FIREFLY® Pedicle Screw Navigation Guide

K181399 · Mighty Oak Medical, Inc. · PQC · Jul 24, 2018 · Orthopedic

Device Facts

Record IDK181399
Device NameFIREFLY® Pedicle Screw Navigation Guide
ApplicantMighty Oak Medical, Inc.
Product CodePQC · Orthopedic
Decision DateJul 24, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2

Indications for Use

The FIREFLY® Pedicle Screw Navigation Guide can be used with any 510(k) cleared, legally marketed, pedicle screw spinal system (for its cleared indications for use) and its respective compatible components for non-cervical open, posterior spinal fixation procedures (TI-S2/S2Al and ilium) intended for fusion, with the additional conditions listed below: - . Pedicle screw's shank is straight along its longitudinal axis (i.e. not curved) - . Pedicle screw's major and minor thread diameters are centered about the longitudinal axis - Pedicle screw's longitudinal axis matches the direction of insertion ● - Pedicle screw is intended to be inserted into a pilot hole ● - Pedicle screw's diameter is larger than the pilot hole created with FIREFLY® ● - Patient's pedicle must be dimensionally adequate to accommodate a pedicle screw, as determined on ● preoperative scan - . Compatible pediale screw spinal system instruments may be used with the FIREELY® Pedicle Screw Navigation Guide - Pedicle sounding probes (a.k.a. feeler/ball-tip probes) may be used to confirm pedicle integrity o - Only OEM pedicle screw spinal system taps specified in the Approved Patient-Specific Surgical Plan may be O quided to tap pilot holes - All other pedicle screw spingl system components and accessories (including non-quided faps) are to be O used, after removal of the FIREFLY® Pedicle Screw Navigation Guide, as directed by the pedicle screw spinal system's instructions for use This device is intended for single use only.

Device Story

FIREFLY® Pedicle Screw Navigation Guide assists surgeons in accurate pedicle screw placement during spinal fixation. Input: patient-specific CT imaging data. Process: surgeon defines surgical trajectories presurgically; system generates patient-specific pedicle screw guides and bone models that fit patient anatomy. Operation: guides are placed on patient anatomy in the OR to direct surgical instruments (taps) for pilot hole creation along pre-planned trajectories. Output: physical guides for instrument alignment. Benefit: improved accuracy of pedicle screw placement. Used in open, posterior spinal procedures by surgeons. Device is single-use.

Clinical Evidence

No clinical data. Performance supported by cadaveric accuracy testing demonstrating that acceptance criteria were met and the device performs as intended.

Technological Characteristics

Patient-specific navigation guides. Materials: Titanium alloy (ASTM F136), stainless steel (ASTM F899), and epoxy resin (Accura ABS White SL 7810). Single-use. Mechanical guidance system based on patient-specific anatomy derived from CT data.

