LOTUS ULTRASONIC SYSTEM FOR CEMENTED IMPLANT EXTRACTION (USCIX)

K072961 · Sra Developments, Ltd. · JDX · Dec 19, 2007 · Orthopedic

Device Facts

Record IDK072961
Device NameLOTUS ULTRASONIC SYSTEM FOR CEMENTED IMPLANT EXTRACTION (USCIX)
ApplicantSra Developments, Ltd.
Product CodeJDX · Orthopedic
Decision DateDec 19, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.4580
Device ClassClass 2
AttributesTherapeutic

Intended Use

The LOTUS Ultrasonic System for Cemented Implant eXtraction System (USCIX) is indicated for the softening of bone cement to facilitate the removal of implants, including pedicle screws and other spinal implants, during revision surgery.

Device Story

USCIX is an ultrasonic surgical system used during revision surgery to facilitate implant removal. The system comprises a power module, a reusable handpiece, an acoustic coupler, and single-use adaptors. It generates longitudinal mode ultrasound at a nominal frequency of 29kHz (range 28-30kHz). The device is operated by a surgeon; ultrasound energy is applied directly to the implant, which transfers energy to the surrounding bone cement, softening it to allow for easier extraction. This method is specifically designed for scenarios with restricted access, such as spinal procedures. The device benefits patients by simplifying the removal of cemented hardware, potentially reducing surgical time and trauma to surrounding bone tissue.

Clinical Evidence

Bench testing only. Performance testing included electromagnetic compatibility (FCC Part 18) and electrical safety (UL 60601-1) to demonstrate equivalence.

Technological Characteristics

Ultrasonic surgical instrument; longitudinal mode ultrasound at 28-30kHz (nominal 29kHz). System includes power module, reusable handpiece, acoustic coupler, and single-use adaptors. Electrical safety per UL 60601-1; EMC per FCC Part 18.

Indications for Use

Indicated for patients undergoing revision surgery requiring removal of cemented implants, including pedicle screws and other spinal implants, by softening bone cement.

Regulatory Classification

Identification

A sonic surgical instrument is a hand-held device with various accessories or attachments, such as a cutting tip that vibrates at high frequencies, and is intended for medical purposes to cut bone or other materials, such as acrylic.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 888.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K 072961 # Section 5 # 510(k) Summary # Contents | Reason for Submission | 2 | |-----------------------------------|---| | Legally Marketed Predicate Device | 2 | | Device Description | 2 | | Intended Use | 2 | | Technological Characteristics | 3 | | Substantial Equivalence Summary | 3 | | Testing | 3 | | Conclusions | 3 | : ・ DEC 19 2007 : . {1}------------------------------------------------ K072961 #### 510(k) Summary LOTUS Ultrasonic System for Cemented Implant eXtraction System (USCIX) Common Name: Classification Name: Product Code: Subsequent Class: Sponsor: Ultrasonic Surgical Instrument Instrument, Surgical, Sonic And Accessory/Attachment JDX LZV SRA Developments Ltd Bremridge House Ashburton Devon TQ13 7JX nK T: +44 1364 652426 F: +44 1364 653589 Contact: Dr. Michael J.R. Young. #### A. REASON FOR SUBMISSION This 510(k) is being filed to obtain clearance to market the LOTUS Ultrasonic System for Cemented Implant eXtraction System (USCIX). #### B. LEGALLY MARKETED PREDICATE DEVICES This premarket notification will demonstrate that the LOTUS Ultrasonic System for Cemented Implant eXtraction System (USCIX) is substantially equivalent to the Orthosonics OSCAR OE3000DB cleared by FDA as K051053. #### C. DEVICE DESCRIPTION The USCIX consists of a power module which generates the ultrasonic energy and provides overall control of the device, a reusable implant removal handpiece, a reusable acoustic coupler and a range of single use adaptors. USCIX employs longitudinal mode ultrasound at nominally 29kHz (Range 28.00-30.00kHz) to remove cemented implants for revision. #### D. INTENDED USE The LOTUS Ultrasonic System for Cemented Implant eXtraction System (USCIX) is indicated for the softening of bone cement to facilitate the removal of implants, including pedicle screws and other spinal implants, during revision surgery. Section 5 {2}------------------------------------------------ ## E. TECHNOLOGICAL CHARACTERISTICS The basic technological characteristics of the USCIX are the same as those of the predicate device. Both systems are designed to use ultrasound to remove an implant from bone cement during revision surgery. The primary difference is that the USCIX system is specifically designed to apply the ultrasound directly to the implant, softening the bone cement encasing it. This approach is particularly advantageous where the access to the bone cement is restricted such as in the spine. The predicate applies the ultrasound to the bone cement via a probe to soften it facilitating the implants removal. The USCIX utilises the same process used by OSCAR when its titanium probe has become accidentally or deliberately embedded in cement. In this case the OSCAR probe is directly equivalent to the USCIX implant (See OSCAR User Manual in Section 13 of this submission). ## F. SUBSTANTIAL EQUIVALENCE SUMMARY USCIX is a medical device, that uses ultrasound to soften bone cement to facilitate the removal of an implant during revision surgery. This is the same function as the predicate device. USCIX has the same technological characteristics as the predicate devices. However, the descriptive characteristics may not be sufficiently precise to assure substantial equivalence. Therefore, performance testing was carried out for some characteristics. The data from this testing are available and are presented in this 510(k). The data do in fact demonstrate equivalence. ## G. TESTING Testing to FCC Part 18 has been carried out at EMC Network (SW) Ltd, Devon UK, refer to Section 17 for data. Electrical testing to UL 60601-1 has been successfully carried out by Underwriters Laboratories and the Certificate of Compliance is included in Section 17. ## H. CONCLUSIONS This premarket notification has demonstrated substantial equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the center for Devices and Radiological Health. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three lines forming its body and wings. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 1 9 2007 SRA Developments, Ltd. % Michael J.R. Young, Ph.D. Managing Director Bremridge House Asburton, Devon, TQ13 7JX United Kingdom Re: K072961 Trade/Device Name: LOTUS Ultrasonic System for Cemented Implant eXtraction System (USCIX) Regulation Number: 21 CFR 888.4580 Regulation Name: Sonic surgical instrument and accessories/attachment Regulatory Class: II Product Code: JDX Dated: October 11, 2007 Received: October 19, 2007 Dear Dr. Young: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2 - Michael J.R. Young, Ph.D. forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Prostmarket, Surveillance at (240) 276-3474. For questions regarding the reporting of device acre venine (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Mark N. Millerson Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): K 072961 Device Name: LOTUS Ultrasonic System for Cemented Implant eXtraction System (USCIX) Indications For Use: The LOTUS Ultrasonic System for Cemented Implant eXtraction System (USCIX) is indicated for the softening of bone cement to facilitate the removal of implants, including pedicle screws and other spinal implants, during revision surgery. Prescription Use ✓ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use_ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) :
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...