BIOFIRE® FILMARRAY® Pneumonia Panel (BIOFIRE Pneumonia Panel); BIOFIRE® FILMARRAY® Pneumonia Panel plus (BIOFIRE Pneumonia Panel plus)

K243222 · Biofire Diagnostics, LLC (Biomerieux) · QDS · Nov 6, 2024 · Pathology

Device Facts

Record IDK243222
Device NameBIOFIRE® FILMARRAY® Pneumonia Panel (BIOFIRE Pneumonia Panel); BIOFIRE® FILMARRAY® Pneumonia Panel plus (BIOFIRE Pneumonia Panel plus)
ApplicantBiofire Diagnostics, LLC (Biomerieux)
Product CodeQDS · Pathology
Decision DateNov 6, 2024
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 866.4001
Device ClassClass 2

Indications for Use

The BIOFIRE® FILMARRAY® Pneumonia Panel (BIOFIRE Pneumonia Panel) is a multiplexed nucleic acid test intended for use with BIOFIRE® FILMARRAY® 2.0 (BIOFIRE 2.0) or BIOFIRE® FILMARRAY® TORCH (BIOFIRE TORCH) systems for the simultaneous detection and identification of multiple respiratory viral and bacterial nucleic acids, as well as select antimicrobial resistance genes, in sputum-like specimens (induced or expectorated sputum, or endotracheal aspirates) or bronchoalveolar lavage (BAL)-like specimens (BAL or mini-BAL) obtained from individuals suspected of lower respiratory tract infection. The BIOFIRE® FILMARRAY® Pneumonia Panel plus (BIOFIRE Pneumonia Panel plus) is a multiplexed nucleic acid test intended for use with BIOFIRE® FILMARRAY® 2.0 (BIOFIRE 2.0) or BIOFIRE® FILMARRAY® TORCH (BIOFIRE TORCH) systems for the simultaneous detection and identification of nucleic acids from Middle East respiratory syndrome coronavirus (MERS-CoV) and multiple respiratory viral and bacterial nucleic acids, as well as select antimicrobial resistance genes, in sputum-like specimens (induced or expectorated sputum, or endotracheal aspirates) or bronchoalveolar lavage (BAL)-like specimens (BAL or mini-BAL) obtained from individuals meeting MERS-CoV clinical and/or epidemiological criteria.

Device Story

The BIOFIRE Pneumonia Panel and Panel plus are multiplexed nucleic acid tests for respiratory pathogens. The device processes sputum or BAL specimens using automated sample preparation within the BIOFIRE 2.0 or TORCH systems. It employs nested, multiplex reverse transcription PCR (RT-PCR) followed by high-resolution melting curve analysis to detect viral/bacterial nucleic acids and antimicrobial resistance (AMR) genes. The system provides automated test interpretation and report generation. The software update (Pouch Module v2.0–v2.1.4/v2.1.1) mitigates false-positive Coronavirus and CTX-M results caused by cross-reactivity with human genomic DNA. Results aid clinicians in diagnosing lower respiratory infections and MERS-CoV (for Panel plus) when used with other clinical/epidemiological data. The device is intended for use in clinical laboratory settings.

Clinical Evidence

Clinical performance was evaluated via reanalysis of prospective study data. The software update resulted in a slight increase in clinical specificity/NPA for the Coronavirus assay: from 98.4% to 98.7% in BAL specimens and from 99.3% to 99.5% in sputum specimens. Four Coronavirus results were reclassified from 'Detected' to 'Not Detected', necessitating updates to expected values. No other clinical performance claims were altered.

Technological Characteristics

Nested multiplex RT-PCR with high-resolution melting analysis. Automated sample processing and interpretation. Compatible with BIOFIRE 2.0 and TORCH systems. Reagents stored at room temperature. Software-based analysis of melt profiles to exclude non-specific human genomic DNA cross-reactivity.

Indications for Use

Indicated for individuals with signs and symptoms of lower respiratory tract infection. Detects/identifies respiratory viral and bacterial nucleic acids and antimicrobial resistance genes in sputum, bronchoalveolar lavage, and tracheal aspirate specimens. Aids diagnosis in conjunction with clinical/laboratory findings.

