← Product Code [IOR](/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/IOR) · K232198

# Reclining wheelchair (YJ-011S 16D, YJ-011S 16DF, YJ-011S 18D, YJ-011S 18DF, YJ-011S 20D, YJ-011S 20DF) (K232198)

_Zhenjiang Assure Medical Equipment Co., Ltd. · IOR · Sep 15, 2023 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/IOR/K232198

## Device Facts

- **Applicant:** Zhenjiang Assure Medical Equipment Co., Ltd.
- **Product Code:** [IOR](/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/IOR.md)
- **Decision Date:** Sep 15, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3850
- **Device Class:** Class 1
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The YJ-011S Manual Wheelchair is to provide mobility to persons limited to a sitting position.

## Device Story

Manual, foldable wheelchair; provides mobility to individuals restricted to sitting position. Features main frame, seat, adjustable footrests, four wheels, hand rims for manual propulsion, and brakes. Includes skirt guards to prevent clothing entanglement. Operated by user or assistant via push handles. Used in indoor/outdoor environments on dry, level surfaces (concrete, blacktop, asphalt). Mechanical design; no electronic or software components. Benefits patient by facilitating independent or assisted mobility.

## Clinical Evidence

No clinical data. Bench testing only. Device performance verified against ISO 7176 standards (stability, brakes, dimensions, strength, ignition resistance) and biocompatibility standards (ISO 10993-5, -10, -23).

## Technological Characteristics

Mechanical wheelchair; Q235 steel frame; foldable cross-brace design. Manual propulsion via hand rims; occupant-operated tire brakes. Dimensions vary by model (16", 18", 20" seat widths). Complies with ISO 7176 series for stability, strength, and dimensions. Biocompatible materials (cytotoxicity, sensitization, irritation testing per ISO 10993).

## Regulatory Identification

A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- Manual Wheelchair (A011) ([K201461](/device/K201461.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 15, 2023

Zhenjiang Assure Medical Equipment Co., Ltd. % Eva Li Consultant Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave. Shanghai. 200122 China

Re: K232198

Trade/Device Name: Reclining wheelchair (YJ-011S 16"D, YJ-011S 16"DF, YJ-011S 18"D, YJ-011S 18"DF, YJ-011S 20"D, YJ-011S 20"DF)

Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I, reserved Product Code: IOR Dated: July 25, 2023 Received: July 25, 2023

Dear Eva Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Tushar Bansal -S

for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K232198

Device Name

Reclining wheelchair (YJ-011S 16"D, YJ-011S 16"DF, YJ-011S 18"DF, YJ-011S 18"DF, YJ-011S 20"DF)

Indications for Use (Describe)

The YJ-011S Manual Wheelchair is to provide mobility to persons limited to a sitting position.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 1. Submitter

Zhenjiang Assure Medical Equipment Co., Ltd. Address: No.297, Chuqiao Road, Zhenjiang City, Jiangsu Province, China Contact Person: Mei Ying Tel: +86-13852941204 Email: shelly@assuremedical.com.cn

Submission Correspondent Contact: Eva Li Phone: + 86 (215) 881-7802 Email: zxfda@sungoglobal.com Shanghai SUNGO Management Consulting Co., Ltd. Room 1401, Dongfang Building, 1500# Century Ave, Shanghai, 200122 CHN

Prepared Date: July 18th,2023

#### 2. Device

Name of Device: Reclining wheelchair Common Name: Manual Wheelchair Model(s): YJ-011S 16"D, YJ-011S 16"DF, YJ-011S 18"D, YJ-011S 18"DF, YJ-011S 20"D, YJ-011S 20"DF

Regulatory Information Classification Name: Mechanical Wheelchair Regulatory Class: I Product code: IOR Regulation Number: 890.3850 Review Panel: Physical Medicine

#### 3. Predicate device:

K201461 Ningbo Shenyu Medical Equipment Co.,Ltd. Manual Wheelchair (A011)

#### 4. Device description

The YJ-011S series is a mechanical wheelchair which is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position. It can be folded for transport by bringing the two sides together. The manual wheelchair incorporates a main frame, a seat, two adjustable footrests and four wheels.

The larger rear wheels have hand rims of slightly smaller diameter projecting just beyond the tire. These allows the user to maneuver the chair by pushing them on without requiring them to grasp the tires. The manual wheelchairs have brakes that bear on the tires of the rear wheels and two push handles at the upper rear of the frame to allow for manual propulsion by an assistant.

{4}------------------------------------------------

#### Main Components:

Main frame, back upholstery, seat upholstery, handgrip, armrest, front wheel, rear wheel, hand rim, crossbar, footplates, brake, leg rest.

