← Product Code [PLR](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PLR) · K221542

# Galileo Vertebral Body Replacement Device (K221542)

_Bright Spine · PLR · Jan 13, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PLR/K221542

## Device Facts

- **Applicant:** Bright Spine
- **Product Code:** [PLR](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PLR.md)
- **Decision Date:** Jan 13, 2023
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

When used as a single-level partial vertebral body replacement device, the Galileo™ devices are indicated for use in the cervical spine (C2-T1) of skeletally mature patients for the partial replacement (i.e., partial vertebrectomy) of a vertebral body resected or excised for the treatment of tumors. or trauma/fracture, or osteomvelitis, or to achieve decompression of the spinal cord and neural tissues in cervical degenerative disorders. The device is intended for use with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft. When the Galleo ™ device is used as a single-level, partial vertebral body replacement device, supplemental fixation may be used.

## Device Story

Galileo Vertebral Body Replacement Device; surgical-grade titanium (Ti-6Al-4V) implant; available in various widths/heights. Device features openings for bone graft placement and cellular flow between graft and patient bone. Used in cervical spine (C2-T1) for partial vertebrectomy; provides structural support during fusion. Implanted by surgeons in clinical settings. Benefits patient by restoring vertebral height and facilitating spinal stabilization following resection for tumors, trauma, or degenerative conditions.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical mechanical performance testing conducted in accordance with ASTM F2077.

## Technological Characteristics

Material: Titanium (Ti-6Al-4V). Form factor: Vertebral body replacement implant with openings for bone graft. Mechanical testing: ASTM F2077. Connectivity: None. Software: None.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Galileo Vertebral Body Replacement Device ([K192145](/device/K192145.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 13, 2023

Bright Spine % Bahram Parvinian Founder & Principal Consultant Lighthouse Regulatory Consulting Group 5801 Nicholson Lane, Apt. 1705 North Bethesda, Maryland 20852

Re: K221542

Trade/Device Name: Galileo Vertebral Body Replacement Device Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: PLR Dated: December 14, 2022 Received: December 19, 2022

Dear Bahram Parvinian:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Katherine D. Kavlock -S

for Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

Submission Number (if known)

K221542

Device Name

Galileo Vertebral Body Replacement Device

#### Indications for Use (Describe)

When used as a single-level partial vertebral body replacement device, the Galileo™ devices are indicated for use in the cervical spine (C2-T1) of skeletally mature patients for the partial replacement (i.e., partial vertebrectomy) of a vertebral body resected or excised for the treatment of tumors. or trauma/fracture, or osteomvelitis, or to achieve decompression of the spinal cord and neural tissues in cervical degenerative disorders. The device is intended for use with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft. When the Galleo ™ device is used as a single-level, partial vertebral body replacement device, supplemental fixation may be used.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| Date Prepared               | May 5, 2022                                                                                                                                                                                                                                                                                                                                                                                                                         |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted By                | Bright Spine LLC<br>Robert Simonson<br>t 561-289-9378<br>r@brightspine.com                                                                                                                                                                                                                                                                                                                                                          |
| Primary Contact             | Bahram Parvinian PhD,<br>Founder and Principal Consultant<br>Lighthouse Regulatory Consulting Group LLC<br>(301)938-7669<br>Bahram@lighthouseregulatory.com                                                                                                                                                                                                                                                                         |
| Trade Name                  | Galileo Vertebral Body Replacement Device                                                                                                                                                                                                                                                                                                                                                                                           |
| Common Name                 | Vertebral body replacement device                                                                                                                                                                                                                                                                                                                                                                                                   |
| Classification<br>Name      | Spinal intervertebral body fixation orthosis                                                                                                                                                                                                                                                                                                                                                                                        |
| Class                       | II                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Product Code                | PLR                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| CFR Section                 | 21 CFR section 888.3060                                                                                                                                                                                                                                                                                                                                                                                                             |
| Device Panel                | Orthopedic                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Primary<br>Predicate Device | Galileo Vertebral Body Replacement Device (K192145)                                                                                                                                                                                                                                                                                                                                                                                 |
| Device<br>Description       | The Galileo Vertebral Body Replacement device is a surgical grade titanium (Ti-<br>6A1-4V) device and is available in various widths and heights. It has openings of<br>various sizes to allow for the placement of bone graft and for the free flow of cells<br>between the bone graft and the bone of the patient. It is intended for partial vertebral<br>body replacement in a single vertebra and to hold bone graft material. |
| Materials                   | Titanium (Ti-6A1-4V)                                                                                                                                                                                                                                                                                                                                                                                                                |
| Intended Use                | Partial cervical vertebral body replacement                                                                                                                                                                                                                                                                                                                                                                                         |

{4}------------------------------------------------

| Substantial<br>Equivalence<br>Claimed to<br>Predicate Device                   | Galileo VBR is substantially equivalent to the predicate device in terms of intended<br>use, design, materials used, mechanical safety and performances.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
|--------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| Indications for Use                                                            | When used as a single-level vertebral body replacement device, the Galileo™<br>devices are indicated for use in the cervical spine (C2-T1) of skeletally mature<br>patients for the partial replacement (i.e., partial vertebrectomy) of a vertebral body<br>resected or excised for the treatment of tumors, or trauma/fracture, or osteomyelitis,<br>or to achieve decompression of the spinal cord and neural tissues in cervical<br>degenerative disorders. The device is intended for use with autograft or allogenic<br>bone graft comprising cancellous and/or corticocancellous bone graft. When the<br>Galileo™ device is used as a single-level, partial vertebral body replacement device,<br>supplemental fixation may be used |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
| Summary of the<br>technological<br>characteristics<br>compared to<br>predicate | Galileo VBR has the same intended use and Indications For Use as K192145. The<br>Galileo VBR includes four additional sizes:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
|                                                                                | A (width) B (height) C (depth)    7 mm 7 mm 12 mm    8 mm 8 mm 12 mm    9 mm 9 mm 12 mm    10 mm 10 mm 12 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
|                                                                                | There were no changes made to the device design, materials, or manufacturing<br>method.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
| Non-clinical Test<br>Summary                                                   | Mechanical performance tests were conducted following the ASTM F2077 (Test<br>methods for intervertebral body fusion devices). All tests were passed.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
| Clinical Test<br>Summary                                                       | Not Applicable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |
| SE conclusion                                                                  | The intended use and technological characteristics of Galileo VBR is the same<br>as the identified predicate. The Galileo VBR has additional sizes compared to<br>the predicate; however, these new device sizes do not result in new questions<br>of safety and effectiveness. Based on the required design verification and<br>validation activities, Galileo VBR is substantially equivalent to the identified<br>predicate.                                                                                                                                                                                                                                                                                                            |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |  |

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PLR/K221542](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PLR/K221542)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