Regulatory Classification

Identification

A multiplex respiratory panel to detect and identify emerging respiratory pathogen(s) and common respiratory pathogens in human clinical specimens is identified as an in vitro diagnostic device intended for the qualitative detection and identification of both emerging and common respiratory pathogens from individuals meeting specific emerging respiratory pathogen clinical and/or epidemiological criteria. For example, clinical signs and symptoms associated with infection of the emerging respiratory pathogen, contact with a probable or confirmed emerging respiratory pathogen case, history of travel to geographic locations where cases of the emerging respiratory pathogen were detected, or other epidemiological links for which testing of the emerging respiratory pathogen may be indicated. A device to detect and identify emerging respiratory pathogen(s) and common respiratory pathogens in human clinical specimens, and in turn to distinguish emerging respiratory pathogen(s) from common respiratory pathogens, is intended to aid in the differential diagnosis of the emerging respiratory pathogen infection, in conjunction with other clinical, epidemiologic, and laboratory data, in accordance with the guidelines provided by the appropriate public health authorities.

Special Controls

*Classification.* Class II (special controls). The special controls for this device are:(1) The intended use for the labeling required under § 809.10 of this chapter must include a description of what the device detects and measures, the specimen types, the results provided to the user, the clinical indications for which the test is to be used, the specific intended population(s), the testing location(s) where the device is to be used (if applicable), and other conditions of use as appropriate. (2) The labeling required under § 809.10 of this chapter must include: (i) A device description, including the parts that make up the device, ancillary reagents required but not provided, and an explanation of the methodology. (ii) Performance characteristics from analytical studies, including cut-off (if applicable), analytical sensitivity ( *i.e.,* limit of detection), inclusivity, reproducibility, interference, cross-reactivity, instrument carryover/cross-contamination (if applicable), and specimen stability.(iii) Detailed instructions for minimizing the risk of potential users' exposure to the emerging respiratory pathogen(s) that may be present in test specimens and those used as control materials. (iv) Detailed instructions for minimizing the risk of generating false positive test results due to carry-over contamination from positive test specimens and/or positive control materials. (v) A warning statement that the interpretation of test results requires experienced healthcare professionals who have training in principles and use of infectious disease diagnostics and reporting of results, in conjunction with the patient's medical history, clinical signs and symptoms, and the results of other diagnostic tests. (vi) A warning statement that culture should not be attempted in cases of positive results for an emerging respiratory pathogen unless a facility with an appropriate level of laboratory biosafety ( *e.g.,* BSL 3 and BSL 3+) is available to receive and culture specimens.(vii) A warning statement that device positive results for one or more common respiratory pathogens do not rule out bacterial infection, or co-infection with other common respiratory pathogens. (viii) A warning statement that respiratory pathogen(s) detected may not be the definite cause of disease. (ix) A warning statement that the use of additional laboratory testing ( *e.g.* bacterial culture, immunofluorescence, x-ray findings) and clinical presentation must be taken into consideration in order to obtain the final diagnosis of a respiratory infection.(x) A limiting statement that device negative results for the common respiratory pathogens do not preclude infection of a respiratory pathogen and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions. (xi) A limiting statement that analyte targets ( *e.g.,* pathogen nucleic acid sequences or other molecular signatures) may persist in vivo, independent of organism viability. Detection of analyte target(s) does not imply that the corresponding pathogen(s) is infectious, nor is the causative agent(s) for clinical symptoms.(xii) A limiting statement that detection of pathogen nucleic acid sequences or other molecular signatures is dependent upon proper specimen collection, handling, transportation, storage and preparation. Failure to observe proper procedures in any one of these steps can lead to incorrect results. There is a risk of false negative values resulting from improperly collected, transported, or handled specimens. (xiii) A limiting statement that there is a risk of false positive values resulting from cross-contamination by target organisms, their nucleic acids or amplified product, or from non-specific signals in the assay. (xiv) A limiting statement that there is a risk of false negative results due to the presence of nucleic acid sequence variants in the pathogen targets of the device. (xv) A limiting statement that device performance was not established in immunocompromised patients. (xvi) A limiting statement that positive and negative predictive values are highly dependent on prevalence. The device performance was established during one or more specific respiratory seasons. The performance for some respiratory pathogens may vary depending on the prevalence and patient population tested. False positive test results are likely when prevalence of disease due to a particular respiratory pathogen is low or non-existent in a community. (xvii) In situations where the performance of the device was estimated based largely on testing pre-selected banked retrospective clinical specimens and/or contrived clinical specimen, a limiting statement that the estimated device performance of that specific pathogen or pathogen subtype may not reflect the performance or prevalence in the intended use population. (xviii) For devices with an intended use that includes detection of emerging respiratory pathogen(s), a limiting statement that testing with the device should not be performed unless the patient meets clinical and/or epidemiologic criteria for testing suspected specimens of the emerging respiratory pathogen. (xix) For devices with an intended use that includes detection of emerging respiratory pathogen(s), a limiting statement that positive results obtained with the device for the emerging respiratory pathogen are for the presumptive identification of that pathogen and that the definitive identification of the emerging respiratory pathogen requires additional testing and confirmation procedures in consultation with the appropriate public health authorities ( *e.g.,* local or state public health departments) for whom reporting is necessary.