The device can be operated indoors, or outdoors on dry, level surfaces composed of concrete, blacktop, or asphalt under normal driving conditions.

| Model                              | YJ-011S 16"D, YJ-011S 16"DF, YJ-011S 18"D, YJ-011S 18"DF, YJ-011S 20"D, YJ-011S 20"DF |        |                                               |                        |
|------------------------------------|---------------------------------------------------------------------------------------|--------|-----------------------------------------------|------------------------|
| Overall dimension                  | length 1340mm* high 1258mm<br>*16"width 625mm<br>*18"width 675mm<br>*20"width 725mm   |        |                                               |                        |
| Folded dimension                   | width 323mm                                                                           |        |                                               |                        |
| Seat width                         | 16"                                                                                   | 410mm  | Seat plane angle                              | 3°                     |
|                                    | 18"                                                                                   | 460mm  |                                               |                        |
|                                    | 20"                                                                                   | 510mm  |                                               |                        |
| Seat depth                         | 460mm                                                                                 |        | Seat height from floor                        | 525mm                  |
| Backrest angle                     | 7°-83.5°                                                                              |        | Backrest height                               | 500mm                  |
| Backrest width                     | 16"                                                                                   | 410mm  | Footrest-to-seat distance                     | 448mm-555mm            |
|                                    | 18"                                                                                   | 460mm  |                                               |                        |
|                                    | 20"                                                                                   | 510mm  |                                               |                        |
| Footrest clearance                 | 72mm                                                                                  |        | Footrest length                               | 200mm                  |
| Armrest-to-seat distance           | 228mm                                                                                 |        | Footrest-leg-angle                            | 95°                    |
| Leg-to-seat-surface angle          | 140°                                                                                  |        | Front-armrest-to-backrest distance            | -D: 320mm<br>-F: 475mm |
| Front-armrest-to-backrest distance | -D: 320mm                                                                             |        | Armrest length                                | -D: 260mm<br>-F: 352mm |
|                                    | -F: 475mm                                                                             |        |                                               |                        |
| Armrest width                      | 55mm                                                                                  |        | Armrest angle                                 | 3°-4°                  |
| Distance between armrests          | 16"                                                                                   | 427mm  | Front location of armrest structure           | 320mm                  |
|                                    | 18"                                                                                   | 477mm  |                                               |                        |
|                                    | 20"                                                                                   | 527mm  |                                               |                        |
| Hand rim diameter                  | Φ580mm                                                                                |        | Material                                      | Q235                   |
| Horizontal location of axle        | 55°                                                                                   |        | Maximum weight bearing capacity of the device | 350LBS                 |
| Weight of the device(net)          | 25.5kg                                                                                |        | Static stability sideways                     | 15.25°                 |
| diameter of front wheels           | Φ195mm                                                                                |        | diameter of rear wheels                       | Φ613mm                 |
| Static stability uphill            | 15.2°                                                                                 |        | Static stability downhill                     | 15.2°                  |
| Minimum turning radius             |                                                                                       | 1170mm |                                               |                        |

The specification table is as below:

{5}------------------------------------------------

510(k) Summary K232198

| Parking brake      |       |
|--------------------|-------|
| Max slope uphill   | 8.35° |
| Max slope downhill | 8.15° |

#### 5. Indication for use

The YJ-011S Manual Wheelchair is to provide mobility to persons limited to a sitting position.

#### 6. Comparison of technological characteristics with the predicate device

| Device                         | Predicate Device                                                                                        | Proposed Device                                                                                                         | Results     |
|--------------------------------|---------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|-------------|
| 510K Number                    | K201461                                                                                                 | K232198                                                                                                                 | --          |
| Manufacturer                   | Ningbo Shenyu Medical<br>Equipment Co., Ltd.                                                            | Zhenjiang Assure Medical<br>Equipment Co., Ltd.                                                                         | --          |
| Proprietary Name               | Manual Wheelchair                                                                                       | Reclining wheelchair                                                                                                    | --          |
| Classification                 | I                                                                                                       | I                                                                                                                       | same        |
| Indications for use            | The A011 Manual wheelchair is to<br>provide mobility to persons<br>limited to a sitting position.       | The YJ-011S reclining wheelchair is<br>to provide mobility to persons<br>limited to a sitting position.                 | same        |
| Design<br>Characteristic       | Manual Operation, Four-Wheels,<br>Foldable, Cross-Brace, Pull-to-<br>Lock, Armrest, Backrest, Foot rest | Manual Operation, Four-Wheels,<br>Foldable, Cross-Brace, Pull-to-<br>Lock, Armrest, Backrest, Foot rest,<br>skirt guard | Similar*1   |
| Brake control                  | occupant-operated brake only                                                                            | occupant-operated brake only                                                                                            | same        |
| Operation                      | For indoor/outdoor use                                                                                  | For indoor/outdoor use                                                                                                  | same        |
| Environment                    | For indoor/outdoor use                                                                                  | For indoor/outdoor use                                                                                                  | same        |
| Control Mode                   | Mechanical                                                                                              | Mechanical                                                                                                              | same        |
| Size(unfold)                   | 1100 (L) *660 (W) * 910mm (H)                                                                           | 1340 (L)* 1258 (H)<br>*16" 625mm(W)<br>*18" 675mm(W)<br>*20" 725mm(W)                                                   | Different*2 |
| Stowage<br>length/width/height | 810 (L) X 320 (W) X 930mm (H)                                                                           | 1340 (L)* 1258 (H)*323mm(W)                                                                                             | different*2 |
| Weight (Total)                 | 16kg(35.2lbs)                                                                                           | 25.5kg                                                                                                                  | different*2 |
| Weight Capacity                | 136Kg(300lbs)                                                                                           | 350LBS                                                                                                                  | different*2 |
| Seat Width                     | 480mm                                                                                                   | 410-510mm                                                                                                               | different*2 |
| Seat height                    | 540mm                                                                                                   | 525mm                                                                                                                   | different*2 |
| Seat depth                     | 420mm                                                                                                   | 460mm                                                                                                                   | different*2 |
| Back type                      | Fixed                                                                                                   | Adjustable                                                                                                              | Different*3 |
| Tires                          | Front: 200mm<br>Rear:610mm                                                                              | Front: 195mm<br>Rear:613mm                                                                                              | different*2 |
| Armrest                        | Flip back armrest                                                                                       | fixed                                                                                                                   | different*3 |
| Foot rest                      | Optional/ swing away Optional/<br>swing away                                                            | Optional/ swing away Optional/<br>swing away                                                                            | same        |

