← Product Code [PBF](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PBF) · K240066

# Fine Osteotomy (K240066)

_Bodycad Laboratories, Inc. · PBF · Feb 9, 2024 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PBF/K240066

## Device Facts

- **Applicant:** Bodycad Laboratories, Inc.
- **Product Code:** [PBF](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PBF.md)
- **Decision Date:** Feb 9, 2024
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Fine Osteotomy™ is a system intended for open- and closed-wedge osteotomies, treatment of bone and joint deformities, fixation of fractures and malalignment caused by injury or disease, such as osteoarthritis, of the distal femur and proximal tibia. Fine Osteotomy™ disposable instrumentation is intended to assist in pre-operative planning and/or in guiding the marking of bone and/or guiding of surgical instruments in non-acute, non-joint replacing osteotomies around the knee. Fine Osteotomy™ is a patient-specific device.

## Device Story

System for pre-operative planning and surgical stabilization of distal femur and proximal tibia; utilizes patient CT images to design patient-specific bone plates, screws, and surgical cutting/marking guides. Surgeon uses patient-specific guides intra-operatively to perform osteotomies and align bone according to pre-surgical plan. Implants (plates/screws) stabilize bone resections or fractures. System configurations include: 1) full system (implants + instruments), 2) instruments only, or 3) implants only. Used in clinical settings by orthopedic surgeons. Benefits include improved surgical accuracy and alignment through patient-specific instrumentation and hardware.

## Clinical Evidence

Bench testing only. Verification and validation included mechanical engineering analysis of updated/new single-use instruments and shelf-life validation. No clinical data presented.

## Technological Characteristics

Materials: Wrought Titanium-6Al-4V ELI (ASTM F136-13) for plates/screws; additively manufactured Nylon-12 for guides/models. Patient-specific, single-use implants and instruments. Provided clean, non-sterile. Mechanical fixation via bone screws and custom-fit plates.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Fine Osteotomy™ ([K231314](/device/K231314.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

February 9, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.

Bodycad Laboratories Inc. Nadine Adia Regulatory Affairs and Quality Director 2035 rue du Haut-Bord Quebec, G1N 4R7 Canada

Re: K240066 Trade/Device Name: Fine OsteotomyTM Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: PBF, HWC, HRS Dated: November 21, 2023 Received: January 9, 2024

Dear Nadine Adia:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Shumaya Ali -S

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

#### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K240066

Device Name

Fine Osteotomy TM

Indications for Use (Describe)

Fine Osteotomy™ is a system intended for open- and closed-wedge osteotomies, treatment of bone and joint deformities, fixation of fractures and malalignment caused by injury or disease, such as osteoarthritis, of the distal femur and proximal tibia.

Fine Osteotomy™ disposable instrumentation is intended to assist in pre-operative planning and/or in guiding the marking of bone and/or guiding of surgical instruments in non-acute, non-joint replacing osteotomies around the knee.

Fine Osteotomy™ is a patient-specific device.

Type of Use (Select one or both, as applicable)

< Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

Prepared on: 2024-01-09

| Contact Details             |                                                                             | 21 CFR 807.92(a)(1) |
|-----------------------------|-----------------------------------------------------------------------------|---------------------|
| Applicant Name              | Bodycad Laboratories Inc.                                                   |                     |
| Applicant Address           | 2035 rue du Haut-Bord Quebec G1N 4R7 Canada                                 |                     |
| Applicant Contact Telephone | 4185271388                                                                  |                     |
| Applicant Contact           | Mrs. Nadine ADIA                                                            |                     |
| Applicant Contact Email     | nadia@bodycad.com                                                           |                     |
| Device Name                 |                                                                             | 21 CFR 807.92(a)(2) |
| Device Trade Name           | Fine Osteotomy™                                                             |                     |
| Common Name                 | Single/multiple component metallic bone fixation appliances and accessories |                     |
| Classification Name         | Orthopaedic Surgical Planning And Instrument Guides                         |                     |
| Regulation Number           | 21 CFR 888.3030                                                             |                     |
| Product Code(s)             | PBF, HRS, HWC                                                               |                     |

| Legally Marketed Predicate Devices |                                                          | 21 CFR 807.92(a)(3) |
|------------------------------------|----------------------------------------------------------|---------------------|
| Predicate #                        | Predicate Trade Name (Primary Predicate is listed first) | Product Code        |
| K231314                            | Fine Osteotomy ™™                                        | PBF                 |

