended to provide cold therapy to body surfaces by pumping and circulating chilled water through a plastic bag or plastic blanket that has been secured to the patient’s anatomy. ## 6. Statement of Technological Characteristics of the Device On the following page is a comparison of the technical specifications of the InCare HOT/ICE Lite Machine and the predicate devices. {1} Hollister Incorporated InCare HOT/ICE Lite Machine | Features/Specs | Proposed device InCare HOT/ICE Lite | InCare HE500 K830757 | InCare System 3 K915196 | | --- | --- | --- | --- | | Intended to provide cold therapy to body surfaces | Yes | Yes | Yes | | Solution Flow Rates GPH=gallons per hour | 4 GPH (nominal) | 15 GPH (nominal) | 19 GPH (nominal) | | Operating Pressure PSI=pounds per square inch | 3.5 PSI (nominal) | 2.5 PSI (nominal) | 4.8 PSI (nominal) | | Liquid media | Water / Alcohol | Water / Alcohol | Water / Alcohol | | Liquid Capacity | 500 ml = 350 ml Water + 150 ml Alcohol | 500 ml = 350 ml Water + 150 ml Alcohol | 650 ml water/alcohol | | Therapy temperature range | 50° to 72° F 10° to 22° C | 40° to 150° F 4° to 40° C | 40° to 150° F 4° to 40° C | | Weight | approx 7.1 lbs | approx 20 lbs | 32 lbs (approximate) | | Power Supply | 120VAC, 136 watts, 50/60Hz | 115 V, 60 Hz, 3 AMPS | 115 VAC, 60 Hz 3 AMPS | ## 7. Performance Testing The InCare HOT/ICE Lite Machine with a HOT/ICE blanket was evaluated for its ability to cool the test subjects skin as measured by a temperature probe on the skin surface. Skin temperature was recorded over a specific period of time. The results indicated that the InCare HOT/ICE Lite Machine in its ability to reduce patient skin temperature. ## 8. Conclusion The InCare HOT/ICE Lite Machine is substantially equivalent to the predicate devices and is safe and effective for its intended use. 37
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...