(xx) For devices with an intended use that includes detection of emerging respiratory pathogen(s), a limiting statement that negative results for the emerging respiratory pathogen, even in the context of device positive results for one or more of the common respiratory pathogens, do not preclude infection with the emerging respiratory pathogen and should not be used as the sole basis for patient management decisions. (xxi) For devices with an intended use that includes detection of emerging respiratory pathogen(s), a limiting statement that negative results for the emerging respiratory pathogen may be due to infection of the emerging respiratory pathogen at a specific respiratory tract location that may not be detected by a particular clinical specimen type. A negative result for the emerging respiratory pathogen in an asymptomatic individual does not rule out the possibility of future illness and does not demonstrate that the individual is not infectious. (xxii) For devices with an intended use that includes detection of emerging respiratory pathogen(s), a limiting statement that a nationally notifiable Rare Disease of Public Health Significance caused by an emerging respiratory pathogen must be reported, as appropriate, to public health authorities in accordance with local, state, and federal law. (3) Design verification and validation must include: (i) Performance results of an appropriate clinical study ( *e.g.,* a prospective clinical study) for each specimen type, and, if appropriate, results from additional characterized samples. The clinical study must be performed on a study population consistent with the intended use population and must compare the device performance to results obtained using FDA-accepted comparator methods or to expected negative results if the infection is not generally expected in the intended use population. Clinical specimens evaluated in the study must contain relevant organism concentrations applicable to the specimen type(s) and the targeted analyte(s). Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses.(ii) For devices with an intended use that includes detection of emerging respiratory pathogen(s) for which an FDA recommended panel is available, design verification and validation must include the performance results of an analytical study testing an FDA recommended reference panel of characterized samples that contain the emerging respiratory pathogen. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses. (iii) An appropriate risk mitigation strategy, including a detailed description of all procedures and methods, for the post-market identification of genetic mutations and/or novel respiratory pathogen isolates or strains ( *e.g.,* regular review of published literature and annual in silico analysis of target sequences to detect possible mismatches. The required documentation for this device must also include all of the results, including any findings, from the application of this post-market mitigation strategy.(iv) For devices with an intended use that includes detection of multiple common respiratory pathogens, in addition to detecting emerging respiratory pathogen(s) in human clinical specimens, a detailed description of the identity, phylogenetic relationship, or other recognized characterization of the common respiratory pathogens that the device is designed to detect is addressed. Also, address in detail how the device results might be used in a diagnostic algorithm and other measures that might be needed for a laboratory diagnosis of respiratory tract infection. Perform an evaluation of the device compared to a currently appropriate and FDA accepted comparator method. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses. (v) A detailed device description, including the parts that make up the device, ancillary reagents required but not provided, and a detailed explanation of the methodology, including molecular target(s) for each analyte, design of target detection reagents, rationale for target selection, limiting factors of the device ( *e.g.,* saturation level of hybridization and maximum amplification and detection cycle number), internal and external controls, and computational path from collected raw data to reported result (*e.g.,* how collected raw signals are converted into a reported signal and result), as applicable and appropriate.(vi) A detailed description of the device software, including software applications and hardware-based devices that incorporate software. (vii) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiate between the Influenza A virus subtypes in human clinical specimens, in addition to detecting emerging respiratory pathogen(s), a detailed description of the identity, phylogenetic relationship, or other recognized characterization of the Influenza A and B viruses that the device is designed to detect, a description of how the device results might be used in a diagnostic algorithm and other measures that might be needed for a laboratory identification of Influenza A or B virus and of specific Influenza A virus subtypes, and a description of the clinical and epidemiological parameters that are relevant to a patient case diagnosis of Influenza A or B and of specific Influenza A virus subtypes. Perform an evaluation of the device compared to a currently appropriate and FDA accepted comparator method. Detailed documentation must be kept of that study and its results, including the study protocol, study report for the proposed intended use, testing results, and results of all statistical analyses. (4) For devices with an intended use that includes detection of Influenza A and Influenza B viruses and/or detection and differentiate between the Influenza A virus subtypes in human clinical specimens, in addition to detecting emerging respiratory pathogen(s), the labeling required under § 809.10 of this chapter must include the following: (i) Where applicable, a limiting statement that performance characteristics for Influenza A were established when Influenza A/H3 and A/H1-2009 (or other pertinent Influenza A subtypes) were the predominant Influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary. (ii) Where applicable, a warning statement that reads if infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. (iii) Where the device results interpretation involves combining the outputs of several targets to get the final results, such as a device that both detects Influenza A and differentiates all known Influenza A subtypes that are currently circulating, the device's labeling required under § 809.10(b)(9) of this chapter must include a clear interpretation instruction for all valid and invalid output combinations, and recommendations for any required follow up actions or retesting in the case of an unusual or unexpected device result. (iv) A limiting statement that if a specimen yields a positive result for Influenza A, but produces negative test results for all specific influenza A subtypes intended to be differentiated ( *e.g.,* H1-2009 and H3), this result requires notification of appropriate local, state, or federal public health authorities to determine necessary measures for verification and to further determine whether the specimen represents a novel strain of Influenza A.