DYNAMIC COMPRESSION SHAPE MEMORY ALLOY (SMA) STAPLE SYSTEM

K012081 · Medtronic Sofamor Danek, Inc. · JDR · Oct 1, 2001 · Orthopedic

Device Facts

Record IDK012081
Device NameDYNAMIC COMPRESSION SHAPE MEMORY ALLOY (SMA) STAPLE SYSTEM
ApplicantMedtronic Sofamor Danek, Inc.
Product CodeJDR · Orthopedic
Decision DateOct 1, 2001
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The DYNAMIC COMPRESSION SMA Staple System is indicated for use in: (1) Hand and foot bone fragment and osteotomy fixation and joint arthrodesis, (2) Fixation of proximal tibial metaphysis osteotomy, and (3) Fixation of soft tissue to bone such as in anterior cruciate ligament reconstruction.

Device Story

Bone fixation device fabricated from Nitinol shape memory alloy; available in two and four prong configurations. Principle of operation: material is pliant at room temperature, allowing prong straightening for insertion; upon reaching body temperature, prongs deflect inward to original shape. Inward deflection provides active compression across osteotomy or arthrodesis sites and creates a claw effect to prevent migration. Used by surgeons in clinical settings for orthopedic procedures. Benefits include stable fixation and dynamic compression to facilitate bone healing.

Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and design.

Technological Characteristics

Material: Nitinol (Shape Memory Alloy). Configuration: Two and four prong staples. Mechanism: Temperature-dependent shape memory effect for active compression. Sterilization: Not specified.

Indications for Use

Indicated for hand and foot bone fragment fixation, osteotomy fixation, joint arthrodesis, proximal tibial metaphysis osteotomy fixation, and soft tissue-to-bone fixation (e.g., ACL reconstruction).

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K012081 # OCT - 1 2001 # DYNAMIC COMPRESSION Shape Memory Alloy (SMA) Staple System 510(k) Summary of Safety and Effectiveness June 2001 Medtronic Sofamor Danek, Inc. USA I. Company: 1800 Pyramid Place Memphis, TN 38132 (901) 396-3133 ### Proposed Proprietary Trade Name: DYNAMIC COMPRESSION Shape Memory II. Alloy (SMA) Staple System #### III. Product Description The DYNAMIC COMPRESSION SMA Staple is fabricated from Nitinol Shape Memory Alloy. This bone fixation device consists of both two and four prong staples and is intended to achieve compression in fixation of bones in the hand, foot, tibia and ankle. Nitinol can change its configuration (shape) based on temperature. At room temperature, the material is pliant and the staple prongs can be straightened. This allows for easy insertion into the bone. After the staple has reached body temperature, the staple prongs deflect inward to their original shape. This inward deflection causes staple retention and compression across the osteotomy or arthrodesis site. This inward bending creates a claw effect, and helps prevent the staple from backing out of the bone. #### IV. Indications The DYNAMIC COMPRESSION SMA Staple System is indicated for use in: (1) Hand and foot bone fragment and osteotomy fixation and joint arthrodesis, (2) Fixation of proximal tibial metaphysis osteotomy, and (3) Fixation of soft tissue to bone such as in anterior cruciate ligament reconstruction. #### V. Substantial Equivalence Documentation was provided which demonstrated the DYNAMIC COMPRESSION SMA Staple System to be substantially equivalent to other previously cleared shape memory alloy staples. {1}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/1/Picture/1 description: The image is a black and white circular logo. The logo features a stylized image of an eagle with its wings spread. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged around the top half of the circle. The text is in a sans-serif font and is evenly spaced around the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT - 1 2001 Dr. Richard W. Treharne Senior Vice President, Regulatory Affairs Medtronic Sofamor Danek 1800 Pyramid Place Memphis, Tennessee 38132 Re: K012081 Trade Name: Dynamic Compression Shape Memory Alloy (SMA) Staple System Regulation Number: 888.3030 Regulatory Class: II Product Code: JDR Dated: June 29, 2001 Received: July 3, 2001 Dear Dr. Treharne: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. {2}------------------------------------------------ Page 2 - Dr. Richard W. Treharne If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Mark McMillan Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ 510(k) Number (if known): K012081 DYNAMIC COMPRESSION Shape Memory Alloy Staple System Device Name: ### Indications for Use: The DYNAMIC COMPRESSION SMA Staple System is indicated for use in: (1) Hand and foot bone fragment and osteotomy fixation and joint arthrodesis, (2) Fixation of proximal tibial metaphysis osteotomy, and (3) Fixation of soft tissue to bone such as in anterior cruciate ligament reconstruction. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Evaluation (ODE) Prescription Use y 2 (Per 21 CFR 801.109) OR Over-the-counter Use No (Optional 1-2-96) Mark N. Mulkerin Division Sion-C Division of General, Restorative and Neurological Devices 510(k) Number _ KO12081
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...