IN-TAC BONE ANCHOR SYSTEM
Device Facts
| Record ID | K964953 |
|---|---|
| Device Name | IN-TAC BONE ANCHOR SYSTEM |
| Applicant | Influence, Inc. |
| Product Code | HXJ · Orthopedic |
| Decision Date | Feb 13, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.4540 |
| Device Class | Class 1 |
| Attributes | Therapeutic |
Indications for Use
The In-Tac Bone Anchor System is intended for soft tissue fixation to the pubic bone by means of bone anchors threaded with suture. The In-Tac Bone Anchor System is indicated for the treatment of stress type (female) urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency.
Device Story
System consists of bone anchor and inserter; used for soft tissue fixation to pubic bone in transvaginal sling procedures. Inserter utilizes force-measuring spring to ensure anchor deployment only upon sufficient insertion force. Modifications from predicate include pre-threaded sutures, integrated retractable anchor shield, and stronger grade stainless steel inserter. Operated by physicians in clinical setting. Device provides mechanical fixation of tissue to bone; assists in treating stress urinary incontinence.
Clinical Evidence
Bench testing only; no clinical data provided. Biocompatibility testing confirmed material safety.
Technological Characteristics
Components: bone anchor and inserter. Materials: stainless steel (inserter), biocompatible materials. Features: pre-threaded sutures, retractable anchor shield, force-measuring spring for deployment control. Mechanical fixation principle.
Indications for Use
Indicated for female patients with stress urinary incontinence caused by urethral hypermobility and/or intrinsic sphincter deficiency.
Regulatory Classification
Identification
An orthopedic manual surgical instrument is a nonpowered hand-held device intended for medical purposes to manipulate tissue, or for use with other devices in orthopedic surgery. This generic type of device includes the cerclage applier, awl, bender, drill brace, broach, burr, corkscrew, countersink, pin crimper, wire cutter, prosthesis driver, extractor, file, fork, needle holder, impactor, bending or contouring instrument, compression instrument, passer, socket positioner, probe, femoral neck punch, socket pusher, reamer, rongeur, scissors, screwdriver, bone skid, staple driver, bone screw starter, surgical stripper, tamp, bone tap, trephine, wire twister, and wrench.
Predicate Devices
- UroTac Bone Anchor System (K962372)
Related Devices
- K990160 — MICROTAC-BONE ANCHOR SYSTEM · Influence, Inc. · Feb 16, 1999
- K992277 — STAPLETAC BONE ANCHOR SYSTEM · Influence, Inc. · Jul 30, 1999
- K962372 — UROTAC BONE ANCHOR INSERTER · Influence, Inc. · Aug 19, 1996
- K972622 — STRAIGHT-IN BONE SCREW FIXATION SYSTEM · Influence, Inc. · Oct 9, 1997
- K020446 — MODIFICATION TO STRAIGHT-IN BONE SCREW FIXATION SYSTEM · American Medical Systems, Inc. · Mar 13, 2002