(5) The manufacturer must perform annual analytical reactivity testing of the device with contemporary influenza strains. This annual analytical reactivity testing must meet the following criteria: (i) The appropriate strains to be tested will be identified by FDA in consultation with the Centers for Disease Control and Prevention (CDC) and sourced from CDC or an FDA designated source. If the annual strains are not available from CDC, FDA will identify an alternative source for obtaining the requisite strains. (ii) The testing must be conducted according to a standardized protocol considered and determined by FDA to be acceptable and appropriate. (iii) By July 31 of each calendar year, the results of the last 3 years of annual analytical reactivity testing must be included as part of the device's labeling. If a device has not been on the market long enough for 3 years of annual analytical reactivity testing to have been conducted since the device received marketing authorization from FDA, then the results of every annual analytical reactivity testing since the device received marketing authorization from FDA must be included. The results must be presented as part of the device's labeling in a tabular format, which includes the detailed information for each virus tested as described in the certificate of authentication, either by: (A) Placing the results directly in the device's labeling required under § 809.10(b) of this chapter that physically accompanies the device in a separate section of the labeling where the analytical reactivity testing data can be found; or (B) In the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public website where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's website that discusses the device, must provide a prominently placed hyperlink to the web page containing this information and must allow unrestricted viewing access. (6) If one of the actions listed at section 564(b)(1)(A)-(D) of the FD&C Act occurs with respect to an influenza viral strain, or if the Secretary of Health and Human Services (HHS) determines, under section 319(a) of the Public Health Service Act, that a disease or disorder presents a public health emergency, or that a public health emergency otherwise exists, with respect to an influenza viral strain: (i) Within 30 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation, the manufacturer must have testing performed on the device with those viral samples in accordance with a standardized protocol considered and determined by FDA to be acceptable and appropriate. The procedure and location of testing may depend on the nature of the emerging virus. (ii) Within 60 days from the date that FDA notifies manufacturers that characterized viral samples are available for test evaluation and continuing until 3 years from that date, the results of the influenza emergency analytical reactivity testing, including the detailed information for the virus tested as described in the certificate of authentication, must be included as part of the device's labeling in a tabular format, either by: (A) Placing the results directly in the device's labeling required under § 809.10(b) of this chapter that physically accompanies the device in a separate section of the labeling where analytical reactivity testing data can be found, but separate from the annual analytical reactivity testing results; or (B) In a section of the device's label or in other labeling that physically accompanies the device, prominently providing a hyperlink to the manufacturer's public website where the analytical reactivity testing data can be found. The manufacturer's home page, as well as the primary part of the manufacturer's website that discusses the device, must provide a prominently placed hyperlink to the web page containing this information and must allow unrestricted viewing access.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION # SPECIAL 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ## I Background Information: A 510(k) Number K243222 B Applicant BioFire Diagnostics, LLC (bioMerieux) C Proprietary and Established Names BIOFIRE FILMARRAY Pneumonia Panel (BIOFIRE Pneumonia Panel); BIOFIRE FILMARRAY Pneumonia Panel plus (BIOFIRE Pneumonia Panel plus) D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | QDS | Class II | 21 CFR 866.4001 - A Multiplex Respiratory Panel To Detect And Identify Emerging Respiratory Pathogen(S) And Common Respiratory Pathogens In Human Clinical Specimens | MI - Microbiology | | QDP | Class II | 21 CFR 866.3985 - Device to detect and identify microorganisms and associated resistance marker nucleic acids directly in respiratory specimens | MI - Microbiology | ## II Review Summary: This 510(k) submission contains information/data on modifications made to the submitter's own Class II device requiring a 510(k). The following items are present and acceptable. Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. 2. Submitter's statement that the INDICATIONS FOR USE/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for an update to the BIOFIRE Pneumonia Panel and Pneumonia Panel plus pouch module software to mitigate issues with false positive Coronavirus assay results. 4. Comparison Information (i.e., similarities and differences) to the submitter's legally marketed predicate device including, labeling, intended use, and physical characteristics. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis. b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied. The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared device. K243222 - Page 2 of 2
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...