Indications for Use

Indicated for patients undergoing non-cervical open, posterior spinal fixation procedures (T1-S2/S2AI and ilium) requiring fusion, where pedicle screws are used. Requires adequate pedicle dimensions determined by preoperative CT scan. Contraindicated for curved pedicle screws or scenarios where screw diameter is not larger than the pilot hole.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in smaller blue letters below. July 24, 2018 Mighty Oak Medical Inc. Mr. Mark A. Wylie Vice President of Quality and Regulatory 750 West Hampden Avenue, Suite 120 Englewood, Colorado 80110 Re: K181399 Trade/Device Name: FIREFLY® Pedicle Screw Navigation Guide Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: POC Dated: May 29, 2018 Received: May 29, 2018 Dear Mr. Wylie: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Ronald P. Jean -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known # K181399 #### Device Name FIREFLY® Pedicle Screw Navigation Guide #### Indications for Use (Describe) The FIREFLY® Pedicle Screw Navigation Guide can be used with any 510(k) cleared, legally marketed, pedicle screw spinal system (for its cleared indications for use) and its respective compatible components for non-cervical open, posterior spinal fixation procedures (TI-S2/S2Al and ilium) intended for fusion, with the additional conditions listed below: - . Pedicle screw's shank is straight along its longitudinal axis (i.e. not curved) - . Pedicle screw's major and minor thread diameters are centered about the longitudinal axis - Pedicle screw's longitudinal axis matches the direction of insertion ● - Pedicle screw is intended to be inserted into a pilot hole ● - Pedicle screw's diameter is larger than the pilot hole created with FIREFLY® ● - Patient's pedicle must be dimensionally adequate to accommodate a pedicle screw, as determined on ● preoperative scan - . Compatible pediale screw spinal system instruments may be used with the FIREELY® Pedicle Screw Navigation Guide - Pedicle sounding probes (a.k.a. feeler/ball-tip probes) may be used to confirm pedicle integrity o - Only OEM pedicle screw spinal system taps specified in the Approved Patient-Specific Surgical Plan may be O quided to tap pilot holes - All other pedicle screw spingl system components and accessories (including non-quided faps) are to be O used, after removal of the FIREFLY® Pedicle Screw Navigation Guide, as directed by the pedicle screw spinal system's instructions for use This device is intended for single use only. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. {3}------------------------------------------------ # 510(K) SUMMARY Image /page/3/Picture/2 description: The image shows the logo for Mighty Oak Medical. The logo has the words "Mighty Oak Medical" in brown and green lettering. Above the words is a graphic of three green blades of grass with a brown curved line underneath. FIREFLY® Pedicle Screw Navigation Guide # Submitter: Mighty Oak Medical Inc. 750 W. Hampden Ave., Suite 120 Englewood, CO 80110 (720) 398-9703 Contact: Mark A. Wylie, VP of Quality and Requlatory Date Prepared: 01June 2018 Device Trade Name: FIREFLY® Pedicle Screw Navigation Guide Common Name: Pedicle Screw Placement Guide Device Classification: Class II Regulation, Name: 221 CFR 888.3070, Thoracolumbosacral pedicle screw system Device Product Code: POC # Predicate Device(s): FIREFLY® Pedicle Screw Navigation Guide Primary Predicate: K162419 (S.E. 10/28/2016) ## Description The FIREFLY® Pedicle Screw Navigation Guide is intended to assist in the accurate placement of pedicle screws. It consists of single-use components designed for treatment of a specific patient as well as reusable non-patient-specific components. The FIREFLY® Pedicle Screw Navigation Guide uses Patient-Specific Pedicle Screw Guides that fit on the patient's anatomy to guide surgical instruments in line with trajectories chosen presurgically, by the surgeon, based on the patient's CT imaging data. Navigation guides are intended to guide instruments to create pilot holes in the pedicles for placing pedicle screws following the Approved Patient-Specific Surgical Plan. Patient-Specific Bone Models may also be provided. The purpose of this traditional 510(k) is to expand the FIREFLY® indications for use to allow use at an additional spinal level. The FIREFLY® design is the same as was originally cleared in K143222 and K162419. {4}------------------------------------------------ # Indications for Use The FIREFLY® Pedicle Screw Navigation Guide can be used with any 510(k) cleared, legally marketed, pedicle screw spinal system (for its cleared indications for use) and its respective compatible components for non-cervical open, posterior spinal fixation procedures (T1-S2/S2AI and ilium) intended for fusion, with the additional conditions listed below: - · Pedicle screw's shank is straight along its longitudinal axis (i.e. not curved) - Pedicle screw's major and minor thread diameters are centered about the longitudinal axis - Pedicle screw's longitudinal axis matches the direction of insertion - · Pedicle screw is intended to be inserted into a pilot hole - Pedicle screw's diameter is larger than the pilot hole created with FIREFLY® - Patient's pedicle must be dimensionally adequate to accommodate a pedicle screw, as determined on preoperative scan - Compatible pedicle screw spinal system instruments may be used with the FIREFLY® Pedicle Screw Navigation Guide - o Pedicle sounding probes (a.k.a. feeler/ball-tip probes) may be used to confirm pedicle integrity - o Only OEM pedicle screw spinal system taps specified in the Approved Patient-Specific Surgical Plan may be quided to tap pilot holes - o All other pedicle screw spinal system components and accessories (including non-quided taps) are to be used, after removal of the FIREFLY® Pedicle Screw Navigation Guide, as directed by the pedicle screw spinal system's instructions for use This device is intended for single use only. # Materials The patient-contacting components of the FIREFLY® Pedicle Screw Navigation Guide are manufactured from titanium alloy (ASTM F136), various stainless steels (ASTM F899), and epoxy resin (Accura ABS White SL 7810). ## Performance Data Additional cadaveric accuracy testing of the FIREFLY® Pedicle Screw Navigation Guide was performed. The results demonstrated that the acceptance criteria were met and that the FIREFLY® Pedicle Screw Navigation Guide's performance is adequate to perform as intended. ## Technological Characteristics The subject FIREFLY® Pedicle Screw Navigation Guide possesses the same technological characteristics as the predicate device. These include: - Performance - Manufacturing process ● - Sterilization ● - Biocompatible materials - Basic design . Technological characteristics which are different have been supported with descriptive information and/or performance data. Therefore the fundamental scientific technology of FIREFLY® is the same as the previously cleared device. ## Conclusion The FIREFLY® Pedicle Screw Navigation Guide possesses the same intended use and technological characteristics as the predicate device. Therefore the FIREFLY® Pedicle Screw Navigation Guide is substantially equivalent for its intended use.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...