{6}------------------------------------------------

## 510(k) Summary

| K232198            |                                                                                                                                       |                                                                                                                                       |           |
|--------------------|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Rear Axle Position | Single                                                                                                                                | Single                                                                                                                                | Same      |
| Frame Construction | Foldable frame<br>Push inward from left and right<br>sides to fold                                                                    | Foldable frame<br>Push inward from left and right<br>sides to fold                                                                    | Same      |
| Safety Feature     | Manual Wheel Lock                                                                                                                     | Manual Wheel Lock                                                                                                                     | Same      |
| Performance        | Comply with:<br>ISO7176-1<br>ISO7176-3<br>ISO7176-5<br>ISO7176-7<br>ISO7176-8<br>ISO7176-11<br>ISO7176-13<br>ISO7176-15<br>ISO7176-16 | Comply with:<br>ISO7176-1<br>ISO7176-3<br>ISO7176-5<br>ISO7176-7<br>ISO7176-8<br>ISO7176-11<br>ISO7176-13<br>ISO7176-15<br>ISO7176-16 | Same      |
| Biocompatibility   | Comply with:<br>ISO10993-1<br>ISO10993-5<br>ISO10993-10                                                                               | Comply with:<br>ISO10993-5<br>ISO10993-10<br>ISO10993-23                                                                              | Similar*4 |

#### Discussion :

| Similar*1:   | Compared to the predicate device, the subject device adds two skirt guards which installed to the sides of the seat frame under the arm rests to provide a barrier between the occupant and the wheels. This feature can prevent occupant's clothes from getting caught in the wheels. This feature will not raise any new risk of safety or effectiveness.                                                                                                         |
|--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Different*2: | Compared to the predicate device, the subject device has different value on the unfold size, stowage size, device weight, capacity, seat width, seat height, seat depth, tire size. However, the subject has passed the <ISO 7176-7-1998 Part7: Measurement of seating and wheel dimensions > and <ISO 7176-5-2008 Part 5: Determination of dimensions, mass and maneuvering space>, so the above different will not raise any new risk of safety or effectiveness. |
| Different*3: | Compared to the predicate device, the subject device's back is adjustable, and the arm can't be flipped. These differences will not raise any new risk of safety or effectiveness.                                                                                                                                                                                                                                                                                  |
| Similar*4    | The stimulation test in ISO 10993-23 replaces the stimulation test in ISO 10993-10:2010. It will not raise any new risk of safety or effectiveness.                                                                                                                                                                                                                                                                                                                 |

#### 7. Summary of Non-Clinic Performance Testing Performance Testing

Non-clinical tests were conducted to verify that the subject device met all design specifications and was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: ISO 7176-1:2014, Wheelchairs - Part 1: Determination of static stability ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes

{7}------------------------------------------------

ISO 7176-5: 2008 Wheelchairs - Part 5: Determination of overall dimensions, mass and maneuvering space

ISO 7176-7: 1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions ISO 7176-8: 2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strengths

ISO 7176-11: 2012 Wheelchairs - Part 11: Test dummies

ISO 7176-13: 1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces

ISO 7176-15: 1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling

ISO 7176-16:2012 Wheelchairs - Part 16: Resistance to ignition of postural support devices ISO 7176-22:2014 Wheelchairs — Part 22: Set-up procedures

## Biocompatibility

Biocompatibility testing was conducted in accordance with the 2020 FDA guidance Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process." Testing included:

- · Cytotoxicity (ISO 10993-5:2009)
- Sensitization (ISO 10993-10:2010)
- · Irritation (ISO 10993-23:2021)

The testing supports the biocompatibility of the patient-contacting device materials that were shown to be non-cytotoxic, non-irritating, and non-sensitizing.

#### 8. Clinical Test Conclusion

No clinical study is included in this submission.

#### 9. Conclusion

The conclusion drawn from the nonclinical tests demonstrates that the subject device in this 510(k) submission is as safe and effective as the legally marketed predicate device cleared under K201461.

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/IOR/K232198](https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/IOR/K232198)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
