← Product Code [INK](/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/INK) · K771018

# TRANS 2000 (K771018)

_Hamilton Industries, Inc. · INK · Jul 14, 1977 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/INK/K771018

## Device Facts

- **Applicant:** Hamilton Industries, Inc.
- **Product Code:** [INK](/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/INK.md)
- **Decision Date:** Jul 14, 1977
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3690
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine

## Regulatory Identification

A powered wheeled stretcher is a battery-powered table with wheels that is intended for medical purposes for use by patients who are unable to propel themselves independently and who must maintain a prone or supine position for prolonged periods because of skin ulcers or contractures (muscle contractions).

## Special Controls

(1) Appropriate analysis and nonclinical testing must demonstrate that the safety controls are adequate to ensure safe use of the device and prevent user falls from the device in the event of a device failure; 
(2) Appropriate analysis and nonclinical testing must demonstrate the ability of the device to withstand the rated user weight load with an appropriate factor of safety; 
(3) Appropriate analysis and nonclinical testing must demonstrate the longevity of the device to withstand external forces applied to the device and provide the user with an expected service life of the device; 
(4) Appropriate analysis and nonclinical testing must demonstrate proper environments of use and storage of the device to maximize the longevity of the device; 
(5) Appropriate analysis and nonclinical testing (such as outlined in appropriate FDA-recognized consensus standards) must validate electromagnetic compatibility and electrical safety; 
(6) Appropriate analysis and nonclinical testing (such as outlined in appropriate FDA-recognized consensus standards) must validate that the skin-contacting components of the device are biocompatible; 
(7) Appropriate analysis and nonclinical testing (such as outlined in appropriate FDA-recognized consensus standards) must validate the software life cycle and that all processes, activities, and tasks are implemented and documented; 
(8) Appropriate analysis and nonclinical testing must validate that the device components are found to be nonflammable; 
(9) Appropriate analysis and nonclinical testing (such as outlined in appropriate FDA-recognized consensus standards) must validate that the battery in the device performs as intended over the anticipated service life of the device; 
(10) Adequate labeling is provided to the user to document proper use and maintenance of the device to ensure safe use of the device in the intended use environment; and 
(11) Appropriate risk assessment including, but not limited to, evaluating the dimensional limits of the gaps in hospital beds, and mitigation strategy to reduce entrapment.

---

**Source:** [https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/INK/K771018](https://fda.innolitics.com/submissions/PM/subpart-d%E2%80%94physical-medicine-prosthetic-devices/INK/K771018)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