## Device Description Summary

Fine Osteotomy™ is a system for planning osteotomies of the distal femur and proximal tibia, and for stabilizing the bone with bone screws and a patient-specific bone plate that fits the patient's anatomy " consists of patient-specific surgical planning and instrument guides designed from images of the patient-specific bone plate designed from the patient's images, compression and/or locking bone screws, and class 1 reusable manual instruments. The bone plate is a patient-specific, single-use implant; the surgical planning and instrument guides are patient-specific, single-use, and discarded after surgery. Fine Osteotomy™ is offered in three configurations: 1) as a system of patient specific implants and single use instruments for performing osteotomies and implanting hardware to stabilize the resection, 2) as patient specific single use instruments alone for performing osteotomies, and 3) as a patient specific bone plate and screws for stabilizing a bone resection or fracture.

When used as a system, Fine Osteotomy™ enables the surgeon to perform and stabilize the bone around the knee that matches the pre-surgical plan using the patient-specific cutting quides and bone planning quides and resection instruments are used alone, Fine Osteotomy™ enables the surgeon to perform an osteotomy around the knee the presurgical plan using the patient-specific cutting guident's CT images. When the bone plate and screws are used alone, Fine Osteotomy™ enables the surgeon to stabilize fractured or resected bone per the pre-surgical plan using the patient's CT images in design of the Bodycad plate and use of the bone models intra operatively to quide placement of the implants and alignment of bone. The Fine Osteotomy System is provided clean, not sterile to the user.

Materials: Wrought Titanium-6Alum ELI Alloy (Ti6A4V ELJ; ASTM F136-13) for the bone plates and screws, additively manufactured Nylon-12 for patient specific, single use resection guides and models.

## Intended Use/Indications for Use

21 CFR 807.92(a)(5)

21 CFR 807.92(a)(4)

{4}------------------------------------------------

Fine Osteotomy™ is a system intended for open- and closed-wedge osteotomies, treatment of bone and joint deformities, fixation of fractures and malalignment caused by injury or disease, such as osteoarthritis, of the distal femur and proximal tibia.

Fine Osteotomy™ disposable instrumentation is intended to assist in pre-operative planning and/or in quiding the marking of bone and/ or guiding of surgical instruments in non-acute, non-joint replacing osteotomies around the knee.

Fine Osteotomy™ is a patient-specific device.

## Indications for Use Comparison

#### 21 CFR 807.92(a)(5)

21 CFR 807.92(a)(6)

(Identical)

Fine Osteotomy™ is a system intended for open- and closed-wedge osteotomies, treatment of bone and joint deformities, fixation of fractures and malalignment caused by injury or disease, such as osteoarthritis, of the distal femur and proximal tibia.

Fine Osteotomy™ disposable instrumentation is intended to assist in pre-operative planning and/or in guiding the marking of bone and/ or quiding of surgical instruments in non-acute, non-joint replacing osteotomies around the knee.

Fine Osteotomy™ is a patient-specific device.

## Technological Comparison

The subject device, Fine Osteotomy™, has the same indications for use, is manufactured with the same materials and processes, and has the same technological features as the primary predicate device which is Fine Osteotomy cleared in K2313 14. Except for updates and additions to single use instrumentation and shelf-life, the technological characteristics of the subject and primary predicate devices are identical. Comparison of the technological characteristics and indicated use of the subject and primary predicate devices demonstrate that the subject device is substantially equivalent to the primary predicate device, both of which are Bodycad's Fine Osteotomy.

#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

Verification and validation (V&V) activities included the following:

• Mechanical Engineering Analysis of updated and new single use instruments demonstrated no new worst case scenarios, and

• Modification of Shelf-Life Referential for Fine Osteotomy Systems

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PBF/K240066](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/PBF/K240066)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
